Ethicon, Inc. recalls

30 recalls recorded since 2018, most recently on 8 September 2026. 15 are listed as ongoing. Status and classification are as published by the FDA.

Recalls30
Ongoing (FDA)15
Terminated or completed15
Class I (FDA)0

Ongoing Class II Medical devices

Recall started
8 September 2026
Reason
Silk sutures, indicated for use in cardiovascular, ophthalmic and neurological procedures, packaged with alcohol-based fluid, not part of product design, that if discovered prior to use may cause surgical delay. If used, fluid could cause reduced suture tensile strength or knot security, tissue irritation/inflammatory tissue response, and treatment failure, including wound dehiscence/bleeding.
Distribution
Worldwide - US Nationwide distribution in the states of SC, PA, TX, CA, ID, LA, MO, AL, NC, UT, IL, MA, OH, AZ, TN, OK, NE, ME, IN, WI, FL, CO, WY, NH, NY, NC, MN and the country of CA.
Initiated
Voluntary: Firm initiated
Recall number
Z-0153-2027

Ongoing Class II Medical devices

Recall started
15 August 2025
Reason
Potential for barb non-engagement.
Distribution
Worldwide - US Nationwide distribution in the states of AR, AZ, CA, IL, IN, KS, MA, MD, MI, MO, MN, NC, NH, NM, TX, VA, WA, WI and the countries of Singapore and Malaysia.
Initiated
Voluntary: Firm initiated
Recall number
Z-2605-2025

Ongoing Class II Medical devices

Recall started
20 December 2024
Reason
Sutures may have an open seal on the primary packaging, due to a manufacturing issue on a specific machine, and if issue is undetected could lead to breach in sterility, could introduce pathogens to the patient and cause infection.
Distribution
Worldwide - US Nationwide distribution including in the states of AL, AR, AZ, CA, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, LA MA, MD, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI and the countries of AT, AU, BE, CA, CH, CN, CZ, DE, DK, FI, FR, GB, HU, IE,
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1096-2025, Z-1098-2025, Z-1097-2025

Ongoing Class II Medical devices

Recall started
8 October 2024
Reason
Due a component mix-up in manufacturing, the affected products may contain incorrect needle type/size and suture length.
Distribution
U.S.: AK, AL, AR, AZ, CA, CO,CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, WV, and WY. O.U.S.: Not provided.
Initiated
Voluntary: Firm initiated
Recall number
Z-0487-2025

Ongoing Class II Medical devices

Recall started
25 September 2024
Reason
Complaints were received regarding needle pull-off and suture degradation. Analysis on returned product from this lot confirmed that some sutures were visually degraded in the package.
Distribution
International distribution to the country of China.
Initiated
Voluntary: Firm initiated
Recall number
Z-0319-2025

Ongoing Class II Medical devices

Recall started
10 May 2024
Reason
Issue on a specific packaging machine resulted in a hole in the primary packaging, product sterility compromised and could result in infection .
Distribution
Worldwide distribution - US Nationwide including PR and the countries of AR, AU, AU, BE, BR, CA, CA, CL, CN, CO, CR, EC, IL, JP, JP, KW, MX, NZ, PA, PR, PR, SA, SG, UY, ZA.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2147-2024, Z-2148-2024, Z-2144-2024, Z-2146-2024, Z-2145-2024, Z-2149-2024

Ongoing Class II Medical devices

Recall started
23 April 2024
Reason
Some applicator tips from these lots had open seals.
Distribution
US Nationwide distribution in the states of AR, CA, FL, GA, ID, IL, IN, KY, MA, MD, MN, MO, NC, NE, NJ, PA, TN, TX, WI.
Initiated
Voluntary: Firm initiated
Recall number
Z-1881-2024

Ongoing Class II Medical devices

Recall started
8 March 2024
Reason
Degraded and unusable upon removal from the foil pouch due to exposed environmental conditions during transit and storage, compromising sterility
Distribution
CO GA IA MI NJ NY NY PA SC WI
Initiated
Voluntary: Firm initiated
Recall number
Z-1533-2024

Ongoing Class II Medical devices

Recall started
10 August 2023
Reason
Internal testing on returned product from this lot confirmed that some PDS" II (polydioxanone) Sutures from this lot did not meet tensile strength requirement.
Distribution
US Nationwide. Global Distribution.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2616-2023, Z-2617-2023

