Cytrellis Biosystems, Inc. recalls

2 recalls recorded since 2018, most recently on 23 June 2023. 1 is listed as ongoing. Status and classification are as published by the FDA.

Recalls2
Ongoing (FDA)1
Terminated or completed1
Class I (FDA)0

Ongoing Class II Medical devices

Recall started
23 June 2023
Reason
Potential failure of a bearing adhesive joint that can occur due to an assembly issue of the X/Y stage of the handpiece. If a broken glue bond were to occur, it could cause an overlap of cores, potentially leading to prolonged healing and irregularities in patient skin texture post-treatment, and/or scarring
Distribution
Nationwide
Initiated
Voluntary: Firm initiated
Recall number
Z-2393-2023

Completed Class II Medical devices

Recall started
7 April 2023
Reason
Irregular core patterns could cause potential overlap of cores at the 8% setting, potentially leading to prolonged healing, irregularities in patient skin texture post-treatment, and/or scarring.
Distribution
Nationwide
Initiated
Voluntary: Firm initiated
Recall number
Z-1648-2023

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.