Covidien, LP recalls

96 recalls recorded since 2018, most recently on 24 June 2026. 63 are listed as ongoing. Status and classification are as published by the FDA.

Recalls96
Ongoing (FDA)63
Terminated or completed33
Class I (FDA)11

Ongoing Class II Medical devices

Recall started
24 June 2026
Reason
Product may contain a perforation or tear in the sterile barrier packaging. This issue may result in biocontamination of the sterile sutures, which may lead to infection, loss of vision if in ophthalmic utilization, or prolonged surgery.
Distribution
US Nationwide and the countries of Australia, Austria, Belgium, Bulgaria, Canada, Canary Islands, Chile, China, Colombia, Croatia, Cyprus, Czech Republic, Denmark, Estonia, Finland, France, Germany, Greece, Guadeloupe, Guatemala, Honduras, Hong Kong, Hungary, Iceland, India, Ireland, Israel, Italy,
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2919-2026, Z-2922-2026, Z-2918-2026, Z-2921-2026, Z-2920-2026

Ongoing Class II Medical devices

Recall started
13 May 2026
Reason
The system's 24-hour usage limit results in catheters incorrectly flagged as expired, making them unavailable for clinical procedures.
Distribution
Worldwide - U.S. Nationwide distribution including in the states of AK, AZ, CA, CO, CT, DC, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, NE, NH, NJ, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, VT, WA, WI, and WV. The countries of Argentina, Australia, Austria, Belgium, Canada, Col
Initiated
Voluntary: Firm initiated
Recall number
Z-2950-2026

Ongoing Class II Medical devices

Recall started
23 April 2026
Reason
Temperature probe devices lack FDA clearance.
Distribution
US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, FL, GA, HI, IL, LA, MA, MD, MI, MN, MO, MS, NC, NE, NJ, NY, OH, OK, OR, PA, PR, TN, TX, UT, VA, WA, WV.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2273-2026, Z-2272-2026, Z-2271-2026

Ongoing Class II Medical devices

Recall started
23 April 2026
Reason
Due to product labeling incorrectly including the symbol indicating non-DEHP (Bis(2-ethylhexyl) phthalate as the product includes 30 to 35% DEHP.
Distribution
U.S. Nationwide distribution including in the states and territories of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and Virgin Islands. The c
Initiated
Voluntary: Firm initiated
Recall number
Z-2653-2026

Ongoing Class II Medical devices

Recall started
15 April 2026
Reason
Due to customer complaint regarding incorrect display box labeling.
Distribution
Worldwide - U.S.: Nationwide distribution in the state of TX and the countries of France, French, Polynesia, Italy, Pakistan, Poland, Portugal, Spain, and Taiwan.
Initiated
Voluntary: Firm initiated
Recall number
Z-2261-2026

Ongoing Class II Medical devices

Recall started
15 April 2026
Reason
The presence of holes in the pouch may compromise the package s ability to maintain the sterile barrier and protect the product from environmental exposure.
Distribution
Domestic: Nationwide distribution; International: Austria, Canada, Denmark, France, Italy;
Initiated
Voluntary: Firm initiated
Recall number
Z-2281-2026

Ongoing Class II Medical devices

Recall started
30 January 2026
Reason
Due to complaints and investigations stating that alarms from the primary patient bedside monitor was not alarming/being transmitted and received on the remote patient monitoring system.
Distribution
Worldwide - U.S. Nationwide distribution in the states of AK, AL, AZ, CA, FL, GA, IL, IN, KY, LA, MI, MN, MS, NC, NY, OH, OK, OR, SC, TN, TX, and VA. The country of Spain.
Initiated
Voluntary: Firm initiated
Recall number
Z-1557-2026

Ongoing Class II Medical devices

Recall started
25 September 2025
Reason
A manufacturing issue could lead to a malformed cover tube press which can lead to internal components' movement during firing - leading to the potential for uncontrolled motion.
Distribution
Worldwide distribution - US Nationwide and the countries of Austria, Belgium, Chile, Estonia, Finland, France, Germany, Greece, Hungary, Israel, Italy, Japan, Kuwait, Malaysia, Netherlands, Poland, Portugal, Romania, Saudi Arabia, Slovakia, Spain, United Kingdom.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0486-2026, Z-0485-2026

Ongoing Class II Medical devices

Recall started
23 June 2025
Reason
Bedside SpO2 Patient Monitoring System alarms not heard/recognized, resulting in: treatment delay, lack of low oxygen saturation response, respiratory failure, arrhythmia. Highlighting operator's manual: 1)For patient Safety don't silence/decrease alarm volume, 2)Sensor off alarm (no perfusion detected) is low priority alarm, 3)Set monitors to Homecare Mode outside of hospital/professional setting
Distribution
Worldwide - US Nationwide distribution including in the states of MS, TX, FL, SC, VA, AZ, PA, MN, OH, CA, MI, CO, KS, IN, IL, TN, MO, IA, OK, WI, AK, NY, MD, AL, WV, GA, NC, AR, WA, SD, KY, DE, NJ, UT, NV, MA, MT, GU, NE, LA, HI, OR, RI, NH, ND, NM, ME, CT, DC, WY and the countries of Algeria, Andor
Initiated
Voluntary: Firm initiated
Recall number
Z-2151-2025

Ongoing Class I Medical devices

Recall started
21 May 2025
Reason
Ventilator Printed Circuit Board Assembly may have two separate capacitors that may fail, which may result in the ventilator either shutting down during use, thus necessitating use of an alternate form of ventilation, or the shutdown alert alarm fails to alarm effectively during shut down, which may result in respiratory failure, hypoventilation, low oxygen saturation, hypoxia, treatment delay.
Distribution
Worldwide - US Nationwide distribution including in the states of MA, CT, KS, OH, PR, NH, GA, WI, MS, VA, CO, TX, MN, AL, IA, NY, UT, NM, KY, RI, SD, PA, IA, CA, MO, HI, MI, IL, MT, NJ, TN, SC, FL, IN, AZ, ID, LA, WA, NC, MD, WY, NE, NV, ND, ME, AK, DE, OK, OR and the countries of Japan, Canada, S
Initiated
Voluntary: Firm initiated
Recall number
Z-1915-2025

