Recall Record

Minnetonka, MN, United States

Deerfield Imaging, Inc. recalls

9 recalls recorded since 2018, most recently on 17 February 2022. None are listed as ongoing. Status and classification are as published by the FDA.

Recalls9
Ongoing (FDA)0
Terminated or completed9
Class I (FDA)0

Terminated Class II Medical devices

Recall started
17 February 2022
Terminated
15 May 2024
Reason
Reports have been received that after normal or routine lateral forces are applied to the head fixation device, the split block clamp may deflect (rotationally around the axis) and not return to within acceptable distance from its prior position after the force is removed. In order to reduce deflection at the table mount location, users have tightened the hand knob screw to a degree that it is then difficult to untighten it later in the clinical workflow. Rotational movement at the table attachment point could result in HFD movement going unnoticed, which could result in a situation where surgical or medical intervention is necessary, a situation where navigation needs to be reestablished, or a situation resulting in extended time under anesthesia for the patient. It may also be difficult to loosen the mount clamp component after use, potentially leading to ergonomic injury or a workaround during equipment teardown.
Distribution
Worldwide distribution - US Nationwide distribution in the states of AL, FL, GA, MA, MN, NC, NY, OH, and TX. The countries of Sweden and Switzerland.
Initiated
Voluntary: Firm initiated
Recall number
Z-0934-2022

Terminated Class II Medical devices

Recall started
25 August 2021
Terminated
30 April 2024
Reason
ORT200/300 rotational lock may fail to function such that it cannot change states between locked and unlocked, due to a mechanical hardware issue.
Distribution
Worldwide distribution - US Nationwide distribution FL, NH, TX, OH, CA, OR, UT, TN, NC and the countries of Canada, Australia, Sweden, Japan.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2585-2021, Z-2584-2021, Z-2586-2021

Terminated Class II Medical devices

Recall started
26 November 2019
Terminated
4 August 2020
Reason
There is a potential hardware error for the iMRX System with a generator A100. A tolerance issue in the power supply of the generator control above a specific value may cause generator A100 failure. As a result, the X-ray tube will not be supplied with the required voltage and X-ray will not be possible
Distribution
CT, MA, VA. International distribution: Australia, Canada, and Qatar.
Initiated
Voluntary: Firm initiated
Recall number
Z-0762-2020

Terminated Class II Medical devices

Recall started
26 June 2019
Terminated
29 October 2020
Reason
Breaking of the HFD200 Drive Knob force indicator due to over-torqueing of the drive knob beyond the maximum allowable pinning force specification may result in the clinical staff unknowingly pinning the patient's skull with insufficient force; the clinical staff unknowingly pinning the patient's skull with excessive force; or the clinical staff identifying the issue and not being able to continue with the procedure.
Distribution
TN, NH, FL
Initiated
Voluntary: Firm initiated
Recall number
Z-0163-2020

Terminated Class II Medical devices

Recall started
6 November 2018
Terminated
25 July 2021
Reason
Table may drift in the roll position while in use
Distribution
Worldwide distribution - US Nationwide and countries of China, Qatar, Canada, Australia, Japan, France, Germany, Sweden, Italy.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0645-2019, Z-0648-2019, Z-0646-2019, Z-0647-2019

Terminated Class II Medical devices

Recall started
4 April 2018
Terminated
22 July 2020
Reason
The DC power to the IFOSS collision detector control panel at your facility is being adversely affected by the magnetic field. Our investigation shows the magnetic field is causing the IFOSS control panel to malfunction.
Distribution
Worldwide Distribution - US Nationwide in the states of OH, FL, NH, TX, Sweden, Canada
Initiated
Voluntary: Firm initiated
Recall number
Z-1645-2018

Terminated Class II Medical devices

Recall started
23 February 2018
Terminated
12 September 2019
Reason
The Operating Room Table may drift while in use, which could potentially result in unintended movement of the table.
Distribution
US, Canada, China, Australia, Sweden, Germany
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1266-2018, Z-1264-2018, Z-1265-2018, Z-1263-2018

Terminated Class II Medical devices

Recall started
23 February 2018
Terminated
27 August 2019
Reason
The rotational locking mechanisms on the ORT 200/300 tables fail to operate such that it cannot change states between locked and unlocked.
Distribution
US, Canada, China, Sweden, Germany, Japan, Qatar, France, Audtralia
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1217-2018, Z-1218-2018

Terminated Class II Medical devices

Recall started
20 February 2018
Terminated
27 August 2019
Reason
The rotational locking mechanisms on the ORT 100/ T2X tables can become disengaged or unlocked while in use which could potentially result in unintended movement of the table.
Distribution
US, Canada, Australia, China, Qatar
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0976-2018, Z-0975-2018

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.