Draeger Medical, Inc. recalls

17 recalls recorded since 2018, most recently on 15 October 2024. 8 are listed as ongoing. Status and classification are as published by the FDA.

Recalls17
Ongoing (FDA)8
Terminated or completed9
Class I (FDA)10

Ongoing Class I Medical devices

Recall started
15 October 2024
Reason
The Atlan anesthesia workstation either indicated a failure of the piston ventilator before use or suffered a failure of the mechanical ventilation during use.
Distribution
No Atlan A300 or A300XL devices were imported into the US. Devices were direct shipped from Draegerwerk Ag and Co. KGaA to the following countries: Argentina, Australia, Austria, Azerbaijan, Bahrain, Belgium, Brazil, Bulgaria, Cambodia, Canada, Chili, China, Colombia, Croatia, Czech Republic, Egypt,
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0668-2025, Z-0667-2025

Ongoing Class I Medical devices

Recall started
15 October 2024
Reason
The Atlan anesthesia workstation either indicated a failure of the piston ventilator before use or suffered a failure of the mechanical ventilation during use.
Distribution
US Nationwide distribution in the states of AL, AZ, CA, FL, GA, IA, IL, KY, MA, ME, ND, NE, NH, NJ, NY, OH, PA, TN, TX, WI, WY. International distribution: Argentina, Australia, Austria, Azerbaijan, Bahrain, Belgium, Brazil, Bulgaria, Cambodia, Canada, Chili, China, Columbia, Croatia, Czech Republi
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0305-2025, Z-0304-2025

Ongoing Class I Medical devices

Recall started
20 March 2024
Reason
Unexpected shutdown while operating on battery power.
Distribution
Domestic: Nationwide Distribution; Foreign: Angola, Argentina, Australia, Austria, Bahrain, Belarus, Belgium, Bolivia, Botswana, Brazil, Brunei Dar-se- S, Bulgaria, Canada, Chili, China, Columbia, Costa Rica, Croatia, Czech Republic, Denmark, Ecuador, Egypt, Finland, France, French Polynesia, French
Initiated
Voluntary: Firm initiated
Recall number
Z-1512-2024

Ongoing Class I Medical devices

Recall started
12 July 2023
Reason
Presence of polyether polyurethane (PE-PUR) and emission of 1,3-Dichloropropan-2-ol that exceed the acceptable uptake level during continuous use (>30 days) in pediatric patients.
Distribution
US Nationwide Distribution.
Initiated
Voluntary: Firm initiated
Recall number
Z-2248-2023

Ongoing Class I Medical devices

Recall started
12 June 2023
Reason
Draeger has become aware of instances where the device stopped working when in use when transitioning from battery to AC operation.
Distribution
Distributed Nationwide. OUS Distribution: Afghanistan, Algeria, Andorra, Angola, Antigua/Barbados, Argentina, Australia, Austria, Azerbaijan, Bahrain, Belarus, Belgium, Bolivia, Brunei Dar-es-S, Bulgaria, Burkina-Faso, Cambodia, Canada, Cayman Islands, Chile, China, Christmas Island, Columbia, Congo
Initiated
Voluntary: Firm initiated
Recall number
Z-2018-2023

Terminated Class I Medical devices

Recall started
17 April 2023
Terminated
3 September 2024
Reason
Glued connections of the breathing circuit can become loose before or during ventilation resulting in partial or complete detachment of pieces.
Distribution
Domestic: US Nationwide Distribution.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1440-2023, Z-1459-2023, Z-1439-2023, Z-1444-2023, Z-1458-2023, Z-1449-2023, Z-1442-2023, Z-1447-2023, Z-1454-2023, Z-1446-2023, Z-1441-2023, Z-1453-2023 and 10 more

Ongoing Class I Medical devices

Recall started
23 January 2023
Reason
The magnets used in the swivel frame/tube set and the headgear clips might impact medical devices (implanted as well as not implanted) and metallic implants.
Distribution
US Nationwide Distribution
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1138-2023, Z-1137-2023

Ongoing Class I Medical devices

Recall started
16 May 2022
Reason
Possible occlusion of filters due to manufacturing error.
Distribution
Worldwide distribution - US Nationwide distribution in the states of AZ, CA, CO, CT, FL, GA, IL, IN, KY, LA, MA, MD, MI, MN, NC, ND, NV, NY, OH, PA, TN, TX, WA, and WV. The countries of Austria, Belgium, Bulgaria, Canada, Columbia, Croatia, Czech Republic, Denmark, France, Germany, India, Indonesia,
Initiated
Voluntary: Firm initiated
Recall number
Z-1211-2022

