Cook Inc. recalls

56 recalls recorded since 2018, most recently on 16 September 2026. 17 are listed as ongoing. Status and classification are as published by the FDA.

Recalls56
Ongoing (FDA)17
Terminated or completed39
Class I (FDA)8

Ongoing Class II Medical devices

Recall started
16 September 2026
Reason
The female Luer lock adapter component in the affected device lots may not have been correctly seated due to misalignment and inadequate torquing during manufacturing. Potential events that may occur as a result of this issue include increased procedural time and inadequate treatment.
Distribution
US States: FL, IL, NM, TX, WA. Canada, France, Germany, Italy, Spain, Switzerland.
Initiated
Voluntary: Firm initiated
Recall number
Z-0163-2027

Ongoing Class I Medical devices

Recall started
2 April 2026
Reason
Marker bands may be at an increased risk of cracking/breakage. Potential adverse events that may result include increased procedural time, harms associated with device fragmentation/separation, and vessel injury. In a worst-case scenario, device fragmentation and separation may cause life-threatening harm or death.
Distribution
US distribution nationwide. OUS distribution to Brunei, Brazil, Canda, China, Colombia, Costa Rica, Hong Kong, Israel, India, Japan, South Korea, Macau, New Zealand, Singapore, El Salvador, Taiwan, Uruguay
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1934-2026, Z-1935-2026, Z-1936-2026

Ongoing Class II Medical devices

Recall started
5 March 2026
Reason
Products from the affected device lots were labeled with expiration dates that exceed the true shelf life.
Distribution
Worldwide distribution.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1778-2026, Z-1785-2026, Z-1784-2026, Z-1791-2026, Z-1780-2026, Z-1777-2026, Z-1789-2026, Z-1788-2026, Z-1775-2026, Z-1790-2026, Z-1779-2026, Z-1786-2026 and 6 more

Ongoing Class II Medical devices

Recall started
26 November 2025
Reason
Cook Medical identified that products supplied in the affected device lots may have been manufactured out of specification. Impacted products may have been cut to the incorrect length, not correctly trimmed, and not inspected properly.
Distribution
Worldwide distribution.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0993-2026, Z-0995-2026, Z-0994-2026

Ongoing Class I Medical devices

Recall started
15 May 2025
Reason
Cook Medical identified that catheters supplied in the affected device lots may experience tip separation.
Distribution
Worldwide distribution.
Initiated
Voluntary: Firm initiated
Recall number
Z-1902-2025

Ongoing Class II Medical devices

Recall started
17 April 2025
Reason
Cook Medical identified that devices from the affected device lots may contain the incorrectly sized introducer sheath.
Distribution
Worldwide distribution - US Nationwide and the countries of Belgium, Germany, Spain, France, Ireland, Italy Netherlands, Switzerland, South Africa, Slovenia.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1786-2025, Z-1787-2025

Ongoing Class II Medical devices

Recall started
17 April 2025
Reason
Cook Medical identified that devices from the affected device lots may contain the incorrectly sized embolization coil.
Distribution
worldwide - No US distribution.
Initiated
Voluntary: Firm initiated
Recall number
Z-1762-2025

Ongoing Class II Medical devices

Recall started
14 October 2024
Reason
Devices from the affected device lots have labels that state the incorrect expiration dates.
Distribution
International Distribution to: Korea only
Initiated
Voluntary: Firm initiated
Recall number
Z-0499-2025

Ongoing Class II Medical devices

Recall started
8 October 2024
Reason
Affected device lots have labels that state the incorrect expiration dates.
Distribution
Worldwide - US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0347-2025, Z-0346-2025

Ongoing Class II Medical devices

Recall started
16 September 2024
Reason
Cook Medical is conducting removal of a lot of Check-Flo Performer Introducer Set due to complaints that the white caps that connect the shaft to the hub of these products may disconnect.
Distribution
International distribution to the country of Australia.
Initiated
Voluntary: Firm initiated
Recall number
Z-0045-2025

Ongoing Class II Medical devices

Recall started
15 May 2024
Reason
Product labels state the incorrect expiration dates that extends beyond their true expiration date
Distribution
MA Foreign: Spain, and Netherlands
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2222-2024, Z-2223-2024

