Fresenius Kabi USA, LLC recalls

44 recalls recorded since 2018, most recently on 10 August 2026. 24 are listed as ongoing. Status and classification are as published by the FDA.

Recalls44
Ongoing (FDA)24
Terminated or completed20
Class I (FDA)24

Ongoing Class I Drugs

Recall started
10 August 2026
Reason
Presence of particulate matter:An internal investigation at the firm found the product to contain glass particles.
Distribution
Nationwide in the USA
Initiated
Voluntary: Firm initiated
Recall number
D-0827-2026

Ongoing Class I Drugs

Recall started
4 August 2026
Reason
Labeling: Label Mixup: MicroVault labeled as Morphine 2 mg/1 mL, contains a Prefilled Syringe of Dilaudid 0.5 mg/0.5 mL.
Distribution
Nationwide within the USA.
Initiated
Voluntary: Firm initiated
Recall number
D-0788-2026

Ongoing Class II Drugs

Recall started
16 July 2026
Reason
Presence of Particulate Matter: Hair was found in products
Distribution
Nationwide within in the USA
Initiated
Voluntary: Firm initiated
Recall numbers
D-0727-2026, D-0728-2026, D-0726-2026, D-0729-2026, D-0725-2026

Ongoing Class III Drugs

Recall started
14 May 2026
Reason
Failed Impurities/Degradations Specifications
Distribution
US Nationwide.
Initiated
Voluntary: Firm initiated
Recall number
D-0581-2026

Ongoing Class I Medical devices

Recall started
6 May 2026
Reason
Potential software anomaly that can cause the pump to report a false battery-health value (typically 69%) and trigger a battery-depletion alarm that will shut down the pump unless it is reconnected to AC mains power.
Distribution
US Nationwide distribution in the states of CA, CO, FL, GA, GA, ID, IL, MD, MI, MN, MO, MS, NE, NJ, NV, OK, OR, SC, TX, VA, WA, WI.
Initiated
Voluntary: Firm initiated
Recall number
Z-2223-2026

Ongoing Class I Medical devices

Recall started
6 May 2026
Reason
Fresenius Kabi has been made aware via customer complaints that dropping or severe jarring of the Ivenix LVP may cause damage to internal components that may not be apparent or readily visible and the devices therefore remained in service. More specifically, drops or jarring on a hard surface can cause the touchscreen input cable to loosen, resulting in unintended touchscreen behavior on the LVP, including random screen touches and/or lack of screen responsiveness to user input. Under the reported failure condition, unintended touch inputs may register at arbitrary screen locations without user interaction. These inputs occur intermittently and without a predictable pattern.
Distribution
Domestic: CA, CO, FL, GA, ID, IL, MD, MI, MI , MN, MO, MS, NE, NJ, NV, OK, OR, SC, TX, VA, WA, WI, WI;
Initiated
Voluntary: Firm initiated
Recall number
Z-2224-2026

Ongoing Class II Drugs

Recall started
11 March 2026
Reason
Lack of Assurance of Sterility
Distribution
US Nationwide , Alaska, and Puerto Rico.
Initiated
Voluntary: Firm initiated
Recall numbers
D-0427-2026, D-0424-2026, D-0432-2026, D-0426-2026, D-0428-2026, D-0430-2026, D-0435-2026, D-0437-2026, D-0436-2026, D-0433-2026, D-0425-2026, D-0429-2026 and 2 more

Ongoing Class II Medical devices

Recall started
4 February 2026
Reason
Potential for the Cassette Loading Lever to break.
Distribution
US Nationwide distribution in the states of CA, CO, FL, GA, ID, IL, MA, MD, ME, MI, MN, MS, NE, NJ, NV, OK, OR, SC, TX, VA, WA, WI.
Initiated
Voluntary: Firm initiated
Recall number
Z-2274-2026

Ongoing Class II Medical devices

Recall started
21 November 2025
Reason
Emphasizing instructions for LVP duration programming located in the IFU.
Distribution
US Nationwide distribution in the states of CA, CO, FL, GA, ID, IL, MD, MI, MI, MN, MS, NE, NJ, NV, OK, OR, SC, TX, VA, WA, WI, WI.
Initiated
Voluntary: Firm initiated
Recall number
Z-1048-2026

Ongoing Class I Medical devices

Recall started
14 November 2025
Reason
Software version 5.10.1 and earlier contain anomalies that have the potential to cause serious patient harm or death.
Distribution
US (Domestic) distribution: CA, CO, FL, GA, ID, IL, MD, MI, MN, MS, NE, NJ, NV, OK, OR, SC, TX, VA, WA & WI.
Initiated
Voluntary: Firm initiated
Recall number
Z-0885-2026

