DePuy Orthopaedics, Inc. recalls

24 recalls recorded since 2018, most recently on 12 March 2024. 4 are listed as ongoing. Status and classification are as published by the FDA.

Recalls24
Ongoing (FDA)4
Terminated or completed20
Class I (FDA)0

Ongoing Class II Medical devices

Recall started
12 March 2024
Reason
Product incorrectly labelled.
Distribution
Worldwide - US Nationwide distribution in the states of CA, IL, NJ, WI and the country of Canada.
Initiated
Voluntary: Firm initiated
Recall number
Z-1456-2024

Ongoing Class II Medical devices

Recall started
15 February 2023
Reason
The subject product lots are being recalled because they received a higher than specified irradiation dose. This exceeds the validated range for exposure to gamma radiation of these devices and may result in changes to the implant material properties.
Distribution
Worldwide distribution - US Nationwide and the countries of Canada, Poland, Germany, India, Australia, & New Zealand
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1263-2023, Z-1266-2023, Z-1264-2023, Z-1265-2023, Z-1267-2023

Ongoing Class II Medical devices

Recall started
29 June 2022
Reason
Manufactured with a Galvanized High Carbon Steel spring instead of the correct 302 Stainless steel spring. has the potential to oxidize and corrode in the ATTUNE Measured Sizing & Rotation Guide, resulting in surgical delay, debris in surgical field, Osteolysis, Poor Joint Mechanics, Soft Tissue Irritation, Pain
Distribution
Nationwide Foreign: Argentina, Australia, Austria, Bahrain, Belgium, Brazil, Canada, Chile, China, Croatia, Czech Republic, Denmark, Finland, France, Germany, Greece, Hong Kong, India, Ireland, Israel, Italy, Japan, Kuwait, Lithuania, Mauritius, Mexico, Morocco, Netherlands, New Zealand, Norway, P
Initiated
Voluntary: Firm initiated
Recall number
Z-1419-2022

Ongoing Class II Medical devices

Recall started
18 May 2022
Reason
The Cathcart Ball has an inherent +5mm offset which is not currently specified in its Instructions for Use or Surgical Technique Guide. The STG has been updated to reflect the clarification that the +5mm offset should be taken into consideration when using a Cathcart head in a construct for hemi-hip arthroplasty procedures.
Distribution
Worldwide distribution - US Nationwide and the countries of Australia, Austria, Belgium, Canada, Denmark, Finland, Germany, Greece, Guam, Hong Kong, Ireland, Netherlands, New Zealand, Norway, South Africa, Spain, Sweden, Switzerland, United Kingdom.
Initiated
Voluntary: Firm initiated
Recall number
Z-1375-2022

Terminated Class II Medical devices

Recall started
11 March 2022
Terminated
18 April 2024
Reason
System software v1.5.1 has a system software issue related to the Daylight Savings Time (DST) change that can cause a system error, requiring the user to restart the system and potentially cause a delay in treatment
Distribution
Worldwide distribution - US Nationwide and the countries of Israel, Australia & New Zealand.
Initiated
Voluntary: Firm initiated
Recall number
Z-1029-2022

Terminated Class II Medical devices

Recall started
5 January 2022
Terminated
22 December 2023
Reason
May have adhesive residue on the posterior surface of the femoral augment, if undetected prior to implantation, soft tissue irritation and/or infection may occur.
Distribution
US Nationwide distribution in the states of TN, TX.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0631-2022, Z-0632-2022

Terminated Class II Medical devices

Recall started
7 December 2021
Terminated
17 May 2024
Reason
Stems may have an undersized spigot bore diameter therefore separating the proximal body from the stem may not be possible without surgical delay and/or bone damage
Distribution
Worldwide distribution - US Nationwide and the countries of Austria, Australia, Belgium, Canada, Denmark, Estonia, Finland, France, Germany, Greece, Ireland, Israel, Italy, Netherlands, New Zealand, Norway, Poland, South Africa, Sweden, Switzerland, Thailand, United Kingdom.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0487-2022, Z-0490-2022, Z-0486-2022, Z-0488-2022, Z-0485-2022, Z-0489-2022, Z-0483-2022, Z-0484-2022