Ongoing Class II Medical devices

Recall started
1 March 2024
Reason
Packaging may have low seal strength and not meet peel strength specifications, compromising device sterility
Distribution
Worldwide distribution - US Nationwide and the countries of AT, AU, BR, CA, CH, DE, ES, GB, HK, IE, IL, IN, IT, NZ, SG.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1528-2024, Z-1530-2024, Z-1529-2024, Z-1527-2024, Z-1526-2024, Z-1532-2024, Z-1531-2024, Z-1525-2024

Ongoing Class I Medical devices

Recall started
7 April 2023
Reason
There were reports from customers that the tracheostomy in the sets and trays is not making a secure connection with the 15mm cap and other 15mm circuit components and accessories. This resulted in an unsecure connection between the device connector and circuit components, cap, or accessories.
Distribution
Worldwide distribution - US Nationwide.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1428-2023, Z-1427-2023

Ongoing Class II Medical devices

Recall started
16 August 2022
Reason
Cook Medical identified that devices from the affected device lots may have a complete breach of the chevron seal of the packaging. Therefore, the sterility of affected devices may be compromised.
Distribution
Domestic distribution nationwide. Foreign distribution worldwide.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1833-2022, Z-1835-2022, Z-1844-2022, Z-1836-2022, Z-1825-2022, Z-1840-2022, Z-1842-2022, Z-1824-2022, Z-1832-2022, Z-1831-2022, Z-1830-2022, Z-1843-2022 and 12 more

Ongoing Class II Medical devices

Recall started
19 July 2022
Reason
Product did not meet the acceptance criteria for packaging testing, which included 3-year age acceleration testing, therefore, the sterility of the device may be compromised.
Distribution
Nationwide Foreign: AR AT AU BE BG CA CA CH CL CN CZ DE DK ES FR GB GR GU HU IL IT JO JP KR KW MX MY NC NL PE PL PR PT PY RU SE SI TR UY ZA
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1715-2022, Z-1708-2022, Z-1709-2022, Z-1718-2022, Z-1717-2022, Z-1711-2022, Z-1706-2022, Z-1710-2022, Z-1713-2022, Z-1712-2022, Z-1707-2022, Z-1716-2022 and 1 more

Completed Class II Medical devices

Recall started
25 May 2022
Reason
The products may contain excess coating material on the outside and/or interior of the needle component. Potential adverse events that may occur if an affected product is used include a delay in the procedure, a prolonged procedure, or particulate entering the bloodstream which may result in pulmonary embolism or ischemia.
Distribution
US Nationwide. Brazil, Canada,
Initiated
Voluntary: Firm initiated
Recall number
Z-1362-2022

Ongoing Class II Medical devices

Recall started
5 May 2022
Reason
The loading cartridges included with these devices may contain a small, unintended, stainless-steel cannula.
Distribution
Domestic distribution nationwide. Foreign distribution worldwide.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1241-2022, Z-1242-2022, Z-1240-2022

Terminated Class II Medical devices

Recall started
12 November 2021
Terminated
26 April 2024
Reason
Flexor Check-Flo Introducer Ansel Modification mislabeled. Products packaged with a label stating 6FR may contain a 7FR device and vice versa
Distribution
AL, AZ, CA, FL, GA, KS, MI, MO, MN, NC, NY, OK, TX Foreign: BE, CH, DE, ES, FR, GB,GU, IT, ZA
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0388-2022, Z-0387-2022

Ongoing Class II Medical devices

Recall started
13 October 2021
Reason
Specific lots may contain black debris on the exterior of the inner stylet and/or on the interior of the outer cannula. Local inflammatory reactions may occur if an affected product is used.
Distribution
Worldwide distribution - US Nationwide distribution in the states of AL, CA, IN, LA, MS, MT, NC, NY, OR, PA, WA and the countries of France, Italy, and Mexico.
Initiated
Voluntary: Firm initiated
Recall number
Z-0414-2022

Terminated Class II Medical devices

Recall started
8 October 2021
Terminated
14 February 2024
Reason
Transseptal needles may contain rust on the interior and/or exterior of the needle, which may not be detected by the user. Potential adverse events if an affected product is used ranges from increased procedural time to inflammatory reactions, which at their most severe include systemic reactions potentially leading to permanent impairment or death.
Distribution
Domestic distribution nationwide. Foreign distribution worldwide.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0223-2022, Z-0222-2022

