Waukesha, WI, United States

GE Healthcare, LLC recalls

114 recalls recorded since 2018, most recently on 31 July 2026. 50 are listed as ongoing. Status and classification are as published by the FDA.

Recalls114
Ongoing (FDA)50
Terminated or completed64
Class I (FDA)8

Ongoing Class II Medical devices

Recall started
31 July 2026
Reason
Under certain conditions historical imaging data migrated from a legacy system may become incorrectly associated with a different patient. If this occurs, studies or images may be displayed under the wrong patient record, which could lead to misdiagnosis or a delay in patient care.
Distribution
Worldwide - US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-3169-2026, Z-3168-2026, Z-3166-2026, Z-3171-2026, Z-3170-2026, Z-3167-2026

Ongoing Class II Medical devices

Recall started
27 February 2026
Reason
Whole-Body MR scanners used to produce images of the inside of the human body to aid diagnosis of disease may have been manufactured/serviced with ferrous (steel) fittings instead of non-ferrous (brass) fittings on the magnet rear and if system is accessed to perform service while the magnet is ramped, the magnetic field could attract the ferrous fittings potentially resulting in personnel injury.
Distribution
Worldwide - US Nationwide distribution in the states of CA, MA, NJ, NY, OH, SD, TX, VA, PR and the countries of Sweden, Switzerland, United Kingdom, Vietnam, China, Norway, Portugal, Spain, France, Germany, Greece, Korea, Republic of
Initiated
Voluntary: Firm initiated
Recall number
Z-1767-2026

Ongoing Class II Medical devices

Recall started
16 January 2026
Reason
Under certain workflows, patient information shown in the viewer may not match the images displayed on Centricity Universal Viewer Zero Footprint Client (ZFP) with versions ZFP v6.0: SP11.6, SP11.6.0.1 and SP11.6.1 on systems with Patient Timeline installed,
Distribution
Worldwide distribution - US Nationwide and the countries of Belarus, Brazil, Canada, China, France, Georgia, Germany, India, Indonesia, Ireland, Israel, Italy, Japan, Korea (Republic Of), Kuwait, Lebanon, Malaysia, Netherlands, Philippines, Poland, Russia, Saudi Arabia, Singapore, South Africa, Spai
Initiated
Voluntary: Firm initiated
Recall number
Z-1475-2026

Ongoing Class II Medical devices

Recall started
3 November 2025
Reason
GE HealthCare has become aware that the Patient History tab in ViewPoint 6 versions 6.15.3 and 6.15.4 ultrasound software may not include complete patient medical history information.
Distribution
Worldwide
Initiated
Voluntary: Firm initiated
Recall number
Z-0935-2026

Ongoing Class II Medical devices

Recall started
19 February 2025
Reason
There is a potential security vulnerability in Centricity Universal Viewer (UV), Centricity PACS-IW (PACS-IW), Centricity Radiology RA600 (RA600), & Centricity Cardiology CA1000 (CA1000) where the service login credentials are able to be identified which could allow a malicious actor with these credentials to access the system and potentially manipulate patient data.
Distribution
Worldwide distribution.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1487-2025, Z-1485-2025, Z-1484-2025, Z-1486-2025, Z-1483-2025

Ongoing Class II Medical devices

Recall started
3 October 2024
Reason
GE HealthCare has become aware of an issue in Centricity Universal Viewer Zero Footprint Client (ZFP) where there is a potential security vulnerability which could allow a malicious actor to access the system and potentially manipulate patient data.
Distribution
Worldwide distribution - United States Nationwide and the countries of Algeria, Argentina, Brazil, China, Dominican Republic, France, Germany, India, Indonesia, Ireland, Israel, Japan, Kuwait, Malaysia, Mexico, Panama, Qatar, Russia, Saudi Arabia, Singapore, Spain, United Kingdom.
Initiated
Voluntary: Firm initiated
Recall number
Z-0318-2025

Ongoing Class II Medical devices

Recall started
29 August 2024
Reason
GE HealthCare has become aware of an issue in Centricity Universal Viewer Zero Footprint Client (ZFP) version v6.0 SP11.x where the latest report is not shown by default to the user. The issue may occur when a study has multiple DICOM structured reports (i.e. preliminary, final, and/or addendum) stored in different series.
Distribution
Worldwide distribution.
Initiated
Voluntary: Firm initiated
Recall number
Z-0039-2025

Ongoing Class II Medical devices

Recall started
8 August 2024
Reason
The National Institute of Standards and Technology (NIST) has identified in the National Vulnerability Database (NVD) that NexGen Healthcare Mirth Connect, prior to version 4.4.1, has a potential vulnerability allowing the deserialization of untrusted data (CVE-2023-43208). Mirth products may have been installed and configured with your ViewPoint / ViewPoint 6 system by GE HealthCare. This vulnerability could allow a malicious actor to access the system and potentially manipulate patient data.
Distribution
US Nationwide distribution including in the states of Alabama, Alaska, Arizona, Arkansas, California, Colorado, Connecticut, Delaware, District of Columbia, Florida, Georgia, Idaho, Illinois, Indiana, Iowa, Kansas, Kentucky, Louisiana, Maine, Maryland, Massachusetts, Michigan, Minnesota, Mississippi
Initiated
Voluntary: Firm initiated
Recall numbers
Z-3146-2024, Z-3145-2024

Ongoing Class II Medical devices

Recall started
5 August 2024
Reason
GE HealthCare has become aware of an issue where image storage is not possible when Centricity PACS reaches the maximum number of unique image identifiers (Ckeys). If this occurs, it can result in acquisition of a partial study.
Distribution
Worldwide distribution.
Initiated
Voluntary: Firm initiated
Recall number
Z-3214-2024

Ongoing Class II Medical devices

Recall started
23 July 2024
Reason
Footprint Client (ZFP) versions v6.0 SP9.x and SP10.x where the latest addended report is not shown by default to the user. The issue occurs when the addendum is created on the same day as the original report but at a different time, and only when launched on the following browsers. " Internet Explorer 10 (IE10) " Internet Explorer 11 (IE11) " Firefox.
Distribution
Worldwide distribution.
Initiated
Voluntary: Firm initiated
Recall number
Z-3160-2024

Ongoing Class II Medical devices

Recall started
14 September 2023
Reason
GE HealthCare has become aware of an issue where images from two different patients may be contained in a single study, when stored in Centricity PACS-IW V3.7.x, Centricity PACS-IW with Universal Viewer V5.0 and Centricity Universal Viewer V6.0 with PACS-IW foundation.
Distribution
US and Austria, Azerbaijan, Bosnia and Herzegovina, Brazil, Canada, Chile, Colombia, France, Germany, India, Ireland, Italy, Kuwait, Malaysia, Netherlands, Netherlands Antilles, Panama, Poland, Russia, Saudi Arabia, Slovenia, South Africa, Spain, Switzerland, United Kingdom, Uruguay, Venezuela, Viet
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0139-2024, Z-0140-2024, Z-0138-2024

Ongoing Class II Medical devices

Recall started
8 September 2023
Reason
When using Universal Viewer or Centricity Universal Viewer with Workflow Manager (WFM), if a third-party reporting application is used to launch exams, it could display information on a different patient than WFM and the Viewer.
Distribution
Nationwide distribution to AR, AZ, CA, CO, DE, FL, GA, IL, IN, ME, MI, MN, MO, NE, NJ, NM, NV, NY, OH, OK, PA, SC, SD, TN, TX. International distribution to Austria, Canada, France, Germany, Hong Kong, Israel, Italy, Philippines, Singapore, Thailand, Turkey, Vietnam.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0143-2024, Z-0142-2024

Ongoing Class II Medical devices

Recall started
28 March 2023
Reason
GE HealthCare has become aware of an issue where information from two different patients can be mismatched when correcting patient or study information.
Distribution
Worldwide Distribution
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1500-2023, Z-1501-2023, Z-1502-2023

