Recall Record

Marietta, GA, United States

Erbe USA Inc recalls

2 recalls recorded since 2018, most recently on 26 June 2026. 2 are listed as ongoing. Status and classification are as published by the FDA.

Recalls2
Ongoing (FDA)2
Terminated or completed0
Class I (FDA)1

Ongoing Class II Medical devices

Recall started
26 June 2026
Reason
A strong pull on the high-pressure transfer hose of cryosurgical unit may lead to pneumatic short circuit between he gas inlet and the connector resulting in CO2 leaking into the return path, leading to excessive pressure causing the cryoprobe to rupture.
Distribution
US Nationwide distribution including in the states and territories of Alabama, Alaska, Arizona, Arkansas, California, Colorado, Connecticut, Delaware, District of Columbia, Florida, Georgia, Hawaii, Idaho, Illinois, Indiana, Iowa, Kansas, Kentucky, Louisiana, Maine, Maryland, Massachusetts, Michig
Initiated
Voluntary: Firm initiated
Recall number
Z-2938-2026

Ongoing Class I Medical devices

Recall started
12 February 2026
Reason
Probes may rupture/burst during activation
Distribution
US Nationwide distribution, including Puerto Rico.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1566-2026, Z-1568-2026, Z-1567-2026

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.