Delfi Medical Innovations, Inc. recalls

1 recall recorded since 2018, most recently on 3 April 2019. None are listed as ongoing. Status and classification are as published by the FDA.

Recalls1
Ongoing (FDA)0
Terminated or completed1
Class I (FDA)0

Terminated Class II Medical devices

Recall started
3 April 2019
Terminated
22 July 2022
Reason
Delfi Medical Innovations Inc has discovered that tourniquets cuffs in certain lots of the VariFit Sterile Single Use Thigh Cuffs, Single Port, Single Bladder models may have been manufactured with an inadequate perimeter weld on the cuff. Once applied to a patients thigh and inflated to a pressure to occlude blood flow in the thigh distal to the cuff, after a period of time the cuff may leak to the extent that the tourniquet instrument supplying air pressure to the cuff may not be able to adequately control and regulate pressure in the cuff. This may result in some blood flow in the thigh distal to the cuff.
Distribution
US: Ohio OUS: Austrailia
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1901-2020, Z-1902-2020