Ongoing Class II Medical devices

Recall started
15 July 2022
Reason
Inaccurate Distance and Area measurements with use of Global Stack viewport.
Distribution
Distribution was made to AR, AZ, CA, CO, DE, FL, GA, IA, ID, IL, IN, KS, LA, MA, MD, ME, MI, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, VT, WA, WI, and WV. There was government distribution and no military distribution. Foreign distribution was made to Argentina, A
Initiated
Voluntary: Firm initiated
Recall number
Z-1566-2022

Ongoing Class II Medical devices

Recall started
24 June 2022
Reason
Under rare conditions, SIGNA Premier gradient coils could result in elevated acoustic noise during scanning. After prolonged periods of elevated acoustic noise, this could potentially lead to hearing loss.
Distribution
Worldwide distribution - US Nationwide distribution in the states of AL, AZ, CA, MI, MN, NY, and WI. There was government distribution but no military distribution. The countries of China, France, Hong Kong, India, Italy, Japan, Korea, Kuwait, Norway, Oman, Saudi Arabia, Slovakia, Spain, Sweden, Sw
Initiated
Voluntary: Firm initiated
Recall number
Z-1430-2022

Terminated Class II Medical devices

Recall started
14 June 2022
Terminated
16 January 2024
Reason
The Event Notification Manager (ENM) functionality for certain products does not process notifications for study modifications performed on post verified exams.
Distribution
Worldwide distribution - US Nationwide distribution in the states of CO, NV, OH, PA, WA and the countries of Brazil, Canada, China, France, Germany, Israel, Italy, Japan, Oman, Spain, Switzerland, and United Kingdom.
Initiated
Voluntary: Firm initiated
Recall number
Z-1466-2022

Ongoing Class II Medical devices

Recall started
8 June 2022
Reason
Potential to display inaccurate measurements on images in Centricity Universal Viewer Zero Footprint Client (ZFP)
Distribution
Worldwide distribution.
Initiated
Voluntary: Firm initiated
Recall number
Z-1333-2022

Ongoing Class II Medical devices

Recall started
2 June 2022
Reason
As a result of unclear guidance in the service manual, the table can drop during a specific service activity if the wrong screws are removed by the service personnel without the appropriate table supports in place resulting in injury to service personnel.
Distribution
US Nationwide distribution including PR. There was also government/military distribution. The countries of Afghanistan, Albania, Algeria, Angola, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Plurinational State of Bolivia, Bosnia and Herzegovina
Initiated
Voluntary: Firm initiated
Recall number
Z-1511-2022

Terminated Class I Medical devices

Recall started
1 June 2022
Terminated
21 October 2024
Reason
There is a potential reversal of the O2 and air cylinder pressure transducer connections.
Distribution
Worldwide Distribution. US nationwide, Australia, Austria, Belgium, Bolivia, Brazil, Canada, China, Colombia, Costa Rica, Croatia, Czech Republic, Denmark, Dominican Republic, Egypt, Finland, France, Germany, Greece, Hong Kong, Hungary, Iceland, India, Indonesia, Ireland, Italy, Japan, Kenya, Korea
Initiated
Voluntary: Firm initiated
Recall number
Z-1389-2022

Ongoing Class II Medical devices

Recall started
27 May 2022
Reason
The base of the anesthesia system can have a crack and could potentially fracture resulting in a tip or overbalance of the anesthesia system. This can result in potential injury if it falls on a person.
Distribution
Distribution was made to AZ, CA, CO, CT, FL, IA, ID, IL, IN, KY, LA, MA, MD, ME, MI, MO, MT, NC, NE, NV, NY, OH, PA, SC, TN, TX, VA, and WV. There was also military distribution but no government distribution. Foreign distribution was made to Australia, Canada, China, Estonia, Finland, France, G
Initiated
Voluntary: Firm initiated
Recall number
Z-1427-2022

Ongoing Class II Medical devices

Recall started
10 May 2022
Reason
The CARESAPE Central Station (CSCS) V2 can shut down due to a potential power supply component failure. This can lead to loss of patient monitoring at the central station.
Distribution
Worldwide distribution - US Nationwide. There was also military/government distribution and the countries of Algeria, Argentina, Australia, Austria, Bahamas, Bahrain, Bangladesh, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Burkina Faso, Cameroon, Canada, Chile, China, Colombia, Costa
Initiated
Voluntary: Firm initiated
Recall number
Z-1265-2022

Ongoing Class II Medical devices

Recall started
28 April 2022
Reason
Potential for Distance and Area measurements to display inaccurate measurement values when performed on magnified images, and specific to ZFP, Distance and Area measurements can display inaccurate measurement values when performed on lossy images that are scaled down from their original resolution.
Distribution
Worldwide distribution. Nationwide distribution to AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY. International distribution to Argentina, Aust
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1231-2022, Z-1229-2022, Z-1230-2022, Z-1232-2022, Z-1233-2022

Ongoing Class II Medical devices

Recall started
25 April 2022
Reason
Backup batteries can fail earlier than their estimated life.
Distribution
Worldwide distribution - US Nationwide and the countries of Algeria, Australia, Austria, Bangladesh, Belgium, Brazil, China, Columbia, Costa Rica, Cyprus, Czech Republic, Denmark, Egypt, France, Germany, Greece, Hong Kong, India, Indonesia, Iraq, Italy, Japan, Kazakhstan, Kenya, Korea, Kuwait, Lao P
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1285-2022, Z-1286-2022

Ongoing Class I Medical devices

Recall started
18 April 2022
Reason
Insufficient battery backup power resulting in premature shutdown of the ventilator when not connected to AC mains power supply.
Distribution
Worldwide Distribution: US (nationwide) to AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, NE, NH, NJ, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, WI and OUS (internationally) to Australia, Austria, Bangladesh, Belgium, Brazil, Bulgaria, Canada, Chile, Chi
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1227-2022, Z-1226-2022

Terminated Class II Medical devices

Recall started
16 March 2022
Terminated
4 May 2026
Reason
Two scenarios may cause edits to measurements and diagnosis statements can be lost after a test is Signed in the MUSE NX web client.
Distribution
US and Australia, Austria, Bahrain, Belgium, Brazil, Canada, Denmark, England, France, Germany, Ireland, Italy, Korea, Malaysia, Netherlands, Norway, Poland, Qatar, Saudi Arabia, Singapore, Spain, Sweden, Taiwan, Turkey
Initiated
Voluntary: Firm initiated
Recall number
Z-0919-2022

Ongoing Class II Medical devices

Recall started
4 March 2022
Reason
During the de-installation of an MR system magnet, if all rigging and hardware that attach to the magnet for transportation are not properly installed and secured, the magnet may fall, leading to potential injury.
Distribution
Worldwide Distribution: US (nationwide), including Puerto Rico and the U.S. Virgin Islands; and OUS (Foreign) was made to Albania, Algeria, Argentina, Armenia, ASEAN HQ, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Bhutan, Bolivia, Bosnia and Herzegovina, Botswana, Brazil,
Initiated
Voluntary: Firm initiated
Recall number
Z-0983-2022

Ongoing Class II Medical devices

Recall started
18 February 2022
Reason
There is potential for the images to be flipped left to right.
Distribution
Worldwide Distribution. US nationwide, Argentina, Australia, Azerbaijan, Bosnia and Herzegovina, Brazil, Canada, China, Denmark, Ecuador, Ethiopia, France, Germany, Greece, Hong Kong, India, Iran, Ireland, Italy, Japan, Korea, Kuwait, Lebanon, Mexico, Netherlands, New Zealand, Pakistan, Poland, Po
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0857-2022, Z-0859-2022, Z-0858-2022, Z-0860-2022

Ongoing Class II Medical devices

Recall started
28 January 2022
Reason
If the CARESCAPE Central Station v2.0 is used with an unapproved keyboard, the audio can be muted resulting in loss of audible alarms.
Distribution
Worldwide distribution - US nationwide, including Puerto Rico. There was also government/military distribution. The countries of Algeria, Argentina, Australia, Austria, Bahrain, Bangladesh, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Canada, Chile, China, Colombia, Costa Rica, Cote d'lvoire,
Initiated
Voluntary: Firm initiated
Recall number
Z-0808-2022