Terminated Class II Medical devices

Recall started
28 September 2021
Terminated
28 February 2024
Reason
Affected lots may be manufactured incorrectly with the radiopaque marker band located just below the Check-Flo proximal fitting instead of the distal tip. The issue may not be identified until the device is placed in the patient and viewed under fluoroscopy, leading to increased procedural time.
Distribution
Domestic distribution to AR AZ CT FL GA IL LA MA ME MN ND NJ NY OH PA TX WA. Foreign distribution to Germany, Finland, Italy, Netherlands, and Switzerland.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0186-2022, Z-0185-2022

Terminated Class II Medical devices

Recall started
4 January 2021
Terminated
16 September 2021
Reason
Manufactured with incorrect material, which may result in the stent being stiffer and potentially susceptible to degradation could potentially lead to breakage/fragmentation.
Distribution
FL, KY, MD, NS, NY, PA, TX Foreign: Canada, Switzerland
Initiated
Voluntary: Firm initiated
Recall number
Z-0960-2021

Terminated Class I Medical devices

Recall started
24 November 2020
Terminated
7 September 2022
Reason
There is an increased likelihood of the introducer sheath separating at the proximal bond site. If separation occurs during use, it could result in life-threatening adverse events. The potential adverse events that may occur include, but are not limited to increased procedural time, intervention to retrieve a separated segment, embolization occluding blood flow to a vital organ, vessel injury, and hemorrhage.
Distribution
US Nationwide Distribution including states of: AK, CT, IA, LA, MO, NH, OK, TN, WI, AL, DC, ID, MA, MS, NJ, OR, TX, WV, AR, DE, IL, MD, MT, NM, PA, UT, WY, AZ, FL, IN, ME, NC, NV, RI, VA, CA, GA, KS, MI, ND, NY, SC, VT, CO, HI, KY, MN, NE, OH, SD, WA and Global Distribution to: AU, SG, Austria, Belg
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0598-2021, Z-0606-2021, Z-0599-2021, Z-0596-2021, Z-0608-2021, Z-0591-2021, Z-0610-2021, Z-0607-2021, Z-0605-2021, Z-0601-2021, Z-0600-2021, Z-0609-2021 and 9 more

Terminated Class I Medical devices

Recall started
22 October 2020
Terminated
2 September 2021
Reason
Affected product was dropped during manufacturing, which may result in a bent wire guide and/or the presence of loose foreign matter on the wire guide.
Distribution
US distribution to AZ, IA, IN, MA, and ND.
Initiated
Voluntary: Firm initiated
Recall number
Z-0399-2021

Terminated Class II Medical devices

Recall started
31 August 2020
Terminated
1 July 2021
Reason
The indwell time for the Percutaneous Neonatal Pigtail Nephrostomy Set and Pediatric Nephrostomy Stent Set is decreasing from 4 months to 4 weeks. The current IFU is in process of being updated and will be provided with orders placed following implementation.
Distribution
Worldwide distribution - US Nationwide.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0106-2021, Z-0105-2021

Terminated Class II Medical devices

Recall started
11 August 2020
Terminated
28 July 2021
Reason
May contain excess glue within the handle of the delivery system, resulting in an inability to fully deploy the graft via the standard method or troubleshooting deployment method provided in the IFU
Distribution
Foreign Only: Canada SWEDEN AUSTRIA Belgium CZECH REPUBLIC DENMARK ESPA¿A FINLAND France GERMANY GREECE HUNGARY IRELAND ISRAEL Italia Nederland NORWAY POLAND PORTUGAL REUNION SERBIA SLOVENIA SOUTH AFRICA SWEDEN UNITED ARAB EMIRATES UNITED KINGDOM AUSTRALIA Hong Kong J
Initiated
Voluntary: Firm initiated
Recall number
Z-2964-2020

