Fairfield, NJ, United States
Datascope Corp. recalls
39 recalls recorded since 2018, most recently on 6 February 2026. 32 are listed as ongoing. Status and classification are as published by the FDA.
Recalls39
Ongoing (FDA)32
Terminated or completed7
Class I (FDA)14
Ongoing Class III Medical devices
- Recall started
- 6 February 2026
- Reason
- The Battery Charging Station is an optional dual-bay charger used to charge the Cardiosave IABP Lithium- Ion batteries when they are not being used to operate the IABP. Getinge has identified that a protruding screw in the left battery bay limits full insertion of certain batteries and prevents proper charging. The right bay of the Battery Charging Station is not affected by this issue.
- Distribution
- Worldwide - US Nationwide distribution in the states of FL, DC, GA, IL, IN, MD, MI, NC, TX, WA, WI and the countries of Australia, Hong Kong, Israel, Japan, Norway.
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-1587-2026
- Battery Charging Station; Model: 0998-00-0802;
Ongoing Class II Medical devices
- Recall started
- 23 January 2026
- Reason
- The firm has identified that the battery runtime and cycle specifications contained in the devices Instructions for Use (IFU) need to be updated.
- Distribution
- Worldwide distribution - US Nationwide and the countries of Afghanistan, Albania, Algeria, Argentina, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Barbados, Belarus, Belgium, Bolivia, Bosnia And Herzegovina, Brazil, Bulgaria, Cambodia, Canada, Chile, China, Colombia, Costa Rica, Cyprus, Czec
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-1487-2026, Z-1486-2026
- CS300 IABP. Software Version CS300 IABP C.01.
- CS100 IABP. Software Version CS100 IABP Q.01.
Ongoing Class II Medical devices
- Recall started
- 4 November 2025
- Reason
- The IFU addendum revises the Preventative Maintenance schedule to align with the update introduced in the Service Manual as of June 2023.
- Distribution
- Domestic: Nationwide Distribution; Foreign: Albania, Algeria, Angola, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Bosnia And Herzegovina, Botswana, Brazil, Brunei, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, C¿te d Ivoire, Croatia, Cyprus, Cze
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-0917-2026, Z-0916-2026
- Cardiosave Rescue. Intra-Aortic Balloon Pump system.
- Cardiosave Hybrid. Intra-Aortic Balloon Pump system.
Ongoing Class III Medical devices
- Recall started
- 23 October 2025
- Reason
- The IFU addendum updates the Vibration and Shock Table to reference the correct standards.
- Distribution
- Domestic: Nationwide distribution. International: Albania, Algeria, Angola, Argentina, Armenia, Republic of, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Bosnia And Herzegovina, Botswana, Republic of, Brazil, Brunei, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, C¿
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-0932-2026, Z-0931-2026
- Cardiosave Rescue
- Cardiosave Hybrid
Ongoing Class II Medical devices
- Recall started
- 17 September 2024
- Reason
- There is discrepant labeling on the inner and outer packaging of the device.
- Distribution
- International distribution to the countries of Germany, Spain , Finland, Great Britain, Hong Kong, Ireland, Italy, Lithuania, Moldova, Mexico, Norway, New Zealand, Poland, Sweden, Singapore.
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-0311-2025
- MEGA 7.5Fr. 40cc Intra-Aortic Balloon with Accessories, Model Number 0684-00- 0295-01
Ongoing Class I Medical devices
- Recall started
- 8 August 2024
- Reason
- Firm has developed a software correction (Revision D.00) outside the US to mitigate product problems that pose a risk of hemodynamic instability and failure of transmission of clinical data in Cardiosave Hybrid and Cardiosave Rescue Intra-Aortic Balloon Pump devices.
- Distribution
- International distribution to the countries of UNITED ARAB EMIRATES, ARGENTINA , AUSTRIA, AUSTRALIA, BELGIUM, BULGARIA, BAHRAIN, BRAZIL, CANADA, SWITZERLAND, CHILE, CHINA, COLOMBIA, CZECH REPUBLIC, GERMANY, DENMARK, ECUADOR, EGYPT, SPAIN, FINLAND, FRANCE, UNITED KINGDOM, INDONESIA, IRELAND, INDIA, I
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-3134-2024, Z-3135-2024
- CARDIOSAVE HYBRID INTRA-AORTIC BALLOON PUMP (IABP). Electromechanical system used to inflate and deflate intra-aortic balloons.
- CARDIOSAVE RESCUE INTRA-AORTIC BALLOON PUMP (IABP). Electromechanical system used to inflate and deflate intra-aortic balloons.
Ongoing Class I Medical devices
- Recall started
- 8 August 2024
- Reason
- Firm has developed a software correction (Revision D.01) for outside the US to mitigate product problems that pose a risk of hemodynamic instability and failure of transmission of clinical data in Cardiosave Hybrid and Cardiosave Rescue Intra-Aortic Balloon Pump devices.