Ongoing Class I Medical devices

Recall started
26 February 2025
Reason
Tracheostomy tube with cuff reusable inner cannula has a flange that may disconnect from the outer cannula of the device, which could cause respiratory failure, aspiration, unspecified tissue injury, aspiration, respiratory tract infection, bronchospasm, a delay to treatment and/or death.
Distribution
OUS distribution only to countries of: France, Germany, Ireland, Italy, Lithuania, Portugal, Saudi Arabia, Slovakia, Slovenia, South Africa, Spain, Taiwan
Initiated
Voluntary: Firm initiated
Recall number
Z-1390-2025

Ongoing Class II Medical devices

Recall started
19 September 2024
Reason
Specific lots of the Chameleon PTA Balloon Catheter failed to meet in-house balloon pressure testing.
Distribution
Worldwide Distribution: US (Nationwide) and OUS (foreign) to countries of: Austria, Canada, Canary Islands, China, France, French Guiana, Germany, Greece, Guadeloupe, India, Ireland, Israel, Italy, Japan, Netherlands, Portugal, Singapore, Spain and Turkey.
Initiated
Voluntary: Firm initiated
Recall number
Z-0257-2025

Ongoing Class I Medical devices

Recall started
9 July 2024
Reason
Battery management system within Laryngoscope devices may deplete below the deign threshold which may result in thermal event followed by risk of explosion.
Distribution
Worldwide - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, Puerto Rico, Virgin Islan
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2636-2024, Z-2637-2024

Ongoing Class II Medical devices

Recall started
9 July 2024
Reason
IFU update to address device's battery handling information.
Distribution
Worldwide - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, Puerto Rico, US Virgin I
Initiated
Voluntary: Firm initiated
Recall number
Z-2717-2024

Ongoing Class I Medical devices

Recall started
24 June 2024
Reason
Any single affected ventilator should only be used for the 10- year labelled service life, but if longer ventilatory support is required using more than one ventilator over time, the use of these devices for more than 14 years of cumulative duration may pose harm to health due to a volatile organic compound (VOC), 2-propanol, 1,3-dichloro released from some components in the ventilator gas pathway
Distribution
US: KY, OH, GA, MN, IN,UT, IL, NJ, NY, PR, LA, MD, MO, TN, MA, ME, CA, SC, AR, TX, WI, PA, CO, MI, NM, RI, VA, DC, VT, NC, NH, WV, NV, IA. OUS: Argentina, Australia, Bangladesh, Belgium, Botswana, Brazil, Bulgaria, Cambodia, Canada, Canary Islands, Chile, China, Colombia, Costa Rica, Czech Republic
Initiated
Voluntary: Firm initiated
Recall number
Z-2446-2024

Ongoing Class II Medical devices

Recall started
25 April 2024
Reason
Due to saline conductivity issues, their is a potential that dilation catheters may provide inaccurate esophageal diameter measurements that may lead to various harms.
Distribution
Worldwide - US Nationwide distribution including in the states of AK, AL, AZ, CA, CO, CT, DC, FL, GA, IA, IL, KS, KY, LA, MA, MD, MI, MN, MO, NC, NE, NJ, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, WI and the countries of Argentina, Australia, Austria, Bahrain, Belgium, Canada, Denmark, Germa
Initiated
Voluntary: Firm initiated
Recall number
Z-2038-2024

Ongoing Class II Medical devices

Recall started
15 April 2024
Reason
Potential damage to the cartridge can result in poor staple formation and/or incomplete staple line, may cause a delay in surgery
Distribution
Worldwide - US Nationwide distribution in the state of NJ and the countries of Australia, Japan, New Zealand, Saudi Arabia, Singapore, South Africa, Taiwan, Thailand.
Initiated
Voluntary: Firm initiated
Recall number
Z-1891-2024

Ongoing Class II Medical devices

Recall started
15 March 2024
Reason
Products may not be heparin coated as indicated on the label-leads to a potential loss of benefit due to the absence of the heparin coating
Distribution
Worldwide distribution - US Nationwide and the countries of Austria, Belgium, Canary Islands, Denmark, Faroe Islands, France, Germany, Greece, Iceland, Ireland, Italy, Mayotte, Norway, Poland, Reunion, South Africa, Spain, Sweden, United Kingdom.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1634-2024, Z-1635-2024

Ongoing Class II Medical devices

Recall started
28 February 2024
Reason
The potential for trocar seal disengagement when using mesh products incorrectly with the device.
Distribution
Worldwide distribution - US Nationwide and the countries of Albania, Algeria, Aruba, Australia, Austria, Belgium, Brazil, Bulgaria, Canada, Canary Islands, Colombia, Croatia, Czech Republic, Denmark, Estonia, Finland, France, French Guiana, French Polynesia, Germany, Gibraltar, Greece, Guadeloupe, H
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1470-2024, Z-1468-2024, Z-1467-2024, Z-1471-2024, Z-1466-2024, Z-1465-2024, Z-1469-2024

Ongoing Class II Medical devices

Recall started
21 February 2024
Reason
Packs including 5 laparotomy sponges may contain incorrect product with a blue loop, X-ray detectable feature, and ethylene oxide treatment; however, this incorrect product wasn't steam pre-treated prior to sterilization, and has no RF tag, which if non-sterile, may lead to infection, tissue trauma, and sepsis, and if in the scannable area, would lead a "CLEAR" display.
Distribution
US Nationwide distribution including in the states of AK, AL, AZ, CA, CO, FL, GA, IA, IL, IN, KS, LA, MA, MD, MI, MN, MO, MS, NC, NE, NJ, NM, NY, OH, OK, PA, RI, TN, TX, UT, VA, WA, WI, WV.
Initiated
Voluntary: Firm initiated
Recall number
Z-1330-2024

Ongoing Class II Medical devices

Recall started
14 February 2024
Reason
Under certain firing conditions, reloads were found to articulate in an uncontrolled manner potentially causing disruption to the staple line. The issue is related to components in impacted reloads that are not fully secure, which leads to uncontrolled articulation. The issue can occur using either the Endo GIA" Ultra Universal Stapler or Signia" Stapling System.
Distribution
Worldwide distribution - US Nationwide and the countries of France, Hong Kong, Israel and Japan.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1457-2024, Z-1458-2024

Ongoing Class II Medical devices

Recall started
25 January 2024
Reason
One lot of product was exposed to more than the approved number of Ethylene Oxide (EO) sterilization cycles. This may decrease the tensile strength of the sutures over time, which could result in harms such as wound dehiscence, hemorrhage, blood loss, bleeding, tissue breakdown, peritonitis, unspecified infection, vision loss (when used in ophthalmic application), and/or prolonged surgery.
Distribution
International distribution in the countries of Portugal and Italy.
Initiated
Voluntary: Firm initiated
Recall number
Z-1697-2024

