Dannoritzer Medizintechnik GmbH & Co. KG recalls

1 recall recorded since 2018, most recently on 6 August 2018. None are listed as ongoing. Status and classification are as published by the FDA.

Recalls1
Ongoing (FDA)0
Terminated or completed1
Class I (FDA)0

Terminated Class III Medical devices

Recall started
6 August 2018
Terminated
22 July 2019
Reason
The Manual cleaning process and the usability of the reprocessing instruction by the user had not been validated. The IFU was modified to remove the manual cleaning method and revised the instructions for use for clarity and readability.
Distribution
US Nationwide in the States of OH, MI, MA, CA, OK, TN, GA, FL, IL, NC, TX, MO.
Initiated
Voluntary: Firm initiated
Recall number
Z-0062-2019

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.