Terminated Class II Medical devices

Recall started
30 July 2020
Terminated
4 November 2021
Reason
Universa Soft Ureteral Stent Sets were incorrectly placed in the outer boxes labeled as Universa Firm Ureteral Stent Set ;pouch label reflects the packaged product.
Distribution
International distribution to the country of Australia.
Initiated
Voluntary: Firm initiated
Recall number
Z-2971-2020

Terminated Class II Medical devices

Recall started
2 June 2020
Terminated
5 May 2022
Reason
Identified products may contain a damaged bushing within the delivery system, which could potentially result in difficulty or an inability to fully deploy the graft via the standard method provided in the IFU.
Distribution
Foreign distribution to the countries of Canada, Chile, Argentina, Colombia, Great Britain, AUSTRIA Belgium. CZECH REPUBLIC DENMARK ESPA¿A France GERMANY GREECE HUNGARY ICELAND IRELAND ISRAEL Italia MARTINIQUE Nederland NORWAY POLAND PORTUGAL SERBIA SOUTH AFRICA SWEDEN UNITED KI
Initiated
Voluntary: Firm initiated
Recall number
Z-2546-2020

Terminated Class II Medical devices

Recall started
30 April 2020
Terminated
29 March 2021
Reason
The Pressure Monitoring Set and Tray may contain excess coating material on the outside and/or interior of the needle component which could cause occlusion.
Distribution
Worldwide distribution including distribution to the following states: Alaska, Alabama, Arkansas, Arizona, California, Colorado, Connecticut, Washington DC, Delaware, Florida, Georgia, Hawaii, Iowa, Idaho, Illinois, Indiana, Kansas, Kentucky, Louisiana, Massachusetts, Maryland, Maine, Michigan, Minn
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2203-2020, Z-2204-2020

Terminated Class II Medical devices

Recall started
15 April 2020
Terminated
15 April 2024
Reason
The firm received complaints on five devices from one specific lot where the stent dislodged from the balloon delivery system prior to deployment. Use of affected product may lead to increased procedural time and additional intervention to remove a stent that dislodged within an unintended location.
Distribution
Domestic distribution to OH, IA, VA, and NC. Foreign distribution to Canada.
Initiated
Voluntary: Firm initiated
Recall number
Z-1940-2020

Terminated Class II Medical devices

Recall started
6 March 2020
Terminated
18 December 2020
Reason
The printed expiration dates are incorrect.
Distribution
The products were distributed to the following US states: CO, DE, MD, MN, NC, NV, and WV. The products were distributed to the following foreign countries: Canada.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1566-2020, Z-1565-2020

Terminated Class II Medical devices

Recall started
16 January 2020
Terminated
12 November 2020
Reason
The firm has identified factors which contributed to failures relating to thrombus formation and/or lumen occlusion reports, and is highlighting to customers key points of the IFU pertaining to the prevention of these factors. In addition, the firm has updated the Planning and Sizing Worksheet to include associated information.
Distribution
Foreign distribution only. Distributed worldwide.
Initiated
Voluntary: Firm initiated
Recall number
Z-1355-2020

Terminated Class II Medical devices

Recall started
27 December 2019
Terminated
23 November 2020
Reason
Specific lots of the Dawson-Mueller Drainage Catheter were not manufactured to specification, which could lead to leakage from the Mac-Loc hub assembly.
Distribution
Nationwide domestic distribution. Foreign distribution to Australia, Belgium, Canada, China, Colombia, Denmark, France, Germany, Italy, Korea, Netherlands, New Zealand, Peru, Russia, Spain, Sweden, and United Kingdom.
Initiated
Voluntary: Firm initiated
Recall number
Z-0815-2020

Terminated Class I Medical devices

Recall started
8 November 2019
Terminated
1 February 2021
Reason
A manufacturing error may cause the radiopaque marker bands to be too loose on certain CXC3.0 CrossCath Support Catheters (compatible with 0.014 wire guides) and too tight on certain CXC3.4 CrossCath Support Catheters (compatible with 0.018 wire guides).
Distribution
Distribution to US states of AL, CA, CT, FL, GA, IL, MA, MD, MI, MO, NC, NJ, NY, OH, OK, PA, SC, TN, TX, and VA, and Brazil.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0588-2020, Z-0590-2020, Z-0589-2020, Z-0587-2020