Ongoing Class II Medical devices

Recall started
23 December 2021
Reason
The magnetic resonance systems could potentially have a cryogen ventilation system that does not meet the venting requirements.
Distribution
Distribution was nationwide, including Puerto Rico and the Virgin Islands. There was also government and military distribution. Foreign distribution was made to Albania, Algeria, Argentina, Armenia, Australia, Austria, Azerbaidjan, Bahrain, Bangladesh, Belarus, Belgium, Bhutan, Bolivia, Bosnia and
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0600-2022, Z-0604-2022, Z-0612-2022, Z-0599-2022, Z-0591-2022, Z-0608-2022, Z-0592-2022, Z-0609-2022, Z-0615-2022, Z-0613-2022, Z-0619-2022, Z-0605-2022 and 20 more

Terminated Class II Medical devices

Recall started
4 November 2021
Terminated
24 July 2026
Reason
The firm identified that the Vscan Extend device will not power on because of depleted battery, due to the device not being turned off prior to battery removal or when being connected to a computer without AC power.
Distribution
Domestic: AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT. NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY, PR, and DC. International: Albania, Australia, Austria, Bahrain, Bangladesh, Belgium, Bosnia, Braz
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0522-2022, Z-0520-2022, Z-0521-2022, Z-0523-2022, Z-0524-2022, Z-0518-2022, Z-0519-2022

Terminated Class II Medical devices

Recall started
15 October 2021
Terminated
17 August 2026
Reason
There is a potential to display incomplete patient imaging study.
Distribution
Worldwide distribution - US Nationwide distribution in the states of AR, AZ, CA, CO, DE, FL, GA, IL, IN, LA, MA, MD, ME, MI, MN, MO, MS, NC, NE, NJ, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VT, WA, and WV. The countries of Argentina, Aruba, Australia, Belgium, Bolivia, Brazil, Canada, Chile,
Initiated
Voluntary: Firm initiated
Recall number
Z-0329-2022

Terminated Class II Medical devices

Recall started
26 August 2021
Terminated
4 May 2026
Reason
Under certain conditions, some slices may be missing which can lead to a gap of anatomy in the 3D Volume images.
Distribution
Distribution in United States to AZ, CA, FL, GA, IL, LA, MA, MD, MN, NC, NH, NJ, NY, OH, OK, TX, VA, WA, and Puerto Rico OUS distribution to: Australia, Belgium, Canada, Chile, Colombia, France, Germany, Italy, Japan, Korea, Poland, Saudi Arabia, Spain, Sweden, Turkey, UAE, and UK
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0134-2022, Z-0140-2022, Z-0141-2022, Z-0133-2022, Z-0137-2022, Z-0135-2022, Z-0138-2022, Z-0139-2022, Z-0136-2022, Z-0132-2022

Terminated Class II Medical devices

Recall started
12 August 2021
Terminated
22 August 2024
Reason
A software anomaly exists in the Centricity Universal Viewer study management feature in which study changes are not propagated to either the Centricity Enterprise Archive (EA) or another Vendor Neutral Archive (VNA).
Distribution
AR, IL, ME, MS, NE, NJ, NY, and TN Belgium, Canada, Costa Rica, France, Germany, India, Japan, Korea, Kuwait, Saudi Arabia, Singapore, Switzerland, Switzerland, Turkey and United Kingdom
Initiated
Voluntary: Firm initiated
Recall number
Z-0012-2022

Ongoing Class II Medical devices

Recall started
6 August 2021
Reason
Image acquisition failures and synchronization failure with the Centricity Enterprise Archive
Distribution
AK, AZ, CA, CO, DE, FL, GA, IA, IN, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NJ, NM, NV, NY, OH, OK, OR, PA, SD, TX, UT, WA, WI, WV and WY. Austria, Brazil, Canada, France, Germany, Ghana, India, Ireland, Israel, Italy, Kuwait, Netherlands, Panama, Poland, Russia, Saudi Arabia, Slovenia, South Afric
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2503-2021, Z-2502-2021

Ongoing Class II Medical devices

Recall started
2 August 2021
Reason
The flow sensors built prior to June 2021 could have damaged tubes with small punctures or cuts that could cause leaks resulting in incorrect anesthesia machine tidal volumes potentially leading to over-delivery of volume to the patient.
Distribution
Distribution was nationwide, including PR. Foreign distribution was made to Albania, Algeria, Argentina, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belgium, Bermuda, Bolivia, Bosnia and Herzegovina, Brazil, Brunei, Bulgaria, Cambodia, Canada, Chile, China, Colombia, Costa Rica, Cote D'lvo
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2596-2021, Z-2595-2021

Ongoing Class II Medical devices

Recall started
30 June 2021
Reason
When adding one or more intended scan groups names must contain only letters, numbers (0-9), or a space or the user interface will not update to confirm the intended group(s) has been added to the acquisition.
Distribution
US
Initiated
FDA Mandated
Recall numbers
Z-2023-2021, Z-2022-2021, Z-2021-2021

Ongoing Class II Medical devices

Recall started
30 June 2021
Reason
The accumulated dose is incorrectly displayed in certain situations. Thisoccurs during an interventional exam using the SmartStep option if the table height is adjusted after the exam starts. In this situation, the SmartStep display will show a value that is higher (up to 100 times) than the actual dose delivered.
Distribution
Worldwide and US Nationwide Distribution
Initiated
FDA Mandated
Recall number
Z-0242-2022

Ongoing Class II Medical devices

Recall started
30 June 2021
Reason
The accumulated dose is incorrectly displayed in certain situations. There are two situations that will result in the SmartStep displayed dose being inaccurately high.
Distribution
Worldwide - US Nationwide distribution.
Initiated
FDA Mandated
Recall numbers
Z-0815-2022, Z-0816-2022

Terminated Class II Medical devices

Recall started
24 June 2021
Terminated
14 June 2023
Reason
An unexpected column motion while attempting to park the column on the AMX Navigate mobile X-ray system.
Distribution
Domestic Distribution Only
Initiated
Voluntary: Firm initiated
Recall number
Z-2590-2021

Ongoing Class II Medical devices

Recall started
14 June 2021
Reason
Improperly loaded software options may result in additional X -ray radiation exposure to the patient.
Distribution
3,665 (US = 956; OUS = 2,709)
Initiated
FDA Mandated
Recall number
Z-1900-2021

Ongoing Class II Medical devices

Recall started
4 June 2021
Reason
GE Healthcare has recently become aware of an issue on the affected products listed below where the "Patient Orientation" button may inadvertently be clicked when intending to click on the "Save RX" button. This changes the prescribed patient orientation on the system prior to running the initial 3-Plane Localizer Scan (See Figure 1 for reference to buttons.) Selecting and saving a patient orientation that does not match the patient's actual position may result in incorrectly annotated and/or flipped images.
Distribution
Worldwide distributions.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2052-2021, Z-2061-2021, Z-2057-2021, Z-2055-2021, Z-2065-2021, Z-2064-2021, Z-2060-2021, Z-2059-2021, Z-2063-2021, Z-2056-2021, Z-2058-2021, Z-2053-2021 and 2 more

Ongoing Class II Medical devices

Recall started
21 May 2021
Reason
Masimo SpO2 Saturation Values can become frozen after an extended length of use without a power down.
Distribution
worldwide distribution US nationwide, Albania, Algeria, Argentina, Australia, Austria, Bahamas, Bahrain, Belgium, Bermuda, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Fij
Initiated
Voluntary: Firm initiated
Recall number
Z-2139-2021

Terminated Class II Medical devices

Recall started
19 April 2021
Terminated
31 August 2023
Reason
The IGS system can experience a single vertical line defect where the vertical line divides and horizontally shifts live monitor images into two unequal image parts within the monitor display screen.
Distribution
Worldwide distribution - US Nationwide distribution in the states of CA, CO, FL, GA, KY, LA, MA, MO, MS, NC, NJ, NV, NY, PA, SC, TN, TX, VA, WA, and WI. Government distribution was made and potentially military distribution was made. The countries of Algeria, Argentina, Canada, China, Dominican
Initiated
Voluntary: Firm initiated
Recall number
Z-1864-2021

