Newton, KS, United States
Full Vision Inc recalls
2 recalls recorded since 2018, most recently on 12 May 2023. 2 are listed as ongoing. Status and classification are as published by the FDA.
Recalls2
Ongoing (FDA)2
Terminated or completed0
Class I (FDA)0
Ongoing Class II Medical devices
- Recall started
- 12 May 2023
- Reason
- Treadmill malfunction: drive PCB can misread the signals and stop or reverse and then accelerate forward.
- Distribution
- USA (Nationwide), Australia, Bahrain, Canada, Chile, Colombia, Costa Rica, Dominican Republic, FINLAND, Guatemala, HONG KONG, India, Indonesia, Israel, Mexico, NEW ZEALAND, Nicaragua, Panama, PARAQUAY, Peru, Philippines, Saudi Arabia, Singapore, SOUTH KOREA, Thailand, Ukraine, United Arab Emirates,
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-2058-2023, Z-2055-2023, Z-2052-2023, Z-2057-2023, Z-2056-2023, Z-2053-2023, Z-2054-2023
- TRACKMASTER, TREADMILL GE T2100-ST2 220V CHINA, MODEL NUMBER 317-07927GE CHINA
- TRACKMASTER, TREADMILL TMX428CP 220V, MODEL NUMBER 317-07929
- TRACKMASTER, TREADMILL TMX428 110V, MODEL NUMBER 317-07926
- TRACKMASTER, TREADMILL GE T2100-ST2 220V, MODEL NUMBER 317-07927GE
- TRACKMASTER, TREADMILL GE T2100-ST1 110VV, MODEL NUMBER 317-07926GE
- TRACKMASTER, TREADMILL TMX428 220V, MODEL NUMBER 317-07927
- TRACKMASTER, TREADMILL TMX428CP 110V, MODEL NUMBER 317-07928
Ongoing Class II Medical devices
- Recall started
- 31 August 2022
- Reason
- This has been identified a rare occurrence, under a specific scenario, where the circuit is found in a latch-up condition on the PCB that controls the safety function of the Magnetic Safety Tether, that when the magnet is pulled loose as during a fall, the treadmill may fail to stop.
- Distribution
- Wi, GA, NY, Mexico
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-0367-2023, Z-0365-2023, Z-0368-2023, Z-0370-2023, Z-0369-2023, Z-0366-2023, Z-0364-2023
- FULL VISION INC. TRACKMASTER, TREADMILL TMX428CP 220V, Model #317-07929
- FULL VISION INC. TRACKMASTER,TREADMILL TMX428 220V, Model #317-07927
- FULL VISION INC. TRACKMASTER, TREADMILL GE T2100-ST1 110V, Model #317-07926GE
- FULL VISION INC. TRACKMASTER, TREADMILL GE T2100-ST2 220V, Model #317-07927GE CHINA
- FULL VISION INC. TRACKMASTER, TREADMILL GE T2100-ST2 220V, Model #317-07927GE
- FULL VISION INC. TRACKMASTER, TREADMILL TMX428CP 110V, Model #317-07928
- FULL VISION INC. TRACKMASTER, TREADMILL TMX428 110V, Model #317-07926
A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.