Terminated Class II Medical devices

Recall started
9 September 2021
Terminated
9 August 2023
Reason
Packaging may include screws that are not intended to be used with the 5mm augments, may cause delay in surgery
Distribution
US Nationwide distribution in the states of CA, LA, MN, MO, NC, NE, OR, PA, WA, WI.
Initiated
Voluntary: Firm initiated
Recall number
Z-0074-2022

Terminated Class II Medical devices

Recall started
10 August 2021
Terminated
17 May 2023
Reason
Software: A coding error associated with the Fast3D Segmentation software. During the scanning process the three images required of the hip, knee and ankle may not be aligned, causing the anatomic landmark locations to potentially result in limb malalignment.
Distribution
Distribution to US states of CA, CO, FL, IL, IN, NY, MI, MN, TX, and WA, Poland and UK.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2398-2021, Z-2395-2021, Z-2397-2021, Z-2396-2021, Z-2400-2021, Z-2399-2021, Z-2394-2021

Terminated Class II Medical devices

Recall started
1 July 2021
Terminated
1 February 2023
Reason
Package labeled as 16mm x 80mm contained a 16mm x 130mm stem instead and result in obtaining a replacement
Distribution
International distribution in the countries of Canada, France.
Initiated
Voluntary: Firm initiated
Recall number
Z-2348-2021

Terminated Class II Medical devices

Recall started
7 May 2021
Terminated
9 March 2023
Reason
Fifty (50) individual units from Lot D20111130 were distributed with the incorrect part number and size etched on the product.
Distribution
Domestic: CA, CT, FL, GA, IA, IN, LA, ME, MN, MO, MT, NC, NE, NY, OH, OK, OR, PA, SC, TN, WI. Foreign: Bermuda
Initiated
Voluntary: Firm initiated
Recall number
Z-1845-2021

Terminated Class II Medical devices

Recall started
22 February 2021
Terminated
11 October 2023
Reason
Globally the company is removing 37 lots affected lots for MBT & ATTUNE (S+) Tibial Bases due to potential damaged packaging. These implant components are utilized in Total Knee Arthroplasty. The affected tibial bases are designed to be used in cemented applications.
Distribution
Worldwide distribution - US Nationwide distribution including in the states of AL, AZ, CA, CO, CT, DC, DE, FL, GA, ID, IL, KS, KY, LA, MA, MD, ME, MI, MN, MO, MT, NC, NE, NH, NJ, NY, OH, OK, OR, PA, SC, TN, TX, VA, VT, WA, WI, WV and the countries of Brazil, Austria, Belgium, Denmark, Estonia, Franc
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1314-2021, Z-1316-2021, Z-1315-2021

Terminated Class II Medical devices

Recall started
8 January 2021
Terminated
17 January 2024
Reason
The taper dimensions may be out of specification due to a production issue during the manufacturing of these lots.
Distribution
Worldwide distribution - US Nationwide distribution in the states of AK, CA, FL, IN, MI, MN, MS, NC, NY, OH, PA, TX, VA and the countries of Canada, China, Germany, Puerto Rico, Switzerland.
Initiated
Voluntary: Firm initiated
Recall number
Z-1103-2021

Terminated Class II Medical devices

Recall started
15 December 2020
Terminated
14 November 2023
Reason
Certain Pinnacle Cup devices may potentially exhibit an oversized "minor diameter", which could lead the Apex HE to thread through the shell of the cup without stopping or to protrude internally as a result of "cross-threading". If cross-threading occurs, physicians will not feel the positive stop of the Apex HE against the cup.
Distribution
Worldwide distribution - US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall number
Z-0802-2021

Terminated Class II Medical devices

Recall started
3 January 2020
Terminated
12 August 2021
Reason
The device will not connect to the Specialist 2 Locating Outrigger which renders them unusable in surgery.
Distribution
Domestic distribution to IA, MA, NY, OH, and PA. Foreign distribution to Canada, China, Hungary, United Kingdom, and Singapore. ***Updated 7/9/20*** Expanded notice includes 6 new countries: Belgium, France, Germany, Maldives, New Zealand, and UAE
Initiated
Voluntary: Firm initiated
Recall number
Z-1169-2020

