Fort Defiance Industries, LLC recalls

3 recalls recorded since 2018, most recently on 8 July 2026. 2 are listed as ongoing. Status and classification are as published by the FDA.

Recalls3
Ongoing (FDA)2
Terminated or completed1
Class I (FDA)0

Ongoing Class II Medical devices

Recall started
8 July 2026
Reason
The Automated Field Steam Sterilizer's water recovery system 120V Amphenol connector may experience an intermittent loss of ground continuity, which could result in an unreliable fault current return path.
Distribution
Worldwide - US Nationwide distribution in the states of MD, VA, NJ, TN and the countries of Canada, Lithuania, Spain.
Initiated
Voluntary: Firm initiated
Recall number
Z-2983-2026

Ongoing Class II Medical devices

Recall started
19 August 2021
Reason
The rear heater plate and cover assembly that contains the electrical wiring for the heaters does not have sufficient bonding to ground.
Distribution
Distributed nationwide to PA, MD, TX, NJ, OH, TN, UT, VA and internationally to Canada.
Initiated
Voluntary: Firm initiated
Recall number
Z-0121-2022

Terminated Class II Medical devices

Recall started
24 May 2018
Terminated
27 June 2023
Reason
During long-term storage (i.e. military depot), the interaction of stagnant water and dissimilar metals in the plumbing of the P2131 device produces deposits in the water system that can potentially interfere with the rear sight gauge and / or the water liquid level switch.
Distribution
US Nationwide Distribution in the states of PA. VA. TX
Initiated
Voluntary: Firm initiated
Recall number
Z-2411-2018

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.