Ongoing Class II Medical devices

Recall started
31 March 2023
Reason
Suture breakage during intra-operative use could result in poor performance of the impacted product because the intended benefit of tissue approximation and/or ligation may not be achieved
Distribution
International distribution in the countries of United Kingdom, Ireland and Turkey.
Initiated
Voluntary: Firm initiated
Recall number
Z-1571-2023

Ongoing Class II Medical devices

Recall started
16 November 2022
Reason
knot tensile strengths were observed to be significantly below the minimum knot tensile, will result in breakage prior to use or intra-operatively rather than premature post-operative breakage
Distribution
International (OUS) only: Canada, Japan, Australia and New Zealand
Initiated
Voluntary: Firm initiated
Recall number
Z-0480-2023

Ongoing Class II Medical devices

Recall started
20 September 2022
Reason
Packages may contain a PS-1 cutting edge needle and not the intended SH taper point needle may result in more tissue trauma than is anticipated for taper point needles, which may necessitate repair
Distribution
International distribution of the country of China.
Initiated
Voluntary: Firm initiated
Recall number
Z-0150-2023

Ongoing Class II Medical devices

Recall started
8 August 2022
Reason
Product may contain the incorrect material which could impact treatment.
Distribution
Distribution in Japan only
Initiated
Voluntary: Firm initiated
Recall number
Z-1773-2022

Ongoing Class II Medical devices

Recall started
24 March 2022
Reason
Affected product lots may contain the incorrect needle type/size and suture length due to component mix-up.
Distribution
International distribution in the country of China.
Initiated
Voluntary: Firm initiated
Recall number
Z-1053-2022

Terminated Class II Medical devices

Recall started
4 January 2021
Terminated
29 November 2023
Reason
Product distributed with a red tape as part of the sterile packaging which could impact the seal and result in a sterility breach
Distribution
US Nationwide
Initiated
Voluntary: Firm initiated
Recall number
Z-0953-2021

Terminated Class II Medical devices

Recall started
23 October 2020
Terminated
21 December 2022
Reason
The firm is recalling the PROCEED Surgical Mesh device, because it is a sterile device, and a customer found a hair inside of the primary packaging of a single unit.
Distribution
The impacted lot (295 units) was distributed to 27 countries worldwide excluding the US. However, the firm is only recalling the 15 units that were distributed in South Korea due to South Korea's Ministry of Food and Drug Safety (MFDS) requirements.
Initiated
Voluntary: Firm initiated
Recall number
Z-0533-2021

Terminated Class II Medical devices

Recall started
9 September 2020
Terminated
7 November 2023
Reason
Product in the foil package for VICRYL V960 may not match the foil label and sales unit carton. The correct VICRYL suture is size 10-0, 4 inches (10 cm) in length, has monofilament construction, and a single needle. The incorrect VICRYL suture is size 6-0, 18 inches (45 cm) in length, has braided construction, and two needles
Distribution
International distribution including the countries of Australia, Canada, Japan, and New Zealand
Initiated
Voluntary: Firm initiated
Recall number
Z-0116-2021

Ongoing Class II Medical devices

Recall started
10 June 2020
Reason
Lots of size 8-0 Blue twisted Silk Suture may not meet the stringent full shelf life of 5 years
Distribution
No domestic distribution. Foreign distribution to Algeria Angola Australia Austria Azerbaijan Bahrain Belgium Bosnia and Herzegovina Canada Chad Chile Czechia Egypt Finland France Gabon Germany Greece Hong Kong Iceland India Iran Ireland Israel Italy Japan Jordan Kenya Ku
Initiated
Voluntary: Firm initiated
Recall number
Z-2486-2020

Terminated Class II Medical devices

Recall started
3 January 2020
Terminated
13 August 2021
Reason
Some sutures in the lot may exhibit a small section of braid fraying, which can lead to reduced tensile strength at that point.
Distribution
Domestic distribution nationwide. US and Puerto Rico only.
Initiated
Voluntary: Firm initiated
Recall number
Z-0989-2020

Terminated Class II Medical devices

Recall started
28 August 2019
Terminated
11 February 2021
Reason
The product do not meet certain internal testing specifications.
Distribution
International: Australia, Hong King, India, Japan, Korea, Singapore, Thailand
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1334-2020, Z-1335-2020, Z-1336-2020