Terminated Class II Medical devices

Recall started
19 April 2021
Terminated
14 January 2026
Reason
If a certain sequence of events occur, the pdf test report that belongs to one patient will appear in a different patient's record when viewed in an Electronic Medical Record (EMR) or a similar system.
Distribution
Worldwide distribution - US Nationwide. Government/military distribution was also made, and the countries of Algeria, Argentina, Australia, Austria, Azerbaijan, Bahamas, Bahrain, Bangladesh, Belgium, Bermuda, Brazil, Brunei Darussalam, Cambodia, Canada, Cayman Islands, Chile, China, Colombia, Costa
Initiated
Voluntary: Firm initiated
Recall number
Z-1927-2021

Terminated Class II Medical devices

Recall started
19 March 2021
Terminated
11 August 2023
Reason
When the Cross Enterprise Display option is enabled on the Centricity Universal Viewer and Workflow Manager option is used as the PACS worklist launching the viewer, the patient jacket may potentially include a different patient s comparison exam(s).
Distribution
Distributed nationwide to AR, CA, CO, DE, FL, IL, IN, LA, ME, NE, NJ, NV, NY, PA, SC, TX and internationally to Canada.
Initiated
Voluntary: Firm initiated
Recall number
Z-1533-2021

Completed Class II Medical devices

Recall started
19 March 2021
Reason
GE Healthcare has become aware of a potential issue on the Revolution CT system where a collimator blade does not find its appropriate home position. This issue could result in the delivered dose of x-ray radiation being either larger or smaller than its prescribed value.
Distribution
Worldwide distribution - US Nationwide
Initiated
FDA Mandated
Recall number
Z-2004-2021

Ongoing Class II Medical devices

Recall started
7 February 2021
Reason
GE Healthcare has become aware of a potential issue on 800 family NM/CT Scanners that could cause a shift in CT radiation exposure range of up to 5cm compared to the intended radiation exposure range of the planned scan under specific workflows. The issue occurs only on a hybrid whole-body continuous F3 protocol with Zoom < 1 where the scan range is set on the Smart Console. In some cases, this may also necessitate a re-scan of the patient, which would expose the patient to additional X-ray radiation.
Distribution
Worldwide distribution - US Nationwide
Initiated
FDA Mandated
Recall numbers
Z-1546-2021, Z-1547-2021, Z-1545-2021

Terminated Class II Medical devices

Recall started
17 December 2020
Terminated
9 January 2023
Reason
GE Healthcare has become aware of a potential issue on Revolution Apex, Revolution CT with Apex Edition, Revolution CT, and Revolution CT ES systems where the table pinch protector could be damaged or missing leading to exposed table screws which could cause a laceration due to a sharp edge.
Distribution
Worldwide - US Nationwide Distribution in the states of AL, AR, AZ, CA, CO, CT, FL, GA, HI, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MO, MS, NC, NE, NJ, NV, NY, NH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, WV; In the countries of Algeria, Australia, Belarus, Belgium, Bosnia and Herzegovina, Braz
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1130-2021, Z-1129-2021

Terminated Class II Medical devices

Recall started
10 December 2020
Terminated
21 August 2026
Reason
GE Healthcare has become aware that rotor bearing screws were found loose on one detector in the field, leading to the release of one of the four rotor bearings in the detector.
Distribution
Worldwide
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0855-2021, Z-0856-2021, Z-0857-2021, Z-0853-2021, Z-0854-2021, Z-0858-2021

Terminated Class II Medical devices

Recall started
10 December 2020
Terminated
18 November 2022
Reason
GE Healthcare has become aware of a potential safety issue associated with the improper installation of the Precision 600FP monitor suspension. A partial detachment of the monitor suspension could result in potential injury to a patient or operator.
Distribution
US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, FL, GA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MO, MS, MT, NC, NJ, NM, NV, NY, OH, PA, SC, TN, TX, UT, VA, WA, WI, WV
Initiated
Voluntary: Firm initiated
Recall number
Z-1101-2021

Terminated Class II Medical devices

Recall started
23 November 2020
Terminated
29 August 2023
Reason
A potential for protocol(s) with unintended patient orientation or scan setting when using Intelligent Protocoling
Distribution
Worldwide distribution - US Nationwide distribution including in the states of FL and TX. The countries of Belgium, France, and Spain.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0694-2021, Z-0695-2021

Terminated Class II Medical devices

Recall started
27 October 2020
Terminated
3 May 2024
Reason
Incorrect patient identification and/or patient demographic errors.
Distribution
Domestic Distribution: AL,AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI,MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY. International Distribution: AUSTRALIA, AUSTRIA, BELGIUM, Canada, CZECH REP
Initiated
Voluntary: Firm initiated
Recall number
Z-0543-2021

Terminated Class II Medical devices

Recall started
21 September 2020
Terminated
28 December 2022
Reason
There is a potential for a smudge artifact that could be suspect for pathology in some images due to incorrect settings.
Distribution
Worldwide distribution - US Nationwide distribution including in the states of FL, GA, IL, MI, NC, NE, NJ, PA, TX, VA, WA, and WI. The countries of Australia, Austria, Belgium, Canada, China, Denmark, Estonia, Finland, France, Japan, Kuwait, Norway, Saudi Arabia, Spain, Sweden, Switzerland, Thailand
Initiated
Voluntary: Firm initiated
Recall number
Z-0306-2021

Ongoing Class II Medical devices

Recall started
17 September 2020
Reason
Flow sensors could have damaged tubes in the form of small punctures or cuts.
Distribution
U.S. states where product was distributed - AL, AR, AZ, CA, CO, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MO, MS, MT, NC, NE, NJ, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, WI Foreign countries impacted - Australia, Canada, Czech Republic, Estonia, France, Germany, India, Indonesia, Is
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0315-2021, Z-0311-2021, Z-0314-2021, Z-0312-2021, Z-0317-2021, Z-0316-2021, Z-0313-2021

Terminated Class II Medical devices

Recall started
8 September 2020
Terminated
28 April 2021
Reason
Potential electrical arc flash within the Power Gradient & RF (PGR) Cabinet.
Distribution
Worldwide distribution - US Nationwide distribution including in the states of FL, KY, NC, NJ, NM, NY, TX, UT and WA. The countries of Canada, Hungary, India, Korea, China, Russia, Taiwan, Puerto Rico and Ubekistan.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0130-2021, Z-0127-2021, Z-0129-2021, Z-0128-2021, Z-0131-2021

Terminated Class II Medical devices

Recall started
14 August 2020
Terminated
6 May 2021
Reason
Rotor bearing screws were found loose on detector.
Distribution
USA : AR CT FL GA IL KS KY LA MA MD MI MN MO NE NJ NV NY OH OK RI SC TN TX VA VT WA WI OUS: ARGENTINA AUSTRIA BELGIUM BRAZIL Canada CHINA FRANCE GERMANY GREECE INDIA ISRAEL ITALY JAPAN Korea (Republic of) NETHERLANDS NORWAY PANAMA POLAND PORTUGA
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2991-2020, Z-2994-2020, Z-2997-2020, Z-2998-2020, Z-2993-2020, Z-2995-2020, Z-2996-2020, Z-2992-2020

Terminated Class II Medical devices

Recall started
14 July 2020
Terminated
23 September 2024
Reason
The oxygen sensor for certain CARESCAPE R860 devices has a potential issue that may result in an inaccurate display of FiO2 value from what is being delivered.
Distribution
Domestic Distribution: USA International Distribution: Australia, Bahrain, Bermuda, Canada, Egypt, France, Germany, Honduras, India, Iraq, Italy, Lebanon, Mexico, Panama, Peru, Russia, Saudi Arabia, Singapore, South Africa, Spain, Taiwan and United Arab Emirates.
Initiated
Voluntary: Firm initiated
Recall number
Z-2883-2020

