Diagnostic Hybrids, Inc. recalls

4 recalls recorded since 2018, most recently on 4 April 2017. None are listed as ongoing. Status and classification are as published by the FDA.

Recalls4
Ongoing (FDA)0
Terminated or completed4
Class I (FDA)0

Terminated Class II Medical devices

Recall started
4 April 2017
Terminated
28 August 2020
Reason
There is a potential for the slope of the standard curve to fail to meet the assay validation requirements for the assay.
Distribution
The products were distributed to the following US states: NY, TX, CO, UT. The products were distributed to the following foreign countries: Switzerland.
Initiated
Voluntary: Firm initiated
Recall number
Z-1496-2018

Terminated Class III Medical devices

Recall started
26 October 2016
Terminated
28 August 2020
Reason
There may be a loss of separation between Calibrators A and B which may impact the assay calibration at the lower end of the assay.
Distribution
The product was distributed to the following countries: Singapore, Switzerland
Initiated
Voluntary: Firm initiated
Recall number
Z-1520-2018

Terminated Class II Medical devices

Recall started
14 July 2016
Terminated
28 August 2020
Reason
There is a possibility of low volume and/or leaking standard bottles.
Distribution
The products were distributed to the following US states: CA, CT, GA, IL, KY, MA, MD, MO, PA, TX, WA, and WI. The products were distributed to the following foreign countries: Australia, Brazil, Canada, China, Costa Rica, India, Japan, Mexico, The Philippines, Russia, South Korea, South Africa, Sw
Initiated
Voluntary: Firm initiated
Recall number
Z-1797-2018

Terminated Class III Medical devices

Recall started
3 November 2015
Terminated
28 August 2020
Reason
A component of the kit was found to contain bacterial contamination.
Distribution
Worldwide Distribution - USA (nationwide) to the states of : CO, CT, IA, IL, MN, MO, NC, NY, OH TX, and UT., and to the countries of : Switzerland, India, Hong Kong, and Japan.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1499-2018, Z-1498-2018, Z-1497-2018