DRG Instruments GmbH recalls

3 recalls recorded since 2018, most recently on 6 June 2019. None are listed as ongoing. Status and classification are as published by the FDA.

Recalls3
Ongoing (FDA)0
Terminated or completed3
Class I (FDA)0

Terminated Class II Medical devices

Recall started
6 June 2019
Terminated
11 May 2020
Reason
Salivary Progesterone assay may observe a higher percentage of samples with low results. Saliva samples in the low measuring range of the kit {< 20 pg/ml) were assayed and producing a significantly lower result.
Distribution
Distribution to Canada
Initiated
Voluntary: Firm initiated
Recall number
Z-2410-2019

Terminated Class II Medical devices

Recall started
21 May 2019
Terminated
29 May 2020
Reason
Salivary assay may observe a higher percentage of samples with 0 pg/ml testosterone or results which are too low. Saliva samples in the low measuring range of the kit(< 50 pg/ml) were assayed significantly lower
Distribution
Worldwide Distribution - US Nationwide CA, MI Foreign: Canada, Brazil, China, Poland
Initiated
Voluntary: Firm initiated
Recall number
Z-2159-2019

Terminated Class II Medical devices

Recall started
14 July 2016
Terminated
4 August 2020
Reason
Lower recovery of salivary estradiol levels due to poor differentiation between kit Standard 0 and Standard 1.
Distribution
NJ; Germany, Belgium, Spain & Germany
Initiated
Voluntary: Firm initiated
Recall number
Z-2836-2018

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.