Elekta Instrument AB recalls

5 recalls recorded since 2018, most recently on 25 September 2024. 4 are listed as ongoing. Status and classification are as published by the FDA.

Recalls5
Ongoing (FDA)4
Terminated or completed1
Class I (FDA)2

Ongoing Class I Medical devices

Recall started
25 September 2024
Reason
Elekta has become aware that Disposable Biopsy Needles can contain some microscopic debris on the inside of the biopsy needle. The material in the debris is stainless steel, same material as the biopsy needle. No debris has been found on the outer parts of the biopsy needles.
Distribution
US: AZ, TX, FL, CA, NY, GA, VA, NJ, MO, PA, WA,
Initiated
Voluntary: Firm initiated
Recall number
Z-0227-2025

Ongoing Class I Medical devices

Recall started
15 March 2024
Reason
Elekta has become aware that Disposable Biopsy Needles (911933) from one batch (837838839) can contain some microscopic debris on the inside of the biopsy needle. The material in the debris is stainless steel, same material as the biopsy needle. No debris has been found on the outer parts of the biopsy needles. The sterility of the biopsy needles has not been affected. This issue has been reported from one site.
Distribution
US and Australia, Austria, Belgium, France, Germany, Greece, Italy, Netherlands, Spain, Sweden, Switzerland
Initiated
Voluntary: Firm initiated
Recall number
Z-1397-2024

Ongoing Class II Medical devices

Recall started
1 March 2023
Reason
The incorrect IFU was distributed with the devices.
Distribution
US Nationwide distribution in the states of AZ, CA, CO, FL, GA, ID, IL, KS, MA, MD, MI, MN, MO, MS, NE, NH, NJ, NM, NV, NY, OH, OK, PA, SC, TN, TX, UT, WA, WI, and WV. There was also government distribution.
Initiated
Voluntary: Firm initiated
Recall number
Z-1309-2023

Terminated Class II Medical devices

Recall started
10 June 2020
Terminated
27 August 2024
Reason
The locking mechanism at the interface of the Instrument Carrier and the Leksell Vantage Arc may not function properly.
Distribution
MA, IL, NH, NC, UT, OH, FL, MN, TX, NY, LA, AZ, MT, WV, Austria, Croatia, Estonia, Finland, France, Germany, Hong Kong, India, Italy, Japan, Kuwait, Lithuania, Netherlands, Norway, Portugal, South Africa, Spain, Sweden, Switzerland, United Kingdom
Initiated
Voluntary: Firm initiated
Recall number
Z-2442-2020

Ongoing Class II Medical devices

Recall started
27 June 2018
Reason
The margin tool in Leksell GammaPlan 11.1 systematically overestimates margin in certain areas of the volume.
Distribution
Worldwide Distribution - US Nationwide distribution to GA, MD, NJ, PA, PR, WA. International distribution to Australia, Belgium, France, Germany, Hong Kong, Italy, Japan, Korea, Democratic People's Republic of, Netherlands, Romania, Russian Federation, Spain, Turkey, United Kingdom.
Initiated
Voluntary: Firm initiated
Recall number
Z-1412-2019

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.