Helsinki, Finland
GE Healthcare Finland Oy recalls
5 recalls recorded since 2018, most recently on 26 September 2025. 5 are listed as ongoing. Status and classification are as published by the FDA.
Recalls5
Ongoing (FDA)5
Terminated or completed0
Class I (FDA)1
Ongoing Class II Medical devices
- Recall started
- 26 September 2025
- Reason
- Potential loss of monitoring on affected monitors when powered by certain batteries. Affected batteries can lose capability to hold charge and provide back up power. If this situation occurs, it will cause loss of patient monitoring and potentially delay recognizing patient status changes requiring treatment.
- Distribution
- Worldwide - US Nationwide distribution.
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-0455-2026, Z-0454-2026, Z-0456-2026, Z-0457-2026, Z-0453-2026
- B105P, B125P, B105M, B125M, B155M Patient Monitors; Model nos. 6160000-002 6160000-003 6160000-004 6160000-005 6160000-101 6160000-102 6160000-103 6160000-104 6160000-105 6160001-063 6160001-064 6160001-065 Monitor, Ph
- CARESCAPE ONE, Model no. 2087075-300 - shipped with potentially affected batteries or affected batteries installed as a spare part Monitor, Physiological, Patient (With Arrhythmia Detection or Alarms Updated 3/19 to ad
- Portrait VSM Model nos. 6660000-666 Monitor, Physiological, Patient (With Arrhythmia Detection or Alarms
- BATTERY FLEX-3S2P 10.8V 18650 LI-ION SMBUS used with affected patient monitors Model No 2062895-001 Manufacturer Part Number U80296-1R01, revision S
- Carescape B450, Model no. 5805686 - shipped with potentially affected batteries or affected batteries installed as a spare part Monitor, Physiological, Patient (With Arrhythmia Detection or Alarms Record updated 3/19/2
Ongoing Class II Medical devices
- Recall started
- 5 April 2024
- Reason
- GE HealthCare has become aware of the potential for loss of monitoring on CARESCAPE Canvas 1000, CARESCAPE Canvas Smart Display, and certain versions of CARESCAPE B850 and CARESCAPE B650 monitors if, following the replacement of the CPU timekeeper battery, a mains power loss occurs during active monitoring. In the unlikely event that this occurs, it can result in the displays or monitors not turning on, after the mains power is restored. This can result in a delay in recognizing changes in patient condition until a replacement monitor is connected.
- Distribution
- Worldwide distribution.
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-2168-2024, Z-2170-2024, Z-2172-2024, Z-2171-2024, Z-2169-2024
- CARESCAPE Canvas 1000, Model Numbers: a) 5865770; b) 5865770-01040169; c) 5865770-01054142; d) 5865770-01054696; e) 5865770-01054698; f) 5865770-01054721; g) 5865770-01054740; h) 5865770-01055202; i) 5865770-01
- CARESCAPE Canvas Smart Display, Model Numbers: a) 5514418, b) 5514418 / MAEP2, c) 5514418-01059642, d) 5514418-01080402, e) 5514418-01081317, f) 5514418-01086007, g) 5514418-01086838, h) 5514418-01088182, i)
- CARESCAPE B850 Model Numbers: a) MBC303, Model Numbers: 1) 2078633-001, 2) 2078633-010, 3) 2078633-012, 4) 2078633-017, 5) 2078633-028, 6) 2095802-001-54069330, 7) 2095802-001-554922, 8) 2095802-001-554923,
- CARESCAPE B650: Model Numbers: a) MBB313, Model Numbers: 1) 2095801-001, 2) 2095801-001-01018416, 3) 2095801-001-01023150, 4) 2095801-001-01023151, 5) 2095801-001-01023152, 6) 2095801-001-01032962, 7) 2095801-
- CARESCAPE Canvas Smart Display, Model Numbers: a) 5700023, b) 5700023-02, c) 5939590-102; monitor, physiological, patient
Ongoing Class I Medical devices
- Recall started
- 19 May 2023
- Reason
- There is a potential reduction of energy reaching patient during defibrillation, potential contact with unintended voltage, or inaccurate measurement.
- Distribution
- Worldwide
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-2028-2023, Z-2027-2023, Z-2034-2023, Z-2032-2023, Z-2026-2023, Z-2029-2023, Z-2033-2023, Z-2031-2023, Z-2030-2023
- TruSignal Sensitive Skin Sensor, REF TS-SE-3; Oximeter
- TruSignal AllFit Sensor: a) REF TS-AF-10, 10 pcs; B) REF AF-25, 25 pcs; Oximeter
- TruSignal Integrated Ear Sensor with Ohmeda Connector REF TS-E4-H; Oximeter
- TruSignal Integrated Ear Sensor with Datex Connector, REF TS-E4-GE; Monitor, physiological, patient(with arrhythmia detection or alarms)
- TruSignal Adult/Pediatric Sensor; Oximeter, Ear: a) REF TS-AP-10, 10 pcs; b) REF TS-AP-25, 25 pcs
- TruSignal Wrap Sensor, REF TS-W-D; Oximeter
- TruSignal Integrated Ear Sensor with Datex Connector, REF TS-E4-N; Oximeter
- TruSignal Integrated Ear Sensor with GE Connector REF TS-E2-GE; Oximeter, Ear
- and 1 more product lines
Ongoing Class II Medical devices
- Recall started
- 4 September 2018
- Reason
- When multiple CARESCAPEMonitor B450 or B850 units are connected to the same network and a network overload occurs for a prolonged time, the monitors may simultaneously restart as designed. The monitor restart will not be completed until the network issue has been corrected.
- Distribution
- Worldwide Distribution-US Nationwide
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-0095-2019, Z-0096-2019
- GE Healthcare CARESCAPE Monitor B450 Product Usage: Intended uses of CARESCAPE B450 in 510(k) countries [UR_CRSCP-885]: The CARESCAPE Monitor B450 is a multi-parameter patient monitor intended for use in multiple are
- GE Healthcare CARESCAPE Monitor B850 Product Usage: Intended uses of CARESCAPE B850 V2 in 510(k) countries [UR_CRSCP-1046]: The CARESCAPE Monitor B850 is a multi-parameter high acuity patient monitor intended for u
Ongoing Class II Medical devices
- Recall started
- 25 May 2018
- Reason
- When multiple CARESCAPE Monitor B650 units are connected to the same network and a network overload occurs for a prolonged time, the monitors may simultaneously restart as designed. The monitor restart will not be completed until the network issue has been corrected.
- Distribution
- worldwide
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-2340-2018
- GE Healthcare CARESCAPE Monitor B650
A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.