Recall Record

Golden, CO, United States

Epilog Laser Corp. recalls

1 recall recorded since 2018, most recently on 8 August 2024. 1 is listed as ongoing. Status and classification are as published by the FDA.

Recalls1
Ongoing (FDA)1
Terminated or completed0
Class I (FDA)0

Ongoing Class II Medical devices

Recall started
8 August 2024
Reason
a. A small number of Fusion Pro 24 units were released with an incorrectly placed Interlock device. As a result, the system can be operated with the front-loading door in an open position. b. Fails to comply with Class 2 designation, indicated in Part 6 of Laser Product Report. An incorrectly placed Interlock device can allow Class 4 laser radiation fields during operation.
Distribution
US Nationwide Distribution
Initiated
FDA Mandated
Recall number
Z-0524-2025

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.