Ongoing Class I Drugs

Recall started
6 November 2025
Reason
Microbial Contamination of Sterile Products; out of limit results obtained for endotoxin testing.
Distribution
Nationwide within the United States as well as AK, HI, and PR.
Initiated
Voluntary: Firm initiated
Recall number
D-0182-2026

Ongoing Class II Medical devices

Recall started
3 November 2025
Reason
Downstream Occlusion alarms may occur during procedures using low flow infusion rates when a check valve* is added to the Ivenix Administration Set.
Distribution
US Nationwide distribution in the sates of CA, CO, FL, GA, ID, IL, MD, MI, MN, MS, NE, NJ, NV, OK, OR, SC, TX, UT, VA, WA, WI.
Initiated
Voluntary: Firm initiated
Recall number
Z-0848-2026

Ongoing Class I Medical devices

Recall started
3 November 2025
Reason
Incorrect assembly of specific lot of LVP Primary Administration Set that could result in serious patient injury.
Distribution
US Nationwide distribution in the states of IL, MD, MI, OK, SC, TX, VA & WI.
Initiated
Voluntary: Firm initiated
Recall number
Z-0586-2026

Ongoing Class II Medical devices

Recall started
3 September 2025
Reason
Potential issue that can cause the device to register "phantom" touches in the lower-left corner of the touchscreen during operation.
Distribution
US Nationwide distribution in the state of Minnesota.
Initiated
Voluntary: Firm initiated
Recall number
Z-0378-2026

Ongoing Class II Medical devices

Recall started
24 June 2025
Reason
Potential for external cassette leaks
Distribution
US Nationwide distribution in the states of CA, CO, DE, FL, GA, ID, IL, MA, MD, MI, MN, MS, NJ, NV, OK, OR, SC, TX, VA, WA, WI.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2235-2025, Z-2232-2025, Z-2234-2025, Z-2233-2025

Ongoing Class I Medical devices

Recall started
12 May 2025
Reason
Mis-assembly error of Blood Products Administrations Sets where the 200-micron drip chamber filter may be incorrectly positioned and unable to filter out large blood particulates.
Distribution
US distribution: CA, CO, DE, GA, IL, MD, MI, MN, MS, NC, NJ, NV, OK, SC, TX, VA, WA & WI.
Initiated
Voluntary: Firm initiated
Recall number
Z-1869-2025

Ongoing Class I Medical devices

Recall started
10 January 2025
Reason
Large Volume Pump Software, version 5.9.2 and earlier has potential for the following anomalies: 1. The pump may become nonfunctional if during an alarm condition the Pause Audio option is repeated 70 time or more. 2. If a secondary infusion is started at the exact moment a primary infusion completes and VTBI reaches 0, it will switch to primary. The primary infusion will infuse at the previously programed primary rate and continue until the infusion is stopped or the bag is empty.
Distribution
US distribution to CA, CO, GA, ID, MI, MS, NE, NJ, NV, OK, TX, UT, VA, WA, WI.
Initiated
Voluntary: Firm initiated
Recall number
Z-1019-2025

Ongoing Class I Medical devices

Recall started
6 December 2024
Reason
Potential pneumatic valve failure that may cause the LVP to stop an active infusion and/or could prevent the pump from being used.
Distribution
US distribution to states of: ID, MN, NV, TX, VA, WI.
Initiated
Voluntary: Firm initiated
Recall number
Z-0763-2025

Ongoing Class I Medical devices

Recall started
30 August 2024
Reason
The software has anomalies that have the potential to cause alarms, nonfunctioning pump, or unresponsive screen while continuing therapy. These could cause serious patient harm or death.
Distribution
US Nationwide.
Initiated
Voluntary: Firm initiated
Recall number
Z-3153-2024

Ongoing Class I Medical devices

Recall started
1 August 2024
Reason
A defect may cause an uncontrolled medication flow condition which may result in a dose greater than prescribed for the patient (i.e., medication overdose).
Distribution
US Distribution to states of: CO, ID, NJ, UT.
Initiated
Voluntary: Firm initiated
Recall number
Z-2721-2024

Ongoing Class I Medical devices

Recall started
7 March 2024
Reason
Software has anomalies that have the potential to cause underdose, overdose, or delay in therapy which could lead to serious patient harm or death.
Distribution
US Distribution: CA, CO, ID, MI, NE, NJ, NV, OK, TX, UT, VA, WA, & WI.
Initiated
Voluntary: Firm initiated
Recall number
Z-1484-2024