Terminated Class II Medical devices

Recall started
22 June 2020
Terminated
19 August 2024
Reason
CARESCAPE ONE may not provide visual and audible alarms for Ventricular Fibrillation (V Fib), if V Fib occurs at the time CARESCAPE ONE is docked to a CARESCAPE B450/B650/B850 host monitor.
Distribution
Worldwide distribution - US Nationwide distribution including in the states of MT, NC, TX and WI. The countries of Australia, Austria, Bahrain,Belgium, Bolivia, Brazil ,Canada, Chile, Check Republic, Denmark, Ecuador, Egypt, Estonia, Finland, France, French Guiana, Germany, Greece, Hong Kong, Hungar
Initiated
Voluntary: Firm initiated
Recall number
Z-2750-2020

Ongoing Class II Medical devices

Recall started
18 May 2020
Reason
The CARESCAPE Patient Data Module (PDM) is used with the following monitors: CARESCAPE B450/B650/B850, SOLAR 8000M/I and Transport Pro. If Pace detection is turned ON, on the monitor, and if an Automated External Defibrillator (AED) is used to perform defibrillation, the low frequency signals deployed by the AED, cause the PDM to issue a false flat line ECG waveform and alarm Asystole to be displayed on the monitor.
Distribution
Worldwide Distribution
Initiated
Voluntary: Firm initiated
Recall number
Z-2424-2020

Ongoing Class II Medical devices

Recall started
8 May 2020
Reason
GEHC Surgery became aware that some users of the 9800 and 9900 fluoroscopic mobile C-Arms (referred to as the system here on) within the United States had requested that a Field Service Engineer (FSE) change the dose value on the live update screen from AKR / CAK to DAP. All systems sold to customers in the US that leave GEHC Surgery (the place of manufacture) are configured to display AKR / CAK when they leave the factory. An FSE changing the dose display, per the customer request, to display DAP instead does not comply to 21CFR 1020.32(k), which requires fluoroscopic equipment manufactured on or after June 10, 2006 to display continuous updates of AKR and CAK values.
Distribution
Nationwide
Initiated
FDA Mandated
Recall number
Z-2872-2020

Terminated Class II Medical devices

Recall started
28 April 2020
Terminated
23 September 2024
Reason
Potential for certain Giraffe Incubators and Giraffe OmniBeds to deliver a different environmental oxygen level than what is displayed by the device, if a certain optional component referred to as the Servo Oxygen Module is installed.
Distribution
Worldwide Distribution. US nationwide including US Virgin Islands, Australia, Austria, Bahrain, Belgium, Bermuda, Bulgaria, Burundi, Canada, Chile, China, Croatia, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Estonia, Finland, Fra
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2236-2020, Z-2235-2020

Terminated Class II Medical devices

Recall started
2 April 2020
Terminated
6 May 2024
Reason
During Giraffe Shuttle charging, the batteries emit low concentrations of gases (outgassing) producing a strong sulfur odor.
Distribution
US Nationwide distribution in the states of AZ, CA, CO, DE, FL, GA, ID, IL, IN, LA, MD, MI, MS, NC, ND, NM, NV, NY, OH, Ok, OR, PA, SC, TN, TX, VA, WA, WI and WV.
Initiated
Voluntary: Firm initiated
Recall number
Z-2111-2020

Terminated Class II Medical devices

Recall started
23 December 2019
Terminated
11 December 2020
Reason
It was identified that due to a potential installation workflow issue, the MR system date could be set incorrectly. The system's date and time setting are used to populate the DICOM header information on images. This could result is an inaccurate date recorded on the images.
Distribution
Worldwide distribution - US nationwide distribution and countries of Australia, China, France, Germany, India, Italy, Japan, Netherlands, Oman, Spain, Sweden, Switzerland, Turkey, and United Kingdom.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0875-2020, Z-0876-2020, Z-0874-2020, Z-0877-2020

Terminated Class II Medical devices

Recall started
16 December 2019
Terminated
5 October 2023
Reason
Centricity Universal Viewer measurements saved into a DICOM Grayscale Presentation State are incorrect in subsequent views for exams containing series with different pixel sizes and may lead to a potential misdiagnosis.
Distribution
Distributed nationwide; Internationally distributed to Mexico and 24 additional countries.
Initiated
Voluntary: Firm initiated
Recall number
Z-0819-2020

Ongoing Class II Medical devices

Recall started
3 December 2019
Reason
An issue with the Lunar iDXA control panel may cause the start button to not release and potentially contribute to an accidental radiation occurrence.
Distribution
US Nationwide Distribution
Initiated
FDA Mandated
Recall numbers
Z-1709-2020, Z-1710-2020, Z-1712-2020, Z-1711-2020

Terminated Class I Medical devices

Recall started
25 November 2019
Terminated
30 November 2023
Reason
GE Healthcare has become aware that there is a potential for a loose cable connection inside specific manufactured anesthesia devices. This would cause a loss of mechanical ventilation and the system will provide high priority audio and visual alarms. Loss of mechanical ventilation could lead to hypoxia if the clinician does not intervene.
Distribution
Worldwide Distribution: US (Nationwide) and countries of: Albania, Algeria, Argentina, Australia, Bahrain, Bangladesh, Belgium, Bolivia, Brazil, Canada, Chile, China, Colombia, Croatia and Hrvatska, Czech Republic, Dominican Republic, Ecuador, Egypt, Estonia, Finland, France, Germany, Greece, Guatem
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0814-2020, Z-0813-2020

Terminated Class II Medical devices

Recall started
12 November 2019
Terminated
2 March 2026
Reason
When connected to the Mission Critical (MC) and /or Information Exchange (IX) networks, certain versions of the CARESCAPE Telemetry Server, ApexPro Telemetry Server, CARESCAPE Central Station (CSCS) version 1 and Central Information Center (CIC) systems were identified to have vulnerabilities to a cyber-attack.
Distribution
Worldwide distribution.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0725-2020, Z-0726-2020, Z-0727-2020

Terminated Class I Medical devices

Recall started
11 October 2019
Terminated
15 August 2024
Reason
Giraffe Incubator, Giraffe OmniBed, Giraffe Incubator Carestation, and Giraffe OmniBed Carestation Bedside panels and portholes can appear closed without being latched. Potential patient fall.
Distribution
worldwide
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0540-2020, Z-0538-2020, Z-0541-2020, Z-0539-2020

Terminated Class I Medical devices

Recall started
27 September 2019
Terminated
26 September 2023
Reason
Displayed End Tidal Oxygen (EtO2) and Fraction of Inspired Oxygen (FiO2) values may be incorrect which could cause up to 50% measurement error in the EtO2/FiO2 values.
Distribution
US OUS: Australia, Austria, Belgium, Bolivia, Brazil, Canada, Chile, China, Colombia, Costa Rica, Czech Republic, Finland, France, Germany, Hong Kong, India, Indonesia, Ireland, Italy, Japan, Korea (Republic of), Malaysia, Mexico, Myanmar, Netherlands, New Zealand, Peru, Poland, Republic of Serbia,
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0757-2020, Z-0760-2020, Z-0756-2020, Z-0759-2020, Z-0758-2020, Z-0754-2020, Z-0755-2020

Terminated Class II Medical devices

Recall started
16 September 2019
Terminated
16 December 2020
Reason
Potential for failure of the patient leakage current test. There is a potential that if another device with electrical connectivity (e.g., ablation device, ECG monitor) fails, then the CLab II Plus Amplifier will not prevent an electric current from completing a circuit, which could cause an electrical shock to a patient.
Distribution
Nationwide distribution to CA, CT, FL, IN, MD, MI , MN, MO, NC, NE, NJ, OH, PA, TX, WI. International distribution to Canada, France, India, Japan, Korea (Republic Of), Mexico, Spain, Taiwan.
Initiated
Voluntary: Firm initiated
Recall number
Z-0257-2020