- Distribution
- International distribution to the countries of UNITED ARAB EMIRATES, ARGENTINA , AUSTRIA, AUSTRALIA, BELGIUM, BULGARIA, BAHRAIN, BRAZIL, CANADA, SWITZERLAND, CHILE, CHINA, COLOMBIA, CZECH REPUBLIC, GERMANY, DENMARK, ECUADOR, EGYPT, SPAIN, FINLAND, FRANCE, UNITED KINGDOM, INDONESIA, IRELAND, INDIA, I
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-3137-2024, Z-3136-2024
- CARDIOSAVE RESCUE INTRA-AORTIC BALLOON PUMP (IABP). Electromechanical system used to inflate and deflate intra-aortic balloons.
- CARDIOSAVE HYBRID INTRA-AORTIC BALLOON PUMP (IABP). Electromechanical system used to inflate and deflate intra-aortic balloons.
Ongoing Class I Medical devices
- Recall started
- 31 July 2023
- Reason
- Users were reporting that the device was not charging as expected. It was discovered that users were unaware that the Cardiosave console was not completely inserted into the hospital cart. If the console is not fully inserted back into the cart the battery(ies) will not charge.
- Distribution
- Worldwide - US Nationwide distribution.
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-2431-2023, Z-2432-2023
- Cardiosave Hybrid Intra-Aortic Balloon Pumps (IABP), Part Numbers 0998-00-0800-31, 0998-UC-0800-31, 0998-00-0800-32, 0998-00-0800-33, 0998-UC-0800-33, 0998-00-0800-34, 0998-00-0800-35, 0998-00-0800-45, 0998-00-0800-52, 0
- Cardiosave Rescue Intra-Aortic Balloon Pumps (IABP), Part Numbers 0998-00-0800-75, 0998-00-0800-83, 0998-00-0800-85
Ongoing Class I Medical devices
- Recall started
- 31 July 2023
- Reason
- Users were reporting instances of "Gas Loss in IAB Circuit" and "Gas Gain in IAB Circuit" alarms while providing therapy. An internal investigation of the complaints determined that there is a potential trigger for these alarms that was not listed in the IFU: Patient Movement (coughing, general movement, and swallowing). Blood in the circuit, kins and abrasions of the tubing/catheter and certain patient conditions could also trigger these alarms. The HHE also identified hardware issues related to these alarms.
- Distribution
- Worldwide - US Nationwide distribution.
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-2435-2023, Z-2436-2023
- Cardiosave Hybrid Intra-Aortic Balloon Pumps (IABP), Part Numbers 0998-00-0800-31, 0998-UC-0800-31, 0998-00-0800-32, 0998-00-0800-33, 0998-UC-0800-33, 0998-00-0800-34, 0998-00-0800-35, 0998-00-0800-45, 0998-00-0800-52, 0
- Cardiosave Rescue Intra-Aortic Balloon Pumps (IABP), Part Numbers 0998-00-0800-75, 0998-00-0800-83, 0998-00-0800-85
Ongoing Class I Medical devices
- Recall started
- 31 July 2023
- Reason
- Users reported "System Over Temperature" alarms associated with a loss of pumping and/or the Cardiosave system entering Standby mode.
- Distribution
- Worldwide - US Nationwide distribution.
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-2433-2023, Z-2434-2023
- Cardiosave Hybrid Intra-Aortic Balloon Pumps (IABP), Part Numbers 0998-00-0800-31, 0998-UC-0800-31, 0998-00-0800-32, 0998-00-0800-33, 0998-UC-0800-33, 0998-00-0800-34, 0998-00-0800-35, 0998-00-0800-45, 0998-00-0800-52, 0
- Cardiosave Rescue Intra-Aortic Balloon Pumps (IABP), Part Numbers 0998-00-0800-75, 0998-00-0800-83, 0998-00-0800-85, 0998-UC-0800-75, 0998-UC-0800-83, 0998-UC-0800-85,
Ongoing Class I Medical devices
- Recall started
- 31 July 2023
- Reason
- Users were identifying autofill failure conditions on the devices causing pump stops.
- Distribution
- Worldwide - US Nationwide distribution.
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-2438-2023, Z-2437-2023
- Cardiosave Rescue Intra-Aortic Balloon Pumps (IABP), Part Numbers 0998-00-0800-75, 0998-00-0800-83, 0998-00-0800-85
- Cardiosave Hybrid Intra-Aortic Balloon Pumps (IABP), Part Numbers 0998-00-0800-31, 0998-UC-0800-31, 0998-00-0800-32, 0998-00-0800-33, 0998-UC-0800-33, 0998-00-0800-34, 0998-00-0800-35, 0998-00-0800-45, 0998-00-0800-52, 0
Ongoing Class II Medical devices
- Recall started
- 31 July 2023
- Reason
- Users were experiencing a failure in the IAB Fiber Optic Sensor input on the IABP when inserting the Intra-Aortic Balloon Fiber Optic connector.
- Distribution
- US Nationwide. Global Distribution.