Ongoing Class II Medical devices

Recall started
9 January 2024
Reason
manufacturing defects to sensor cable connectors may result in error codes or non-functioning channel
Distribution
US: AK AZ CA CT FL GA IA IN LA MA MD MI MN MO NC NY OH OK PA SC TN TX VA WA WI OUS: Australia Austria Belgium Canary Islands Chile Czech Republic Denmark Finland France Germany Guatemala Indonesia Ireland Italy Japan Korea, Republic Of Kuwait Lithuania Mayotte New Zealand Nicaragua Poland Portugal R
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1731-2024, Z-1732-2024

Ongoing Class II Medical devices

Recall started
21 December 2023
Reason
Specific lots of sutures were sterilized with gamma doses that exceeded the range approved. The extra gamma levels may decrease the strength of the sutures over time, which could result in harms such as wound dehiscence and hemorrhage/blood loss/ bleeding at a critical level.
Distribution
Worldwide US Nationwide distribution including in the states of AL, AZ, CA, FL, GA, HI, IA, IL, IN, MD, MI, MN, MO, MS, NC, NE, NJ, NV, NY, OH, OK, OR, PA, PR, TN, TX, VA, WA, WV and the countries of Australia, Bahrain, Belgium, Bulgaria, Canary Islands, China, Cyprus, Czech Republic, Finland, Fra
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0844-2024, Z-0845-2024

Ongoing Class II Medical devices

Recall started
16 November 2023
Reason
Packaging of the device may cause a low seal strength in the side seal of the foil pouch potentially resulting in a sterile/moisture barrier breach, compromising the product sterility
Distribution
US Nationwide distribution in the states of CA, CO, FL, IL, TX.
Initiated
Voluntary: Firm initiated
Recall number
Z-0986-2024

Ongoing Class II Medical devices

Recall started
16 November 2023
Reason
Due to software issue, New systems may exhibit an error which may prohibit use of new systems resulting in system being inoperable
Distribution
Worldwide distribution - US: Nationwide and surrounding territories.
Initiated
Voluntary: Firm initiated
Recall number
Z-0631-2024

Ongoing Class I Medical devices

Recall started
28 September 2023
Reason
Defective video laryngoscopes that did not pass quality tests established for the product, that were designated for destruction, were stolen, and subsequently distributed for sale. Defective device use may result in failed laryngoscopy and/or intubation.
Distribution
Unknown
Initiated
Voluntary: Firm initiated
Recall number
Z-0144-2024

Ongoing Class II Medical devices

Recall started
22 August 2023
Reason
Some catheter kits labeled as 23 cm implant length incorrectly included catheters of 28 cm implant length; the actual catheter implant length is identifiable based upon the correct labeling on the catheter body.
Distribution
Worldwide distribution - US Nationwide and the countries of Austria, Belgium, Brazil, China, Czech Republic, Finland, France, Germany, Italy, Netherlands, Norway, Poland, Portugal, Puerto Rico, Spain, Thailand, United Kingdom.
Initiated
Voluntary: Firm initiated
Recall number
Z-2632-2023

Ongoing Class II Medical devices

Recall started
22 August 2023
Reason
Inadequate internal fixation of the power inlet port to the cuff pressure manager device causes the external power inlet to became dislodged within the device housing after several uses, which prevents the device from being charged, and could lead to treatment delay.
Distribution
US: AL, NJ, MA, MI, NY, NC, CO, TX, FL, KY, UT, CA, AZ, KS, IL, TN. OUS: El Salvador, Sweden, Canary Islands
Initiated
Voluntary: Firm initiated
Recall number
Z-2651-2023

Ongoing Class II Medical devices

Recall started
10 August 2023
Reason
Laryngoscope blades may contain an ineffective application of anti-fog agent which may result in condensation accumulating causing the screen to be blurred or obscured.
Distribution
US: AK AL AR AZ CA CO CT DE FL GA IA IL IN KS LA MA MD MI MO MT NC ND NH NJ NM NY OH OK OR PA SD TN TX UT VA WA WI WY OUS: Australia Austria Azerbaijan Bahamas Belgium Croatia Czech Republic Denmark Finland France French Guiana Germany Guadeloupe Hong Kong Hungary Ireland Israel Italy Kazakhstan Ko
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0168-2024, Z-0169-2024

Ongoing Class I Medical devices

Recall started
28 June 2023
Reason
Catheter center lumen was found to have an occlusion in the tip of the catheter; the source of the occlusion an excessive MDX, a silicone-based lubricant which coats the catheter tip. An occluded, partially or fully, and/or uncured or excessive MDX may dislodge from the catheter lead to full catheter obstruction resulting in delay to treatment and partial obstruction resulting in reduced flow, or particulate dislodgement that may result in delay to treatment, hemolysis, embolism/embolus or thrombosis/thrombus
Distribution
US Nationwide - Worldwide Distribution Foreign: Argentina Austria Bahrain Bangladesh Belgium Brazil Brunei Darussalam Canada Cayman Islands China Costa Rica Czech Republic Denmark Dominican Republic Finland France French Polynesia Germany Greece Hungary India Iran, Islamic Republic Of Ireland Ital
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2372-2023, Z-2388-2023, Z-2391-2023, Z-2383-2023, Z-2371-2023, Z-2384-2023, Z-2379-2023, Z-2390-2023, Z-2376-2023, Z-2369-2023, Z-2378-2023, Z-2380-2023 and 13 more

Ongoing Class II Medical devices

Recall started
2 June 2023
Reason
Cell Collection Device may be at increased risk of the sponge detaching from the string during removal of the device from the patient, which could lead to device fragments in patient, obstruction, airway obstruction, secondary intervention, secondary intervention (with the primary procedure), supraglottic airway obstruction, and aspiration.
Distribution
Worldwide - US Nationwide distribution in the states of NY, MN, VA, WI and the countries of Canada, Netherlands, Poland, United Kingdom.
Initiated
Voluntary: Firm initiated
Recall number
Z-2282-2023

Ongoing Class II Medical devices

Recall started
22 May 2023
Reason
Single ILLUMISITE Platform Console was converted from a 220-240VAC model to a 120VAC configuration using a non-supplied power cord and non-supplied fuses in the field by Medtronic Field Service Engineer.
Distribution
International distribution in the country of Taiwan.
Initiated
Voluntary: Firm initiated
Recall number
Z-2008-2023

