Irvine, CA, United States
Endologix recalls
2 recalls recorded since 2018, most recently on 6 May 2020. None are listed as ongoing. Status and classification are as published by the FDA.
Recalls2
Ongoing (FDA)0
Terminated or completed2
Class I (FDA)2
Terminated Class I Medical devices
- Recall started
- 6 May 2020
- Terminated
- 3 May 2024
- Reason
- A material weakness adjacent to the polymer fill channel may become compromised during pressurization with liquid polymer. Polymer leaks may occur during implantation of abdominal stent graft system. Clinical events related to polymer leaks may be systemic and/or aneurysm related.
- Distribution
- Worldwide DIstribution: U.S (nationwide).: PA, AZ, NY, FL, HI, CA, OR, IL, MS, AR, WI, KY, TX, TN, MO, NJ, MA, MT, VA, MI, CO, WV, NC, NH, NV, CT, ME, WY, OH, OK, IN, UT, AL, WA, GA, SC, KS, LA, NM, MD, DC, NE, SD, RI, IA, VT, ND. and Countries of:: Argentina, Australia, Chile, Hong Kong, Japan, Mal
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-2263-2020
- Ovation iX Abdominal Stent Graft Systems, REF: TV-AB2080-I, TV-AB2080-J, TV-AB2380-I, TV-AB2380-J, TV-AB2680-I, TV-AB2680-J, TV-AB2980-I, TV-AB2980-J, TV-AB3480-I, TV-AB3480-J
Terminated Class I Medical devices
- Recall started
- 31 July 2018
- Terminated
- 9 April 2024
- Reason
- Endologix is notifying physicians of patient tailored surveillance recommendations as well as general warnings/precautions for interventions to or through an existing AFX device. This is a follow-up notification to the one that was sent out in December 2016. The updates are related to the observation of Type III endoleaks.
- Distribution
- U.S.: PA, VA, CA, AZ, IN, GA, FL, IL, OK, MN, OH, KY, MS, NC, MI, AL, CT, LA, SC, TN, NY, WI, MD, WV, AR, NV, NJ, TX, CO, NE, NM, MA, IA, AK, OR, ME, SD, MO, MT, KS, DE, WA, UT, NH, ND, DC, RI, HI, WY, ID, VT, PR; Foreign (OUS): Canada, Argentina, Australia, Brazil, CHILE , Hong Kong, Japan, So
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-0008-2019, Z-0006-2019, Z-0007-2019, Z-0009-2019
- ENDOLOGIX AFX/AFX2, Proximal (Aortic) Extension Endograft, Rx Only, Sterile EO, AFX Proximal Accessory, Infrarenal Model #/REF: A22-22/C55, A22-22/C75, A22-22/C95, A25-25/C55, A25-25/C75
- ENDOLOGIX AFX Endovascular AAA System, Stent Graft System, Rx Only, Sterile EO, REF: BA22-120/I20-40, BA22-100/I20-40, BA22-80/I20-40, BA22-60/I20-40, BA22-40/I20-40
- ENDOLOGIX AFX2 Bifurcated Endograft System, Stent Graft System, Rx Only, Sterile EO, REF: BEA22-120/I20-40, BEA22-100/I20-40, BEA22-80/I20-40, BEA22-60/I20-40, BEA22-40/I20-40, BEA22-1
- ENDOLOGIX AFX/AFX2, Limb (Iliac) Extension Endograft, Rx Only, Sterile EO, AFX Limb Accessory Model #/REF: I16-16/C55, I16-16/C55F, I16-16/C88, I20-13/C70F, I20-13/C88F, I16-16/C55, I16-16/C5
A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.