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-2487-2023, Z-2486-2023
- Cardiosave Rescue Intra-Aortic Balloon Pumps (IABP), Part Numbers 0998-00-0800-75, 0998-00-0800-83, 0998-00-0800-85
- Cardiosave Hybrid Intra-Aortic Balloon Pumps (IABP), Part Numbers 0998-00-0800-31, 0998-UC-0800-31, 0998-00-0800-32, 0998-00-0800-33, 0998-UC-0800-33, 0998-00-0800-34, 0998-00-0800-35, 0998-00-0800-45, 0998-00-0800-52, 0
Ongoing Class II Medical devices
- Recall started
- 23 June 2023
- Reason
- During IAB catheter insertion, the introducer dilator may fracture at the hub, leaving the body of the introducer dilator housed within the sheath. Risks to health include damage to the femoral artery, descending aorta, or embolization; surgical intervention and delay of IAB therapy may also result. Firm has received 10 complaints, including 3 serious adverse events and 1 patient death.
- Distribution
- Domestic distribution nationwide. International distribution worldwide.
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-2223-2023, Z-2217-2023, Z-2216-2023, Z-2221-2023, Z-2218-2023, Z-2219-2023, Z-2224-2023, Z-2220-2023, Z-2222-2023
- LINEAR 7.5Fr.IAB, Intra-Aortic Balloon Catheter and Accessories, used to provide counter pulsation therapy in the aorta Product Description
- SENSATION 7 Fr. IAB, Intra-Aortic Balloon Catheter and Accessories, used to provide counter pulsation therapy in the aorta Product Description Par
- YAMATO PLUS IAB; intra-aortic balloon catheter and accessories used to provide counter pulsation therapy in the aorta. Product Description
- SENSATION PLUS 7.5Fr IAB, Intra-Aortic Balloon Catheter and Accessories, used to provide counter pulsation therapy in the aorta Product Description
- TRANS-RAY IAB, Intra-Aortic Balloon Catheter and Accessories, used to provide counter pulsation therapy in the aorta Product Description Part Number 7FR. 34cc TRAN
- MEGA 7.5Fr.IAB , Intra-Aortic Balloon Catheter and Accessories, used to provide counter pulsation therapy in the aorta Product Description
- REINFORCED INTRODUCER SET, Intra-Aortic Balloon Catheter and Accessories, used to provide counter pulsation therapy in the aorta Product Description
- MEGA 8Fr. IAB, Intra-Aortic Balloon Catheter and Accessories, used to provide counter pulsation therapy in the aorta Product Description
- and 1 more product lines
Ongoing Class I Medical devices
- Recall started
- 5 June 2023
- Reason
- IABP may lose the ability to charge batteries in one or both bay slots. Therapy may be interrupted if batteries fail to charge and the device is disconnected from AC power. Low battery alarms may alert the User to the issue prior to interruption of therapy.
- Distribution
- Worldwide Distribution: US (Nationwide) and OUS (foreign) to countries of: ALBANIA ALGERIA, ANGOLA, ARGENTINA, AUSTRALIA, AUSTRIA, AZERBAIJAN, BAHRAIN, BANGLADESH, BARBADOS, BELARUS, BELGIUM, BOSNIA and HERZEGOVINA, BRAZIL, BRUNEI, BULGARIA, CANADA, CHILE, CHINA, COLOMBIA, COSTA RICA, C¿te d'Ivoire,
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-2149-2023, Z-2150-2023
- Cardiosave Hybrid Intra-Aortic Balloon Pump (IABP), Model numbers 0998-00-0800-31 0998-00-0800-32 0998-00-0800-33 0998-00-0800-34 0998-00-0800-35 0998-00-0800-45 0998-00-0800-52 0998-00-0800-53 0998-00-0800-55
- Cardiosave Rescue Intra-Aortic Balloon Pump (IABP), Model numbers 0998-00-0800-75 0998-00-0800-83 0998-00-0800-85
Ongoing Class I Medical devices
- Recall started
- 5 June 2023
- Reason
- An unexpected shutdown of the IABP may occur due to failures of tantalum capacitors in the Power Management Board and/or Solenoid Board. This issue may lead to an unexpected interruption of therapy.
- Distribution
- Worldwide Distribution- US (Nationwide) and OUS (foreign) countries of: ALBANIA, ALGERIA, ANGOLA, ARGENTINA, AUSTRALIA, AUSTRIA, AZERBAIJAN, AZERBAIJAN , BAHRAIN, BANGLADESH, BARBADOS, BELARUS, BELGIUM, BOSNIA and HERZEGOVINA, BRAZIL, BRUNEI, BULGARIA, CANADA, CHILE, CHINA, COLOMBIA, COSTA RICA, C¿t
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-2151-2023, Z-2152-2023
- Cardiosave Hybrid Intra-Aortic Balloon Pump (IABP), Model numbers 0998-00-0800-31 0998-00-0800-32 0998-00-0800-33 0998-00-0800-34 0998-00-0800-35 0998-00-0800-45 0998-00-0800-52 0998-00-0800-53 0998-00-0800-55
- Cardiosave Rescue Intra-Aortic Balloon Pump (IABP), Model numbers 0998-00-0800-75 0998-00-0800-83 0998-00-0800-85
Ongoing Class III Medical devices
- Recall started
- 4 May 2023
- Reason
- The digital IFUs provided with the Cardiosave IABP during the software update to version D.00 incorrectly annotated compliance to standard 60601-1-12:2014 within the updated IFU.
- Distribution
- International distribution in the country of Germany.