Terminated Class II Medical devices

Recall started
21 September 2018
Terminated
22 April 2020
Reason
There is the potential for debris/material to be found behind the O-rings in the neck trials.
Distribution
Worldwide Distribution - US Nationwide Distributed to accounts in AL AZ CA CO FL GA HI IA IL IN KY LA MA MD ME MI MN MS NC ND NH NJ NV NY OH OR PA SC TN TX UT VA WA, and WI. International distribution worldwide.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1127-2019, Z-1128-2019, Z-1129-2019

Terminated Class II Medical devices

Recall started
18 July 2018
Terminated
17 April 2019
Reason
This unit may be missing the screw/collet Assembly
Distribution
The devices were not distributed in the United States. The devices were distributed to the following foreign countries: Great Britain.
Initiated
Voluntary: Firm initiated
Recall number
Z-2949-2018

Terminated Class II Medical devices

Recall started
21 June 2018
Terminated
6 April 2020
Reason
Two lots of femoral stems were labeled as Size 11. However, they are actually Size 12. Potential adverse effects include poor joint mechanics, improper fit, loosening, and additional surgical procedure to correct. It may also result in an intraoperative fracture which could represent a surgical delay and additional procedures.
Distribution
Accounts in AL, CA, CO, GA< IL, LA, ME, NC, NH, NY, OR, TX, UT, and WA. Foreign distribution to Belgium, Croatia, Cyprus, Denmark, Germany, Indonesia, Korea, Sweden, and Tunisia.
Initiated
Voluntary: Firm initiated
Recall number
Z-2521-2018

Terminated Class II Medical devices

Recall started
16 April 2018
Terminated
2 April 2019
Reason
There may be burrs on the extraction hole threads.
Distribution
The products were distributed to the following US states: FL, MO, and PA.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1704-2018, Z-1703-2018

Terminated Class II Medical devices

Recall started
21 December 2017
Terminated
1 April 2020
Reason
The SIGMA¿ HP PFJ Cemented Trochlear Implants, a standalone component of the partial knee system is being recalled based on elevated revision rates observed as part of the company s post market surveillance process.
Distribution
Worldwide Distribution - USA (nationwide) Distribution including Puerto Rico and to the states of : AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, and
Initiated
Voluntary: Firm initiated
Recall number
Z-1219-2018

Terminated Class II Medical devices

Recall started
28 November 2017
Terminated
27 March 2020
Reason
The screw in specific lots of the GLOBAL UNITE Anatomic Body and GLOBAL UNITE Fracture Body was inverted during assembly to the body, which will cause the humeral stem to sit proud and may cause surgical delays.
Distribution
Distribution Level: The device removal is extended to the companies US Distributors/Sales Offices and Medical Facilities that may have used, purchased, or received the affected devices. Approximately 53 units distributed to firm's international affiliates.
Initiated
Voluntary: Firm initiated
Recall number
Z-1012-2018

Terminated Class II Medical devices

Recall started
31 August 2017
Terminated
12 March 2020
Reason
The affected lots are being recalled because the epiphysis may not assemble to the stem, which may cause a surgical delay.
Distribution
Nationally
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0834-2018, Z-0833-2018

Terminated Class II Medical devices

Recall started
26 July 2017
Terminated
3 April 2019
Reason
Potential for Intra-operative breakage of driver tips
Distribution
Nationwide Distribution
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0849-2018, Z-0848-2018

Terminated Class II Medical devices

Recall started
12 July 2017
Terminated
20 May 2019
Reason
Incorrect device in the package. A package that was labeled as the CORAIL COXA VARA HIGH OFFSET STEM Collared Size 9 (Part No. 3L93709 / Lot No. 5291990), contained the CORAIL HIGH OFFSET STEM Collarless SIZE 14 (Part No. L20314 / Lot No. 5292130).
Distribution
Nationally; US Consignees for this recall are the affected are US Distributors and US Medical Facilities (Hospitals/User Facilities) that may have purchased, used, or received the affected device.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0594-2018, Z-0593-2018

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.