Waltham, MA, United States
EVIDENT SCIENTIFIC INC recalls
1 recall recorded since 2018, most recently on 22 July 2024. 1 is listed as ongoing. Status and classification are as published by the FDA.
Recalls1
Ongoing (FDA)1
Terminated or completed0
Class I (FDA)0
Ongoing Class II Medical devices
- Recall started
- 22 July 2024
- Reason
- A broken fuse inside the laser that was caused by a short circuit at PIN 4 of the DOOR CTRL. This short circuit was caused by a badly positioned cable that made contact with the housing/frame and caused a short circuit and destroyed the fuse in the laser. This caused the shutter to remain in its most recent state; either in an open or closed position. Once the fuse was destroyed the shutter could not be operated.
- Distribution
- US
- Initiated
- FDA Mandated
- Recall number
- Z-2416-2024
- SlideView VS200 slide viewer,
A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.