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-1907-2023
- Cardiosave Hybrid Intra-Aortic Balloon Pumps (IABP), Model Number 0998-00-0800-55, Part Number 0998-00-0800-XX
Ongoing Class II Medical devices
- Recall started
- 4 April 2023
- Reason
- The power cord supplied (Type J plug end) for the Cardiosave Hybrid IABP to Brazil is incompatible with the Brazilian Type N power receptacle.
- Distribution
- International distribution in the country of Brazil.
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-1462-2023
- MAQUET CARDIOSAVE hybrid Intra-Aortic Balloon Pump, Model Number 0998-XX-0800-32
Ongoing Class II Medical devices
- Recall started
- 28 February 2023
- Reason
- There is an issue related to the Safety Disk not meeting a performance requirement that may impact Intra-Aortic Balloon catheter (IAB) displacement volume. This potential failure of an impacted Safety Disk may lead to slightly reduced IAB inflation during therapy or a reduction in the degree of augmentation provided to the patient during therapy.
- Distribution
- Distribution list not yet provided.
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-1277-2023, Z-1276-2023, Z-1278-2023, Z-1279-2023
- Cardiosave Hybrid Intra-Aortic Balloon Pump (IABP), Model Numbers Model Numbers 0998-00-0800-75, 0998-00-0800-83, 0998-00-0800-85
- Cardiosave Hybrid Intra-Aortic Balloon Pump (IABP), Model Numbers 0998-00-0800-31, 0998-00-0800-32, 0998-00-0800-33, 0998-00-0800-35, 0998-00-0800-45, 0998-00-0800-52, 0998-00-0800-53, 0998-00-0800-55, and 0998-00-0800-6
- Cardiosave Intra-Aortic Balloon Pump (IABP) Safety Disk, Model Number 0202-00-0140
- Cardiosave Intra-Aortic Balloon Pump (IABP) Pneumatic Interface Module (PIM) , Model Number 0997-00-1178
Ongoing Class II Medical devices
- Recall started
- 7 February 2023
- Reason
- There have been reported failures of the high pressure helium regulator, which may cause a helium leak in the Cardiosave Hospital Cart. In instance of helium regulator failure, a Pump Console s internal reservoir of helium will not be replenished when docked into an impacted Hospital Cart. This may result in an insufficient amount of helium within the internal reservoir.
- Distribution
- US Nationwide. Global Distribution.
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-1230-2023, Z-1229-2023
- Cardiosave Rescue Intra-Aortic Balloon Pump (IABP), Model Numbers 0998-00-0800-75, 0998-00-0800-83, and 0998-00-0800-85
- Cardiosave Hybrid Intra-Aortic Balloon Pump (IABP), Model Numbers 0998-00-0800-31, 0998-00-0800-32, 0998-00-0800-33, 0998-00-0800-34, 0998-00-0800-35, 0998-00-0800-45, 0998-00-0800-52, 0998-00-0800-53, 0998-00-0800-55, 0
Ongoing Class II Medical devices
- Recall started
- 7 February 2023
- Reason
- Firm has received reports of damaged, worn, or torn O-rings on the Cardiosave Pump Console quick disconnect fitting, which results in helium tank leaks. If the helium supply is depleted, therapy will be interrupted.
- Distribution
- Nationwide US distribution. Worldwide international distribution.
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-1231-2023, Z-1232-2023
- Cardiosave Hybrid Intra-Aortic Balloon Pump. Model Numbers: D998-00-0800-31, D998-00-0800-32, D998-00-0800-33, D998-00-0800-35, D998-00-0800-36, D998-00-0800-45, D998-00-0800-52, D998-00-0800-53, D998-00-0800-55, D998-00
- Cardiosave Rescue. Model Number: 0998-00-0800-83.
Ongoing Class I Medical devices
- Recall started
- 7 February 2023
- Reason
- An unexpected shutdown of the IABP may occur due to loss of communication between the Executive Processor PCBA and the Video Generator PCBA. This may result in an interruption to therapy which may threaten the hemodynamic stability of the supported patient as the user is left unaware to the status of the Cardiosave IABP.
- Distribution
- Worldwide Distribution: US (nationwide) including states of: AK, AL, AR, AZ, CA, CO, CT, DC, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NY, NV, OH, OK, OR, PA, SC, SD, TN, TX, VA, WA, WI, and WV; and OUS (Foreign) countries of: ALBANIA, ALGERIA, ANGO
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-1147-2023, Z-1148-2023
- Cardiosave Hybrid Model Nos. 0998-00-0800-31 0998-00-0800-32 0998-00-0800-33 0998-00-0800-34 0998-00-0800-35 0998-00-0800-45 0998-00-0800-52 0998-00-0800-53 0998-00-0800-55 0998-00-0800-65 0998-UC-0800-31*
- Cardiosave Rescue Model Nos. 0998-00-0800-75 0998-00-0800-83 0998-00-0800-85
Ongoing Class I Medical devices
- Recall started
- 7 February 2023
- Reason
- An unexpected shutdown of the IABP may occur due to a failure of the connection between the Coiled Cord cable and the Cable Assembly backplane to the Coiled Cord cable which provides the communication between the display head and base unit.
