Waltham, MA, United States
Dentsply IH, Inc. recalls
1 recall recorded since 2018, most recently on 24 July 2025. 1 is listed as ongoing. Status and classification are as published by the FDA.
Recalls1
Ongoing (FDA)1
Terminated or completed0
Class I (FDA)0
Ongoing Class II Medical devices
- Recall started
- 24 July 2025
- Reason
- The defect in the thread area will not allow the device to fully engage with an implant or analog (implant replica used for prosthetic modeling). The threads will only engage approximately 2 rotations as opposed to full engagement which is approximately 5 rotations. This equates to approximately 1mm difference in depth of engagement.
- Distribution
- Worldwide - US Nationwide distribution in the states of IL, LA, MA, MN, MO, ND, NJ, NM, OH, OK, PA, TX, WV and the countries of AT, AU, BE, CA, DE, DK, ES, FI, FR, IT, JP, NL, NO, PL, QA, SE, UA, UK.
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-2422-2025, Z-2421-2025
- Brand Name: Atlantis IO FLO Product Name: Atlantis IO FLO-S Kit Model/Catalog Number: A04S / 6802034 Product Description: The Atlantis intraoral feature locating object for suprastructures (IO FLO-S) is intended to su
- Brand Name: Atlantis IO FLO-S Product Name: Atlantis IO FLO-S Kit Model/Catalog Number: A04B / 68020033 Software Version: n/a Product Description: The Atlantis intraoral feature locating object for suprastructures (I
A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.