Bloomington, IN, United States
Cook Inc. recalls
56 recalls recorded since 2018, most recently on 16 September 2026. 17 are listed as ongoing. Status and classification are as published by the FDA.
Recalls56
Ongoing (FDA)17
Terminated or completed39
Class I (FDA)8
Ongoing Class II Medical devices
- Recall started
- 16 September 2026
- Reason
- The female Luer lock adapter component in the affected device lots may not have been correctly seated due to misalignment and inadequate torquing during manufacturing. Potential events that may occur as a result of this issue include increased procedural time and inadequate treatment.
- Distribution
- US States: FL, IL, NM, TX, WA. Canada, France, Germany, Italy, Spain, Switzerland.
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-0163-2027
- Williams Cystoscopic Injection Needles, used for cystoscopic injection into the urethra, bladder neck, and bladder wall, Reference Part Number 090001-S26, Order Number (GPN) G16112
Ongoing Class I Medical devices
- Recall started
- 2 April 2026
- Reason
- Marker bands may be at an increased risk of cracking/breakage. Potential adverse events that may result include increased procedural time, harms associated with device fragmentation/separation, and vessel injury. In a worst-case scenario, device fragmentation and separation may cause life-threatening harm or death.
- Distribution
- US distribution nationwide. OUS distribution to Brunei, Brazil, Canda, China, Colombia, Costa Rica, Hong Kong, Israel, India, Japan, South Korea, Macau, New Zealand, Singapore, El Salvador, Taiwan, Uruguay
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-1934-2026, Z-1935-2026, Z-1936-2026
- Centimeter Sizing Catheter RPN N5.0-35-100-P-10S-PIG-CSC-20 GPN G11916 Catheters for use in angiographic procedures are available in a variety of French sizes, endhole sizes, lengths, materials and designs (e.g., po
- Aurous Centimeter Sizing Catheter RPN N5.0-35-100-P-10S-PIG-CSC-20-01 GPN G47308 Sizing catheters are single-use, sterile catheters with gold marker bands. These catheters are for use in angiographic procedures and
- Beacon Tip Centimeter Sizing Catheter RPN/GPN NR5.0-35-70-P-10S-0-CSC-20 G31213 NR5.0-35-100-P-10S-0-CSC-20 G31214 NR5.0-35-100-P-10S-PIG-CSC-20 G31216 NR5.0-35-70-P-10S-VCF-CSC-20 G31220 NR5.0-35-100-P-10S-VCF
Ongoing Class II Medical devices
- Recall started
- 5 March 2026
- Reason
- Products from the affected device lots were labeled with expiration dates that exceed the true shelf life.
- Distribution
- Worldwide distribution.
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-1778-2026, Z-1785-2026, Z-1784-2026, Z-1791-2026, Z-1780-2026, Z-1777-2026, Z-1789-2026, Z-1788-2026, Z-1775-2026, Z-1790-2026, Z-1779-2026, Z-1786-2026 and 6 more
- COOK MEDICAL Arterial Pressure Monitoring Tray, Reference Part Number APMY-2.5-15-0-2.5 , Order Number G58412; Reference Part Number APMY-4.0-21J-1-12, Order Number G58419
- Melker Universal Cuffed Emergency Cricothyrotomy Catheter Set: Reference Part Number C-TCCSB-500-UNI, Order Number G32193; Reference Part Number C-TCCSB-500-UNI-01, Order Number G47665.
- COOK MEDICAL Cook¿ Strange Bile Duct Stone Exploration Set: Reference Part Number C-CSRC-7.5D-40-STRANGE, Order Number G09108
- COOK MEDICAL Wayne Pneumothorax Set: Reference Part Number C-UTPT-1020-WAYNE-IMH, Order Number G56532; Reference Part Number C-UTPT-1400-WAYNE-112497-IMH, Order Number G56535.