- Distribution
- Worldwide Distribution: US (nationwide) and OUS (Foreign) ARGENTINA, AUSTRALIA, AUSTRIA, AZERBAIJAN, BAHRAIN, BELARUS, BELGIUM, BRAZIL, BRUNEI , BULGARIA, CANADA, CHILE, CHINA, CZECH REPUBLIC, ECUADOR, EGYPT, EL SALVADOR, FINLAND, FRANCE, GERMANY, GHANA, HONG KONG, ICELAND, INDIA, INDONESIA, IRAN, I
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-1146-2023, Z-1145-2023
- Cardiosave Rescue Intra-Aortic Balloon Pumps (IABP), Containing Coiled Cord cable (part number 0012-00-1801, Model Numbers 0998-00-0800-75, 0998-00-0800-83, 0998-00-0800-85
- Cardiosave Hybrid Intra-Aortic Balloon Pumps (IABP), Containing Coiled Cord cable (part number 0012-00-1801, Model Numbers 0998-00-0800-31, 0998-00-0800-32, 0998-00-0800-33, 0998-00-0800-34, 0998-00-0800-35, 0998-00-0800
Ongoing Class I Medical devices
- Recall started
- 19 December 2022
- Reason
- The Cardiosave IABP may shut down unexpectedly due to blood entering in the Cardiosave IABP when therapy is provided with a compromised intra-aortic balloon catheter.
- Distribution
- Worldwide Distribution - US Nationwide and the countries of Albania, Angola, Argentina, Australia, Austria, AZERBAIJAN, BAHRAIN, BANGLADESH, Barbados, Belarus, Belgium, BOSNIA and HERZEGOVINA, Brazil, Brunei, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, C¿te d'Ivoire, Croatia, CZECH REPUBLI
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-0885-2023, Z-0886-2023
- Cardiosave Hybrid Intra-Aortic Balloon Pump (IABP), Model Numbers 0998-00-0800-31, 0998-00-0800-32, 0998-00-0800-33, 0998-00-0800-34, 0998-00-0800-35, 0998-00-0800-36, 0998-00-0800-45, 0998-00-0800-52, 0998-00-0800-53, 0
- Cardiosave Rescue Intra-Aortic Balloon Pump (IABP), Model Numbers 0998-00-0800-75, 0998-00-0800-83, 0998-00-0800-85, 0998-UC-0800-75, 0998-UC-0800-83, 0998-UC-0800-85.
Ongoing Class II Medical devices
- Recall started
- 5 August 2022
- Reason
- The balloon volume is incorrectly described as 34cc instead of 40cc on one of the three internal labels enclosed for hospital use. A labelling error may result in user confusion and/or delay in initiating therapy. Therapy delays may impact those patients who are more hemodynamically unstable.
- Distribution
- No US distribution. Foreign distribution to Argentina, Brazil, Canada, Chile, Columbia, Ecuador, Egypt, France, Indonesia, Israel, Italy, Jordan, Kenya, Malaysia, Mexico, Netherlands, Pakistan, Poland, Portugal, South Korea, Russia, Spain, Turkmenistan, United Arab Emirates, United Kingdom, Uruguay.
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-0131-2023
- LINEAR 7.5Fr. 40cc IAB with Accessories (APA), Model No. 0684-00-0480-02. intra-aortic balloon catheter.
Ongoing Class II Medical devices
- Recall started
- 5 August 2022
- Reason
- Certain lots containing undersized dilator.
- Distribution
- Awaiting domestic distribution. Foreign: Australia, Belgium, Canada, Czech Republic, France, Germany, Hong Kong, Iceland, Ireland, Italy, Netherlands, New Zealand, Norway, Poland, Portugal, Saudi Arabia, Slovenia, South Africa, Spain, Sweden, Switzerland, United Arab Emirates, United Kingdom.
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-1685-2022, Z-1686-2022, Z-1687-2022
- MEGA 8Fr 50cc Intra-Aortic Balloon Catheter
- Sensation Plus 8Fr 50cc Intra-Aortic Balloon Catheter
- Insertion Kit for use with SENSATION PLUS 8F. 50cc Intra-Aortic Balloon Catheter
Ongoing Class I Medical devices
- Recall started
- 15 November 2021
- Reason
- Firm is initiating a correction due to the possibility of fluid ingress. Fluid entering the device may short various electronic components thus leading to system shutdown, which could result in unsafe hemodynamic instability.
- Distribution
- Worldwide distribution - US Nationwide.
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-0337-2022, Z-0336-2022
- Cardiosave Rescue IntraAortic Balloon Pump. Model Numbers: 0998-00-0800-75, 0998-UC-0800-75, 0998-00-0800-83, 0998-UC-0800-83, 0998-00-0800-85, 0998-UC-0800-85.
- Cardiosave Hybrid IntraAortic Balloon Pump. Model Numbers: 0998-00-0800-31, 0998-UC-0800-31, 0998-00-0800-32, 0998-UC-0800-32, 0998-00-0800-33, 0998-UC-0800-33, 0998-00-0800-34, 0998-UC-0800-34, 0998-00-0800-35, 0998-UC-
Ongoing Class I Medical devices
- Recall started
- 21 September 2021
- Reason
- A Cardiosave Lithium-Ion Battery Pack used during protocol testing failed to meet the minimum runtime requirement per Getinge internal Product Specification. These nonconforming batteries were inadvertently released to customers.
