CooperSurgical, Inc. recalls

38 recalls recorded since 2018, most recently on 11 August 2026. 16 are listed as ongoing. Status and classification are as published by the FDA.

Recalls38
Ongoing (FDA)16
Terminated or completed22
Class I (FDA)2

Ongoing Class I Medical devices

Recall started
11 August 2026
Reason
Potential to exhibit a loose connection between the flexible tubing and the nasal cannula, during assembly or while in use, which may result in loss of CPAP pressure, possibly reducing the effectiveness of treatment and lead to decreased oxygenation until connection is restored.
Distribution
Nationwide distribution to CT, FL, GA, IA, IL, IN, MD, MI, MS, NJ, NY, OH, PA, VA. International distribution to Brazil and Peru.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-3140-2026, Z-3139-2026

Ongoing Class II Medical devices

Recall started
11 June 2025
Reason
The HSG Procedure Trays and Endosee System Convenience Kits contain non-sterile iodine pouches, however the peel away lid labels incorrectly identify the iodine pouches as sterile.
Distribution
US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and Washington DC.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2137-2025, Z-2136-2025, Z-2135-2025

Ongoing Class II Medical devices

Recall started
15 November 2024
Reason
The warming surfaces or incubator well temperatures of the listed products may exceed the desired set temperature. If temperatures exceed the desired set value, the device will alarm appropriately.
Distribution
States: NC, NJ, TX, VA, IL, NY, CA, PA Please note: The countries below, excluding Iraq, all consider these devices to be "lab equipment", not medical devices and are not subject to classification or recall guidelines. Countries: AT, AU, BE, BH, BR, CA, CH, CN, CZ, DE, EG, ES, ET, FR, GB, GH, HU,
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0688-2025, Z-0689-2025, Z-0692-2025, Z-0687-2025, Z-0690-2025, Z-0691-2025

Ongoing Class II Medical devices

Recall started
31 July 2024
Reason
During the manufacturing process the devices were laser-marked with the incorrect part number.
Distribution
Worldwide distribution - US Nationwide and the countries of Armenia, Bahrain, Canada, Hong Kong, India, New Zealand, Qatar, Sweden.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-3189-2024, Z-3188-2024

Ongoing Class II Medical devices

Recall started
16 February 2024
Reason
Low concentration of sodium pyruvate in the wash could compromise gamete development
Distribution
Worldwide distribution - US Nationwide.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-3298-2024, Z-3296-2024, Z-3297-2024

Ongoing Class II Medical devices

Recall started
20 December 2023
Reason
The firm has become aware of a sudden increase in complaints for three associated lots. Performance issues may lead to impaired embryo development prior to the blastocyst stage.
Distribution
Worldwide - US Nationwide including in the states of AL, AZ, CA, CO, FL, GA, IL, IN, IA, KS, KY, LA, MD, MA, MI, MO, NV, NJ, NM, NY, NC, OH, OK, OR, PA, RI, TN, TX, UT, VT, VA, WA, WV and the countries of Canada, Brazil, Mexico, Argentina, Hong Kong, Japan, Turkey, Netherlands, Belgium, Spain, Taiwa
Initiated
Voluntary: Firm initiated
Recall number
Z-1089-2024

Ongoing Class II Medical devices

Recall started
29 August 2023
Reason
Incorrect component descriptions within the "Contents" section on the Tyvek lid for individual kit trays.
Distribution
US Nationwide distribution in the state of MO.
Initiated
Voluntary: Firm initiated
Recall number
Z-0985-2024

Ongoing Class II Medical devices

Recall started
13 February 2023
Reason
It has come to CooperSurgical's attention that the affected Product may contain a medium other than the Global Total LP single step medium which does not contain the protein source required for embryo culture and development.
Distribution
Worldwide - US Nationwide distribution in the states of NY, UT and the countries of Argentina, Belgium, Bulgaria, Ecuador, France, Germany, Hong Kong, Italy, Malaysia, Mexico, Peru, Singapore, Spain, Switzerland, Thailand, Turkey, the United Kingdom.
Initiated
Voluntary: Firm initiated
Recall number
Z-1348-2023

