Fresh Roast Systems Inc recalls
1 recall recorded since 2018, most recently on 15 April 2026. 1 is listed as ongoing. Status and classification are as published by the FDA.
Recalls1
Ongoing (FDA)1
Terminated or completed0
Class I (FDA)0
Ongoing Class II Medical devices
- Recall started
- 15 April 2026
- Reason
- A defect has been identified in the manufacture of the Fresh Roast Systems ColorTrack instrument model number BENCH R-100 which could under very rare specific misuse circumstances potentially allow human access to infrared laser light levels in excess of the safety Classification of the product.
- Distribution
- U.S.
- Initiated
- FDA Mandated
- Recall number
- Z-2280-2026
- Fresh Roast Systems ColorTrack. Model Name: ColorTrack. Model Number: BENCH R-100