- Distribution
- Worldwide distribution. US nationwide, Australia, Brazil, Bahrain, Belgium, Canada, China, Czech Republic, Egypt, France, Germany, India, Ireland, Italy, Japan, Kuwait, Netherlands, Norway, Pakistan, Saudi Arabia, Singapore, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, T
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-0091-2022
- Cardiosave Li-Ion Battery Pack, Part No. 0146-00-0097. Component of the Cardiosave Hybrid Intra-Aortic Balloon Pump (IABP) with the below Model Numbers. 1. Cardiosave Hybrid. Model Numbers: 0998-00-0800-31, 0998-00-080
Ongoing Class II Medical devices
- Recall started
- 8 September 2021
- Reason
- The Cardiosave IABP may unexpectedly shut down when the device is running on AC power, only one battery is installed in the IABP, and the battery is physically removed while the battery is being charged. This occurs during a very specific set of conditions.
- Distribution
- Nationwide domestic distribution. Foreign distribution worldwide.
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-2615-2021, Z-2616-2021
- Cardiosave Hybrid Intra-Aortic Balloon Pump (IABP), Model Nos. 0998-00-0800-31, 0998-00-0800-32, 0998-00-0800-33, 0998-00-0800-34, 0998-00-0800-35, 0998-00-0800-45, 0998-00-0800-52, 0998-00-0800-53, 0998-00-0800-55, 0998
- Cardiosave Rescue Intra-Aortic Balloon Pump (IABP), Model Nos. 0998-00-0800-75, 0998-00-0800-83, 0998-00-0800-85
Ongoing Class II Medical devices
- Recall started
- 8 September 2021
- Reason
- The Helium Indicator on the Cardiosave display may over represent the amount of helium in unit in the presence of Electromagnetic Interference.
- Distribution
- Worldwide distribution - US Nationwide distribution.
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-0075-2022, Z-0076-2022
- Cardiosave Hybrid Intra-Aortic Balloon Pump (IABP), Model Nos. 0998-00-0800-31, 0998-00-0800-32, 0998-00-0800-33, 0998-00-0800-34, 0998-00-0800-35, 0998-00-0800-45, 0998-00-0800-52, 0998-00-0800-53, 0998-00-0800-55, 0998
- Cardiosave Rescue Intra-Aortic Balloon Pump (IABP), Model Nos. 0998-00-0800-75, 0998-00-0800-83, 0998-00-0800-85
Ongoing Class III Medical devices
- Recall started
- 22 March 2021
- Reason
- There are cybersecurity vulnerabilities in a widely used low-level TCP/IP software library that may result in a loss of communication to the Hospital Information System/Clinical Information System (HIS/CIS).
- Distribution
- Worldwide distribution - US Nationwide distribution.
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-1507-2021, Z-1506-2021
- Cardiosave Rescue IABP - Product Usage: used to inflate and deflate intra-aortic balloons. It provides temporary support to the left ventricle via the principle of counterpulsation, Model Numbers 0998-00-0800-75, 0998-00
- Cardiosave Hybrid IABP - Product Usage: used to inflate and deflate intra-aortic balloons. It provides temporary support to the left ventricle via the principle of counterpulsation, Model Numbers 0998-00-0800-32, 0998-00
Ongoing Class II Medical devices
- Recall started
- 27 July 2020
- Reason
- Potential Endotoxin Contamination
- Distribution
- Worldwide distribution - US Nationwide distribution and the countries of Argentina, Austrailia Austria Belgium Brazil Canada Chile Colombia Czech Republic Denmark Finland France Germany Hong Kong India Ireland Italy Japan Korea Mexico Netherlands New Zealand Norway Paraguay Poland Portugal Russia, S
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-2839-2020, Z-2842-2020, Z-2843-2020, Z-2833-2020, Z-2837-2020, Z-2838-2020, Z-2846-2020, Z-2831-2020, Z-2847-2020, Z-2841-2020, Z-2832-2020, Z-2845-2020 and 5 more
- Datascope Intra-Aortic Ballon Catheters (IABs)-Linear 7.5Fr 34cc IAB Kit P/N: 0684-00-0479-01, 0684-00-0479-01C, 0684-00-0479-01U, 0684-00-0479-02, 0684-00-0479-02U, 0684-00-0479-07, 0684-00-0479-08, 0684-00-0479-10 - Pr
- Datascope Intra-Aortic Ballon Catheters (IABs)-Trans-Ray 7Fr 40cc IAB (Japan) - Product Usage: Indications For Use: Acute Coronary Syndrome Cardiac and Non-Cardiac Surgery Complications of Heart Failure. Kit P/N: 0
- Datascope Intra-Aortic Ballon Catheters (IABs)-TRANS-RAY PLUS 7.5Fr 35cc IAB (Japan) Kit P/N: 0684-00-0607 - Product Usage: Indications For Use: Acute Coronary Syndrome Cardiac and Non-Cardiac Surgery Complications of H
