Draeger Medical Systems, Inc. recalls

6 recalls recorded since 2018, most recently on 14 January 2025. 5 are listed as ongoing. Status and classification are as published by the FDA.

Recalls6
Ongoing (FDA)5
Terminated or completed1
Class I (FDA)1

Ongoing Class II Medical devices

Recall started
14 January 2025
Reason
Edifier speakers (shipped in or after April 2023) may not annunciate when configured in analog mode which can lead to delayed treatment to the patient.
Distribution
Worldwide - US Nationwide distribution in the states of AL, AS, AZ, CA, CO, FL, GA, IA, ID, IL, IN, KS, MA, NE, NY, OK, PA, SD, TN, TX, VA, WA and the countries of Argentina, Australia, Austria, Belgium, Botswana, Brazil, Canada, Chile, China, Colombia, Croatia, Cyprus, Czech Republic, Egypt, Finlan
Initiated
Voluntary: Firm initiated
Recall number
Z-1066-2025

Ongoing Class II Medical devices

Recall started
13 May 2024
Reason
Device is not in full compliance to Type CF requirements of IEC 60601-1 and IEC 60601-2-34 standards. Therefore, there is a potential risk of electric shock. Health consequences may include electrical shock, burns, and cardiac arrhythmia.
Distribution
US Nationwide distribution. International distribution to Afghanistan, Argentina, Armenia, Australia, Austria, Belarus, Belgium, Bolivia, Botswana, Brazil, Brunei Dar-es-S, Bulgaria, Canada, Chile, China, Columbia, Costa Rica, Croatia, Czech Republic, Egypt, Estonia, Finland, France, French Guiana,
Initiated
Voluntary: Firm initiated
Recall number
Z-2189-2024

Ongoing Class II Medical devices

Recall started
16 May 2023
Reason
The Resuscitaire Infant Radiant Warmer with the optional scale could display inaccurate weight values.
Distribution
US Nationwide distribution including in the states of AL, AR, AZ, CA, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MD, MI, MN, MO MS, MT, NC, NE, NH, NJ, NV, NY, OH, OR, PA, PR, SD, TN, TX, VT, WA, and WI.
Initiated
Voluntary: Firm initiated
Recall number
Z-1992-2023

Ongoing Class II Medical devices

Recall started
5 April 2023
Reason
The software on the Infinity CentralStation drops peaks on narrow waveforms causing the Infinity M300 to fail to meet Frequency and Impulse Response requirements. In this condition, ECG waveforms with unusually narrow and/or high frequency QRS may be displayed or printed with QRS amplitudes intermittently lower than actual.
Distribution
Worldwide - US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1472-2023, Z-1471-2023

Ongoing Class II Medical devices

Recall started
20 May 2022
Reason
The firm has initiated this recall to replace component on devices distributed containing a component within the oxygen supply system that is not rated to the working pressure of the system.
Distribution
No US distribution. Foreign distribution to BY, CL, DE, ES, GB, HU, IN, IT, KW, MX, SA, SV, TW.
Initiated
Voluntary: Firm initiated
Recall number
Z-1365-2022

Terminated Class I Medical devices

Recall started
15 May 2018
Terminated
8 February 2023
Reason
Users have misinterpreted the display for out of range measurement indicated by the blinking" ---" to mean a zero measurement.
Distribution
Worldwide Distribution: US (nationwide) including states of: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY; and countries of: Algeria, Argentin
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2046-2018, Z-2047-2018

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.