Ongoing Class I Medical devices

Recall started
24 March 2023
Reason
A manufacturing error, resulted in a less than specified diameter of adult flexible tracheostomy tube connectors, which results in unsecure connections with 15mm caps and other 15mm circuit components and accessories. An unsecure connection could lead to respiratory failure, dyspnea, treatment delay, tissue injury, or bleeding.
Distribution
Worldwide - US Nationwide distribution including in the states of PR, CA, FL, MN, PA, AZ, IL, TN, OH, MI, GA, KS, HI, MD, AK, NY, NE, IA, OK, VA, ND, MO, NC, SC, KY, OR, IN, WA, TX, WI, CT, MA, WV, CO, VT, NJ, SD, AR, AL, MS, LA, DE, MT, WY, DC, UT, NV, NH, RI, ID, NM, ME and the countries of Algeri
Initiated
Voluntary: Firm initiated
Recall number
Z-1357-2023

Ongoing Class II Medical devices

Recall started
22 March 2023
Reason
There is a manufacturing non-conformance that may result in a metal shaving (burr) on the inside of the VersaOneTM Reusable Positioning Cannula, located where the Cannula and Body components are welded together.
Distribution
Worldwide - US Nationwide distribution in the states of NC and the countries of Belgium, Denmark, France, Germany, India, Italy, Netherlands, Spain, Sweden.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1464-2023, Z-1465-2023, Z-1463-2023

Ongoing Class II Medical devices

Recall started
1 March 2023
Reason
Medtronic was made aware of a supplier calibration issue during the contract sterilization process that may lead to unsterile product. Use of affected product may lead to an unspecified infection or, in the case of sutures used in the eye, pose a loss of vision risk.
Distribution
Worldwide Distribution: US (nationwide) to States of: AR, CA, FL, GA, LA, MA, NC, NY, OH, WA; and countries of: Canada, Colombia, El Salvador, Indonesia, Nepal, Puerto Rico, Taiwan, and Thailand.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1387-2023, Z-1388-2023, Z-1390-2023, Z-1389-2023

Ongoing Class II Medical devices

Recall started
10 February 2023
Reason
Due to potential leaks associate with the Oral/Nasal Tracheal Tube Cuffed inflation system.
Distribution
U.S.: AK, AZ, CA, FL, GA, IA, IL, IN, KS MD, MI, MS, NC, NE, NJ, NY, OH, OK, PA, SC, TX, UT, VA, WA, and WV O.U.S.: Canada
Initiated
Voluntary: Firm initiated
Recall number
Z-1318-2023

Ongoing Class II Medical devices

Recall started
25 January 2023
Reason
Affected lots have the potential for a broken sled vane, which may cause the reload to misfire leading to non-functional staple line closure, transecting tissue without forming staples, and tissue hang-up. These conditions may be associated to a delay to treatment, unspecified infection, hemorrhage/blood loss/bleeding, failure to anastomose, peritonitis, sepsis, pneumothorax, tissue trauma, or death.
Distribution
Worldwide distribution - US Nationwide distribution in the states of Arkansas, California, Florida, Georgia, Louisiana, Massachusetts, New York, North Carolina, Ohio, Washington and the countries of Australia, Austria, Belgium, Canada, Canary Islands, Denmark, Finland, France, Germany, Ireland, It
Initiated
Voluntary: Firm initiated
Recall number
Z-1228-2023

Ongoing Class II Medical devices

Recall started
11 January 2023
Reason
Collagen film was placed on the opposite side of the mesh may cause delay to treatment/therapy, adhesions, erosion/migration, pain, fistula, hernia (recurrence), and failure of implant
Distribution
US Distribution to states of: CA, GA, NY and OUS Foreign distribution to countries of: Armenia, Belgium, France, Germany, Israel, Japan, Luxembourg, Romania, Spain, Switzerland and United Kingdom.
Initiated
Voluntary: Firm initiated
Recall number
Z-1183-2023

Ongoing Class I Medical devices

Recall started
7 December 2022
Reason
There is a potential leaking condition within the hub of affected devices, due to an inter-lumen void in the hub component, which may result in unanticipated fluid return during use.
Distribution
Worldwide distribution. US nationwide including Guam, USVI, and Puerto Rico; Australia, Austria, Belgium, Brazil, Canada, Chile, China, Colombia, Costa Rica, Czech Republic, Denmark, Dominican Republic, Ecuador, Finland, France, French Polynesia, Germany, Greece, Hungary, Iceland, Italy, Jamaica, K
Initiated
Voluntary: Firm initiated
Recall number
Z-0877-2023

Ongoing Class II Medical devices

Recall started
2 December 2022
Reason
Sterile Barrier Breach of Breather Pouch may cause infection or vision loss for ophthalmic procedures in cases where there is direct patient contact
Distribution
US Nationwide Distribution and OUS Foreign distribution to countries of: Albania Argentina Armenia Australia Austria Azerbaijan Belarus Belgium Bosnia And Herzegovina Brazil Bulgaria Cambodia Cameroon Canada Canary Islands Chile China Colombia Croatia Czech Republic Denmark Egypt Finland France Geor
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1178-2023, Z-1172-2023, Z-1182-2023, Z-1181-2023, Z-1177-2023, Z-1176-2023, Z-1174-2023, Z-1180-2023, Z-1173-2023, Z-1175-2023, Z-1179-2023

Ongoing Class II Medical devices

Recall started
23 November 2022
Reason
The product is labeled with the incorrect expiration date.
Distribution
Foreign Distribution: Hong Kong
Initiated
Voluntary: Firm initiated
Recall number
Z-0955-2023

Ongoing Class II Medical devices

Recall started
28 September 2022
Reason
Medtronic has received 210 reports of needles breaking during endoscopic suturing. The potential harm(s) include a delay in treatment/therapy, the potential for a portion of the needle to remain in the patient, foreign body reaction, allergic reaction, tissue injury and unintended radiation exposure from additional imaging for both intra-op and postoperative scenarios.
Distribution
Domestic distribution nationwide. Foreign distribution to Australia Austria Belgium Canada Canary Islands Chile Croatia Cyprus Denmark Egypt Finland France Germany Greece Hungary Ireland Israel Italy Jordan Kazakhstan Kuwait Lithuania Luxembourg Mexico Netherlands New Zealand Norway Poland Portugal
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0200-2023, Z-0201-2023

Ongoing Class II Medical devices

Recall started
19 August 2022
Reason
Medtronic has confirmed six reports from customers outside the US where the ventilator became inoperable due to loss of gas supply caused by manufacturing error of specific turbine components.
Distribution
Domestic distribution to one customer in Washington, D.C. Foreign distribution worldwide.
Initiated
Voluntary: Firm initiated
Recall number
Z-1775-2022

