Recall Record

Stockholm, N/A, Sweden

ELEKTA SOLUTIONS AB recalls

2 recalls recorded since 2018, most recently on 17 September 2025. 1 is listed as ongoing. Status and classification are as published by the FDA.

Recalls2
Ongoing (FDA)1
Terminated or completed1
Class I (FDA)0

Ongoing Class II Medical devices

Recall started
17 September 2025
Reason
When appending a care plan that contains one or more wave medication orders, the occurrence and frequency of the appended orders may not match the intended schedule.
Distribution
Worldwide - U.S. Nationwide distribution in the state of NM and the countries of Canada, Malta, and South Africa.
Initiated
Voluntary: Firm initiated
Recall number
Z-0503-2026

Terminated Class II Medical devices

Recall started
28 September 2021
Terminated
14 June 2024
Reason
There is a risk that the Diode D1 in the Modulator can fail in a destructive manner when removed from the machine, and that the protective tube surrounding the diode may fail to contain the ejected matter.
Distribution
Distribution was to ND and NC. There was no military/government distribution.
Initiated
Voluntary: Firm initiated
Recall number
Z-0115-2022

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.