Cytrellis Biosystems, Inc. recalls
2 recalls recorded since 2018, most recently on 23 June 2023. 1 is listed as ongoing. Status and classification are as published by the FDA.
Recalls2
Ongoing (FDA)1
Terminated or completed1
Class I (FDA)0
Ongoing Class II Medical devices
- Recall started
- 23 June 2023
- Reason
- Potential failure of a bearing adhesive joint that can occur due to an assembly issue of the X/Y stage of the handpiece. If a broken glue bond were to occur, it could cause an overlap of cores, potentially leading to prolonged healing and irregularities in patient skin texture post-treatment, and/or scarring
- Distribution
- Nationwide
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-2393-2023
- ellacor System with Micro-Coring Technology-Indicated for use by medical professionals for the treatment of moderate and severe wrinkles in the mid and lower face in adults aged 22 years or older with Fitzpatrick skin ty
Completed Class II Medical devices
- Recall started
- 7 April 2023
- Reason
- Irregular core patterns could cause potential overlap of cores at the 8% setting, potentially leading to prolonged healing, irregularities in patient skin texture post-treatment, and/or scarring.
- Distribution
- Nationwide
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-1648-2023
- Cytrellis ellacor System with Micro-Coring Technology-Dermal Micro-Coring System System intended for the treatment of moderate and severe wrinkles in the mid and lower face. Model #: Gen 4
A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.