CTL Medical Corporation recalls

2 recalls recorded since 2018, most recently on 21 August 2019. None are listed as ongoing. Status and classification are as published by the FDA.

Recalls2
Ongoing (FDA)0
Terminated or completed2
Class I (FDA)0

Completed Class II Medical devices

Recall started
21 August 2019
Reason
Reports received of the device fracturing during insertion.
Distribution
Distribution was made to AZ, FL, IN, KS, MA, MI, NY, TN, and TX.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2502-2020, Z-2504-2020, Z-2503-2020, Z-2501-2020

Terminated Class II Medical devices

Recall started
5 July 2018
Terminated
28 April 2021
Reason
the anti-toggle ring of the hex driver may separate from the driver shaft. If this occurs during surgery, the anti-toggle ring can be pressed into the pedicle screw head and cause separation of the screw from the housing upon tightening. if the anti-toggle ring remains in the pedicle crew head and can interfere with final tightening of the pedicle screw construct. this can lead to failure of the set crew and relies of the rod from the construct.
Distribution
distributor and user level
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1030-2019, Z-1031-2019

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.