G & H Wire Company recalls

2 recalls recorded since 2018, most recently on 21 December 2023. 1 is listed as ongoing. Status and classification are as published by the FDA.

Recalls2
Ongoing (FDA)1
Terminated or completed1
Class I (FDA)0

Ongoing Class II Medical devices

Recall started
21 December 2023
Reason
Latex elastic bands were labeled as non-latex elastic bands, an allergic rx can occur due to patient latex allergy
Distribution
Worldwide - US Nationwide distribution in the states of UT, MO, WV, TX, NJ and the countries of Canada, France, Germany, New Zealand, South Africa, Thailand.
Initiated
Voluntary: Firm initiated
Recall number
Z-0976-2024

Terminated Class II Medical devices

Recall started
21 January 2021
Terminated
20 September 2021
Reason
The text on the patient pack is correct and the product itself remains unchanged. FlexMedics does not expect device failures; Instead, there is an opportunity for misuse of the device if the label is not read. If the band is too small, the treatment may be more aggressive. It is important to note that elastic band force decreases over time. If the band is too small and still applied, the band could break. Both circumstances could result in temporary patient discomfort.
Distribution
Worldwide distribution - US Nationwide distribution in the states of CA, CT, FL, IL, KY, MI, MN, MS, MT, NH, NJ, NY, NC, TN, TX, UT, VT, WA, WV, AR, IL, and the countries of Spain, Netherlands, New Zealand, Germany, France, Czech Republic, Greece, Poland, Thailand, Belgium, Denmark, United Kingdom,
Initiated
Voluntary: Firm initiated
Recall number
Z-1164-2021

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.