- COOK MEDICAL Blue Rhino¿ G2-Multi Percutaneous Tracheostomy Introducer Tray: Reference Part Number C-PTISY-100-HC-G-NA, Order Number G57703; Reference Part Number C-PTISY-100-UNL-HC-G-NA, Order Number G57706; Refere
- COOK MEDICAL Approach CTO-18 Micro Wire Guide, Reference Part Number CMW-14-190-18G, Order Number G50788; Reference Part Number CMW-14-300-18G, Order Number G50792
- COOK MEDICAL Spectrum¿ Central Venous Tray: Reference Part Number C-UQLMYJ-1001J-RSC-ABRM-HC-FST-A-RD, Order Number G44800; Reference Part Number C-UTLMYJ-701J-ABRM-HC-IHI-FST-A-RD, Order Number G44809; Reference Pa
- COOK MEDICAL Spectrum¿ Central Venous Catheter Set: Reference Part Number C-UDLM-401J-ABRM-HC, Order Number G43983; Reference Part Number C-UDLM-401J-ABRM-HC-CAH, Order Number G24754; Reference Part Number C-UDLM-40
- and 10 more product lines
Ongoing Class II Medical devices
- Recall started
- 26 November 2025
- Reason
- Cook Medical identified that products supplied in the affected device lots may have been manufactured out of specification. Impacted products may have been cut to the incorrect length, not correctly trimmed, and not inspected properly.
- Distribution
- Worldwide distribution.
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-0993-2026, Z-0995-2026, Z-0994-2026
- Flexor Check-Flo Introducers and Sets, Reference Part Numbers: 1) KCFW-9.0-18/38-45-RB-ANL0-HC; 2) KCFW-9.0-18/38-45-RB-ANL1-HC; 3) KCFW-10.0-35-45-RB-HFANL0-HC; 4) KCFW-10.0-35-45-RB-HFANL1-HC; 5) KCFW-12.0-3
- R¿sch-Uchida Transjugular Liver Access Sets, Reference Part Numbers: RUPS-100; The R¿sch-Uchida Transjugular Liver Access Set is intended for transjugular liver access in diagnostic and interventional procedures.
- Ring Transjugular Intrahepatic Access Sets, Reference Part Numbers: 1) RTPS-100; 2) RTPS-100-10.0; The Ring Transjugular Intrahepatic Access Set is intended for transjugular liver access in diagnostic and interven
Ongoing Class I Medical devices
- Recall started
- 15 May 2025
- Reason
- Cook Medical identified that catheters supplied in the affected device lots may experience tip separation.
- Distribution
- Worldwide distribution.
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-1902-2025
- Beacon Tip 5.0 Fr Angiographic catheters are designed to facilitate diagnostic and therapeutic procedures. Catheters are available in various configurations. Configurations include different shaft lengths and tip curves
Ongoing Class II Medical devices
- Recall started
- 17 April 2025
- Reason
- Cook Medical identified that devices from the affected device lots may contain the incorrectly sized introducer sheath.
- Distribution
- Worldwide distribution - US Nationwide and the countries of Belgium, Germany, Spain, France, Ireland, Italy Netherlands, Switzerland, South Africa, Slovenia.
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-1786-2025, Z-1787-2025
- Check=Flo Performer Introducer, 10.0Fr, Reference Part Number(RPN) REF RCFW-10.0-0-38-30-RB, Order Number (GPN) REF G08686,
- Check=Flo Performer Introducer, 14.0Fr, Reference Part Number(RPN) REF RCFW-14.0-0-38-30-RB, Order Number (GPN) REF G08957,
Ongoing Class II Medical devices
- Recall started
- 17 April 2025
- Reason
- Cook Medical identified that devices from the affected device lots may contain the incorrectly sized embolization coil.
- Distribution
- worldwide - No US distribution.
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-1762-2025
- Tornado Embolization Microcoil, intended for embolization of selective vessel supply to arteriovenous malformations and other vascular lesions, RPN MWCE-18S-3/2-TORNADO (GPN G08261), RPN MWCE-18S-3/2-TORNADO-081800 (GPN
Ongoing Class II Medical devices
- Recall started
- 14 October 2024
- Reason
- Devices from the affected device lots have labels that state the incorrect expiration dates.
