DATEX--OHMEDA, INC. recalls
3 recalls recorded since 2018, most recently on 3 September 2024. 2 are listed as ongoing. Status and classification are as published by the FDA.
Recalls3
Ongoing (FDA)2
Terminated or completed1
Class I (FDA)2
Ongoing Class I Medical devices
- Recall started
- 3 September 2024
- Reason
- GE HealthCare has become aware that for certain Giraffe OmniBed and Giraffe OmniBed Carestation devices, the screw that secures the doors that cover the warmer heaters may not have been torqued to specification. This may lead to the doors becoming loose.
- Distribution
- Worldwide distribution.
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-0047-2025, Z-0048-2025
- GE Heathcare Giraffe OmniBed; neo natal incubator
- GE Heathcare Giraffe OmniBed Carestation, Model Number 2082844-001-XXX; neo natal incubator
Completed Class I Medical devices
- Recall started
- 6 June 2023
- Reason
- Prior to current established manufacturing practices, higher levels of airborne formaldehyde exposure from newly manufactured incubators could be seen during clinical use that would rapidly decrease over a week. Affected customers were notified in June 2023 to run incubators for a week to potentially off gas any residual airborne formaldehyde. The general scientific consensus is that sensory irritation of the eyes is the most sensitive endpoint with respect to exposure to airborne formaldehyde and is the adverse effect clinically experienced at the lowest concentration.
- Distribution
- worldwide
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-0141-2025, Z-0140-2025
- GE Healthcare Giraffe Incubator Carestation CS1, Infant Incubator. Model Numbers: 1) 2082844-002-01040417; 2) 2082844-002-01044956; 3) 2082844-002-01058424; 4) 2082844-002-01069017; 5) 2082844-002-01089666;
- GE Healthcare Giraffe Omnibed Carestation CS1, combination infant incubator and infant warmer. Model Numbers: 1) 2082844-001-01040418; 2) 2082844-001-01046784; 3) 2082844-001-01052324; 4) 2082844-001-01052572;
Ongoing Class II Medical devices
- Recall started
- 19 May 2023
- Reason
- GE HealthCare has become aware of the potential that an incorrect type of secondary latch could have been installed on the northeast (NE) and/or northwest (NW) bedside panels for certain Giraffe OmniBed and Giraffe OmniBed Carestation devices
- Distribution
- Worldwide distribution.
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-2100-2023, Z-2102-2023, Z-2101-2023
- Giraffe OmniBed Carestation CS1, Model Numbers: a. 2082844-001-XXX, b. 2074920-001-XXXXXX, c. M1050205, d. M1135822, e. M1145475, f. M1145476, g. M1235957, h. M1149051, i. M1146098; Labeled as the following
- Beside panel FRU (Field Replacement Units) Kits
- Giraffe OmniBed, with the following Model Numbers: 6650-0004-901, 6650-0001-901, 6650-0003-901, 6650-0002-901, 6650-0094-901, 6650-0001-9901, 6650-0029-901, 6650-0080-901, 6650-0032-901, 6650-0077-901, 6650-001-901, 6650