FUJIFILM Healthcare Americas Corporation recalls

14 recalls recorded since 2018, most recently on 22 July 2026. 14 are listed as ongoing. Status and classification are as published by the FDA.

Recalls14
Ongoing (FDA)14
Terminated or completed0
Class I (FDA)0

Ongoing Class II Medical devices

Recall started
22 July 2026
Reason
Software issue may cause the ruler measurement to be inaccurate.
Distribution
Worldwide - US Nationwide distribution in the states of NY, VA, KS, OR, OH, and the country of Australia.
Initiated
Voluntary: Firm initiated
Recall number
Z-3030-2026

Ongoing Class II Medical devices

Recall started
9 January 2026
Reason
It was found that the contacts of the electromagnetic contactors used inside the cabinet of the X-ray high-voltage generator were welded, and the current might continue to be supplied to the power supply circuit even if the power of the device is cut off. As a result, it cannot be denied that the circuit in question overheats and damages the surrounding parts. It may also cause smoke.
Distribution
US Nationwide distribution in the states of AK, CA, CT, FL, IL, NH, OH, WI.
Initiated
Voluntary: Firm initiated
Recall number
Z-1299-2026

Ongoing Class II Medical devices

Recall started
15 September 2025
Reason
It was found that the shaft for mounting the X-ray tube unit on the support may break. In addition, it was found that the protective parts might not work effectively only when the shaft was broken while being operated toward the front while lifting.
Distribution
Domestic: AK, CA, FL, IL, MA, NH, OH, WI.
Initiated
Voluntary: Firm initiated
Recall number
Z-0044-2026

Ongoing Class II Medical devices

Recall started
11 June 2025
Reason
Devices had an unapproved slabbing software function enabled for use.
Distribution
Worldwide distribution - US Nationwide and the countries of Canada, Chile, Ecuador.
Initiated
Voluntary: Firm initiated
Recall number
Z-2219-2025

Ongoing Class II Medical devices

Recall started
11 February 2025
Reason
The incorrect computed patient age is showing in VX for patients less than 3 months old.
Distribution
Domestic: AR, AZ, CA, CO, CT, FL, GA, HI, IL, IN, KS, KY, LA, MA, MN, MO, MS, NC, NJ, NY, OH, OK, OR, TN, TX, VT, WA, WI, WV, WY. International: Australia, Belgium, Brazil, France, Germany, Portugal, UK
Initiated
Voluntary: Firm initiated
Recall number
Z-1407-2025

Ongoing Class II Medical devices

Recall started
11 April 2024
Reason
Bolts on the CH-200 tube may rupture due to fatigue failure, causing the tube support to fall, could result in patient injury
Distribution
US Nationwide distribution in the states of AK, CA, FL, IL, NE, NH, OH, TN, WI.
Initiated
Voluntary: Firm initiated
Recall number
Z-1898-2024

Ongoing Class II Medical devices

Recall started
11 April 2024
Reason
AC adapters of Noblus can heat up, and cause smoke in the worst case, when an excessive mechanical stress is applied to the socket for power cable (frequent insertion/removal of the power cable).
Distribution
US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall number
Z-1884-2024

Ongoing Class II Medical devices

Recall started
10 January 2024
Reason
Measurements on a Secondary Capture 2D image, that does not have pixel spacing in the DICOM header, when combined with a Breast Tomo Series results in incorrect measurements.
Distribution
Worldwide distribution - US Nationwide, including Puerto Rico and the countries of Australia, Belgium, Brazil, Canada, Colombia, Ecuador, France, Germany, India, Italy, Mexico, Namibia, Netherlands, Portugal, Qatar, Spain, Thailand, Turkey, UAE, United Kingdom.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1271-2024, Z-1268-2024, Z-1270-2024, Z-1269-2024, Z-1273-2024, Z-1272-2024

Ongoing Class II Medical devices

Recall started
3 January 2024
Reason
The LV Mass (2D Bullet) equation may be calculated incorrectly, resulting in a variance in the mass of the left ventricle (LV Mass). If used as a main factor for diagnosis, there is a risk of misdiagnosis or incorrect treatment plan of a patient, resulting in long-term health consequences or serious deterioration of health.
Distribution
US domestic distribution nationwide. International distribution to Israel.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1171-2024, Z-1172-2024

Ongoing Class II Medical devices

Recall started
23 November 2023
Reason
When the key switch is turned on, the exposure conditions are set to the default values (85kV, 1.6 mAs, approx. 0.13mGy@ SID 1m) intended for adult chest imaging. In addition, if exposure conditions are set from the console while the key switch is off, the set exposure conditions will not be reflected.
Distribution
US Nationwide and Worldwide
Initiated
FDA Mandated
Recall number
Z-3158-2024

Ongoing Class II Medical devices

Recall started
27 February 2023
Reason
Arm which holds the tube head in place is susceptible to breaking, if the device breaks it could potentially cause serious harm or injury to the patient and/or user.
Distribution
US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall number
Z-1470-2023

Ongoing Class II Medical devices

Recall started
12 July 2022
Reason
Error 00003050 or 00003052 may occur during scanning, the captured images will not be reconstructed and the Raw Data will not be displayed in the work list.result, may result in the need for additional scanning to capture the images properly for reconstruction, thereby exposing the patient to additional radiation
Distribution
US Nationwide Distribution
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1750-2022, Z-1748-2022, Z-1749-2022

Ongoing Class II Medical devices

Recall started
14 October 2021
Reason
Diagnostic ultrasound system with the specified software version and used in conjunction with a Fujifilm (formally Hitachi) transducer does not display the measurement results correctly with two cardiology measurement functions: Mitral Regurgitation Flow (MR Flow) Measurement; Mitral Valve (MV) Measurement. In addition there is an error in the instruction manual.
Distribution
Nationwide
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0190-2022, Z-0191-2022, Z-0189-2022

Ongoing Class II Medical devices

Recall started
14 October 2021
Reason
When the ultrasound system is used in combination with a Fujifilm (formally Hitachi) transducer, there is an error in the calculation formula of Effective Regurgitant Orifice Area (EROA) when you perform MR Flow combined with Left Ventricular Outflow Tract Flow (LVOT Flow) after performing an echocardiogram. The incorrect value is displayed. In addition, the error appears in the instruction manual
Distribution
US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall number
Z-0192-2022

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.