Terminated Class II Medical devices

Recall started
21 October 2019
Terminated
30 June 2020
Reason
The potential compromise of the sterility of products from an undersealed chevron in the packaging pouches of Torcon NB Advantage Catheters.
Distribution
Worldwide distribution - US Nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, IL, IN, KS, LA, MA, MD, Ml, MN, MO, MS, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SD, TN, TX, UT, VA, WA, WI, WV, and PR. Countries of AU, BE, CA, CN, CZ, DK, FR, DE, GT, HK, IT, KR, LU, MX
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0435-2020, Z-0471-2020, Z-0487-2020, Z-0483-2020, Z-0479-2020, Z-0456-2020, Z-0441-2020, Z-0472-2020, Z-0452-2020, Z-0426-2020, Z-0427-2020, Z-0439-2020 and 59 more

Terminated Class II Medical devices

Recall started
3 October 2019
Terminated
13 May 2020
Reason
Bent transfer catheter tips may cause difficulty or inability in inserting the transfer catheter into the guiding catheter. Potential adverse events that may occur if an affected product is used include increased procedural time, the need for a repeat embryo transfer procedure, or the need for a patient to undergo an additional IVF cycle.
Distribution
No US distribution. The products were distributed to the following foreign countries: Belgium, Denmark, Germany, Italy, Netherlands, Norway, Poland, Republic of Macedonia, Slovenia, Spain, and Switzerland.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0415-2020, Z-0416-2020

Terminated Class II Medical devices

Recall started
17 September 2019
Terminated
21 May 2020
Reason
Lidstock perforation line may be over the sterile area of the tray, when it should be on the lip of the tray, potentially compromising the sterility of the device.
Distribution
US Nationwide distribution in the states of CA, FL, IN, MI, MT, NC, NY, OH.
Initiated
Voluntary: Firm initiated
Recall number
Z-0506-2020

Terminated Class I Medical devices

Recall started
24 May 2019
Terminated
31 October 2019
Reason
Multiple complaints were received for balloons bursting below the rated burst pressure. Potential adverse events that may occur if an affected product is used include a delay in the procedure, additional intervention, vessel injury, and balloon fragmentation in the patient.
Distribution
Worldwide Distribution - US Nationwide The affected product was distributed to the following states: FL, OH, WA. The products were distributed to the following foreign countries: Belgium, Denmark, Finland, Germany, Italy, Netherlands, Norway, Poland, South Africa, Spain, Sweden, Switzerland, Un
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1758-2019, Z-1756-2019, Z-1757-2019

Terminated Class II Medical devices

Recall started
2 May 2019
Terminated
23 August 2019
Reason
The products may have been manufactured without the bottom seal on the outer packaging, compromising the sterility of the product.
Distribution
No distribution in the US. Products were distributed to the following foreign countries: China.
Initiated
Voluntary: Firm initiated
Recall number
Z-1780-2019

Terminated Class II Medical devices

Recall started
26 April 2019
Terminated
8 November 2019
Reason
may contain a damaged gray safety lock knob, which could potentially result in difficulty or an inability to fully deploy the graft via the standard or troubleshooting method provided in the Instructions for Use (IFU). Potential adverse events that may occur if an affected product is used include a prolonged procedure and open surgical intervention.
Distribution
No US distribution. Product distributed to the following foreign countries: Australia, Austria, Belgium, Brazil, Canada, Canary Islands, Chile Columbia, Costa Rica, Czech Republic, Denmark, Finland, France, Germany, Greece, Ireland, Israel, Italy, Japan, Morocco, Netherlands, Norway, Poland, Port
Initiated
Voluntary: Firm initiated
Recall number
Z-1844-2019

Terminated Class II Medical devices

Recall started
22 April 2019
Terminated
5 May 2020
Reason
Kwart RetroInject Stent Set (Multi-Length) and Salle Intraoperative Pyeloplasty Stent Set Instructions for Use (IFU) did not include a warning for stent knotting.
Distribution
Nationwide Foreign: Australia Canada Austria Belgium Denmark Germany Spain Finland France Martinique French Polynesia Switzerland Ireland Italy Netherlands Poland Sweden United Kingdom Hong Kong New Zealand Chile Colombia Israel Russian Federation Czech Republic Greece Jo
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1369-2019, Z-1370-2019

