GE Medical Systems, LLC recalls

29 recalls recorded since 2018, most recently on 5 June 2026. 24 are listed as ongoing. Status and classification are as published by the FDA.

Recalls29
Ongoing (FDA)24
Terminated or completed5
Class I (FDA)0

Ongoing Class II Medical devices

Recall started
5 June 2026
Reason
GE HealthCare has become aware that in certain Optima XR200/220/240amx systems that have had a column replacement completed, functional testing of the safety pawl may not have been performed to confirm proper assembly. If a failure of the internal counterpoise cable occurs, the Column Safety mechanism will not work, causing the arm with X-ray tube and collimator to drop downward on the column track. This could result in patient injury during an exam.
Distribution
Worldwide - US Nationwide and the countries of Brazil, Canada, Korea, Mexico, Saudi Arabia, Spain, UK.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2726-2026, Z-2727-2026, Z-2728-2026

Ongoing Class II Medical devices

Recall started
1 May 2026
Reason
GE HealthCare has become aware of a potential issue involving the canopy soffit beneath the control panel end of the device. Under certain conditions, for example, impact during transport, one or more of the fasteners that secure the soffit can disengage. If this occurs, a fastener could fall into the patient compartment. In rare circumstances, if multiple fasteners disengage, the canopy soffit can become loose or detach. These conditions could potentially result in patient injury.
Distribution
Worldwide - US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2477-2026, Z-2475-2026, Z-2476-2026

Ongoing Class II Medical devices

Recall started
20 April 2026
Reason
GE HealthCare has become aware of a potential issue affecting an electrical cable on certain IGS systems equipped with Gaia generators. This issue could lead to a risk of electrical shock to service personnel performing installation, maintenance, or troubleshooting activities inside the C-FRT cabinet while power is ON.
Distribution
Worldwide - US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2314-2026, Z-2317-2026, Z-2318-2026, Z-2315-2026, Z-2316-2026

Ongoing Class II Medical devices

Recall started
26 March 2026
Reason
GE HealthCare has become aware of a potential security vulnerability impacting AW Server deployed via Edison Health Link (EHL) based CT Smart Subscription used in conjunction with certain Revolution Apex, Revolution Ascend, and Revolution CT systems.
Distribution
Worldwide - US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2153-2026, Z-2149-2026, Z-2155-2026, Z-2151-2026, Z-2154-2026, Z-2152-2026, Z-2150-2026

Ongoing Class II Medical devices

Recall started
3 February 2026
Reason
There is a potential issue that can lead to discordance between the simulated needle trajectory/tip position and the actual needle trajectory/tip position.
Distribution
Worldwide distribution - US Nationwide.
Initiated
Voluntary: Firm initiated
Recall number
Z-1590-2026

Ongoing Class II Medical devices

Recall started
30 January 2026
Reason
There is a potential cybersecurity vulnerability affecting certain versions of Centricity Universal Viewer. User login credentials may be exposed on the local client workstation, which could allow an unauthorized individual to potentially impact system availability and/or manipulate data.
Distribution
Worldwide - US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1570-2026, Z-1571-2026, Z-1569-2026

Ongoing Class II Medical devices

Recall started
24 October 2025
Reason
Some Senographe Pristina systems either have a missing or an incorrect X-Ray Warning label, which is a nonconformance to FDA Regulation 21 CFR ¿1020.30(j).
Distribution
U.S. and OUS.
Initiated
FDA Mandated
Recall number
Z-0875-2026

Ongoing Class II Medical devices

Recall started
18 September 2025
Reason
The Ultrasound-Guided Attenuation Parameter (UGAP) measurement data may display inaccurate values representing liver steatosis. This could potentially lead to inappropriate clinical decisions impacting overall care.
Distribution
US Nationwide. Global Distribution.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0156-2026, Z-0157-2026, Z-0155-2026