Ongoing Class I Medical devices

Recall started
8 June 2022
Reason
There is a potential leaking condition within the hub of specific chronic dialysis catheters, due to an inter-lumen void in the hub component, which may result in unanticipated fluid return during use.
Distribution
Worldwide distribution - US Nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, DC, GU, & PR. The countries of
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1312-2022, Z-1315-2022, Z-1317-2022, Z-1318-2022, Z-1320-2022, Z-1316-2022, Z-1313-2022, Z-1314-2022, Z-1319-2022

Ongoing Class II Medical devices

Recall started
7 June 2022
Reason
Electrical component failure within the control unit. The observed failure mode is a loss of function of the device when used with handpiece and foot switch.
Distribution
AZ CA CO IL KS MA MI MN MO NY OK PA TN TX TX WA Foreign: United Kingdom Ireland Netherlands Reunion Spain
Initiated
Voluntary: Firm initiated
Recall number
Z-1467-2022

Ongoing Class II Medical devices

Recall started
13 May 2022
Reason
The affected staplers have the potential for the staple guide to not be securely attached to the instrument. The issue could cause the component to disengage and if disengaged, could allow the device to transect tissue without forming staples.
Distribution
Worldwide distribution - US Nationwide.
Initiated
Voluntary: Firm initiated
Recall number
Z-1322-2022

Ongoing Class II Medical devices

Recall started
22 December 2021
Reason
Prior to use on a patient, the device alarm speaker failed to annunciate on power up and the device falsely detected a sensor cable disconnect and reconnect during setup. The monitor detection of a false sensor cable disconnect and reconnect may cause the monitor to discontinue monitoring and transition into the calibration screen without an audible alarm.
Distribution
The devices were distributed to the following US states: CA, CO, FL, IL, KY, MA, OH, TN, and TX.
Initiated
Voluntary: Firm initiated
Recall number
Z-0514-2022

Ongoing Class II Medical devices

Recall started
17 December 2021
Reason
The firm has received six (6) complaint reports from customers indicating difficulty removing the balloon protector sleeve from the catheter (e.g., excessive force required to remove the protector sleeve). If the balloon is damaged while removing the protector sleeve and the catheter is subsequently used in a patient procedure, patient harm may potentially occur, including perforation of vessels, foreign body in patient, embolism, risks associated with surgical intervention, or delay to treatment/therapy.
Distribution
Worldwide distribution - US Nationwide and the countries of Germany, Israel, Italy, Portugal, Spain, Chile, Singapore, and China.
Initiated
Voluntary: Firm initiated
Recall number
Z-0557-2022

Terminated Class II Medical devices

Recall started
6 December 2021
Terminated
11 May 2023
Reason
Customers reported that the device jaws were difficult to open or would not open following application on tissue. Use of a device with this potential assembly issue could result in the potential for bleeding, unintended tissue injury or loss, and delay in treatment while an alternate device is obtained.
Distribution
Worldwide Distribution in the countries of Australia, Austria, Belgium, Czech Republic, France, Germany, Ireland, Kazakhstan, Netherlands, Spain, United Kingdom. No US distribution.
Initiated
Voluntary: Firm initiated
Recall number
Z-0495-2022

Ongoing Class II Medical devices

Recall started
12 November 2021
Reason
Intermittent risk of electric shock to users if an affected charger is used in conjunction with a faulty or bypassed institutions Mains Protective Earth (Ground) circuit
Distribution
Brazil
Initiated
Voluntary: Firm initiated
Recall number
Z-0332-2022

Terminated Class I Medical devices

Recall started
4 November 2021
Terminated
9 July 2024
Reason
Manufacturing assembly error where a capacitor within the ventilator was assembled incorrectly, which may cause the device to become inoperable during use.
Distribution
Worldwide distribution - US Nationwide.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0351-2022, Z-0343-2022, Z-0342-2022, Z-0352-2022, Z-0348-2022, Z-0341-2022, Z-0339-2022, Z-0346-2022, Z-0340-2022, Z-0347-2022, Z-0338-2022, Z-0349-2022 and 4 more

Terminated Class II Medical devices

Recall started
29 September 2021
Terminated
13 February 2025
Reason
The firm identified an error in the device temperature monitoring circuitry which is designed to interrupt energy delivery in the event of a primary cooling system failure. Use of an affected device in conjunction with a primary cooling system failure could lead to the applicator tip overheating and associated heat damage which could result in an allergic/toxic reaction, unintended tissue ablation, tissue burn, and/or a delay of treatment while an alternate device is located.
Distribution
No domestic distribution. Foreign distribution to Hong Kong and UK.
Initiated
Voluntary: Firm initiated
Recall number
Z-0210-2022

Terminated Class II Medical devices

Recall started
23 June 2021
Terminated
28 January 2025
Reason
Potential for a manufacturing assembly error-may result in a non-functional audio indicator of device activation or a non-functional device, may cause unintended thermal burn, hemorrhage, or tissue damage and a delay of treatment while an alternate device is located
Distribution
US Nationwide Distribution: AL, AZ, CT, GA, IN, MA, MD, MN, NY, OH, TX, VA and WA.
Initiated
Voluntary: Firm initiated
Recall number
Z-2240-2021

Ongoing Class II Medical devices

Recall started
28 May 2021
Reason
This voluntary withdrawal is being conducted to implement point-of-use labeling on each device and enhancements to the Instructions for Use (IFU).
Distribution
Worldwide distribution.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2373-2021, Z-2374-2021, Z-2369-2021, Z-2371-2021, Z-2361-2021, Z-2370-2021, Z-2368-2021, Z-2358-2021, Z-2367-2021, Z-2359-2021, Z-2372-2021, Z-2365-2021 and 6 more

Ongoing Class II Medical devices

Recall started
29 April 2021
Reason
The company was notified by their supplier of sterilization services for airway products in Italy of potential deviations from validated parameters for ethylene oxide sterilization.
Distribution
Worldwide distribution - US Nationwide distribution in the states of DE, KY, NY, TX, VA, VT, and WA. Global Distribution.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1868-2021, Z-1867-2021, Z-1866-2021, Z-1869-2021