Terminated Class II Medical devices

Recall started
9 April 2019
Terminated
28 April 2020
Reason
Affected products may have been manufactured with the catheter distal end hole too small
Distribution
Domestically: OH; Internationally: CANADA, SPAIN, ITALY, BELGIUM, GERMANY, UNITED ARAB EMIRATES, GREECE, SWITZERLAND, OMAN, MONACO, SAUDI ARABIA, AUSTRALIA, HONG KONG, INDIA & KOREA
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1725-2019, Z-1724-2019, Z-1722-2019, Z-1723-2019

Terminated Class I Medical devices

Recall started
1 February 2019
Terminated
30 June 2020
Reason
The products were manufactured without a back bevel on the needle tip. This could cause damage to the inside of the introducer during needle insertion.
Distribution
US nationwide and Canada. US distribution to the following states: AZ, FL, IL, IN, KS, MA, MI, MO, PA, SC, TN, and WA. The products were distributed to the following foreign countries: Canada.
Initiated
Voluntary: Firm initiated
Recall number
Z-0979-2019

Terminated Class II Medical devices

Recall started
4 January 2019
Terminated
29 May 2020
Reason
The wire guide may be incorrectly loaded into the wire guide holder, leading to the stiff tip of the wire guide being introduced into the patient instead of the flexible tip.
Distribution
Domestic distribution to nationwide US. Foreign distribution to Australia, Belgium, Bolivia, Canada, Canary Islands, Chile, China, Colombia, Croatia, Czech Republic, Denmark, France, French Polynesia, Germany, Guatemala, Honduras, Hungary, India, Ireland, Italy, Japan, Korea, Luxembourg, Malay
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0977-2020, Z-0962-2020, Z-0966-2020, Z-0972-2020, Z-0970-2020, Z-0967-2020, Z-0975-2020, Z-0974-2020, Z-0964-2020, Z-0971-2020, Z-0963-2020, Z-0959-2020 and 8 more

Terminated Class II Medical devices

Recall started
21 December 2018
Terminated
30 April 2020
Reason
The products were manufactured with longer than specified sheaths.
Distribution
The products were distributed to the following US states: CA, CT, GA, MA, NY, PA, VA, and WI.
Initiated
Voluntary: Firm initiated
Recall number
Z-0851-2019

Terminated Class II Medical devices

Recall started
19 December 2018
Terminated
28 April 2020
Reason
There is a potential for difficulty in sliding down the retention mechanism on the sets manufactured with a specific extension spring lot.
Distribution
Worldwide Distribution: US (nationwide) to of states: AL, AZ, CA, GA, HI, IL, IN, KY, LA, MI, MN, MO, NC, NJ, NY, OH, PA, RI, TN, TX, and UT; and countries of: Austria, Belgium, Canada, Denmark, France, Germany, Great Britain, Hungary, Iceland, Italy, Netherlands, Norway, Spain, and Switzerland.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0968-2019, Z-0967-2019

Terminated Class II Medical devices

Recall started
31 May 2018
Terminated
11 April 2019
Reason
This lot of VentriClear failed endotoxin testing. Potential adverse events include immune responses ranging from non-specific febrile reaction to life-threatening systemic inflammatory response syndrome (SIRS).
Distribution
One distributor in Indiana; product was not further distributed to end users.
Initiated
Voluntary: Firm initiated
Recall number
Z-2791-2018

Terminated Class II Medical devices

Recall started
16 April 2018
Terminated
27 June 2019
Reason
Incorrect product labeling. Product labeled as 6mm x 2cm balloon are packaged with a 4mm x 4 cm balloon label and vice-versa.
Distribution
Worldwide Distribution - US Nationwide in the states of AZ, DE, FL, GA, ID, IL, KY, MI, MO, NH, NJ, PA, RI, TX, UT and the country of Peru
Initiated
Voluntary: Firm initiated
Recall number
Z-1521-2018

Terminated Class II Medical devices

Recall started
29 March 2018
Terminated
9 April 2019
Reason
Cap-adapter assembly of the device may potentially not be adequately tightened, possibly leading to fluid leakage from or air aspiration into the connecting tube.
Distribution
Distributed to accounts in 7 states: AL, CA, GA, NE, NV, NY, and TX. International distribution to Japan, Korea, and Taiwan.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1752-2018, Z-1755-2018, Z-1756-2018, Z-1754-2018, Z-1753-2018, Z-1751-2018