Ongoing Class II Medical devices

Recall started
15 December 2022
Reason
The affected product may contain a medium other than the Biopsy Medium. There is a risk that use of the Product from this lot could cause degradation of the embryo.
Distribution
Global Distribution including countries of: Bahrain, Belgium, Bulgaria, Egypt, France, Georgia, Greece, Hong Kong, Iran, Iraq, Israel, Japan, Jordan, Kazakhstan, Kuwait, Lebanon, Libya, Nigeria, Oman, Pakistan, Poland, Republic of North Macedonia, Saudi Arabia, Slovenia, Spain, Thailand, Turkey, Tur
Initiated
Voluntary: Firm initiated
Recall number
Z-1044-2023

Ongoing Class II Medical devices

Recall started
20 October 2022
Reason
Firm has received 11 complaints involving a loose or detached metal clips, including 3 cases where the metal clip was ingested by the patient, and 3 cases where the metal clip was removed from the patient's mouth. In one case, the patient suffered a laceration to the throat.
Distribution
Worldwide distribution: US (Nationwide) including Puerto Rico; and countries (foreign) of: Australia, Bahrain, Belgium, Bulgaria, Canada, Colombia, Cyprus, Ecuador, Germany, Greece, Hong Kong, Hungary, India, Ireland, Italy, Malaysia, Malta, Netherlands, New Zealand, Saudi Arabia, Singapore, Turkey,
Initiated
Voluntary: Firm initiated
Recall number
Z-0223-2023

Ongoing Class II Medical devices

Recall started
12 August 2022
Reason
There is a potential breach to the sterile barrier containing the Humidifier Bottles due to the packaging of the filter box located inside the bottle packaging. Damage to the sterile barrier may result in use of an unsterilized device, which may cause contamination and degradation or loss of embryo during incubation.
Distribution
Domestic distribution to the following states: AL AZ CA CT DE FL GA HI IL KS MA MI MN MO NJ NY OH OR SC TX VA VT WA WI WV Foreign distribution to Barbados, Canada, Chile, Micronesia Federate, and Panama.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1823-2022, Z-1822-2022

Ongoing Class II Medical devices

Recall started
13 June 2022
Reason
There was translation error in the Spanish version of the IFU where the Spanish version instructs the user to leave the blastocyte to air thaw for 5 minutes rather than the correct 5 seconds as indicated in the English version.
Distribution
Domestic: USA AL, AZ, CA, CT, FL, GA, ID, IL, IN, KS, MA, MD, MI, MN, MO, MT, NC, NE, NJ, NV, NY, OH, OK, OR, PA, SC, TX, UT, VT, WA, WI, International: Belgium, Canada, Croatia, Netherlands, Portugal, Spain
Initiated
Voluntary: Firm initiated
Recall number
Z-1421-2022

Ongoing Class II Medical devices

Recall started
1 April 2022
Reason
Products in this lot may have compromised barrier seals, introducing a risk to sterility.
Distribution
Domestic distribution to AZ, CA, FL, IL, MI, MO, NY, OR, TX, and WV. No foreign distribution.
Initiated
Voluntary: Firm initiated
Recall number
Z-1108-2022

Ongoing Class II Medical devices

Recall started
31 March 2022
Reason
Incorrect bonding material used for assembly of the catheter, may fail, resulting in detachment of component parts, broken components, or leakage. This breakage may result in delay of procedure
Distribution
Nationwide Foreign: BELGIUM CANADA MALAYSIA NETHERLANDS SPAIN SWEDEN SWITZERLAND
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0502-2024, Z-0503-2024, Z-0504-2024

