Daavlin Distributing Company recalls

3 recalls recorded since 2018, most recently on 5 March 2026. 3 are listed as ongoing. Status and classification are as published by the FDA.

Recalls3
Ongoing (FDA)3
Terminated or completed0
Class I (FDA)0

Ongoing Class II Medical devices

Recall started
5 March 2026
Reason
Calibrated Output was not calculating correctly - UV Sensor GAIN being applied was incorrect - GAIN seemed to be "4/5" instead of "7" (counts were at ~1000 instead of ~5000).
Distribution
US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2744-2026, Z-2742-2026, Z-2743-2026

Ongoing Class II Medical devices

Recall started
10 December 2025
Reason
Some of the Calibration/Output Certificates had the outputs reversed.
Distribution
US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall number
Z-2295-2026

Ongoing Class II Medical devices

Recall started
26 April 2022
Reason
Software issue, resuming an interrupted treatment will result in swap of utilized calibration treatment distance 0" inches to "9. Resulting in over or under dosing.
Distribution
Worldwide distribution - US Nationwide distribution in the states of AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NY, OH, OK, OR, PA, SD, TN, TX, VA, WA, WI, WV and the countries of Australia, Canada.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1073-2022, Z-1074-2022