- Datascope Intra-Aortic Ballon Catheters (IABs)-MEGA 8Fr 50cc IAB Kit P/N: 0684-00-0296-01, 0684-00-0296-01U,0684-00-0296-02, 0684-00-0296-02U, 0684-00-0296-10; 0684-00-0498-01, 0684-00-0498-07, 0684-00-0498-08 Produc
- Datascope Intra-Aortic Ballon Catheters (IABs)-Sensation Plus 8Fr 50cc IAB Kit P/N: 0684-00-0576-01, 0684-00-0576-01U, 0684-00-0576-05, 0684-00-0576-09 - Product Usage: Indications For Use: Acute Coronary Syndrome Cardi
- Datascope Intra-Aortic Ballon Catheters (IABs)-Linear 7.5Fr 25cc IAB Kit P/N: 0684-00-0478-01, 0684-00-0478-01C, 0684-00-0478-01U 0684-00-0478-02, 0684-00-0478-02U - Product Usage: Indications For Use: Acute Coronar
- Datascope Intra-Aortic Ballon Catheters (IABs)-YAMATO PLUS-R 7.5Fr 35cc IAB (Japan) Kit P/N: 0684-00-0560-01 - Product Usage: Indications For Use: Acute Coronary Syndrome Cardiac and Non-Cardiac Surgery Complications of
- Datascope Intra-Aortic Ballon Catheters (IABs)-MEGA 7.5Fr 30cc IAB, P/N: 0684-00-0294-01,0684-00-0294-01U, 0684-00-0294-02, 0684-00-0294-02U- Product Usage: Indications For Use: Acute Coronary Syndrome Cardiac and No
- and 9 more product lines
Terminated Class II Medical devices
- Recall started
- 20 November 2019
- Terminated
- 15 June 2021
- Reason
- Potential for compromised sterility in breached pouches of specific lots of Reinforced Introducer Sets Maquet 7 Fr., 7.5 Fr. and 8 Fr. IABs.
- Distribution
- US: AK AL AZ CA CO CT FL GA IA ID IL IN KS KY LA MD MI MN MO MS NC ND NE NJ NY OH OK PA SC SD TN TX VA WA WI WV OUS: Switzerland, Netherlands, South Africa, Canada, Germany, Austria, Spain, Italy, Poland, Bharain, Malaysia
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-0934-2020, Z-0932-2020, Z-0933-2020
- Reinforced Introducer Sets Maquet 8 Fr., Part Number: 068400040310. It is an accessory to be used for percutaneous insertion of MAQUET IntraAortic Balloon Catheters.
- Reinforced Introducer Sets Maquet 7 Fr., Part Number: 0684 00 0403 06. It is an accessory to be used for percutaneous insertion of MAQUET Intra Aortic Balloon Catheters.
- Reinforced Introducer Sets Maquet 7.5 Fr., Part Number: 068400040305. It is an accessory to be used for percutaneous insertion of MAQUET IntraAortic Balloon Catheters.
Terminated Class II Medical devices
- Recall started
- 3 October 2019
- Terminated
- 8 January 2021
- Reason
- One lot of "Low Level Output Cable - Interface to Philips Monitor" was received from the supplier with an incorrect cable in the packaging.
- Distribution
- Distributed to CA, GA, and TX.
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-0266-2020
- Low Level Output Cable Interface to Philips Monitor For Use With Cardiosave and CS300 Intra-Aortic Balloon Pumps (IABPs), Part Number 0012-00-1589-03
Ongoing Class I Medical devices
- Recall started
- 17 June 2019
- Reason
- If battery maintenance is not performed appropriately, the battery may provide less than the minimum run time of operating power. Revised labeling and training are being provided.
- Distribution
- Worldwide distribution. US Nationwide, AFGHANISTAN , HUNGARY , PANAMA , ALGERIA , ICELAND , PARAGUAY , ARGENTINA , INDIA , PERU , AUSTRALIA , INDONESIA , PHILIPPINES , AUSTRIA , IRAN , POLAND , AZERBAIJAN , IRAQ , PORTUGAL , BAHRAIN , IRELAND , QATAR , BANGLADESH , ISRAEL , ROMANIA , BELARUS , ITALY
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-1942-2019, Z-1941-2019, Z-1944-2019, Z-1939-2019, Z-1943-2019, Z-1940-2019
- CS100 Intra-Aortic Balloon Pump, English, 220V; 0998-00-3013-55 For cardiovascular use.
- CS100 Intra-Aortic Balloon Pump, English, 110V; 0998-00-3013-53 For cardiovascular use.
- CS300 Intra-Aortic Balloon Pump, English, 220V; 0998-00-3023-55 For cardiovascular use.
- Cardiosave Hybrid Intra-Aortic Balloon Pump (IABP), 0998-00-0800-53 For cardiovascular use.
- CS300 Intra-Aortic Balloon Pump, English, 110V; 0998-00-3023-53 For cardiovascular use.
- Cardiosave Rescue IABP, 0998-00-0800-83 Intra-Aortic Balloon Pump For cardiovascular use.
Terminated Class II Medical devices
- Recall started
- 28 May 2019
- Terminated
- 25 September 2023
- Reason
- The device was distributed with the outer carton labeled with the manufacture date and the expiry date reversed. This showed the product as expired before it was manufactured.
