Beijing, China

GE Healthcare (China) Co., Ltd. recalls

4 recalls recorded since 2018, most recently on 31 January 2025. 4 are listed as ongoing. Status and classification are as published by the FDA.

Recalls4
Ongoing (FDA)4
Terminated or completed0
Class I (FDA)0

Ongoing Class II Medical devices

Recall started
31 January 2025
Reason
For certain MR systems, a scan can be resumed following a Power Monitor trip when the "6-minute average SAR" is above the limit indicated on the "SAR Display". The user interface may incorrectly enable a scan to proceed earlier than intended. If resumed by the operator, such a scan can result in potential excessive tissue heating, especially if the patient is in contact with a conductive material or has an implant.
Distribution
Domestic US distribution nationwide. International distribution worldwide.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1240-2025, Z-1239-2025, Z-1242-2025, Z-1243-2025, Z-1241-2025

Ongoing Class II Medical devices

Recall started
5 November 2024
Reason
GE HealthCare has become aware of a potential issue on certain Discovery, Optima, Revolution series CT systems, and certain operator console upgrades that can result in rotated CT images. Particularly with helical, cine, and cardiac (if applicable) scan modes, after the first image of an exam, each subsequent image can be rotated by an increasing amount. The amount of progressive image rotation is dependent on the gantry speed and the duration of the scan. The final image in the exam can be rotated up to a maximum of ~56 degrees. The entire anatomy rotates as a whole by equal degrees, without distortion, and the anatomical positioning and z-axis of the images remains correct. Small degrees of image rotation might not be noticed. If image rotation is noticed and necessitates a rescan of the patient, the same image rotation will reoccur.
Distribution
Worldwide distribution.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0603-2025, Z-0602-2025, Z-0607-2025, Z-0600-2025, Z-0609-2025, Z-0604-2025, Z-0612-2025, Z-0614-2025, Z-0606-2025, Z-0605-2025, Z-0611-2025, Z-0608-2025 and 5 more

Ongoing Class II Medical devices

Recall started
24 October 2024
Reason
Steel cables in the Wall Stand for certain Proteus XR/a systems serviced by GE HealthCare have not been replaced at the intervals detailed in the Preventative Maintenance Action List.
Distribution
Worldwide - US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall number
Z-0550-2025

Ongoing Class II Medical devices

Recall started
12 July 2024
Reason
The XR system, Proteus XR/a, does not have a de-installation manual describing process steps for de-installing the devices. De-installation instruction process steps must be followed when de-installing these devices to avoid personnel injury during de-install.
Distribution
Worldwide distribution.
Initiated
Voluntary: Firm initiated
Recall number
Z-2733-2024

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.