Terminated Class I Drugs

Recall started
4 February 2022
Terminated
6 February 2023
Reason
Non-sterility
Distribution
US Nationwide
Initiated
Voluntary: Firm initiated
Recall number
D-0643-2022

Ongoing Class II Medical devices

Recall started
29 September 2021
Reason
Coagulate function may not operate with the use of the Hand Switch, failure of the device to coagulate tissue may lead to a prolonged procedure requiring medical intervention to control any intraoperative bleeding and complete the case.
Distribution
US Nationwide Distribution Foreign: ARMENIA, BOLIVIA, CAMBODIA, CANADA, CHILE, CHINA, COSTA RICA, DOMINICAN REPUBLIC, ECUADOR, ENGLAND, GERMANY, HONG KONG, INDIA, INDONESIA, IRAN, ITALY, JORDAN, KENYA, MALAYSIA, MAURITIUS, MEXICO, NEW ZEALAND, NICARAGUA, NIGERIA, PHILIPPINES, PUERTO RICO, QATAR, R
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0261-2022, Z-0258-2022, Z-0263-2022, Z-0260-2022, Z-0257-2022, Z-0262-2022, Z-0259-2022

Terminated Class II Medical devices

Recall started
30 June 2021
Terminated
21 November 2022
Reason
Units are displaying output readings that differ from the measurements taken using a Gas Analyzer. When the display is reading a target setting of 5%, an external meter is reading between 3% and 12% and the Incubator does not alarm. May result in loss of embryo during storage.
Distribution
Only Foreign: Bahrain, Belgium, Bulgaria, Czech Republic Denmark, Egypt, Estonia, France, Georgia, Germany, Ghana, Great Britain, Greece, Hong Kong, Hungary, India, Israel, Italy, Japan, Jordan, Korea, Kuwait, Lebanon, Netherlands, Norway, Poland, Portugal, Romania, Russia, Saudi Arabia, Singapore,
Initiated
Voluntary: Firm initiated
Recall number
Z-2307-2021

Terminated Class II Medical devices

Recall started
30 April 2021
Terminated
21 July 2023
Reason
The affected Wallach LOOP Electrodes were incorrectly packaged with a Round Electrode inside a package labeled as a Square Electrode.
Distribution
US distribution to: CA, MD, MO, NJ, PA, and TX
Initiated
Voluntary: Firm initiated
Recall number
Z-1847-2021

Ongoing Class II Medical devices

Recall started
23 April 2021
Reason
The affected Koh Cup were incorrectly packaged with a 3.5cm Ultem Koh cup, P/N 34131-04, instead of a 3.0cm Ultem Koh cup, P/N 34131-03 as required.
Distribution
Domestic: UT, OH, DE
Initiated
Voluntary: Firm initiated
Recall number
Z-1686-2021

Terminated Class II Medical devices

Recall started
21 January 2021
Terminated
3 January 2024
Reason
The affected Milex Gellhorn pessaries were incorrectly manufactured with a long stem (2.170 ), instead of a short stem (1.670 or 1.960 nominal height) as required. The length discrepancy does not meet the released print dimensions.
Distribution
Worldwide distribution - US Nationwide distribution including in the states of AK, AZ, CA, FL, IL, KS, MA, MI, NC, NY, PA, TN, TX, WA, & Puerto Rico. The countries of New Zealand, United Kingdom.
Initiated
Voluntary: Firm initiated
Recall number
Z-1152-2021

Terminated Class II Medical devices

Recall started
8 September 2020
Terminated
11 January 2021
Reason
There is a discrepancy between the printing on the Product Tyvek label of the individual component packaging. The Chinese characters on the label indicates that it contains an Equilibration Medium (P/N: 12220001) rather than the Vitrification Medium (P/N: 12230001) as indicated on the remainder of the label.
Distribution
The products were distributed to China.
Initiated
Voluntary: Firm initiated
Recall number
Z-0100-2021