- Distribution
- The products were distributed to the following US states: FL. The products were distributed to the following foreign countries: Australia, Austria, Belgium, Estonia, France, Germany, Ireland, Italy, Kuwait, Lithuania, Malaysia, Mexico, Netherlands, Poland, Russia, Saudi Arabia, Slovakia, Sweden
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-1966-2019
- Sensation Plus 7.5Fr. 40cc Intra-Aortic Balloon Catheter with accessories, Part Number 0684-00-0568-01 Product Usage: The Sensation Plus 7.5Fr. 40cc Intra-Aortic Balloon Catheter with accessories is a device that inc
Terminated Class II Medical devices
- Recall started
- 21 February 2018
- Terminated
- 18 March 2020
- Reason
- The U.S. Instructions For Use for the Mega Intra-Aortic Balloon (IAB) Catheter was missing the Contraindications section.
- Distribution
- US Nationwide Distribution
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-1384-2018, Z-1383-2018, Z-1385-2018
- Intra-Aortic Balloon Catheter Mega 7.5FR 30cc Product Usage: The intra-aortic balloon catheter and accessories are used to provide counter pulsation therapy in the aorta, whereby balloon inflation during diastole and
- Intra-Aortic Balloon Catheter Mega 7.5FR 40cc Product Usage: The intra-aortic balloon catheter and accessories are used to provide counter pulsation therapy in the aorta, whereby balloon inflation during diastole and
- Intra-Aortic Balloon Catheter Mega 8FR 50cc Product Usage: The intra-aortic balloon catheter and accessories are used to provide counter pulsation therapy in the aorta, whereby balloon inflation during diastole and def
Terminated Class II Medical devices
- Recall started
- 19 October 2017
- Terminated
- 24 April 2020
- Reason
- The Better Bladder device provided as part of some custom tubing kits has been recalled by its manufacturer. Seal separation from the housing may allow air to enter the housing and collapse the balloon. This may increase resistance to flow in the venous line and cause a drop in ECMO circuit blood flow.
- Distribution
- Distributed to accounts in AL, AR, CA, DC, FL, IL, LA, MO, NC, PA, SC, TX, and WA.
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-1712-2018, Z-1729-2018, Z-1721-2018, Z-1728-2018, Z-1716-2018, Z-1708-2018, Z-1731-2018, Z-1727-2018, Z-1732-2018, Z-1709-2018, Z-1714-2018, Z-1719-2018 and 16 more
- BEQ-TOP 22702 ECC W/HMOD 30000, Catalog No. 701051179
- BEQ-TOP 22300, Custom Tubing Kit, Catalog No. 701049504
- BEQ-TOP 27610 ECC PACK 3/8, Custom Tubing Kit, Catalog No. 701065622
- BEQ-TOP 22301, Custom Tubing Kit, Catalog No. 701049513
- BEQ-TOP 33700 ECC 3/8" PACK, Custom Tubing Pack with Bioline Coating, Catalog No. 701053845
- Adult ECC Pack BEQ-TOP 22300, Catalog No. 701049504
- BO-TOP 13901 RAPID RESPONSE, Custom Tubing Kit, Catalog No. 701054874
- ECC Small - Respiratory BEQ-TOP 5210, Custom Tubing Kit, Catalog No. 701055604
- and 20 more product lines
Terminated Class II Medical devices
- Recall started
- 27 July 2017
- Terminated
- 7 February 2020
- Reason
- There is a potential for the over-pressure safety (OPS) valve in the tubing kits to not allow flow through.
- Distribution
- The products were distributed to the following US states: IA and FL.
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-2178-2018, Z-2179-2018, Z-2180-2018
- Maquet Cardiopulmonary Bypass Custom Tubing kit BO-TOP 38600 ADULT CPB, Part Number 701056035 Product Usage: The Maquet Cardiopulmonary Bypass Custom Tubing kit is a sterile, non-pyrogenic device for single use only
- Maquet Cardiopulmonary Bypass Custom Tubing kit BEQ-T 4906 PEDIARIC CPB PACK, Part Number 701056041 Product Usage: The Maquet Cardiopulmonary Bypass Custom Tubing kit is a sterile, non-pyrogenic device for single use
- Maquet Cardiopulmonary Bypass Custom Tubing kit BO-TOP 38600 ADULT CPB, Part Number 701056035R01 Product Usage: The Maquet Cardiopulmonary Bypass Custom Tubing kit is a sterile, non-pyrogenic device for single use on
Terminated Class III Medical devices
- Recall started
- 11 March 2016
- Terminated
- 7 June 2019
- Reason
- There is a potential problem with the cuvette where failure of the cuvette would cause an error message on the monitor display of H/S DISCONNECT AT CUVETTE.
- Distribution
- The products were distributed to the following US states: AL, OR, and TX.
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-2106-2018, Z-2104-2018, Z-2105-2018
- Maquet Cardiopulmonary Bypass Custom Tubing kit BEQ-TOP 5909 14 Pack QPID, Part Number 701064992
- Maquet Cardiopulmonary Bypass Custom Tubing kit BEQ-T 5208 ECC SET, Part Number 701054588
- Maquet Cardiopulmonary Bypass Custom Tubing kit BEQ-TOP 15801 Neonatal Special, Part Number 701055490
A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.