Draegar Medical Systems, Inc. recalls

8 recalls recorded since 2018, most recently on 4 March 2022. 5 are listed as ongoing. Status and classification are as published by the FDA.

Recalls8
Ongoing (FDA)5
Terminated or completed3
Class I (FDA)0

Ongoing Class II Medical devices

Recall started
4 March 2022
Reason
Software issue resulting in temporary loss of central monitoring functionality.
Distribution
Worldwide distribution - US Nationwide distribution in the states of AL, AR, AZ, CA, CO, FL, IA, ID, IL, IN, LA, MN, MS, NC, NE, NH, NY, OH, PA, SD, TN, TX, WA, WI, and 58 additional foreign countries.
Initiated
Voluntary: Firm initiated
Recall number
Z-1012-2022

Ongoing Class II Medical devices

Recall started
23 December 2020
Reason
The Infinity M540 patient monitor may randomly reboot due to an error to correctly transmit and read the header data of files in the memory of the device. Under this situation, the device will try to reboot to mitigate the error. The device will be available again for use within 30 seconds. If this error continues and the M540 reboots three times in a time span of 10 minutes, it will enter a fail-state. A fail-state will annunciate itself with a continuous sound to alert the user. The M540 will reset to factory default and the user can manually configure and readmit the patient to continue patient monitoring.
Distribution
Worldwide distribution - US Nationwide distribution including in the states of AL, AZ, CA, CO, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, ME, MI, MN, MO, MS, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, VA, WA, WI, WY and the countries of Argentina, Australia, Austria, Bahrain, Belgium,
Initiated
Voluntary: Firm initiated
Recall number
Z-1104-2021

Ongoing Class II Medical devices

Recall started
9 October 2019
Reason
The devices have potential cybersecurity vulnerabilities, which can include Distributed Denial of Service (DDoS), Spoofing, and Tampering.
Distribution
US nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall number
Z-0372-2020

Ongoing Class II Medical devices

Recall started
13 September 2019
Reason
Cybersecurity vulnerabilities may cause device to reboot, lose alarm functionality, and/or lose communication with cockpit and/or the Infinity Network.
Distribution
US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall number
Z-0258-2020

Ongoing Class II Medical devices

Recall started
8 February 2019
Reason
The firm is recalling their Delta family of patient monitors software due to cybersecurity vulnerabilities, which may cause the device to reboot, lose functionality, and/or lose communication.
Distribution
US Nationwide distribution including Puerto Rico.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0974-2019, Z-0975-2019

Terminated Class II Medical devices

Recall started
28 March 2018
Terminated
21 April 2020
Reason
Software anomaly resulting in the loss of patient settings and stored patient data.
Distribution
Worldwide Distribution - USA (nationwide) and to the countries of : Angola, Argentina, Australia Austria Azerbaijan Bahrain Belgium Bolivia Brazil Brunei Dar-es-S Canada Chile China Colombia Costa Rica Czech Republic Denmark Egypt Finland France Germany Ghana Greece Hong Kong H
Initiated
Voluntary: Firm initiated
Recall number
Z-1813-2018

Terminated Class II Medical devices

Recall started
21 February 2018
Terminated
20 April 2020
Reason
The invasive pressure visual and audible alarms may not activate.
Distribution
Worldwide Distribution - USA (nationwide) and the following countries:Angola, Argentina, Australia, Austria, Azerbaijan, Bahrain, Belgium, Bolivia, Botswana, Brazil, Brunei Dar-es-S, Bulgaria, Canada, Chile, China, Columbia, Costa Rica, Czech Republic, Denmark, Egypt, Finland, France, Germany, Gre
Initiated
Voluntary: Firm initiated
Recall number
Z-1382-2018

Terminated Class II Medical devices

Recall started
4 October 2017
Terminated
15 June 2018
Reason
Dual Hemo MCable Pods with Revision Index (RI) 15 and 16, which are used with the Dr¿ger Infinity Acute Care System, may permit liquid ingress inconsistent with their IPX4 rating, which could potentially result in an incorrect measured value of invasive blood pressure.
Distribution
Nationally
Initiated
Voluntary: Firm initiated
Recall number
Z-0942-2018

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.