- Distribution
- International Distribution to: Korea only
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-0499-2025
- Approach CTO Micro Wire Guide, RPN CMW-14-190-25G, GPN G50789
Ongoing Class II Medical devices
- Recall started
- 8 October 2024
- Reason
- Affected device lots have labels that state the incorrect expiration dates.
- Distribution
- Worldwide - US Nationwide distribution.
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-0347-2025, Z-0346-2025
- Approach Hydro ST Micro Wire Guide; Catalog prefix HMW Reference Part Numbers: HMW-14-135-ST HMW-14-190-ST HMW-14-300-ST
- Approach CTO Micro Wire Guide, Guide Wire, Catalog Prefix CMW; Reference Part Numbers: CMW-14-190-12G CMW-14-190-18G CMW-14-190-25G CMW-14-190-6G CMW-14-300-12G CMW-14-300-18G CMW-14-300-25G CMW-14-300-6G
Ongoing Class II Medical devices
- Recall started
- 16 September 2024
- Reason
- Cook Medical is conducting removal of a lot of Check-Flo Performer Introducer Set due to complaints that the white caps that connect the shaft to the hub of these products may disconnect.
- Distribution
- International distribution to the country of Australia.
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-0045-2025
- Check-Flo Performer Introducer Set, Reference Part Number RCF-5.0-35-45-J, Order Number G27242. Performer Introducers are designed to perform as a guiding sheath and/or introducer sheath and come with a Check Flo valve.
Ongoing Class II Medical devices
- Recall started
- 15 May 2024
- Reason
- Product labels state the incorrect expiration dates that extends beyond their true expiration date
- Distribution
- MA Foreign: Spain, and Netherlands
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-2222-2024, Z-2223-2024
- Ultrathane Mac-Loc Locking Loop Multipurpose Drainage Catheter ULT16.0-38-25-P-6S-MCL-HC--Intended for percutaneous drainage in a variety of drainage applications (e.g., nephrostomy, biliary and abscess), either by direc
- Ultrathane Cook-Cope Type Locking Loop Multipurpose Drainage Catheter-ULT12.0-38-45-P-6S-CLM-RH-ntended for percutaneous drainage in a variety of drainage applications (e.g., nephrostomy, biliary and abscess), either by
Ongoing Class II Medical devices
- Recall started
- 1 March 2024
- Reason
- Packaging may have low seal strength and not meet peel strength specifications, compromising device sterility
- Distribution
- Worldwide distribution - US Nationwide and the countries of AT, AU, BR, CA, CH, DE, ES, GB, HK, IE, IL, IN, IT, NZ, SG.
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-1528-2024, Z-1530-2024, Z-1529-2024, Z-1527-2024, Z-1526-2024, Z-1532-2024, Z-1531-2024, Z-1525-2024
- Peel-Away Introducer - Intended for the percutaneous introduction of balloon, electrode and closed or non-tapered end catheters into central and peripheral vasculature, and for nonvascular use. REFERENCE PART NUMBER
- Dilator - Intended to be used for dilating puncture sites or catheter tracts for percutaneous placement of devices for vascular and non-vascular applications such as in the venous, arterial, biliary and renal systems. 1
- Entuit Secure Adjustable Gastrointestinal Suture Anchor Set - Intended for anchoring the anterior wall of the stomach to the abdominal wall prior to introduction of interventional catheters. REFERENCE PART NUMBER (R
- Yueh Centesis Disposable Catheter Needle - Intended for use in percutaneous fluid aspirations, such as paracentesis or thoracentesis. REFERENCE PART NUMBER (RPN): DTVN-5.0-19-10.0- YUEH ORDER NUMBER (GPN): G09490
- Trocar Needle - Used for dilating puncture sites or catheter tracts for percutaneous placement of devices for vascular and non-vascular applications such as in the venous, arterial, biliary and renal system REFERENC
- Dilator - Coons Taper - Used for dilating puncture sites or catheter tracts for percutaneous placement of devices for vascular and non-vascular applications such as in the venous, arterial, biliary and renal systems.
