Recall Record

Somerset, PA, United States

DeVilbiss Healthcare LLC recalls

3 recalls recorded since 2018, most recently on 9 March 2023. 3 are listed as ongoing. Status and classification are as published by the FDA.

Recalls3
Ongoing (FDA)3
Terminated or completed0
Class I (FDA)0

Ongoing Class II Medical devices

Recall started
9 March 2023
Reason
The instruction manual is being replaced with an updated version that is in alignment with labeling requirements specified in the IEC-60101-1 standard.
Distribution
Worldwide - US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall number
Z-1558-2023

Ongoing Class II Medical devices

Recall started
18 November 2022
Reason
The firm has received complaints of the device becoming hot. The firm has updated the user manual to include warnings that applied parts in contact with the patient may in some situations exceed 41 degrees Celsius (105.8 degrees Fahrenheit), clarification on expected service life, and the need for maintenance to certain components.
Distribution
US Nationwide. Foreign distribution pending.
Initiated
Voluntary: Firm initiated
Recall number
Z-0724-2023

Ongoing Class II Medical devices

Recall started
19 September 2022
Reason
An unapproved adhesive used to manufacture the accumulator tank sub assembly (a part of the patient gas pathway)
Distribution
US Distribution in states of: CA, FL, OH, MI, MN, NY, PA, and TX; and (Foreign) country of: Canada.
Initiated
Voluntary: Firm initiated
Recall number
Z-0229-2023

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.