Fresenius Medical Care Renal Therapies Group, LLC recalls

16 recalls recorded since 2018, most recently on 10 August 2026. 4 are listed as ongoing. Status and classification are as published by the FDA.

Recalls16
Ongoing (FDA)4
Terminated or completed12
Class I (FDA)0

Ongoing Class II Medical devices

Recall started
10 August 2026
Reason
Heightened probability of blood leak from the connection between the Arterial Alpha Clip and the main line or pump segment. Leak during treatment may lead to blood loss and hypovolemia.
Distribution
Worldwide - US Nationwide distribution in the states of AL, CA, CO, FL, GA, IL, KY, LA, MA, MD, MN, MO, NC, NH, NH, NY, OH, PA, TN, TX, VA and the country of Lithuania.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-3087-2026, Z-3084-2026, Z-3085-2026, Z-3086-2026

Ongoing Class II Medical devices

Recall started
29 July 2026
Reason
Communication to recommend reviewing the IFU and ensuring that all treatment preparation and handling steps are performed as instructed.
Distribution
Domestic: Nationwide Distribution;
Initiated
Voluntary: Firm initiated
Recall number
Z-3225-2026

Ongoing Class II Medical devices

Recall started
14 July 2026
Reason
Complaints of blue-green discoloration in product
Distribution
Worldwide distribution - US Nationwide and the country of Mexico.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-3091-2026, Z-3092-2026, Z-3093-2026

Ongoing Class II Medical devices

Recall started
30 June 2026
Reason
Fresenius Medical Care Renal Therapies Group, LLC (FMCRTG) has received complaints regarding flow-, no water-, or temperature-related alarms with the 2008T¿ BlueStar Hemodialysis machines.
Distribution
US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall number
Z-3017-2026

Terminated Class II Medical devices

Recall started
29 May 2019
Terminated
19 November 2020
Reason
Additional Instructions: step-by step handling and inserting the cassette into the Liberty Cycler set. A damaged cassette can cause peritoneal dialysis (PD) solution to leak into the cycler and is a potential risk for contamination.
Distribution
US Nationwide Distribution
Initiated
Voluntary: Firm initiated
Recall number
Z-2151-2019

Terminated Class III Medical devices

Recall started
3 April 2019
Terminated
30 August 2019
Reason
Power Logic Board Fail Message on the 2008T BlueStar Machine with software version 2.72.
Distribution
US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall number
Z-1699-2019

Terminated Class II Medical devices

Recall started
5 November 2018
Terminated
17 April 2023
Reason
Potential for external blood leaks from the dialyzer header
Distribution
US Nationwide
Initiated
Voluntary: Firm initiated
Recall number
Z-0658-2019

Terminated Class III Medical devices

Recall started
9 October 2018
Terminated
5 August 2021
Reason
Discolored powder has confirmed the presence of a foreign substance
Distribution
Distribution US nationwide and Mexico
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0516-2019, Z-0517-2019, Z-0518-2019

Terminated Class II Medical devices

Recall started
13 July 2018
Terminated
17 September 2019
Reason
In the hemodialysis machine during the cleaning/disinfection program, when there is a power failure or interruption, the dialysis program button should be greyed out when the machine is powered up and a mandatory rinse should be required. However, if a power failure or interruption occurs during the cleaning/disinfection program with the auto-start feature enabled, the machine enters the dialysis program even though the dialysis program button is disabled and the mandatory rinse has not been performed. As a result, disinfecting chemical may not be completely removed from the machine before starting a treatment and disinfectant could exist in the dialysate delivered to the patient.
Distribution
US Nationwide in the states of : CA, CT, MI, NC, and NY.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-3195-2018, Z-3194-2018

Terminated Class II Medical devices

Recall started
2 May 2018
Terminated
13 May 2020
Reason
The bottles are mislabeled with an incorrect part number.
Distribution
Worldwide Distribution - US Nationwide in the states: CA, IA, IL, IN, KS, MD, MI, MN, MS, NE, NJ, NV, OK, PA, SD, TX, and WI.
Initiated
Voluntary: Firm initiated
Recall number
Z-2110-2018

Terminated Class II Medical devices

Recall started
24 January 2018
Terminated
11 May 2020
Reason
The recalling firm identified a software issue related to the Patient Line Check (PLC) which may result in an increased risk of Overfill (also known as Increased Intraperitoneal Volume, IIPV). Overfill/IIPV may result in serious injury or death.
Distribution
Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall number
Z-1365-2018

Terminated Class II Drugs

Recall started
20 December 2017
Terminated
14 February 2019
Reason
Lack of Assurance of Sterility: Leakage of the peritoneal dialysis (PD) solution bag.
Distribution
Product was distributed throughout the United States to patients and clinics.
Initiated
Voluntary: Firm initiated
Recall number
D-0604-2018

Terminated Class II Medical devices

Recall started
14 December 2017
Terminated
22 September 2021
Reason
When the recirculation ultrafiltration (UF) Goal is set to a value greater than 200ml in service mode and the user starts a treatment using the SLED (Sustained Low Efficiency Dialysis) program the display will show an invalid message and the UF pump will run at the recirculation UF rate which may be up to 4000ml/hour. The SLED program is required to limit the UF rate to a maximum of 1000 ml/hour.
Distribution
US Distribution to the states of: AR, CA, CO, FL, GA, IL, MA, MD, MN, NC, NY, OH, PA, SC, TN, and TX.
Initiated
Voluntary: Firm initiated
Recall number
Z-1278-2018

Terminated Class II Drugs

Recall started
13 November 2017
Terminated
6 December 2018
Reason
Lack of Assurance of Sterility
Distribution
Product was distributed throughout the United States.
Initiated
Voluntary: Firm initiated
Recall number
D-0827-2018

Terminated Class II Medical devices

Recall started
5 October 2017
Terminated
28 July 2021
Reason
While reviewing documentation for the next software release of the 2008T, an R&D technician identified that setting the UF goal to '0' introduces a discrepancy between the UF rate displayed and the actual UF pump rate. The software anomaly is also applicable to the 2008T Machines that contained the 2.63 Bug Fix per DCAF 17-088. 2008T Upgrade kits that contained the 2.63 Bug Fix are also affected. Additionally, SW version 2.64 (in design freeze) is also impacted by this anomaly.
Distribution
US Nationwide Distribution
Initiated
Voluntary: Firm initiated
Recall number
Z-0829-2018

Terminated Class II Medical devices

Recall started
29 September 2017
Terminated
9 August 2018
Reason
The label specifies the potassium concentration is 2.0 mEq/L. Some of the product in the affected lot was found to have a potassium concentration of 2.25 mEq/L.
Distribution
US Distribution to states of: CA, FL, KS, LA, and TX.
Initiated
Voluntary: Firm initiated
Recall number
Z-1010-2018

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.