Cyberonics, Inc recalls

2 recalls recorded since 2018, most recently on 2 December 2017. None are listed as ongoing. Status and classification are as published by the FDA.

Recalls2
Ongoing (FDA)0
Terminated or completed2
Class I (FDA)0

Terminated Class II Medical devices

Recall started
2 December 2017
Terminated
6 August 2018
Reason
Two Model 3000 Programmers were distributed in error by prior to FDA approval of version 1.0.2.2 software.
Distribution
US Distribution to the states of : CO and ND
Initiated
Voluntary: Firm initiated
Recall number
Z-0555-2018

Terminated Class II Medical devices

Recall started
13 November 2017
Terminated
31 October 2022
Reason
Certain Model 3000 programming events can result in miscalculation of parameters stored in the Models 103, 104, 105, and 106 generators. During these programming events, the miscalculations can lead to: " Delivery of more stimulation than intended, resulting in painful stimulation or other common side effects (Model 106 only); " No stimulation in the case of device disablement (Burst Watchdog Timeout), resulting in no therapy to the patient (Model 106 only); " Delivery of less stimulation than intended, resulting in therapeutic settings not being achieved within device specification (Models 103, 104, 105, or 106); and/or " Delays or absence of the 75% and 50% battery life indicators displayed by the programming software (Models 103, 104, 105, or 106).
Distribution
AL, CA, CO, DE, FL, GA, ID, IL, IN, ME, MO, MS, NC, NJ, NY, PA, TN, TN, TX, UT, WA and WI
Initiated
Voluntary: Firm initiated
Recall number
Z-2255-2018

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.