Ongoing Class II Medical devices

Recall started
22 August 2025
Reason
If a user incorrectly places the Acquisition Module into the Acquisition Module Holder, it can result in excessive bending and wear of the Acquisition Trunk Cable. This is visible as a damaged sheath or exposed wires. If a damaged Acquisition Trunk Cable is not recognized and removed from service, the cable can overheat with continued use.
Distribution
Worldwide Distribution: US (nationwide) and OUS (foreign/Global) to countries of: Australia, Austria, Bahrain, Belgium, Canada, Czech Republic, Denmark, Finland, France, Germany, Hong Kong, Ireland, Israel, Italy, Japan, Korea, Republic of, Kuwait, Malaysia, Netherlands, New Zealand, Norway, Philipp
Initiated
Voluntary: Firm initiated
Recall number
Z-0231-2026

Ongoing Class II Medical devices

Recall started
2 July 2025
Reason
GE HealthCare has become aware of a non-conformance to 21 CFR 1020.30 (k) in certain OEC 9800 and OEC 9900 Elite C-arm Systems. Specifically, impacted systems are missing the full layer of lead tape in the collimator covers, which will lead to a small increase in radiation exposure to the patient and/or operator. The missing lead tape is not detectable by the user.
Distribution
US Nationwide distribution.
Initiated
FDA Mandated
Recall numbers
Z-2342-2025, Z-2341-2025

Ongoing Class II Medical devices

Recall started
20 June 2025
Reason
GE HealthCare has become aware that for certain MR systems (SIGNA Artist, SIGNA Voyager, Discovery MR450 1.5T, Optima MR450w 1.5T, SIGNA Architect, SIGNA Premier, SIGNA UHP, Discovery MR750 3.0T, Discovery MR750w 3.0T, SIGNA Hero, SIGNA PET/MR, SIGNA Artist Evo, SIGNA Voyager AIR (China Only), SIGNA Architect AIR (China Only), SIGNA Premier XT (China Only), SIGNA Premier Elite (China Only), SIGNA Premier MAX (China Only), SIGNA Hero XT (China Only), SIGNA Hero Elite (China Only), SIGNA Hero MAX (China Only), SIGNA PET/MR AIR (China Only), SIGNA PET/MR AIR Plus (China Only), the planned maintenance steps to check the functionality of the caster locks on the MR patient table may not have been performed.
Distribution
Worldwide distribution.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2299-2025, Z-2314-2025, Z-2313-2025, Z-2300-2025, Z-2306-2025, Z-2301-2025, Z-2316-2025, Z-2310-2025, Z-2319-2025, Z-2309-2025, Z-2317-2025, Z-2311-2025 and 10 more

Ongoing Class II Medical devices

Recall started
23 May 2025
Reason
GE HealthCare has become aware that gradient coils for certain MR systems (see affected product list in this letter), under specific conditions, can produce elevated acoustic noise during scanning. As a result, acoustic levels can exceed 99dB, the limit established by International Electrotechnical Commission (IEC 60601-2-33) when using hearing protection with a Noise Reduction Rating (NRR) of 29dB, as currently required in the Operator Manual.
Distribution
Worldwide distribution - US Nationwide and the country of China.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2130-2025, Z-2131-2025, Z-2133-2025, Z-2132-2025

Ongoing Class II Medical devices

Recall started
21 March 2025
Reason
OEC Elite and OEC 3D Mobile C-Arms Exceeding 4% Source to Image Distance in X-ray Field.
Distribution
US and OUS
Initiated
FDA Mandated
Recall numbers
Z-1819-2025, Z-1820-2025

Ongoing Class II Medical devices

Recall started
6 March 2025
Reason
The battery for certain ultrasound systems can potentially develop an internal failure which could result in smoke or fire.
Distribution
Worldwide - US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1555-2025, Z-1556-2025

Ongoing Class II Medical devices

Recall started
18 February 2025
Reason
GE HealthCare has become aware of a potential issue on certain Revolution CT and Revolution Apex series CT systems (please see affected product details and the Appendix) that can result in coolant fluid (glycol) leaking onto the floor near the base of the gantry. The fluid is visible as it has a blue color. It is not corrosive and is not at a harmful temperature. However, there is a potential for injury if the fluid is not noticed and a person slips and/or falls.
Distribution
Worldwide distribution.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1297-2025, Z-1301-2025, Z-1302-2025, Z-1303-2025, Z-1299-2025, Z-1300-2025, Z-1298-2025