- Dilator With AQ Hydrophilic Coating - Intended to be used for dilating puncture sites or catheter tracts for percutaneous placement of devices for vascular and non-vascular applications such as in the venous, arterial, b
- Chiba Biopsy Needle - Used for aspiration biopsy. REFERENCE PART NUMBER (RPN) ORDER NUMBER (GPN): 1) DCHN-18-10.0 G01559 2) DCHN-18-10.0-U G05040 3) DCHN-18-15.0 G00850 4) DCHN-22-10.0 G00852 5) DCHN-22-10.0-U
Ongoing Class I Medical devices
- Recall started
- 7 April 2023
- Reason
- There were reports from customers that the tracheostomy in the sets and trays is not making a secure connection with the 15mm cap and other 15mm circuit components and accessories. This resulted in an unsecure connection between the device connector and circuit components, cap, or accessories.
- Distribution
- Worldwide distribution - US Nationwide.
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-1428-2023, Z-1427-2023
- Blue Rhino G2-Multi Percutaneous Tracheostomy Introducer Tray, Reference Part Numbers C-PTISYJ-100-HC-G-EU-FLEX8.5 (G57725), C-PTISYJ-100-HC-G-NA-FLEX7.5 (G57720), C-PTISYJ-100-HC-G-NA-FLEX8.5 (G57721), C-PTISY-100-HC-G-
- Blue Rhino G2-Multi Percutaneous Tracheostomy Introducer Set, Reference Part Numbers C-PTIS-100-HC-G-EU-FLEX7.5 (G57695), C-PTIS-100-HC-G-EU-FLEX8.5 (G57696), C-PTIS-100-HC-G-NA-FLEX7.5 (G57691), C-PTIS-100-HC-G-NA-FLEX8
Ongoing Class II Medical devices
- Recall started
- 16 August 2022
- Reason
- Cook Medical identified that devices from the affected device lots may have a complete breach of the chevron seal of the packaging. Therefore, the sterility of affected devices may be compromised.
- Distribution
- Domestic distribution nationwide. Foreign distribution worldwide.
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-1833-2022, Z-1835-2022, Z-1844-2022, Z-1836-2022, Z-1825-2022, Z-1840-2022, Z-1842-2022, Z-1824-2022, Z-1832-2022, Z-1831-2022, Z-1830-2022, Z-1843-2022 and 12 more
- Fixed Core Wire Guide (Safe-T-J Curved), Guidewire Reference Part Number/GPN: TSCF-25-180-3 G01602 TSCF-35-180-3 G00902 TSCF-21-260-3 G00455 THSCF-25-260-3 G02384 TSCF-25-145-3 G00466 THSCF-35-80-3 G04069 THSCF-3
- Heavy Double Flexible Tipped Wire Guide, Guidewire Reference Part Number/GPN: THDOC-35-60-0-3 G11949 C-HDOC-18-40-0-2 G09201 C-THDOC-18-40-0-2-HTS G26925
- Vinyl Connecting Tube, Connecting Tube Reference Part Number/GPN CTU14.0-30-ST G02278 CTU14.0-40-ST G02327 CTU14.0-30 G02791 CTU14.0-50-ST G02464 C-CTU14.0-30-ST G02898
- Newton Wire Guide (LT Taper/10cm Distal Flexibility), Guidewire Reference Part Number/GPN: TSFNA-35-145 G00701 TSCFNA-35-145-3 G00561
- Bentson Plus Wire Guide (Extra Body/20cm Distal Flexibility with Softened Tip), Guidewire; Reference Part Number: TSFBP-35-180 (GPN G09782) TSFBP-35-145 (GPN G09735)
- Roadrunner PC Wire Guide (The Firm LT), Guidewire Reference Part Number/GPN RLPC-35-145 G06866
- Rosen Curved Wire Guide, Guidewire Reference Part Number/GPN THSCF-35-220-1.5-ROSEN G01623 THSCF-35-80-1.5-ROSEN G01774 THSCF-35-180-1.5-ROSEN G01264 THSCF-35-145-1.5-ROSEN G01261
- Amplatz extra Stiff Wire Guide, Stiff Guidewire (OUS ONLY); Reference Part Number: THSF-35-145-AES-SGH (GPN G06801) THSF-38-80-AES-SGH (GPN G06874)
- and 16 more product lines
Ongoing Class II Medical devices
- Recall started
- 19 July 2022
- Reason
- Product did not meet the acceptance criteria for packaging testing, which included 3-year age acceleration testing, therefore, the sterility of the device may be compromised.