Terminated Class II Medical devices

Recall started
22 March 2018
Terminated
13 May 2020
Reason
Pump was found to be constructed from crimp connectors that do not meet the fixation wiring requirements and can cause failure of the device to operate or electric shock or burn to the user.
Distribution
Worldwide Distribution -- USA, Argentina, Bolivia, Brazil, Canada, Colombia, Guatemala, Mexico, Panama, Peru, Uruguay, and Venezuela.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1461-2018, Z-1460-2018

Terminated Class II Medical devices

Recall started
31 January 2018
Terminated
17 April 2020
Reason
The interior of the product may not be sterilized to the appropriate Sterility Assurance Level when following the reprocessing instructions in the Instructions for Use.
Distribution
US Nationwide, OUS countries: AE, AT, AU, BE, BG, BR, CA, CH, CL, CN, CO, CR, CZ, DE, DJ, DK, DO, EC, EE, EG, ES, FI, FJ, FR, GB, GR, GT, HK, HN, HR, HU, ID, IE, IL, IN, IT, JO, KZ, LB, LT, LV, MA, MO, MX, MY, NI, NL, NO, NZ, PE, PH, PK, PL, PR, PT, QA, RO, RU, SA, SE, SG, SI, SK, SV, TH, TT, TW, U
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1591-2018, Z-1596-2018, Z-1594-2018, Z-1590-2018, Z-1595-2018, Z-1593-2018, Z-1592-2018

Terminated Class II Medical devices

Recall started
12 January 2018
Terminated
8 May 2020
Reason
The temperature of the incubation lids of the MINC Bench top Incubators may become too high or too low for proper gamete/embryo incubation due to fatigue failure of the flat flex lid heater.
Distribution
Nationwide (K-IMNC-100 -US G46022) . (K-IMNC-100 G20079 manufactured in Australia and distributed worldwide)
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1378-2018, Z-1377-2018

Terminated Class II Medical devices

Recall started
19 December 2017
Terminated
9 April 2019
Reason
Certain lots of in vitro fertilization (IVF) products were distributed without completed finished device quality testing with potential adverse events that may include nonviable gametes and/or nonviable embryos.
Distribution
Worldwide Distribution - U.S. Nationwide in the states of: CA, CO, FL, IN, MA, NE, NJ, OH, TN, TX, and WI. and countries of Belarus, Belgium, Brazil, Canada, Chile, China, Egypt, Italy, Japan, Poland, Saudi Arabia, Spain, and United Kingdom.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1185-2018, Z-1186-2018, Z-1188-2018, Z-1187-2018

Terminated Class II Medical devices

Recall started
19 October 2017
Terminated
16 April 2019
Reason
Cook has identified that the affected lots may contain wire guides that were loaded backwards into the wire guide holder.
Distribution
US Nationwide Distribution
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0830-2018, Z-0831-2018, Z-0832-2018

Terminated Class II Medical devices

Recall started
13 October 2017
Terminated
23 October 2019
Reason
Label does not state that the product is heparin-coated.
Distribution
Nationally
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0954-2018, Z-0959-2018, Z-0948-2018, Z-0950-2018, Z-0951-2018, Z-0966-2018, Z-0961-2018, Z-0949-2018, Z-0956-2018, Z-0965-2018, Z-0953-2018, Z-0947-2018 and 8 more

Terminated Class II Medical devices

Recall started
31 July 2017
Terminated
8 May 2019
Reason
Central venous catheters may not fit over guide wire potentially causing procedure delay.
Distribution
Worldwide distribution - US Nationwide in the states of: AL, CA, CO, FL, GA, IA, IL, KY, LA, MD, MI, MN, MO, NC, NY, OH, OK, OR, PA, RI, TN, TX, UT, VA, WA, WI; the District of Columbia and the countries of Belgium, India Italy, Spain, and United Kingdom.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0633-2018, Z-0632-2018, Z-0630-2018, Z-0631-2018, Z-0634-2018

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.