Terminated Class II Medical devices

Recall started
3 September 2019
Terminated
16 February 2021
Reason
Certain Achilles Express and Achilles Insight systems were shipped with European CEE-7/7 Type Power Cords to countries where three protruding pin plugs are required for protective earth connection.
Distribution
Nationwide distribution to AZ, GA, MT, NC, NJ, NY, OH, WI. International distribution to Australia, Bahrain, Bangladesh, Brazil, Cambodia, Chile, China, Colombia, Cyprus, Hong Kong, India, Iraq, Ireland, Italy, Kuwait, Lebanon, Malaysia, Mauritius, Mexico, MEXICO, Myanmar, Nepal, Oman, Panama, Pe
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0113-2020, Z-0112-2020

Ongoing Class II Medical devices

Recall started
30 August 2019
Reason
May not provide visual and/or audible alarms at the CARESCAPE Central Station or Clinical Information Center monitor for ECG arrhythmias, ECG LEADS FAIL or Pulse Oximetry (SpO2) under certain conditions.
Distribution
Worldwide distribution - US Nationwide including the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, GU, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY. The countries of ALGERIA
Initiated
Voluntary: Firm initiated
Recall number
Z-0175-2020

Ongoing Class II Medical devices

Recall started
9 August 2019
Reason
Intermittently not receiving an audible exposure indication upon completion of an exposure on the Proteus stationary radiographic system.
Distribution
Worldwide - US Nationwide
Initiated
FDA Mandated
Recall number
Z-2449-2019

Ongoing Class II Medical devices

Recall started
26 July 2019
Reason
Fasteners potentially torqued to a value less than the specified value and can potentially cause the canopy to become detached and fall.
Distribution
Worldwide distributions - US Nationwide in the states of CA, CT, FL, ID, IL, KS, MI, NC, NC, NJ, NM, PA, TX, VA. Countries of China, France, Germany, Korea, Republic of China.
Initiated
Voluntary: Firm initiated
Recall number
Z-0281-2020

Terminated Class II Medical devices

Recall started
19 July 2019
Terminated
13 April 2026
Reason
Certain Aespire and Aestiva Anesthesia Systems were noted to have a vulnerability to a cyber-attack when connected to the hospital network. An insufficiently secured terminal server may provide an opportunity for a malicious actor that has already penetrated the hospital network to send fraudulent flow sensor correction parameters.
Distribution
Worldwide
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0115-2020, Z-0116-2020, Z-0117-2020, Z-0114-2020

Terminated Class II Medical devices

Recall started
2 July 2019
Terminated
5 October 2020
Reason
Potential for a transducer mis-alignment in certain transvaginal probes.
Distribution
Distribution in the US to Texas. International distribution to France, Greece and Switzerland.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1964-2019, Z-1965-2019

Terminated Class II Medical devices

Recall started
28 June 2019
Terminated
13 April 2021
Reason
Two issues: 1) The MAC VU360 system may intermittently display an incorrect patient ID or visit number on the screen after scanning the patient s barcode. This may result in the ECG report being assigned to the incorrect patient. 2) The MAC VU360 may have the incorrect patient demographics appear on the patient banner. This may result in the ECG report being assigned to the incorrect patient.
Distribution
Worldwide
Initiated
Voluntary: Firm initiated
Recall number
Z-2060-2019

Terminated Class II Medical devices

Recall started
22 May 2019
Terminated
20 September 2024
Reason
There is a potential that one or more image series (i.e., all images within an image set) may be missing from an exam without a warning displayed in the viewer.
Distribution
Worldwide
Initiated
Voluntary: Firm initiated
Recall number
Z-1781-2019

Ongoing Class II Medical devices

Recall started
21 May 2019
Reason
Overestimation bias in automatically calculated ejection fraction (EF) values while using LVivo EF app on the Vscan Extend product.
Distribution
Worldwide distribution - US nationwide distributed in the states of CA, FL, IL, MA , MD, MI, MN, NY, SC, TX, UT. Countries of Australia, China, Croatia, Czech Republic, Denmark, France, Germany, Greece, Hong Kong, Ireland, Italy, Japan, Korea (Republic Of), Netherlands, Romania, Sweden, Switzerland
Initiated
Voluntary: Firm initiated
Recall number
Z-1839-2019

Terminated Class II Medical devices

Recall started
17 May 2019
Terminated
20 November 2023
Reason
There is a potential that one or more images or image series may be missing from exams without a warning displayed in the viewer.
Distribution
Worldwide Distribution - US Nationwide Austria, Belgium, Brazil Canada Colombia France Germany Ghana India Ireland Israel Italy Kuwait Mexico Netherlands Panama Poland Qatar Russia Saudi Arabia Slovenia South Africa Spain Switzerland Turkey United Arab Emirates United Kingdo
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1885-2019, Z-1886-2019

Terminated Class II Medical devices

Recall started
16 May 2019
Terminated
14 April 2023
Reason
There is the possibility of viewing studies directly from the Enterprise Archive or VNA with incorrect patient images because the updated series or study is not present in the archive.
Distribution
Worldwide Distribution: US (nationwide) to states of: FL, GA, IL, IN, LA, MA, MO, NJ, NY, OR, PA, SC, SD, TX, and WA; and countries of : Brazil, Canada, Chile, France, French Polynesia, Germany, India, Italy, Japan, Kuwait, Lebanon, Singapore, Spain and United Kingdom.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1910-2019, Z-1911-2019

Terminated Class II Medical devices

Recall started
16 May 2019
Terminated
16 December 2020
Reason
Potential slippage of the biopsy positioner when the gantry is rotated during a biopsy procedure.
Distribution
Worldwide distribution - US nationwide distribution in the states of AZ, CA, FL, GA, IL, MI, MN, NC, NJ, NY, PA, TX, and VT, and countries of Algeria, Australia, Belgium, Canada, Chile, Columbia, Czech Republic, Finland, France, Germany, Hungary, Iceland, Japan, Korea (Republic of), Kuwait, Lebanon,
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2131-2019, Z-2132-2019

Terminated Class III Medical devices

Recall started
10 May 2019
Terminated
24 May 2024
Reason
ECG trunk cables and leadwires may reduce the amount of energy reaching the patient during defibrillation, potentially limiting success of defibrillating the patients rhythm.
Distribution
Worldwide Distribution: US (nationwide) including Puerto Rico, and countries of: Australia, Austria, Azerbaijan, Bangladesh, Belarus, Belgium, Bermuda, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Canada, Chile, Colombia, Costa Rica, Cyprus, Czech Republic, Denmark, Ecuador, Egypt, El Salvado
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1878-2019, Z-1877-2019, Z-1875-2019, Z-1879-2019, Z-1876-2019

Terminated Class II Medical devices

Recall started
29 March 2019
Terminated
15 April 2021
Reason
There has been a reported incident of a Distar CRT monitor that has fallen off the monitor wall mount related to product aging.
Distribution
Worldwide Distribution: US (nationwide) to states of: AL, CA, CT, FL, GA, HI, IL, LA, MD, MI, MN, MO, NH, NY, OK, PA, TN, and VA; and countries of: Algeria, Argentina, Belgium, Bosnia & Herzegovina, Brazil, Bulgaria, Canada, China, Costa Rica, Cote D'Ivoire, Cyprus, Czech Republic, Egypt , France, F
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1753-2019, Z-1751-2019, Z-1755-2019, Z-1752-2019, Z-1754-2019

Terminated Class II Medical devices

Recall started
20 March 2019
Terminated
5 October 2020
Reason
During installation of certain Nuclear Medicine systems delivered with dismounted detectors, some detector screws may not have been secured correctly, leading to the possible loss of one detector mounting design redundancy.
Distribution
US and Austria, Canada, Croatia, France, Germany, India, Italy, Japan, Republic of Korea, Netherlands, Poland, Portugal, Slovakia, South Africa, Spain, Sweden, Switzerland, UK
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1264-2019, Z-1265-2019, Z-1259-2019, Z-1266-2019, Z-1261-2019, Z-1263-2019, Z-1262-2019, Z-1258-2019, Z-1260-2019, Z-1257-2019

Terminated Class II Medical devices

Recall started
19 March 2019
Terminated
30 March 2020
Reason
The outer seam of the coil, closest to the system cable, may separate and expose the porous material internal to the coil. This could result in user and/or patient contact with infectious agents since the internal porous material cannot be disinfected.
Distribution
Worldwide Distribution - US Nationwide in the states of AZ, CA, MS, NY, WI, DE, and countries of France, Germany, Japan, Netherlands, Spain, Sweden,
Initiated
Voluntary: Firm initiated
Recall number
Z-0062-2020