Terminated Class II Medical devices

Recall started
23 March 2021
Terminated
15 May 2024
Reason
Potential for the pouch to contain an extra laminate layer, creating the potential for the device sterility to be compromised. The inclusion of an additional laminate layer in the package may also impact complete de-gassing of the sterilization gas, Ethylene Oxide and may result in a potentially increased risk for patient infection and/or potential exposure to Ethylene Oxide retained in the pouch
Distribution
AL, AR, FL, IN, LA, MA , MN, MO, NJ, NY, OR, PA, TX
Initiated
Voluntary: Firm initiated
Recall number
Z-1419-2021

Ongoing Class II Medical devices

Recall started
5 March 2021
Reason
Investigations related to customer complaints received for particulate matter inside of the Surgiwand" II device tubing components identified the root cause as a damaged Y-Connector component which connects the tubing to the device. This damage was determined to generate the particulate matter observed and is created during the assembly process for connecting the tubing component to the Y-Connector.
Distribution
Domestic: AK, AL, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MO, MS, NC, NE, NH, NJ, NM, NY, OH, OR, PA, SC, TX, UT, WA, WI, WV. Foreign: Canada, Japan, Taiwan, Australia, New Zealand, Indonesia, Thailand, Singapore, Malaysia, Philippines, Bahamas, Brazil, Chile, El Salvador
Initiated
Voluntary: Firm initiated
Recall number
Z-1411-2021

Terminated Class II Medical devices

Recall started
4 March 2021
Terminated
12 July 2024
Reason
Potential for the device safety interlock to fail. The safety interlock prevents an empty single-use reload from being erroneously fired a second time. If the safety interlock fails and the empty single-use reload is fired a second time, the stapler may cut tissue but no staples will deploy from the empty reload, potentially resulting in a failed anastomosis, bleeding, tissue trauma, and the potential need to convert to an open procedure.
Distribution
Worldwide distribution - US Nationwide distribution and the countries of Austria, Belgium, Chile, China, Denmark, Finland, France, Germany, Hong Kong, Ireland, Italy, Netherlands, Norway, Poland, Portugal, Spain, Sweden, Switzerland, United Kingdom.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1383-2021, Z-1385-2021, Z-1384-2021

Ongoing Class II Medical devices

Recall started
26 February 2021
Reason
The capsule may fail to attach to the esophageal mucosa, which in turn may lead to the aspiration of the capsule. In the event of capsule aspiration, immediate intervention to remove the capsule is required. Potential outcomes following capsule aspiration include low oxygen saturation, intervention to retrieve the capsule, potential need to intubate the patient, extended hospital stay and a delay of treatment. Thirteen reports of capsule aspiration have been received the last 2 years.
Distribution
Worldwide distribution - US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1334-2021, Z-1335-2021

Terminated Class II Medical devices

Recall started
13 January 2021
Terminated
14 February 2023
Reason
incorrect device is contained in the package. The packaging label indicates that the device is for left side use, the product inside the package is for right side use. Use of an opposite side mesh can result in implant failure and hernia recurrence
Distribution
Worldwide distribution - US Nationwide distribution in the state of MA and the countries of France, French Guiana, Germany, Luxembourg, Spain, United Kingdom.
Initiated
Voluntary: Firm initiated
Recall number
Z-1142-2021

Completed Class II Medical devices

Recall started
23 November 2020
Reason
May contain a damaged internal component which could impede staple deployment and result in incomplete staple formation and a non-functional staple line preventing adequate hemostasis, or leading to anastomotic leak, pneumothorax, tissue damage or other delayed secondary complications including infection, peritonitis, and sepsis
Distribution
Nationwide Foreign: Canada, Albania Armenia Australia Austria Belarus Belgium Bosnia And Herzegovina Cyprus Czech Republic Egypt Estonia France Germany Greece Hungary Iceland Israel Italy Japan Jordan Korea Kuwait Latvia Libyan Arab Jamahiriya Lithuania Malaysia Martini
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0748-2021, Z-0747-2021

Ongoing Class II Medical devices

Recall started
16 November 2020
Reason
The firm has released software update V01.05.02.16 (also known as V1.5.2) in response to customer reports of a false display of the message, "Temperature Exceeds Limits" followed by automatic shutdown of the monitor with no accompanying alarm.
Distribution
Worldwide distribution - US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall number
Z-0515-2021

Ongoing Class II Medical devices

Recall started
16 October 2020
Reason
In-house lab observations and customer reports of overheating and/or thermal damage to the battery pack, create the risk of exposure to smoke or hazardous materials, which may result in burns and/or bronchospasm,
Distribution
Worldwide distribution - US Nationwide distribution and the countries of Argentina, Bolivia, Brazil, Cayman Islands, Chile, Columbia, Costa Rica, Dominican Republic, Ecuador, El Salvador, Guatemala, Jamaica, Mexico, Nicaragua, Panama, Peru, Uraguay, Honduras, Austria, Bahrain, Belgium, Bulgaria,
Initiated
Voluntary: Firm initiated
Recall number
Z-0425-2021

Ongoing Class II Medical devices

Recall started
12 August 2020
Reason
The device staples may not properly form upon application preventing adequate hemostasis. Use of a product with this assembly error may result in incomplete staple formation.
Distribution
Worldwide distribution - US Nationwide and Global Distribution.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0003-2021, Z-0009-2021, Z-0005-2021, Z-0001-2021, Z-0004-2021, Z-0007-2021, Z-0008-2021, Z-0002-2021, Z-0006-2021

Ongoing Class II Medical devices

Recall started
26 June 2020
Reason
This voluntary recall is being conducted following three customer reports of the device cannula detaching during a procedure. In one case, the detached cannula was not retrieved during the procedure. Potential risks associated with a detached cannula include delay in treatment, biliary leak, pain and retained device fragments.
Distribution
world wide distribution
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2769-2020, Z-2768-2020

Ongoing Class II Medical devices

Recall started
12 May 2020
Reason
The product is recalled due to a manufacturing assembly error that may cause the staples to fail to properly advance within the staple cartridge when the instrument handle is compressed.
Distribution
Worldwide distribution.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2252-2020, Z-2254-2020, Z-2255-2020, Z-2253-2020

Terminated Class II Medical devices

Recall started
30 April 2020
Terminated
14 December 2023
Reason
Temperature sensing catheters failed to meet the established calibrated temperature sensing tolerance of +/- 0.2¿C. Failure to correctly measure patient body temperatures could lead to an incorrect diagnosis and insufficient treatment
Distribution
US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1985-2020, Z-1991-2020, Z-1988-2020, Z-1990-2020, Z-1994-2020, Z-1987-2020, Z-1986-2020, Z-1984-2020, Z-1993-2020, Z-1989-2020, Z-1992-2020

