Mawson Lakes, N/A, Australia

Ellex Medical Pty Ltd. recalls

1 recall recorded since 2018, most recently on 16 March 2023. 1 is listed as ongoing. Status and classification are as published by the FDA.

Recalls1
Ongoing (FDA)1
Terminated or completed0
Class I (FDA)0

Ongoing Class II Medical devices

Recall started
16 March 2023
Reason
When using a Laser-Indirect-Ophthalmoscope (LIO) as the delivery device, a scattered laser beam inside the laser cavity system may unintentionally exit through the Slit Lamp output port (SDS) and get emitted from the Objective Lens when the laser is fired.
Distribution
US nationwide/ Worldwide Distribution
Initiated
FDA Mandated
Recall number
Z-1971-2023

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.