Ongoing Class II Medical devices

Recall started
15 March 2019
Reason
There is a potential for loss of the x-ray imaging function when the user changes field of view (FOV) from 30cm to 20cm or from 20cm to 30cm while releasing the fluoroscopy footswitch pedal simultaneously.
Distribution
Nationwide distribution to AR, CA, CA, FL, MI, NM, NV, SC, TN, TX, VA, WA, WI. International distribution to Algeria, Belgium, China, Colombia, France, Hungary, Italy, Japan, Korea (Republic Of), Kuwait, Libya, Pakistan, Poland, Russia, Saudi Arabia, South Africa, Sudan, Turkey, Vietnam.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1201-2019, Z-1199-2019, Z-1200-2019

Terminated Class I Medical devices

Recall started
15 March 2019
Terminated
21 May 2024
Reason
The bedside panels and latch areas on the Giraffe Bedded Warmer and Panda iRes Warmer can crack or break if the unit is moved using the bedside panels instead of the front handle or the maneuvering handle.
Distribution
Worldwide Distribution: US (Nationwide distribution) to states of: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, Guam, Northern Mariana Island
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1847-2019, Z-1846-2019

Terminated Class III Medical devices

Recall started
22 February 2019
Terminated
19 August 2020
Reason
The Integrated ECG cable with 3-lead leadwires can short circuit during defibrillation and conduct 25% of the defibrillation energy away from the patient.
Distribution
Worldwide Distribution - US Nationwide to AK, KS, MA, MA, ME, NJ, NJ, NM, NY, TX. Distributed worldwide to Chile, Finland, France, Germany, Italy, Korea (Republic Of), Russia, Spain, Sweden.
Initiated
Voluntary: Firm initiated
Recall number
Z-1060-2019

Ongoing Class II Medical devices

Recall started
12 February 2019
Reason
GE Healthcare PET/CT Table Stop Without Stopping X-Ray Exposure
Distribution
Nationwide Distribution
Initiated
FDA Mandated
Recall numbers
Z-1750-2019, Z-1749-2019, Z-1745-2019, Z-1747-2019, Z-1746-2019, Z-1748-2019

Ongoing Class II Medical devices

Recall started
19 January 2019
Reason
Additional low dose radiation exposure.
Distribution
Worldwide - US Nationwide Distribution
Initiated
FDA Mandated
Recall number
Z-1951-2019

Terminated Class II Medical devices

Recall started
28 December 2018
Terminated
5 October 2020
Reason
Increased temperature in the probe-head surface with LOGIQ P6 Model BT07 and BT09 scanners when used in specific types of scans and with specific probes.
Distribution
Worldwide distribution.
Initiated
Voluntary: Firm initiated
Recall number
Z-0788-2019

Terminated Class II Medical devices

Recall started
20 December 2018
Terminated
9 April 2021
Reason
When switching back & forth between multiple UV instances in the Windows taskbar, the patient images displayed on the mammo high resolution monitors may not show images of the patient selected on UV, which could result in an incorrect diagnosis.
Distribution
Worldwide Distribution: US (nationwide) in states of: AZ, CA, CO, DE, FL, IA, IL, IN, KS, MA, ME, MI, MN, MO, MS, MT, NE, NJ, NV, NY, OH, PA, PR, SC, SD, TN, TX, VA, VT, and WI; and countries of: Antigua, Australia, Austria, Belgium, Bermuda, Brazil, Canada, Cayman Islands, Chile, Colombia, Cote d'I
Initiated
Voluntary: Firm initiated
Recall number
Z-0793-2019

Ongoing Class II Medical devices

Recall started
30 November 2018
Reason
GE Healthcare has identified that a small number of Nuclear Medicine systems are operating without currently updated software. In a recent event, a patient s finger was injured during the Unload process while the automatic Unload feature was being used; the software on the system was not current.
Distribution
Domestic Distribution: AL, AR, AZ, CA, CO, CT, DC, DE, FL,GA, IA, IL, IN, KY, LA,MA, MD, ME, MI, MN, MO, MS, NC, NE, NM, NJ, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT,VA, WI and WV. Foreign Distribution: Argentina, Australia, Belgium, Brazil, Canada, Colombia, Denmark, France, Germany, Ghana, In
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2258-2020, Z-2259-2020

Terminated Class II Medical devices

Recall started
15 November 2018
Terminated
17 January 2021
Reason
Possible incorrect software version loaded.
Distribution
Worldwide Distribution - US Nationwide AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY Albania, Algeria, Argentina, Armenia, Australia, Austri
Initiated
Voluntary: Firm initiated
Recall number
Z-0693-2019

Terminated Class II Medical devices

Recall started
9 November 2018
Terminated
18 May 2026
Reason
Issue #1 affects Software Versions 6.0 SP7, SP7.0.1, SP7.0.2, SP7.0.3, SP8, SP8.0.1, SP8.0.a and SP8.0.2. When an image is flipped horizontally and vertically, the orientation markers appear correctly on the image to which they were applied. However, once new image rendering happens on this image or series (with scrolling; playing cin¿; maximizing or changing viewports/layouts; or synchronizing viewports and resizing the browser), the new image orientation is correctly maintained but the display of orientation markers can become incorrect. Issue #2 affects Software Versions 6.0 SP7, SP7.0.1, SP7.0.2, SP7.0.3, SP8, SP8.0.1, SP8.0.a, SP8.0.2, SP9 and SP9.0.1. When an image in a series is flipped and then rotated in Zero Footprint Viewer (ZFP), the orientation markers appear correctly on the image to which they were applied. However, once you navigate to another image in the series and the image you navigated to has a different orientation than the prior image from which you navigated, the new image orientation is correctly maintained but the display of orientation markers can become incorrect.
Distribution
Worldwide
Initiated
Voluntary: Firm initiated
Recall number
Z-0641-2019

Terminated Class II Medical devices

Recall started
11 October 2018
Terminated
20 March 2023
Reason
Synchronization failure
Distribution
CA, CO, DE, FL, HI, NY, OK, PA, SD, TX, WA, WI, and WV Belgium, Brazil, Canada, Colombia, France, Germany, Ghana, Ireland, Italy, Mexico, Netherlands, Panama, Poland, Saudi Arabia, South Africa, Spain, Switzerland, and Turkey.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0374-2019, Z-0375-2019

Terminated Class I Medical devices

Recall started
5 October 2018
Terminated
29 April 2024
Reason
The ventilator Inspiratory Safety Guard (ISG) may disconnect from the breathing circuit pathway. As a result, this could create a loss of ventilation which may lead to inadequate oxygenation for patients, increasing the possibility of hypoxia.
Distribution
Worldwide Distribution: US (Nationwide) including Puerto Rico, and to countries of: Albania, Argentina, Australia, Austria, Bangladesh, Bolivia, Bosnia and Herzegovina, Brazil, Cambodia, Canada, Chile, China, Colombia, Dominican Republic, Ecuador, Egypt, El Salvador, Finland, France, Germany, Greec
Initiated
Voluntary: Firm initiated
Recall number
Z-0463-2019

Terminated Class II Medical devices

Recall started
27 September 2018
Terminated
12 April 2021
Reason
Upgraded Aisys anesthesia devices and Aisys CS2 anesthesia devices with PSVPro Spontaneous Breathing Modes do not provide the PSVPro ventilation mode feature of transitioning to SIMV PCV backup ventilation mode. PSVPro Spontaneous Breathing Modes do not provide the feature of transitioning to SIMV PCV backup ventilation mode for a patient who stops spontaneous breathing if the PSVPro ventilation mode is in use prior to the clinician starting the Cycling Procedure.
Distribution
Worldwide distribution - US nationwide in the states of CA, CO, DC, FL, GA, HI, IA, IL, IN, KY, LA, ,MA, MD, MI, MN, MO, MT, NC, NE, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, WV. Countries of Argentina, Australia, Austria, Belgium, Bolivia, Brazil, Canada, China, Colombia, Croatia, Cze
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0411-2019, Z-0412-2019, Z-0413-2019