Terminated Class II Medical devices

Recall started
3 June 2020
Terminated
11 May 2021
Reason
Mislabeled: Printing on the top web or Tyvek lid of the individual product blister tray indicating that it contains a Pipet Current or Part Number MX140 and the display box indicating the correct Os FinderTM Cervical Dilators, Part Number 1176. The product in the individual product package is the correct Os Finder Cervical Dilator
Distribution
Nationwide
Initiated
Voluntary: Firm initiated
Recall number
Z-2498-2020

Terminated Class II Medical devices

Recall started
9 April 2020
Terminated
11 May 2021
Reason
The seal of the sterile pouch may be compromised on one lot of the Milex Cervical Dilators.
Distribution
Distribution to North America including the following states: AZ, CA, CT, DC, DE, FL, GA, IA, IN, MA, MI, MN, MO, NC, NH, NY, OH, OK, PA, SC, TN, TX, UT, VA, WI
Initiated
Voluntary: Firm initiated
Recall number
Z-2112-2020

Terminated Class III Medical devices

Recall started
28 February 2020
Terminated
11 August 2020
Reason
Expiration date discrepancy. Marked as 5 years. Expiration date should be 3 years.
Distribution
International distribution to Japan only.
Initiated
Voluntary: Firm initiated
Recall number
Z-1713-2020

Terminated Class II Medical devices

Recall started
13 December 2019
Terminated
11 May 2021
Reason
Potential for unsealed pouches in two (2) lots of the Endosee Advance Cannula (P/N: ESPX5).
Distribution
US: AL, AR, AZ, CA, CO, CT, DC, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, ND, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SD, TN, TX, UT, VA, WA, WI, WV, & WY. OUS: None
Initiated
Voluntary: Firm initiated
Recall number
Z-0798-2020

Terminated Class II Medical devices

Recall started
16 October 2019
Terminated
30 June 2020
Reason
Potential for the Quantum 2000 Electrosurgery Generator, part numbers 909075, 909075-05 and 909075-05W, to cease to either cut, coagulate, or blend during a procedure when the foot pedal is used for activation.
Distribution
US: AL, AZ, CA, CO, FL, GA, HI, ID, IL, IN, LA, MA, ME, Ml, MT, ND, NJ, NV, NY, OK, PA, SC, SD, TN, TX, UT, VA, WA, & WV. International: China, Indonesia, South Africa, & Tanzania.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0602-2020, Z-0600-2020, Z-0601-2020

Terminated Class III Medical devices

Recall started
5 September 2019
Terminated
2 September 2021
Reason
The touchpad may not work properly after cleaning.
Distribution
Worldwide distribution - including US Nationwide in the states of CA, CO, CT, GA, IL, MI, NC, NJ, PA, & WA, and countries of Canada and Denmark.
Initiated
Voluntary: Firm initiated
Recall number
Z-0029-2020

Terminated Class III Medical devices

Recall started
19 August 2019
Terminated
14 July 2020
Reason
Discrepancy between the correct expiration dates on the individual pouches inside the boxes and incorrect expiration dates on the outer boxes, where the month and day were reversed.
Distribution
USA: AK, AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, NH, NJ, NM, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, VT, WA, WI, & WV. Distribution US nationwide, Canada, Colombia, Japan, & Malaysia
Initiated
Voluntary: Firm initiated
Recall number
Z-2545-2019

Terminated Class II Medical devices

Recall started
8 July 2019
Terminated
28 May 2020
Reason
The firm updated the IFU in April 2019 to clarify that use of the TRANSWARMER Mattress with other heat producing devices, such as an incubator, is prohibited and could lead to serious health consequences, such as skin burns. some products do not contain the updated IFU.
Distribution
Worldwide Distribution - US Nationwide in all 50 states. International distribution to Australia, Brazil, Canada, Ecuador, Germany, Gibraltar, India, Ireland, Kuwait, Macedonia, New Zealand, Oman, Poland, Puerto Rico, Qatar, Romania, Saudi Arabia, South Korea, Switzerland, United Arab Emirates, & Un
Initiated
Voluntary: Firm initiated
Recall number
Z-2364-2019