- Distribution
- Nationwide Foreign: AR AT AU BE BG CA CA CH CL CN CZ DE DK ES FR GB GR GU HU IL IT JO JP KR KW MX MY NC NL PE PL PR PT PY RU SE SI TR UY ZA
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-1715-2022, Z-1708-2022, Z-1709-2022, Z-1718-2022, Z-1717-2022, Z-1711-2022, Z-1706-2022, Z-1710-2022, Z-1713-2022, Z-1712-2022, Z-1707-2022, Z-1716-2022 and 1 more
- Roadrunner UniGlide Hydrophilic Wire Guide REFERENCE PART NUMBER (RPN): HPWAS-35-150 ORDER NUMBER (GPN): G56176. For vascular catheter positioning and exchange in diagnostic and interventional procedures.
- Roadrunner PC Hydrophilic Wire Guide REFERENCE PART NUMBER (RPN):RLPC-35-180 ORDER NUMBER (GPN) G07516. For vascular catheter positioning and exchange in diagnostic and interventional procedures.
- Roadrunner PC Hydrophilic Wire Guide REFERENCE PART NUMBER (RPN): RPC-35-180 ORDER NUMBER (GPN):G07518. For vascular catheter positioning and exchange in diagnostic and interventional procedures.
- Roadrunner UniGlide Hydrophilic Wire Guide REFERENCE PART NUMBER (RPN): HPWS-35-80 ORDER NUMBER (GPN): G56151. For vascular catheter positioning and exchange in diagnostic and interventional procedures.
- Roadrunner UniGlide Hydrophilic Wire Guide REFERENCE PART NUMBER (RPN): HPWS-35-150 ORDER NUMBER (GPN): G56152. For vascular catheter positioning and exchange in diagnostic and interventional procedures.
- Roadrunner UniGlide Hydrophilic Wire Guide REFERENCE PART NUMBER (RPN): HPWA-18-150 ORDER NUMBER (GPN):G56161. For vascular catheter positioning and exchange in diagnostic and interventional procedures.
- Roadrunner PC Hydrophilic Wire Guide REFERENCE PART NUMBER (RPN)RFPC-35-180 ORDER NUMBER (GPN) G07937. For vascular catheter positioning and exchange in diagnostic and interventional procedures.
- Roadrunner UniGlide Hydrophilic Wire Guide REFERENCE PART NUMBER (RPN): HPW-35-150 ORDER NUMBER (GPN):G56149. For vascular catheter positioning and exchange in diagnostic and interventional procedures.
- and 5 more product lines
Completed Class II Medical devices
- Recall started
- 25 May 2022
- Reason
- The products may contain excess coating material on the outside and/or interior of the needle component. Potential adverse events that may occur if an affected product is used include a delay in the procedure, a prolonged procedure, or particulate entering the bloodstream which may result in pulmonary embolism or ischemia.
- Distribution
- US Nationwide. Brazil, Canada,
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-1362-2022
- Pressure Monitoring Set, Single-lumen uncoated central venous catheter. RPN C-PMS-250, GPN G02838
Ongoing Class II Medical devices
- Recall started
- 5 May 2022
- Reason
- The loading cartridges included with these devices may contain a small, unintended, stainless-steel cannula.
- Distribution
- Domestic distribution nationwide. Foreign distribution worldwide.