Terminated Class II Medical devices

Recall started
10 February 2020
Terminated
26 April 2024
Reason
Sterile packaging may potentially be compromised, and use of products may result in increased risk for infection.
Distribution
Nationwide domestic distribution. Foreign distribution to Australia, Canada, Austria Belgium Canary Islands Czech Republic Denmark Finland France Germany Greece Hungary Italy Jordan Kazakhstan Kenya Lithuania Luxembourg North Macedonia Poland Portugal Reunion Romania Russian Fed
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1447-2020, Z-1446-2020

Terminated Class II Medical devices

Recall started
17 January 2020
Terminated
29 July 2022
Reason
Potential for the sterile barrier to be compromised. Products with a compromised sterile barrier may result in a potentially increased risk for infection
Distribution
US Nationwide, Israel, and Japan
Initiated
Voluntary: Firm initiated
Recall number
Z-1264-2020

Terminated Class II Medical devices

Recall started
18 December 2019
Terminated
16 August 2022
Reason
There is a potential for packaging integrity issues impacting the humidity or sterility barrier of specific production lots.
Distribution
US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1293-2020, Z-1292-2020, Z-1275-2020, Z-1280-2020, Z-1296-2020, Z-1290-2020, Z-1300-2020, Z-1276-2020, Z-1277-2020, Z-1289-2020, Z-1278-2020, Z-1291-2020 and 15 more

Ongoing Class II Medical devices

Recall started
17 May 2019
Reason
The device may be missing one of two pin components that maintain alignment of the device jaws, resulting in incomplete staple formation which may lead to bleeding, anastomotic leak, peritonitis, or pneumothorax which can result in the potential for infection and/or sepsis.
Distribution
Nationwide domestic distribution, worldwide foreign distribution.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1827-2019, Z-1835-2019, Z-1832-2019, Z-1826-2019, Z-1825-2019, Z-1834-2019, Z-1830-2019, Z-1828-2019, Z-1836-2019, Z-1833-2019, Z-1829-2019, Z-1831-2019

Terminated Class II Medical devices

Recall started
29 April 2019
Terminated
9 January 2024
Reason
The priming volume values printed on the MAHURKA and Argyle acute catheters and Instructions for Use (IFU) are higher than the volumes required to fill each lumen.
Distribution
Nationwide domestic distribution. Foreign distribution to Canada, Australia, New Zealand, Latin America, and Hong Kong.
Initiated
Voluntary: Firm initiated
Recall number
Z-1955-2019

Terminated Class II Medical devices

Recall started
13 March 2019
Terminated
23 April 2020
Reason
Sterilization method used was not consistent with the labeling and the approved sterilization method
Distribution
Worldwide Distribution: US (Nationwide) and to countries of: Canada, Chile, Costa Rica, Guatemala, Japan, Mexico, Panama, Slovakia, Spain, Switzerland and Turkey.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1211-2019, Z-1212-2019, Z-1213-2019

Terminated Class II Medical devices

Recall started
15 February 2019
Terminated
24 July 2020
Reason
The product does not meet sterility requirements. Product labeled as sterile was distributed without undergoing sterilization prior to distribution.
Distribution
The products were distributed to the following US states: CA, NJ, and WA.
Initiated
Voluntary: Firm initiated
Recall number
Z-1041-2019

Terminated Class II Medical devices

Recall started
15 February 2019
Terminated
17 June 2020
Reason
Potential for the sterile packaging to be compromised
Distribution
US Nationwide Distribution
Initiated
Voluntary: Firm initiated
Recall number
Z-1020-2019

Ongoing Class II Medical devices

Recall started
15 November 2018
Reason
The firm received consumer reports of the ceramic trocar tip of the Emprint ablation antenna disengaging from the needle shaft post-ablation. This can result in the tip remaining in the patient.
Distribution
US Nationwide
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0627-2019, Z-0625-2019, Z-0626-2019

Terminated Class II Medical devices

Recall started
5 November 2018
Terminated
14 August 2020
Reason
The product was released into commercial distribution while subject to importation refusal.
Distribution
The products were distributed to the following US states: AL, AR, AZ, CA, CO, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, NE, NJ, NY, OH, OK, PA, RI, SC, TN, TX, UT, VA, WA, and WI.
Initiated
Voluntary: Firm initiated
Recall number
Z-0656-2019

Terminated Class II Medical devices

Recall started
31 October 2018
Terminated
26 May 2020
Reason
This voluntary recall is being conducted due to incorrect packaging. In one lot of Salem Sump Dual Lumen Stomach Tubes the 10Fr product was packaged in a 16Fr pouch. The use of a 10 Fr Salem Sump Tube when a 16Fr is needed could result in decreased fluid collection, with the potential to affect patient care, particularly in an emergency situation where the correct size is not readily available.
Distribution
Worldwide Distribution - US Nationwide and foreign distribution to Canada.
Initiated
Voluntary: Firm initiated
Recall number
Z-0597-2019

Terminated Class II Medical devices

Recall started
5 October 2018
Terminated
7 May 2020
Reason
The firm is issuing this voluntary removal following receipt of reports of parastomal mesh failure identified several years following parastomal hernia repair using the modified Sugarbaker repair technique. In these reports, product failure led to hernia recurrence requiring additional surgical treatment. Symptoms of hernia recurrence may include discomfort, localized pain-free or painful bulging, and possible changes in the overlying skin. Medtronic has received, worldwide, a total of ten reports of mesh failure following use of the product in the last five years. Patients who have received a Parietex" composite parastomal mesh for the treatment of a parastomal hernia should continue to receive ongoing monitoring by their healthcare providers for the recurrence of a parastomal hernia.
Distribution
Worldwide distribution--US Nationwide
Initiated
Voluntary: Firm initiated
Recall number
Z-0461-2019

Terminated Class I Medical devices

Recall started
18 September 2018
Terminated
1 February 2021
Reason
Software Update: External USB Drive performance and its impact on Graphic User Interface (GUI) functionality and labeling of the scalar waveform displayed on GUI during ventilation
Distribution
Worldwide Distribution - US Nationwide Foreign: Canada Argentina Chile Japan Hong Kong Taiwan Uruguay Austria Austria Azerbaijan Belgium Bosnia And Herzegovina Bulgaria Croatia Czech Republic France Germany Greece Hungary Ireland Israel Italy Kazakhstan Kenya Kuwait Latvia
Initiated
Voluntary: Firm initiated
Recall number
Z-0497-2019

