Epitope Diagnostics, Inc. recalls

1 recall recorded since 2018, most recently on 1 October 2020. 1 is listed as ongoing. Status and classification are as published by the FDA.

Recalls1
Ongoing (FDA)1
Terminated or completed0
Class I (FDA)0

Ongoing Class II Medical devices

Recall started
1 October 2020
Reason
The use of deionized water rather than distilled water during the manufacturing process, may have resulted in accelerated decreased OD signals for the Positive Control value.
Distribution
U.S Distribution.: NH, CO, OR, CA, FL, NY, MD, NJ, IL, NC, MI, TX, PA, MA, OH, NM, VA, SC, HI, and TN.
Initiated
Voluntary: Firm initiated
Recall number
Z-0426-2021