GE Medical Systems China Co., Ltd. recalls

7 recalls recorded since 2018, most recently on 21 April 2026. 6 are listed as ongoing. Status and classification are as published by the FDA.

Recalls7
Ongoing (FDA)6
Terminated or completed1
Class I (FDA)2

Ongoing Class II Medical devices

Recall started
21 April 2026
Reason
Software in development inadvertently made available for customer use. Use of this software version can result in clinical information being associated with the incorrect patient. This could impact clinical decision-making.
Distribution
Worldwide - US Nationwide distribution in the states of Florida and Georgia. The countries of India, Italy, Republic of Korea, and United Kingdom.
Initiated
Voluntary: Firm initiated
Recall number
Z-2306-2026

Ongoing Class I Medical devices

Recall started
14 November 2025
Reason
Potential unexpected shutdown of Carestation 600 and 700 Series Anesthesia Systems containing certain power management boards, if the AC mains power is unplugged or in the event of an AC mains power failure. Anesthesia systems only operate on battery power in a rare event that AC mains power is lost and there is no continuous backup emergency power. If AC power is interrupted, the system will not automatically switch over to the battery supply mode and will reboot. If this issue occurs, a temporary disruption of mechanical ventilation, manual ventilation, and volatile agent delivery may occur. Following the reboot, the system will not return to the previous ventilation settings.
Distribution
Worldwide distribution: US (nationwide) and OUS (foreign) to countries of: Argentina, Australia, Bahamas, Bangladesh, Bosnia and Herzegovina, Brazil, Bulgaria, Cambodia, Canada, China, Costa Rica, Czech Republic, Dominican Republic, Ecuador, Egypt, El Salvador, Finland, France, Germany, Guatemala, H
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0913-2026, Z-0914-2026, Z-0915-2026

Ongoing Class II Medical devices

Recall started
16 May 2025
Reason
GE HealthCare has become aware that the Estimated Fetal Weight (EFW) measurement data feature on the Versana Premier R3 and LOGIQ F R3 series ultrasound systems can display previous obstetric patient data in the Whizz report. This could potentially lead to an incorrect clinical decision due to inaccuracy in the fetal size and growth calculation.
Distribution
Worldwide distribution.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2029-2025, Z-2030-2025

Ongoing Class I Medical devices

Recall started
21 March 2025
Reason
Certain Carestation 620/650/650c and 750/750c Anesthesia Delivery Systems will not provide effective ventilation in Volume Control Ventilation (VCV) mode. In these systems, effective ventilation can be achieved in Pressure Control Ventilation (PCV) or Pressure Control Ventilation Volume Guarantee (PCV-VG) modes or with Manual ventilation.
Distribution
Worldwide distribution. US nationwide including Puerto Rico, Albania, Algeria, Argentina, Armenia, Australia, Austria, Bahrain, Bangladesh, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Cambodia, Canada, Chile, China, Colombia, Costa Rica, Cote d'Ivoire, Croatia, Cyprus, Czech Republi
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1625-2025, Z-1632-2025, Z-1636-2025, Z-1630-2025, Z-1627-2025, Z-1629-2025, Z-1633-2025, Z-1626-2025, Z-1638-2025, Z-1628-2025, Z-1631-2025, Z-1635-2025 and 3 more

Ongoing Class II Medical devices

Recall started
8 September 2023
Reason
There is a potential interruption of data communication between E-modules inserted in the F2-01 Frame and CARESCAPE ONE and CARESCAPE Canvas 1000 patient monitors if the F2-01 Frame has not been powered down within the last 120 days.
Distribution
Worldwide distribution - US Nationwide and the countries of Australia, Austria, Bahrain, Belgium, Bulgaria, Cyprus, Czech Republic, Finland, France, Germany, Ireland, Italy, Kuwait, Netherlands, New Zealand, Oman, Poland, Qatar, Saudi Arabia, Slovakia, Slovenia, Spain, Switzerland, Turkey, United Ar
Initiated
Voluntary: Firm initiated
Recall number
Z-2656-2023

Terminated Class II Medical devices

Recall started
24 July 2018
Terminated
16 December 2020
Reason
On systems equipped with the Auxiliary Common Gas Outlet (ACGO), there is a potential for the auxiliary gas switch to unexpectedly switch from the ACGO position back to the circle position. On systems equipped with the Aux O2 + Air function the switch could unexpectedly switch from the Aux O2 + Air position back to the circle position. These situations could lead to hypoxia.
Distribution
Worldwide
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2954-2018, Z-2955-2018, Z-2956-2018

Ongoing Class II Medical devices

Recall started
11 October 2017
Reason
Possible incomplete seal between the CO2 absorber and the breathing circuit CO2 bypass port assembly of the Carestation 600 Series systems.
Distribution
Distributed domestically to AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, ID, IL, IN, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NH, NJ, NV, NY, OH, OK, PA, TN, TX, UT, VA, VT, WA, WI, WV. Distributed internationally to Algeria, Argentina, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belgium, Bo
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0580-2018, Z-0579-2018

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.