Ongoing Class II Medical devices

Recall started
28 January 2025
Reason
The user can continue to take the next exposure after reaching the Automatic Exposure Control (AEC) limit, without taking any further actions to acknowledge or adjust exposure techniques. In all cases, a message still appears on the screen when the AEC limit is reached.
Distribution
Worldwide Distribution: US (nationwide) including Puerto Rico and OUS (foreign) countries to: Canada, Chile, China, Colombia, Coast Rica, Croatia, Cyprus, Czech Republic, Czechia, Dominican Republic, Egypt, Estonia, Finland, France, Germany, Gibraltar, Greece, Honduras, Hong Kong, Hungary, Iceland,
Initiated
FDA Mandated
Recall numbers
Z-1046-2025, Z-1043-2025, Z-1045-2025, Z-1044-2025

Ongoing Class II Medical devices

Recall started
6 December 2024
Reason
There is a software issue that can result in the loss of patient monitoring on the Portrait Central Viewer Application after 425 days of continuous run-time.
Distribution
Worldwide - US Nationwide distribution in the states of OH, TX and the countries of Australia, China, Czech Republic, France, Italy, Netherlands, Qatar, Singapore, UAE, UK.
Initiated
Voluntary: Firm initiated
Recall number
Z-0918-2025

Ongoing Class II Medical devices

Recall started
25 November 2024
Reason
GE HealthCare has become aware that in certain Optima XR200/220/240amx systems that have had a column replacement performed, some bolts used in the Column Safety mechanism can be missing. If these bolts are missing and a failure of the internal counterpoise cable occurs, the Column Safety mechanism will not work properly causing the arm with X-ray tube and collimator to drop downward on the column track.
Distribution
Worldwide
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0812-2025, Z-0811-2025, Z-0813-2025

Ongoing Class II Medical devices

Recall started
18 November 2024
Reason
Due to a manufacturing issue, certain Tec 6 Plus and Tec 800 series vaporizers can deliver less anesthetic agent than the vaporizer dial setting.
Distribution
US Nationwide. Global Distribution.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0815-2025, Z-0818-2025, Z-0814-2025, Z-0817-2025, Z-0816-2025

Ongoing Class II Medical devices

Recall started
28 October 2024
Reason
Gradient coils for certain MR systems (see affected product list in this letter), under specific conditions, can produce elevated acoustic noise during scanning.
Distribution
Worldwide Distribution: US (nationwide) and OUS (foreign) to countries of: Canada, China, India, Japan, Korea, Latvia, Mexico, Poland, Russia, Saudi Arabia, Spain, and Taiwan.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0519-2025, Z-0517-2025, Z-0518-2025, Z-0520-2025

Ongoing Class II Medical devices

Recall started
14 June 2024
Reason
GE HealthCare has become aware that for certain MR systems, the system predicted B1+RMS value can exceed the B1+RMS user prescribed limit when scanning in Low SAR Mode and the following two conditions are met: 1. 2D FSE T2 FLAIR OR T2 FLAIR Propeller imaging sequence is selected, AND 2. the Optimized T2 FLAIR sequence option is turned off OR not included in the MR configuration. If this occurs, it can result in overheating of an MR conditional implant.
Distribution
Worldwide distribution.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2473-2024, Z-2459-2024, Z-2475-2024, Z-2466-2024, Z-2467-2024, Z-2470-2024, Z-2478-2024, Z-2481-2024, Z-2474-2024, Z-2476-2024, Z-2468-2024, Z-2472-2024 and 14 more

Ongoing Class II Medical devices

Recall started
16 November 2022
Reason
GE Healthcare has become aware that if certain B1x5P / B1x5M patient monitors use "mmHg" or "kPa" as the unit for CO2 measurement, the displayed value for CO2 measurement can be inaccurate when used in a location that is not at or near sea level.
Distribution
Worldwide distribution.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0729-2023, Z-0732-2023, Z-0730-2023, Z-0731-2023, Z-0733-2023

Ongoing Class II Medical devices

Recall started
10 November 2022
Reason
Potential for batteries that have not been replaced at 2 years, as recommended in the Service Manual, they can fail or possibly emit smoke, or catch fire.
Distribution
Worldwide - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, GU, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VI, VT, WA, WI, WV and the countries of Afghan
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0753-2023, Z-0754-2023, Z-0752-2023, Z-0756-2023, Z-0757-2023, Z-0759-2023, Z-0755-2023, Z-0758-2023

