5 recalls recorded since 2018, most recently on 8 May 2026. 4 are listed as ongoing. Status and classification are as published by the FDA.

Recalls5
Ongoing (FDA)4
Terminated or completed1
Class I (FDA)0

Ongoing Class II Medical devices

Recall started
8 May 2026
Reason
GE HealthCare has become aware of a context synchronization issue in AW Server 3.2 ext. 6.5. When a user selects a patient or exam in the AW Server Web Client worklist and launches an interactive application (e.g. Volume Viewer), the application may open the previous patient's exam instead of the intended one. When this issue occurs, there is no system warning or error notification. If the error is not recognized, a clinical user could review, interpret, or report images for the wrong patient which could lead to misdiagnosis, incorrect clinical decisions, resulting in delayed or incorrect treatment.
Distribution
Worldwide distribution - United States Nationwide and the countries of Algeria, Australia, Belarus, Belgium, Benin, Brazil, Bulgaria, Canada, Chile, Colombia, Croatia, Denmark, Finland, France, Germany, Greece, Hungary, India, Indonesia, Iraq, Israel, Italy, Japan, Korea (Republic Of), Lithuania, Mo
Initiated
Voluntary: Firm initiated
Recall number
Z-2553-2026

Ongoing Class II Medical devices

Recall started
5 September 2025
Reason
Firm has identified a security vulnerability in AW Server products. If exploited, a malicious actor could compromise the confidentiality, integrity, and availability of patient data.
Distribution
Domestic US distribution nationwide. International distribution worldwide.
Initiated
Voluntary: Firm initiated
Recall number
Z-0068-2026

Ongoing Class II Medical devices

Recall started
18 March 2024
Reason
GE HealthCare has become aware of the potential that adhesive material might not have been applied to some bolts of certain Innova IGS, Discovery IGS, Interact IGS, Allia IGS and Allia Pulse IGS products. This could result in a potential fall of the Detector/Elevator or X-ray tube of the frontal Gantry assembly.
Distribution
Worldwide distribution - US Nationwide in the states of CA, FL, GA, IN, MA, NC, OK, SC, TN, TX, VA and the countries of Armenia, China, France, Hungary, India, Italy, Japan, Kazakhstan, Lithuania, Mexico, Russia, Spain, Taiwan, Turkey, Vietnam
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1781-2024, Z-1779-2024, Z-1783-2024, Z-1777-2024, Z-1780-2024, Z-1778-2024, Z-1782-2024

Ongoing Class II Medical devices

Recall started
26 December 2023
Reason
X-ray exposure termination audible signal on Senographe Pristina with Efficiency Suite 2 intermittently not emitted per 21CFR 1020.31(j) requirements.
Distribution
Us Nationwide-Worldwide Distribution
Initiated
FDA Mandated
Recall number
Z-0862-2024

Terminated Class II Medical devices

Recall started
29 March 2018
Terminated
15 June 2021
Reason
The Discovery IGS and Innova IGS systems may experience X-ray abort errors which could result in the loss of imaging mode before or during fluoroscopic use.
Distribution
US distribution: AR, CA, CA, FL, GA, IA, IL, NC, NH, NJ, NM, NY, PA, SC, TN, A,WI, WV. International Distribution: Argentina, Brazil, China, Colombia, Egypt, France, India, Indonesia, Iran, Italy, Japan, Kuwait, Lithuania, Myanmar, Palestine, Poland, Russia, Saudi Arabia, Singapore, Spain, Tuni
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2387-2018, Z-2386-2018, Z-2388-2018

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.