Ongoing Class I Medical devices

Recall started
29 November 2023
Reason
The device may experience mechanical interference on the Fluid Valve pins, which will trigger a Pump Problem alarm. The issue may lead to delay or interruption of therapy. Depending on the therapy and duration of delay/interruption, the issue may lead to serious harm or death.
Distribution
US distribution to states of: CA, CO, NJ, UT, and WI.
Initiated
Voluntary: Firm initiated
Recall number
Z-0549-2024

Completed Class II Medical devices

Recall started
10 May 2023
Reason
Retroactively reported corrections from 2023: 1) A software defect may cause an incorrect (Fail-Stop) alarm when an administration set is loaded or coupled while the pump is executing the power-up sequence. May lead to delay in therapy. 2) Alert is not annunciated informing the clinician that the bolus cannot be delivered when the entered bolus dose exceeds the Care Profile Hard Rate Max limit and Rapid Bolus is selected. May lead to over infusion. Both issues were resolved in all fielded product in software version 5.8.0, which was installed in affected units May thru August 2023.
Distribution
US Nationwide distribution to the states of CA, CO, NJ, WI, UT.
Initiated
Voluntary: Firm initiated
Recall number
Z-1283-2024

Ongoing Class I Medical devices

Recall started
10 March 2023
Reason
Fluid ingress that can cause a loss of electrical funtion and failure of the Set ID Sensory, resulting in the prevention of infusion or halt of infusion.
Distribution
Domestic: CA, CO, NJ, WI, & UT. No foreign distribution.
Initiated
Voluntary: Firm initiated
Recall number
Z-1313-2023

Terminated Class II Drugs

Recall started
22 November 2022
Terminated
23 October 2023
Reason
Subpotent Drug: Testing results below the defined limit for the epinephrine portion of this product.
Distribution
Nationwide in the USA
Initiated
Voluntary: Firm initiated
Recall numbers
D-0128-2023, D-0129-2023, D-0127-2023

Completed Class II Medical devices

Recall started
16 September 2022
Reason
Retroactively reported correction from 2022: A software defect may cause an incorrect (Fail-Stop) Alarm when an administration set is loaded with an upstream occlusion present, resulting in a non-recoverable (Fail-Stop) Pump Problem alarm. This may lead to delay of therapy, and patient may experience from no to minimal to serious harm. This issue was resolved in all fielded product in software version 5.2.1, which was installed in September 2022.
Distribution
US Nationwide distribution in the states of WI and NJ.
Initiated
Voluntary: Firm initiated
Recall number
Z-1282-2024

Terminated Class I Medical devices

Recall started
8 September 2022
Terminated
10 September 2024
Reason
The display screen may become frozen and unresponsive to user input, triggering an alarm that is evidenced by a buzzer sound and flashing red LEDs as a secondary notification. However, as the screen goes blank it does not indicate an alarm condition or respond. This issue leads to temporary cessation of flow of a medication or fluid, requiring immediate intervention by the clinician.
Distribution
Domestic distribution to NJ and WI.
Initiated
Voluntary: Firm initiated
Recall number
Z-0002-2023

Ongoing Class II Medical devices

Recall started
22 April 2022
Reason
Start-up time for the flow rate range greater than 50 ml/hr, and less than 200 ml/hr may exceed the specified 6 minutes, depending on the characteristics of air bubble entrapment within the Intermediate Pumping Chamber (IPC) of the administration set
Distribution
US Nationwide distribution in the states of NJ, WI.
Initiated
Voluntary: Firm initiated
Recall number
Z-2635-2023

Terminated Class I Drugs

Recall started
4 March 2022
Terminated
29 September 2023
Reason
Presence of particulate matter: particulates composed of carbon and oxygen with varying amounts of iron and trace amounts of sodium, silicon, chromium, aluminum and cellulose.
Distribution
Nationwide USA
Initiated
Voluntary: Firm initiated
Recall number
D-0750-2022

Terminated Class II Drugs

Recall started
17 September 2021
Terminated
20 June 2023
Reason
Defective container: Cracked vials leading to lack of sterility assurance
Distribution
U.S.A. Nationwide
Initiated
Voluntary: Firm initiated
Recall number
D-0856-2021

Terminated Class II Drugs

Recall started
25 June 2021
Terminated
28 December 2022
Reason
Low out of specification results for epinephrine assay.
Distribution
Distributed Nationwide in the USA
Initiated
Voluntary: Firm initiated
Recall number
D-0651-2021

