Fujifilm Medical Systems U.S.A., Inc. recalls

12 recalls recorded since 2018, most recently on 30 April 2021. 3 are listed as ongoing. Status and classification are as published by the FDA.

Recalls12
Ongoing (FDA)3
Terminated or completed9
Class I (FDA)0

Terminated Class II Medical devices

Recall started
30 April 2021
Terminated
29 July 2024
Reason
Software versions 6.0.4 to 6.2.1 using Advanced Reporting -possibility that a previously assigned internal patient database ID can be reused for a new patient. It was discovered that this value does not increase when a patient merge activity is executed in a specific sequence
Distribution
Worldwide distribution - US Nationwide distribution and the countries of Canada, Bermuda, Israel, Hadassah Medical Center.
Initiated
Voluntary: Firm initiated
Recall number
Z-1775-2021

Ongoing Class II Medical devices

Recall started
2 March 2021
Reason
The wrong patient information may be displayed in the viewer or PowerJacket.
Distribution
Worldwide distribution - US Nationwide distribution including in the states of AZ, CA, CO, CT, DC, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, VA, VT, WA, WI, WV, WY and the countries of United Arab Emirates, Ang
Initiated
Voluntary: Firm initiated
Recall number
Z-1348-2021

Terminated Class II Medical devices

Recall started
18 February 2021
Terminated
24 February 2022
Reason
Potentially defective Chip Cassette units, when used for DCP or AFP-L3% measurement, incorrect DCP and AFP-L3% values may result
Distribution
CA, LA, MN, NC, UT, WI
Initiated
Voluntary: Firm initiated
Recall number
Z-1412-2021

Terminated Class III Medical devices

Recall started
22 December 2020
Terminated
20 July 2021
Reason
FUJIFILM has become aware that incorrect concentration values were printed on the bottle label for all product manufactured as part of Lot EL629. This may result incorrect results when used in testing.
Distribution
US Nationwide distribution including in the states of MO, OR, MD, PA.
Initiated
Voluntary: Firm initiated
Recall number
Z-0932-2021

Terminated Class II Medical devices

Recall started
11 December 2020
Terminated
17 November 2021
Reason
The software does not update measurements and calculations in the Clinical Reporting Application (CRA) when the ventricular trace is changed in the study by a different user.
Distribution
US Nationwide distribution in the state of MD.
Initiated
Voluntary: Firm initiated
Recall number
Z-0918-2021

Terminated Class II Medical devices

Recall started
11 September 2020
Terminated
17 November 2021
Reason
FUJIFILM has become aware of the possibility that certain CT studies may report a much higher 3D Sphere MAX value than expected. This issue is specific to 3D Sphere - other density tools are not affected.
Distribution
US Nationwide distribution including in the state of Minnesota.
Initiated
Voluntary: Firm initiated
Recall number
Z-0282-2021

Ongoing Class II Medical devices

Recall started
11 September 2020
Reason
There is a potential for the wrong patient information may be displayed in the viewer or PowerJacket. Additionally, the incorrect max standard uptake values (SUV) for PET ad CT studies may be returned for Siemens Modality.
Distribution
Software was distributed to medical facilities nationwide throughout the U.S.
Initiated
Voluntary: Firm initiated
Recall number
Z-0878-2021

Ongoing Class II Medical devices

Recall started
22 May 2020
Reason
Lateromedial (LM) and Lateromedial Oblique (LMO) Orientation Markers may be displayed incorrectly on the 3D TOMO slider bar. In addition, the slice location of the image and/or the slice direction could be incorrect as well, and result in misdiagnosis
Distribution
Nationwide
Initiated
Voluntary: Firm initiated
Recall number
Z-2471-2020

Terminated Class II Medical devices

Recall started
19 July 2019
Terminated
28 August 2020
Reason
The graphics driver of the FDR Go PLUS might cause the appearance of a Blue Screen of Death (BSoD)
Distribution
Nationwide
Initiated
Voluntary: Firm initiated
Recall number
Z-2441-2019

Terminated Class II Medical devices

Recall started
15 January 2019
Terminated
8 October 2020
Reason
FUJIFILM Medical Systems U.S.A., Inc., Modality Systems became aware of the possibility that the braking mechanism might not be engaged in all cases when the driving handle has been released.
Distribution
US Nationwide Distribution
Initiated
Voluntary: Firm initiated
Recall number
Z-1040-2019

Terminated Class II Medical devices

Recall started
4 August 2017
Terminated
1 April 2019
Reason
FUJIFILM Medical Systems U.S.A., Inc. (FMSU) identified a potential failure with our Mammography system. The acquisition workstations, FDR¿1000AWS, FDR-2000AWS, FDR-3000AWS, and CR¿IR363AWS assign a unique ID number to every image study but very rarely, with the acquisition workstation software versions, VS.O, VS.1, VS.2 , V6.0, V6.1, and V7.0, the system may assign the same ID number to a new set of images that was already assigned to the previous set of images due to the error in ID number generation logic. If an Image with this error (with duplicate ID number) Is transmitted to PACS, it may overwrite the image already stored on PACS.
Distribution
Worldwide Distribution.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0662-2018, Z-0661-2018, Z-0660-2018, Z-0659-2018

Terminated Class II Medical devices

Recall started
14 July 2017
Terminated
20 July 2018
Reason
There was a case that the swivel arm went down unintentionally when it was moved up/down.
Distribution
US Distribution and Internationally to Mexico.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0386-2018, Z-0385-2018

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.