Helsinki, Finland

GE Healthcare Finland Oy recalls

5 recalls recorded since 2018, most recently on 26 September 2025. 5 are listed as ongoing. Status and classification are as published by the FDA.

Recalls5
Ongoing (FDA)5
Terminated or completed0
Class I (FDA)1

Ongoing Class II Medical devices

Recall started
26 September 2025
Reason
Potential loss of monitoring on affected monitors when powered by certain batteries. Affected batteries can lose capability to hold charge and provide back up power. If this situation occurs, it will cause loss of patient monitoring and potentially delay recognizing patient status changes requiring treatment.
Distribution
Worldwide - US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0455-2026, Z-0454-2026, Z-0456-2026, Z-0457-2026, Z-0453-2026

Ongoing Class II Medical devices

Recall started
5 April 2024
Reason
GE HealthCare has become aware of the potential for loss of monitoring on CARESCAPE Canvas 1000, CARESCAPE Canvas Smart Display, and certain versions of CARESCAPE B850 and CARESCAPE B650 monitors if, following the replacement of the CPU timekeeper battery, a mains power loss occurs during active monitoring. In the unlikely event that this occurs, it can result in the displays or monitors not turning on, after the mains power is restored. This can result in a delay in recognizing changes in patient condition until a replacement monitor is connected.
Distribution
Worldwide distribution.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2168-2024, Z-2170-2024, Z-2172-2024, Z-2171-2024, Z-2169-2024

Ongoing Class I Medical devices

Recall started
19 May 2023
Reason
There is a potential reduction of energy reaching patient during defibrillation, potential contact with unintended voltage, or inaccurate measurement.
Distribution
Worldwide
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2028-2023, Z-2027-2023, Z-2034-2023, Z-2032-2023, Z-2026-2023, Z-2029-2023, Z-2033-2023, Z-2031-2023, Z-2030-2023

Ongoing Class II Medical devices

Recall started
4 September 2018
Reason
When multiple CARESCAPEMonitor B450 or B850 units are connected to the same network and a network overload occurs for a prolonged time, the monitors may simultaneously restart as designed. The monitor restart will not be completed until the network issue has been corrected.
Distribution
Worldwide Distribution-US Nationwide
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0095-2019, Z-0096-2019

Ongoing Class II Medical devices

Recall started
25 May 2018
Reason
When multiple CARESCAPE Monitor B650 units are connected to the same network and a network overload occurs for a prolonged time, the monitors may simultaneously restart as designed. The monitor restart will not be completed until the network issue has been corrected.
Distribution
worldwide
Initiated
Voluntary: Firm initiated
Recall number
Z-2340-2018