Terminated Class II Medical devices

Recall started
17 August 2018
Terminated
10 April 2020
Reason
The firm identified the potential for a device to have an incorrect tissue gap. Use of a device with an incorrect tissue gap may result in incomplete staple formation and/or the inability to remove the device from tissue following application potentially leading to bleeding, anastomotic leak or tissue trauma. There have been no reports of serious injury related to this issue.
Distribution
Worldwide Distribution - US Nationwide in the states of CA, FL, IL, NC, NY, UT. and the countries of Austria, Belgium, Denmark, Finland, France, Germany, Iceland, Italy, Latvia, Netherlands, Norway, Poland, Portugal, Reunion, Spain, Sweden, Switzerland, and United Kingdom.
Initiated
Voluntary: Firm initiated
Recall number
Z-0070-2019

Terminated Class II Medical devices

Recall started
1 August 2018
Terminated
3 June 2020
Reason
Potential for product sterility breach due to a compromised or pinched seal defect
Distribution
US Nationwide Distribution
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0016-2019, Z-0018-2019, Z-0014-2019, Z-0017-2019, Z-0013-2019, Z-0015-2019

Ongoing Class II Medical devices

Recall started
26 July 2018
Reason
Revised replacement instructions for the Li-ion battery packs used in BIS Vista and BIS View Monitoring Systems
Distribution
Nationwide Foreign: Australia, New Zealand
Initiated
Voluntary: Firm initiated
Recall number
Z-2958-2018

Terminated Class II Medical devices

Recall started
11 July 2018
Terminated
3 August 2020
Reason
The temperature sensor catheters may be defective in that they will show a lower body temperature measurement than the actual temperature of the patient.
Distribution
The products were distributed to the following US states: CA, CO, FL, GA, IA, IL, IN, KY, LA, MA, MD, MO, NC, NJ, NY, OH, OR, PA, RI, SC, TN, TX, VA, and WA.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-3207-2018, Z-3205-2018, Z-3203-2018, Z-3208-2018, Z-3206-2018, Z-3204-2018

Terminated Class II Medical devices

Recall started
15 June 2018
Terminated
8 May 2019
Reason
Potential proplylene glycol contamination
Distribution
CA, NY & TX
Initiated
Voluntary: Firm initiated
Recall number
Z-2601-2018

Terminated Class II Medical devices

Recall started
9 April 2018
Terminated
12 May 2020
Reason
There is a potential for improper welding of the yellow staple guide to the instrument. Use of a device with an improperly welded staple guide may result in improper staple formation potentially leading to bleeding or anastomotic leak.
Distribution
Worldwide Distribution - US Nationwide in the US states of : AL, CA, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, NE, NJ, NY, OH, OR, SC, TN, TX, UT, and VA. The devices were distributed to the following foreign countries: UAE, Australia, Austria, Belgium, Bulgaria, Canada, Czech Re
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1527-2018, Z-1528-2018

Terminated Class II Medical devices

Recall started
15 March 2018
Terminated
11 December 2019
Reason
Devices were found with a sterile breach (puncture) in the pouch from a sort of 8,064 packaged devices.
Distribution
Worldwide Distribution in the countries of Australia, Austria, Belgium, Denmark, Finland, France, Germany, Ireland, Italy, Netherlands, Norway, Poland, Saudi Arabia, Spain, Swede, Switzerland and United Kingdom
Initiated
Voluntary: Firm initiated
Recall number
Z-2427-2018

Terminated Class II Medical devices

Recall started
13 December 2017
Terminated
14 April 2020
Reason
Product reportedly shutting down during use.
Distribution
Worldwide Distribution: US (Nationwide).
Initiated
Voluntary: Firm initiated
Recall number
Z-1237-2018

Terminated Class II Medical devices

Recall started
21 November 2017
Terminated
10 July 2020
Reason
Rechargeable lithium-ion batteries with incorrect firmware that are used in certain Puritan Bennett 980 ventilators may not fully charge after installation.
Distribution
Worldwide Distribution - USA (nationwide) Distribution to the states of : AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, PA, RI, SC, TN, TX, UT, VA, WA, WI, WV, WY, and the District of Columbia. and to the cou
Initiated
Voluntary: Firm initiated
Recall number
Z-1208-2018

Terminated Class II Medical devices

Recall started
12 October 2017
Terminated
12 May 2020
Reason
This voluntary recall is being conducted due to the potential for failure to detect the affected Situate" gauze - RF and Xray detectable products by the Situate" detection console due to a manufacturing error.
Distribution
U.S.
Initiated
Voluntary: Firm initiated
Recall number
Z-1130-2018

Terminated Class II Medical devices

Recall started
21 September 2017
Terminated
26 September 2018
Reason
Reports of gastric or esophageal perforation occurring in procedures where the GastriSail" Gastric Positioning device has been used.
Distribution
US Nationwide Distribution
Initiated
Voluntary: Firm initiated
Recall number
Z-0640-2018

Terminated Class II Medical devices

Recall started
15 September 2017
Terminated
19 December 2018
Reason
The device cartridge disengaged during use due to manufacturing error.
Distribution
Internationally, including Japan. No USA Customers
Initiated
Voluntary: Firm initiated
Recall number
Z-0908-2018

Terminated Class II Medical devices

Recall started
13 September 2017
Terminated
28 June 2019
Reason
Potential for failure of the device safety interlock. The safety interlock prevents an empty single-use loading unit from being fired a second time.
Distribution
Nationally
Initiated
Voluntary: Firm initiated
Recall number
Z-0636-2018

Ongoing Class II Medical devices

Recall started
7 August 2017
Reason
The firm received consumer reports of the ceramic trocar tip of the Emprint ablation antenna disengaging from the needle shaft post-ablation. This can result in the tip remaining in the patient.
Distribution
Nationwide including PR, Canada, China
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0560-2018, Z-0559-2018, Z-0561-2018, Z-0556-2018, Z-0558-2018, Z-0557-2018

Terminated Class II Medical devices

Recall started
17 July 2017
Terminated
14 September 2018
Reason
Medtronic is voluntarily recalling two production lots of its Covidien Vistec" x-ray detectable sponges de to possible contamination with human blood caused by handling of the product during the manufacturing process and prior to final packaging and Ethylene Oxide sterilization.
Distribution
U.S., Gov Accts, and foreign countries: PA, PR, CO.
Initiated
Voluntary: Firm initiated
Recall number
Z-0534-2018

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.