Terminated Class II Medical devices

Recall started
13 August 2019
Terminated
11 February 2021
Reason
The product does not meet certain internal strength testing specifications.
Distribution
States: IN MD AR IL GA TN NJ PA NC CA OK CO KS UT KY WA OH OR OUS: Australia Hong King India Japan Korea Singapore Thailand United States
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0048-2020, Z-0047-2020

Terminated Class II Medical devices

Recall started
19 April 2019
Terminated
5 February 2021
Reason
Possibility that a suture raw material containing high endotoxin levels was used in the manufacturing of this lot
Distribution
Worldwide - US Nationwide Distribution - IA, NJ, RI, TX, WV Foreign: Canada
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1841-2019, Z-1842-2019

Terminated Class II Medical devices

Recall started
12 March 2019
Terminated
17 July 2020
Reason
VICRYL Suture, VICRYL Plus Suture, PDS Suture- may contain product that does not match the sales unit carton and foil label.
Distribution
Worldwide distributions - US Nationwide and countries of AUSTRALIA BELGIUM CANADA CHINA ISRAEL JAPAN PUERTO RICO SINGAPORE
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1136-2019, Z-1135-2019, Z-1134-2019

Terminated Class II Medical devices

Recall started
18 December 2018
Terminated
29 May 2020
Reason
The sutures inside the packaging are not the same size or type as indicated on the label.
Distribution
The products were distributed to the following US states: AR, CA, FL, GA, KS, LA, MO, MS, ND, NE, NJ, NY, OH, PA, and WI.
Initiated
Voluntary: Firm initiated
Recall number
Z-0747-2019

Terminated Class II Medical devices

Recall started
8 June 2018
Terminated
25 September 2020
Reason
Products do not meet a tensile strength specification
Distribution
Japan
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2533-2018, Z-2534-2018, Z-2532-2018

Terminated Class II Medical devices

Recall started
7 May 2018
Terminated
8 June 2020
Reason
The needle associated with certain suture product codes has the potential to perforate the packaging due to a packaging design issue. A perforated package has the potential to cause a soft tissue injury to a healthcare professional handling the subject product or a patient infection due to a breach of sterility within the packaging if it goes unnoticed by the healthcare professional.
Distribution
Accounts in AZ, CA, CO, FL, GA, IL, IN, KS, MA, MN, MO, NC, NE, NJ, NY, OH, OR, SC, SD, TN, TX, VA, WA, WI, WV.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2448-2018, Z-2464-2018, Z-2465-2018, Z-2459-2018, Z-2455-2018, Z-2458-2018, Z-2463-2018, Z-2453-2018, Z-2450-2018, Z-2451-2018, Z-2462-2018, Z-2457-2018 and 7 more

Terminated Class II Medical devices

Recall started
7 March 2018
Terminated
30 April 2019
Reason
Products labeled for distribution outside the United States may have been distributed to customers in the United States.
Distribution
The device was distributed to the following states: AK, AL, AR, AZ, CA, CO, FL, GA, IA, IL MD, MI, MN, MO, NC, ND, NJ, NV, NY, OH, PA, TN, TX, WA, WI, and WV. The device was also distributed to Puerto Rico.
Initiated
Voluntary: Firm initiated
Recall number
Z-1534-2018

Terminated Class II Medical devices

Recall started
10 January 2018
Terminated
11 April 2019
Reason
Three (3) lots were distributed with the incorrect expiry date on the label. The lots were labeled with an expiry date of five (5) years from date of manufacture instead of the correct expiry date of two (2) years from date of manufacture.
Distribution
Distribution US Nationwide, Canada, India, and Colombia.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1338-2018, Z-1340-2018, Z-1339-2018

Terminated Class II Medical devices

Recall started
27 November 2017
Terminated
30 April 2019
Reason
Ethicon discovered that specific lots of DERMABOND"PRINEO"System may not dry within the specified time after proper application, and thus may fall off.
Distribution
US Nationwide Distribution
Initiated
Voluntary: Firm initiated
Recall number
Z-1147-2018

Terminated Class II Medical devices

Recall started
17 August 2017
Terminated
8 May 2019
Reason
On August 17, 2017, Ethicon had decided to initiate a medical device recall (removal) of the twenty-five (25) lots of VICRYL RAPIDETM (Polyglactin 910) Suture that were distributed to the field. During finished goods testing, a suture damage was identified on a lot of VICRYL RAPIDETM Suture. During the course of the subsequent investigation, the suture also failed knot tensile strength testing.
Distribution
Nationally
Initiated
Voluntary: Firm initiated
Recall number
Z-0693-2018

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.