Datex-Ohmeda, Inc. recalls

6 recalls recorded since 2018, most recently on 30 July 2026. 4 are listed as ongoing. Status and classification are as published by the FDA.

Recalls6
Ongoing (FDA)4
Terminated or completed2
Class I (FDA)3

Ongoing Class II Medical devices

Recall started
30 July 2026
Reason
Anesthesia Delivery Systems containing certain electronic gas mixer assemblies can automatically transition to Alternate O2 control. In Alternate O2 control, the system delivers 100% O2 to the breathing system enabling continued controlled or manual ventilation and initiates audible and visual alarms. When this issue occurs, anesthetic agent delivery will be interrupted and requires clinician intervention. If this is not provided, the patient may experience decreased anesthetic agent effect.
Distribution
Worldwide - US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-3191-2026, Z-3192-2026, Z-3190-2026, Z-3189-2026

Ongoing Class I Medical devices

Recall started
8 May 2026
Reason
Blender knob shaft can loosen, which can affect the delivered oxygen concentration. The system can deliver an oxygen concentration that does not match the intended concentration set by the user, potentially resulting in hypoxia or hyperoxia in the patient.
Distribution
Pending
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2357-2026, Z-2358-2026, Z-2359-2026

Ongoing Class I Medical devices

Recall started
12 July 2024
Reason
Potential issue that may arise if the limb of a patient breathing circuit is mistakenly connected to the Auxiliary Common Gas Outlet (ACGO) port.
Distribution
Worldwide distribution - US Nationwide and the countries of Aland Islands, ALBANIA, Algeria, Argentina, Australia, Austria, Azerbaijan, BAHAMAS, Bahrain, Bangladesh, BARBADOS, BELARUS, Belgium, Benin, Bermuda, Bolivia, BONAIRE, Bosnia and Herzegovina, Brazil, Brunei, Bulgaria, Cambodia, Cameroon, Ca
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0083-2025, Z-0075-2025, Z-0080-2025, Z-0082-2025, Z-0081-2025, Z-0074-2025, Z-0078-2025, Z-0079-2025, Z-0076-2025, Z-0084-2025, Z-0077-2025

Ongoing Class I Medical devices

Recall started
29 December 2023
Reason
GE HealthCare has become aware of elevated levels of formaldehyde when the EVair or EVair 03 (Jun-Air) optional compressors are used with the CARESCAPE R860 or Engstr¿m Carestation/Pro ventilators, respectively. These elevated results were observed in preliminary testing that was conducted at an elevated room temperature of 40oC (104oF), at the lowest possible flow condition of 2 L/min (worstcase, minimum bias flow with no additional ventilation), and all of the gas being supplied from the compressor (i.e., FiO2 of 21% / no supplemental oxygen). GE HealthCare is continuing to evaluate the root cause for these elevated levels of formaldehyde. While these test conditions are not representative of typical clinical use conditions,GE HealthCare is taking this action to further reduce the potential for patient exposure.
Distribution
Worldwide
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0775-2024, Z-0776-2024, Z-0774-2024

Terminated Class II Medical devices

Recall started
28 June 2018
Terminated
22 February 2019
Reason
The Neonatal Flow Sensor Cable may be missing the connector housing exposing the wires in the cable.
Distribution
Worldwide Distribution: US (nationwide) to: Washington, DC and countries of: Algeria, Bangladesh, Brazil, Chile, China, Colombia, Germany, Greece, Guatemala, Honduras, India, Indonesia, Iran, Iraq, Italy, Kenya, Lebanon, Malaysia, Maldives, Mexico, Morocco, Myanmar, Nepal, Netherlands, Nigeria, Pa
Initiated
Voluntary: Firm initiated
Recall number
Z-2738-2018

Terminated Class II Medical devices

Recall started
8 December 2017
Terminated
29 April 2021
Reason
Potential for the display processor to experience an unexpected failure.
Distribution
Worldwide Distribution - USA (nationwide) Distribution to the states of : AL, AZ, CA, GA, ID, IL, IN, MS, NC, NJ, NK, NV, OH, OK, RI, TX, and WI., and to the countries of : Albania, Argentina, Australia, Bahrain, Bangladesh, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Canada, Chile
Initiated
Voluntary: Firm initiated
Recall number
Z-2266-2018

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.