Terminated Class II Medical devices

Recall started
14 February 2022
Terminated
7 January 2025
Reason
Cybersecurity Improvement Action-not equipped against potential cyber security threats
Distribution
US Nationwide
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0870-2022, Z-0871-2022

Terminated Class II Medical devices

Recall started
16 December 2021
Terminated
17 April 2023
Reason
26 Fabius MRI Anesthesia machines were distributed with internal hoses routing Oxygen to the Oxygen Flush and Oxygen Flow Control interchanged. If the hospital O2 Central supply pressure is not stable, the O2 fresh gas flow could deviate from the adjusted value. If the deviation is not noticed, insufficient oxygenation of the patient may occur.
Distribution
US Nationwide distribution in the states of NY, PA, and WI.
Initiated
Voluntary: Firm initiated
Recall number
Z-0501-2022

Ongoing Class II Medical devices

Recall started
10 September 2021
Reason
Software 2.51 and Lower with Installed CO2 Measurement Option, may result in Restart of the Ventilator, brief cessation of ventilation can occur and loss of PEEP for approx. 8 seconds
Distribution
US Nationwide
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0109-2022, Z-0108-2022

Terminated Class II Medical devices

Recall started
16 April 2021
Terminated
6 February 2024
Reason
Three separate and unrelated problems attributed to the software used in the Evita V600, Evita V800, Babylog VN600 and Babylog VN800: 1. Restart of ventilation unit. 2. Incorrect FiO2 high and FiO2 low alarms. 3. SmartCare/PS (SC/PS) suspends weaning
Distribution
US Nationwide distribution in the states of PA, TX, MD, FL, TX, WA, MS, NY.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1692-2021, Z-1690-2021, Z-1691-2021, Z-1693-2021

Terminated Class II Medical devices

Recall started
29 January 2021
Terminated
1 November 2021
Reason
Y-piece may become detached from the ventilation hose either before or during the ventilation process, a leak will be caused that can hinder ventilation of the patient.
Distribution
US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1180-2021, Z-1179-2021, Z-1177-2021, Z-1178-2021

Terminated Class I Medical devices

Recall started
21 December 2018
Terminated
16 April 2019
Reason
Incorrectly pre-assembled breathing circuit has potential for incorrect connection/short circuit of the breathing circuit
Distribution
US Distribution to states of: AL, CA, CO, CT, FL, KY, MI, NJ, OH, PA, SC & TX.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0744-2019, Z-0740-2019, Z-0737-2019, Z-0742-2019, Z-0739-2019, Z-0736-2019, Z-0738-2019, Z-0743-2019, Z-0741-2019

Terminated Class II Medical devices

Recall started
24 May 2018
Terminated
12 May 2020
Reason
The Draeger anesthesia device may be able to dose 100% N2O. In the event of a fault, the S-ORC module would not prevent setting an N2O flow that would result in a hypoxic mixture from being dosed to the patient. Potential adverse outcomes include death of the patient.
Distribution
US Nationwide Distribution - Healthcare facilities in AK, AL, AZ, CA, CO, FL, GA, IA, IL, IN, MA, MD, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, TN, TX, UT, WA, WI, and WV.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2254-2018, Z-2251-2018, Z-2250-2018, Z-2252-2018, Z-2253-2018

Terminated Class II Medical devices

Recall started
26 April 2018
Terminated
7 September 2018
Reason
The valve assembly material may become corroded due to insufficient cleaning by a supplier. There is a risk of the breathing gas becoming contaminated in the affected devices.
Distribution
Distributed to the states of CT, IL, MN, NV, SD, and WA.
Initiated
Voluntary: Firm initiated
Recall number
Z-1954-2018

Terminated Class I Medical devices

Recall started
1 March 2018
Terminated
16 April 2019
Reason
The automatic ventilation may fail if the position detection of the motor is disturbed. If the ventilator cover is damaged, manual ventilation may also fail.
Distribution
US Nationwide Distribution to accounts in 21 states: AZ, CA, CO, ID, IL, KS, MA, MI, MS, NC, NY, OH, OK, PA, RI, TN, TX, UT, VA, WA, WI; and PR.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1481-2018, Z-1482-2018, Z-1483-2018, Z-1484-2018

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.