Ongoing Class II Medical devices

Recall started
28 October 2022
Reason
There is the potential that after installation of the IAS tool, some studies can remain hidden without user notification. Access to those hidden studies requires GE Healthcare assistance, which can result in a delay in reading the study.
Distribution
Nationwide distribution to CA, CO, DE, FL, IA, IL, MA, MI, MN, MO, MS, MT, ND, NJ, NY, OK, OR, SD, TX, WI, WV, WY. International distribution to Austria, Brazil, France, Germany, Ghana, Ireland, Italy, Kuwait, Poland, Saudi Arabia, South Africa, Spain, Switzerland, United Kingdom.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0722-2023, Z-0721-2023

Terminated Class II Medical devices

Recall started
3 October 2018
Terminated
27 May 2022
Reason
A small area on the bore surface, which is normally accessible to the patient, can get warm enough to cause a serious burn in the event the patient touches the bore and proper padding is not used.
Distribution
Worldwide - US Nationwide Distribution in the state of Utah. International distribution to France, Republic of Korea, and Sweden.
Initiated
Voluntary: Firm initiated
Recall number
Z-0236-2019

Terminated Class II Medical devices

Recall started
7 June 2018
Terminated
18 June 2020
Reason
The latest on-site software version may not have been reinstalled at some sites after service activities were performed resulting in images possibly flipped left/right and/or there can be patient data mismatch.
Distribution
Worldwide Distribution: US (nationwide) including states of:: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV; and countries of: Algeria, Argentina,
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0110-2019, Z-0111-2019, Z-0112-2019

Terminated Class II Medical devices

Recall started
6 June 2018
Terminated
27 May 2022
Reason
GE Medical Systems has discovered a small area on the bore surface of the Discovery MR750w Nuclear Magnetic Resonance Imaging System, normally accessible to the patient, can get warm enough to cause a serious burn in the event the patient touches the bore and proper padding is not used.
Distribution
Worldwide Distribution - US Nationwide and in the states of AK, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, NC, NH, NJ, NV, NY, OH, OK, OR, PA, SC, TN, TX, US, UT, VA, WA, WI, and WY. and the countries of Algeria, Argentina, Australia, Austria, Bahrain, Bel
Initiated
Voluntary: Firm initiated
Recall number
Z-3279-2018

Terminated Class II Medical devices

Recall started
30 March 2018
Terminated
29 March 2024
Reason
There is a potential safety issue with the table transfer release of the MR Surgical Suite System which may cause an unexpected loss of function of the patient transfer mechanism that allows the patient to be moved from the GE MR Surgical Table to the Maquet Magnus Surgical Table.
Distribution
US Nationwide distribution to AZ, CA, CO, FL, IL, KS, MI, MN, NY, OH, PA, TX, UT, VA, WI. Worldwide distribution to Austria, Brazil, China, Czech Republic, France, Germany, India, Italy, Japan, Korea, Kuwait, South Africa, Turkey, Turkey, United Arab Emirates, United Kingdom.
Initiated
Voluntary: Firm initiated
Recall number
Z-2527-2018

Terminated Class II Medical devices

Recall started
7 December 2015
Terminated
8 November 2018
Reason
GE Healthcare has recently become aware of a potential safety issue with the 3.0T 6 Channel Flex Coil used with the 3T MR750W Surgical Suite Scanners. Coil overheating can occur when the device is used in Mode 2 setup. This could lead to a serious patient thermal injury. There have been no injuries reported as a result of this issue. UPDATE 5/27/2016: Recall has been updated to include an additional product which includes the Discovery 3.0T MR750 systems running Application Software DV24 R01, DV25 R02, or DV25.1 M3.
Distribution
Worldwide Distribution- US Distribution including the states of AZ, CA, CO, IL IN, MI, MN, NY, OH, OK, SD, TX, WI, VA, DE, FL, MD, ME, MO, NC, NH, OR, PA, TN, WI, DC, PR and the countries of CANADA, CHINA, CZECH REPUBLIC, ITALY, JAPAN, NETHERLANDS, NEW ZEALAND, PORTUGAL, ROMANIA, SOUTH AFRICA, SWIT
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0201-2019, Z-0553-2016

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.