Terminated Class II Medical devices

Recall started
18 June 2019
Terminated
15 June 2020
Reason
The products' cut, coagulate, or blend function may not operate with the use of the Foot Pedal.
Distribution
Nationwide domestic distribution. International distribution to Canada and Costa Rica.
Initiated
Voluntary: Firm initiated
Recall number
Z-2057-2019

Terminated Class II Medical devices

Recall started
14 June 2019
Terminated
19 May 2020
Reason
The soft, flexible cup of the Advincula Delineator may be unvalidated and could potentiall crack when subjected to excessive compressions during use, leading to serious health consequences, such as lacerations.
Distribution
US distribution to state of: AZ, CA, GA, IN, LA, MA, NC, NH, NJ, NY, OH, PA, VA, and WA..
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1978-2019, Z-1977-2019

Terminated Class II Medical devices

Recall started
23 April 2019
Terminated
24 April 2020
Reason
2.5A slow blow fuses may have been installed in the LEEP RPECISION Generator rather than the appropriate 3.15A slow blow fuses.
Distribution
US Nationwide Distribution in the states of CA, CT, IN, MA, MI, MN, NC, NV, NY, OH, TX, WA and WI.
Initiated
Voluntary: Firm initiated
Recall number
Z-1840-2019

Terminated Class II Medical devices

Recall started
27 March 2019
Terminated
1 May 2020
Reason
ICSI CUSTOM MICROPIPETTES the thirty-degree bend is absent
Distribution
Foreign: Denmark, Italy, Sweden
Initiated
Voluntary: Firm initiated
Recall number
Z-1367-2019

Terminated Class II Medical devices

Recall started
4 September 2018
Terminated
27 May 2020
Reason
The seal of the sterile pouch may be compromised, thereby increasing the risk of infection.
Distribution
The products were distributed to the following US states: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY. The products were distributed
Initiated
Voluntary: Firm initiated
Recall number
Z-0098-2019

Terminated Class II Medical devices

Recall started
4 September 2018
Terminated
15 April 2020
Reason
There is a possibility that the seal of the sterile pouch may be compromised, thereby increasing the risk of infection.
Distribution
Worldwide Distribution n- US Nationwide in the states of: AL, AR, AZ, CA, CT, DC, FL, GA, IA, ID, IL, KS, KY, LA, MA, MD, MN, MO, NC, NE, NJ, NY, OH, OR, PA, SC, TX, VA, WA and WV, and the following foreign countries of: Australia, Canada, Colombia, Czech Republic, Germany, Ireland, Israel, Italy,
Initiated
Voluntary: Firm initiated
Recall number
Z-3278-2018

Terminated Class II Medical devices

Recall started
12 March 2018
Terminated
8 May 2019
Reason
Products have been packaged with an incorrect product size.
Distribution
CA, SC, WI, MI, NH, MD, OH, NJ, Australia, France, Spain
Initiated
Voluntary: Firm initiated
Recall number
Z-1879-2018

Terminated Class II Medical devices

Recall started
12 March 2018
Terminated
8 May 2019
Reason
Products have been packaged with the wrong Trocar length type.
Distribution
Worldwide Distribution - US Nationwide in the states of FL, GA, IN, NJ, and countries of Canada, Austria, Chile, Italy, Singapore
Initiated
Voluntary: Firm initiated
Recall number
Z-1584-2018

Terminated Class II Medical devices

Recall started
6 October 2017
Terminated
6 February 2020
Reason
The 2MHZ prove was incorrectlhttp://cts.fda.gov/division-tracking/images/trash.pngy labeled as a 3MHZ probe
Distribution
Worldwide Distribution - US Nationwide and the countries of ( Canada and England)
Initiated
Voluntary: Firm initiated
Recall number
Z-1131-2018

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.