Terminated Class I Drugs

Recall started
23 December 2020
Terminated
9 June 2023
Reason
Presence of Particulate Matter - found in reserve sample vials at the firm.
Distribution
Nationwide USA and Puerto Rico
Initiated
Voluntary: Firm initiated
Recall number
D-0245-2021

Terminated Class II Drugs

Recall started
17 December 2020
Terminated
20 June 2023
Reason
Presence of Particulate Matter - found in reserve sample vials at the firm.
Distribution
USA Nationwide and Puerto Rico
Initiated
Voluntary: Firm initiated
Recall number
D-0184-2021

Terminated Class II Drugs

Recall started
10 December 2020
Terminated
25 August 2022
Reason
Subpotent Drug: Low out-of-specification assay results for the epinephrine component.
Distribution
Nationwide in the USA
Initiated
Voluntary: Firm initiated
Recall number
D-0172-2021

Terminated Class III Drugs

Recall started
17 November 2020
Terminated
25 August 2022
Reason
Cross Contamination with Other Products: trace amounts of octreotide found during testing
Distribution
Nationwide USA and Puerto Rico
Initiated
Voluntary: Firm initiated
Recall number
D-0105-2021

Terminated Class I Drugs

Recall started
17 November 2020
Terminated
12 September 2022
Reason
Cross Contamination with other products: trace amounts of lidocaine
Distribution
USA Nationwide
Initiated
Voluntary: Firm initiated
Recall number
D-0120-2021

Terminated Class I Drugs

Recall started
22 July 2020
Terminated
12 September 2022
Reason
Cross Contamination with other products: trace amounts of lidocaine
Distribution
U.S.A. Nationwide
Initiated
Voluntary: Firm initiated
Recall number
D-1537-2020

Terminated Class III Drugs

Recall started
13 July 2020
Terminated
19 August 2022
Reason
Labeling Error: Label Error on Declared Strength: Carton label and product insert incorrectly states the quantity of the excipient edetate disodium (EDTA) as 5.4 mg / vial, rather than the actual amount of 18.8 mg / vial.
Distribution
Nationwide within the United States
Initiated
Voluntary: Firm initiated
Recall number
D-1391-2020

Terminated Class I, Class II Drugs

Recall started
20 April 2020
Terminated
16 October 2023
Reason
Presence of Particulate Matter - found in reserve sample vials at the firm.
Distribution
USA Nationwide and Puerto Rico
Initiated
Voluntary: Firm initiated
Recall numbers
D-1296-2020, D-1297-2020

Terminated Class I Drugs

Recall started
28 June 2019
Terminated
20 September 2022
Reason
Presence of Particulate Matter; glass particulates
Distribution
Nationwide USA and Puerto Rico
Initiated
Voluntary: Firm initiated
Recall number
D-1489-2019

Terminated Class III Drugs

Recall started
22 May 2019
Terminated
28 January 2021
Reason
Failed Impurities/Degradation Specifications.
Distribution
Nationwide within the United States
Initiated
Voluntary: Firm initiated
Recall number
D-1386-2019

Terminated Class I Drugs

Recall started
15 November 2018
Terminated
1 May 2023
Reason
Labeling Not Elsewhere Classified: Fresenius Kabi is taking this action due to an incorrect statement on the product insert for product codes 918610 and 918620 indicating that the stoppers do not contain natural rubber latex. The tray label for these two product codes and the vial label for product code 918620 also incorrectly state that the stoppers do not contain latex. The above listed product codes and batches being recalled have stoppers containing natural rubber latex.
Distribution
Distributed Nationwide in the USA and Puerto RIco
Initiated
Voluntary: Firm initiated
Recall numbers
D-0299-2019, D-0298-2019

Terminated Class III Drugs

Recall started
16 January 2018
Terminated
17 September 2019
Reason
Failed impurities/ degradation specifications: Out-of-specification (OOS) result for USP related compound F at the 12 month stability test station.
Distribution
Nationwide USA
Initiated
Voluntary: Firm initiated
Recall number
D-0426-2018

Terminated Class I Drugs

Recall started
27 October 2017
Terminated
5 October 2020
Reason
Labeling: Label MIX-UP. Blister Packages, Labeled as Midazolam injection, USP, 2 mg / 2 ml, Containing Syringes of Ondansetron Injection, USP, 4 mg / 2 mL
Distribution
Nationwide in the USA
Initiated
Voluntary: Firm initiated
Recall number
D-0152-2018

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.