San Diego, CA, United States
Epitope Diagnostics, Inc. recalls
1 recall recorded since 2018, most recently on 1 October 2020. 1 is listed as ongoing. Status and classification are as published by the FDA.
Recalls1
Ongoing (FDA)1
Terminated or completed0
Class I (FDA)0
Ongoing Class II Medical devices
- Recall started
- 1 October 2020
- Reason
- The use of deionized water rather than distilled water during the manufacturing process, may have resulted in accelerated decreased OD signals for the Positive Control value.
- Distribution
- U.S Distribution.: NH, CO, OR, CA, FL, NY, MD, NJ, IL, NC, MI, TX, PA, MA, OH, NM, VA, SC, HI, and TN.
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-0426-2021
- EDI Novel Coronavirus COVID-19 IgG ELISA Kit, REF: KT-1032 Per label: This kit is intended for qualitative detection of human anti-COVID-19 IgG antibody in human serum. This kit is to be used as an aid for the detect