Terminated Class II Medical devices

Recall started
25 September 2018
Terminated
13 July 2020
Reason
Potential electric leakage affecting the user.
Distribution
Worldwide distribution. US Nationwide, Algeria, Australia, Austria, Belgium, Canada, Chile, China, Costa Rica, Cyprus, Czech Republic, Denmark, Finland, France, Germany, Honduras, India, Indonesia, Ireland, Italy, Japan, Kazakhstan, Korea, Lithuania, Mexico, Netherlands, New Zealand, Norway, Pakista
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0480-2019, Z-0483-2019, Z-0479-2019, Z-0481-2019, Z-0482-2019, Z-0484-2019

Terminated Class I Medical devices

Recall started
14 September 2018
Terminated
8 October 2020
Reason
A detector can detach and fall.
Distribution
United States: AK,AL,AR,AZ,CA,CO,CT,DC,DE,FL,GA,HI,IA,ID,IL,IN,KS,KY,LA,Guam,MA,MD,ME,MI,MO,MS,MT,NC,NE,NH,NJ,NM,NV,NY,OH,OK,OR,PA,PR, RI,SC,SD,TN,TX,UT,VA,VT,WA,WI AND WV International: Albania, Argentina, Australia, Azerbaijan, Belgium, Bosnia ,Herzegovina, Brazil, Canada, Chile, China, Costa
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0371-2019, Z-0370-2019, Z-0369-2019

Terminated Class II Medical devices

Recall started
14 September 2018
Terminated
29 May 2019
Reason
There is a potential for an undetected over temperature condition on the trans-esophageal (TEE) probe leading to a possible burn injury.
Distribution
Worldwide Distribution: US (nationwide) to states of: AZ, IL, MA, ME, MN, MT, PA, TX, and WA and countries of: Foreign distribution was made to Austria, Denmark, Finland, and Norway.
Initiated
Voluntary: Firm initiated
Recall number
Z-0588-2019

Ongoing Class II Medical devices

Recall started
11 September 2018
Reason
Patient monitors may restart due to network overload caused by network configuration.
Distribution
Worldwide Distribution: US (nationwide) including Puerto Rico; and countries of: Algeria, Argentina, Australia, Austria, Azerbaijan, Bahamas, Bangladesh, Belgium, Bolivia, Bosnia-Herzegovina, Brazil, Brunei, Bulgaria, Burma (Myanmar), Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republ
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0583-2019, Z-0585-2019, Z-0586-2019, Z-0584-2019, Z-0587-2019

Ongoing Class II Medical devices

Recall started
31 August 2018
Reason
The patient monitors may simultaneously restart as designed if all are connected to the same network and a network overload occurs for a prolonged time.
Distribution
Distribution was made nationwide, including Puerto Rico, Guam, and Samoa. There was also government and military distribution. Foreign distribution was made to Canada, Algeria, Angola, Argentina, Australia, Austria, Azerbaijan, Bahamas, Bahrain, Bangladesh, Belarus, Belgium, Bermuda, Bolivia, Bo
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0521-2019, Z-0524-2019, Z-0523-2019, Z-0519-2019, Z-0520-2019, Z-0522-2019

Terminated Class II Medical devices

Recall started
30 August 2018
Terminated
31 August 2022
Reason
Remote comparison studies displayed may be associated with the wrong patient, without a caution icon present.
Distribution
NY
Initiated
Voluntary: Firm initiated
Recall number
Z-0195-2019

Ongoing Class II Medical devices

Recall started
12 July 2018
Reason
On the Revolution CT systems equipped with the SmartStep Option, the Z location displayed on image annotation and on the gantry display may not be the actual location of the table and result in the system scanning at a location that is not what was confirmed by the operator
Distribution
Worldwide Distribution - US (Nationwide Distribution) and countries of: Australia, Belgium, Canada, Estonia, Finland, France, Germany, Hong Kong, Ireland, Japan, Korea Republic of, Malaysia, Morocco, New Zealand, Norway, Saudi Arabia, Singapore, Slovakia, Sweden, Switzerland, Thailand, Turkey, Unite
Initiated
FDA Mandated
Recall number
Z-0776-2019

Terminated Class II Medical devices

Recall started
24 May 2018
Terminated
8 February 2019
Reason
Some CT and CT/PET systems recently installed with a GE supplied A! electrical panel may not be properly wired to the partial system UPS (Uninteruptible Power Supply).
Distribution
Nationwide distribution to AL, CA, CO, FL, IL, IN, MA, MD, MI, MS, NC, NJ, NY,OH, OK, SC, TN, TX, VA.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2583-2018, Z-2584-2018, Z-2579-2018, Z-2578-2018, Z-2585-2018, Z-2586-2018, Z-2582-2018, Z-2580-2018, Z-2581-2018, Z-2587-2018

Completed Class II Medical devices

Recall started
24 May 2018
Reason
When your system was installed, the Main disconnect switch, EPO button on the A1 panel, and Remote EPO button(s) (outside of the A1 panel) may not have been wired correctly which could result in these not shutting off all power to the entire system as intended, resulting in a potential electric hazard.
Distribution
CA, DC, OH, UT, WI, Canada
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0606-2019, Z-0608-2019, Z-0607-2019

Terminated Class II Medical devices

Recall started
23 February 2018
Terminated
30 September 2024
Reason
A database handling error could occur during the image acquisition process affecting the completeness of acquired images with Centricity PACS-IW. There is a potential that one or more image series (i.e. all images within an image set) may be missing from an exam without indication to the user. While this is rare, this can occur with imaging studies that consist of a very small number of images per series. E.g. CR Thorax exam with 1 image per series.
Distribution
Worldwide Distribution-US Nationwide
Initiated
Voluntary: Firm initiated
Recall number
Z-1302-2018

Terminated Class II Medical devices

Recall started
25 January 2018
Terminated
2 November 2023
Reason
Potential that one or more image series may be missing from an exam without a user warning displayed in the Viewer.
Distribution
Worldwide Distribution - US Nationwide in the states of AK, AZ, CA, CO, DE, FL, GA, HI, IA, ID, IL, LA, MA, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OK, OR. PA. SD, TN, TX, UT, WA, WI, WV, and WY and the countries of Austria, Belgium, Brazil, Canada, Colombia, France, Germany, Ghana, India, Irela
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2101-2018, Z-2102-2018

Terminated Class II Medical devices

Recall started
15 December 2017
Terminated
21 April 2020
Reason
if the cables of the Monica IF24 System are removed by a user, and then incorrectly re-connected the Fetal Heart Rate (FHR) and Maternal Heart Rate (MHR) signals would be displayed incorrectly (swapped).
Distribution
Worldwide - US Nationwide Distribution.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1293-2018, Z-1292-2018

Completed Class II Medical devices

Recall started
30 October 2017
Reason
A potential hazardous situation may occur if the collimator locking handle is not properly placed in the lock position and the warning circuit does not detect that the collimator is unlocked because of the patient proximity to the collimator during a clinical scan.
Distribution
Worldwide
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0703-2019, Z-0704-2019

Terminated Class II Medical devices

Recall started
4 October 2017
Terminated
10 April 2020
Reason
Non-Conforming devices were shipped from the service center without proper evaluation and therefore may be nonconforming.
Distribution
US Distribution to states of: AZ, IL NC, NH and PA.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1199-2018, Z-1198-2018

Terminated Class II Medical devices

Recall started
16 September 2016
Terminated
26 July 2018
Reason
The placing of weight on the VRAD detector, of the Brivo XR385 digital radiographic system, can introduce image artifacts, which can result in patient exam retakes.
Distribution
AK, AL, AZ, CA, CT, FL, GA, HI, IA, IL, KS, KY, LA, MA, MD, MS, NJ, NM, OR, PA, TN, TX, VA, and Guam, and Canada.
Initiated
Voluntary: Firm initiated
Recall number
Z-2523-2018

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.