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-1241-2022, Z-1242-2022, Z-1240-2022
- Nester Embolization Microcoil RPN GPN MWCE-18-5-2-NESTER G52733 MWCE-18-5-3-NESTER G52734 MWCE-18-7-3-NESTER-01 G47338 MWCE-18-7-2-NESTER-01 G47337 MWCE-18-7-5-NESTER-01 G4734
- Tornado Embolization Microcoil, RPN GPN MWCE-18S-5/2-TORNADO G08356 MWCE-18S-6/2-TORNADO G08259 MWCE-18S-3/2-TORNADO-01 G47416 MWCE-18S-4/2-TORNADO-01 G47417 MWCE-18S-3/
- Hilal Embolization Microcoil, RPN MWCE-18-1.0-0-HILAL-01 (GPN G47499); RPN MWCE-18-0.5-0-HILAL-01 (GPN G47498)
Terminated Class II Medical devices
- Recall started
- 12 November 2021
- Terminated
- 26 April 2024
- Reason
- Flexor Check-Flo Introducer Ansel Modification mislabeled. Products packaged with a label stating 6FR may contain a 7FR device and vice versa
- Distribution
- AL, AZ, CA, FL, GA, KS, MI, MO, MN, NC, NY, OK, TX Foreign: BE, CH, DE, ES, FR, GB,GU, IT, ZA
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-0388-2022, Z-0387-2022
- Flexor Check-Flo Introducer Ansel Modification Model Number: KCFW-7.0-18/38-45-RB-ANL0-HC Order Number/GPN: G44155 Intended to introduce therapeutic or diagnostic devices into the vasculature, excluding coronary and ne
- Flexor Check-Flo Introducer Ansel Modification Model Number: KCFW-6.0-18/38-45-RB-ANL0-HC Order Number/GPN: G44154 Intended to introduce therapeutic or diagnostic devices into the vasculature, excluding coronary and n
Ongoing Class II Medical devices
- Recall started
- 13 October 2021
- Reason
- Specific lots may contain black debris on the exterior of the inner stylet and/or on the interior of the outer cannula. Local inflammatory reactions may occur if an affected product is used.
- Distribution
- Worldwide distribution - US Nationwide distribution in the states of AL, CA, IN, LA, MS, MT, NC, NY, OR, PA, WA and the countries of France, Italy, and Mexico.
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-0414-2022
- Skinny Needle with Chiba Tip. RPN 090010, 090010-ET; GPN G14254, G15290. Rigid needle available with EchoTip to provide enhanced visualization of needle tip when used with ultrasonic imaging equipment.
Terminated Class II Medical devices
- Recall started
- 8 October 2021
- Terminated
- 14 February 2024
- Reason
- Transseptal needles may contain rust on the interior and/or exterior of the needle, which may not be detected by the user. Potential adverse events if an affected product is used ranges from increased procedural time to inflammatory reactions, which at their most severe include systemic reactions potentially leading to permanent impairment or death.
- Distribution
- Domestic distribution nationwide. Foreign distribution worldwide.
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-0223-2022, Z-0222-2022
- Transseptal Needle with Catheter. RPN TSN-17-75.0-ENDRYS. GPN G19261. Coaxial set consisting of an outer catheter, curved-tip metal cannula, and a tapered-tip inner needle
- Transseptal Needle, Trocar. RPN TSNC-18-71.0, TSNC-19-56.0. GPN G02364, G02365. Product consists of a transseptal needle and obturator.
Terminated Class II Medical devices
- Recall started
- 28 September 2021
- Terminated
- 28 February 2024
- Reason
- Affected lots may be manufactured incorrectly with the radiopaque marker band located just below the Check-Flo proximal fitting instead of the distal tip. The issue may not be identified until the device is placed in the patient and viewed under fluoroscopy, leading to increased procedural time.
- Distribution
- Domestic distribution to AR AZ CT FL GA IL LA MA ME MN ND NJ NY OH PA TX WA. Foreign distribution to Germany, Finland, Italy, Netherlands, and Switzerland.
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-0186-2022, Z-0185-2022
- Flexor Check-Flo Introducer - Raabe Modification, Introducer Set, RPN KCFW-5.0-38-90-RB-RAABE, GPN G13069 The introducer sheath incorporates a hydrophilic coated Flexor shaft with a hemostasis valve and is provided with
- Flexor Check-Flo Introducer, Introducer Set, RPN KCFW-6.0-35-45-RB, GPN G09908 The introducer sheath incorporates a hydrophilic coated Flexor shaft with a hemostasis valve and is provided with a dilator.
Terminated Class II Medical devices
- Recall started
- 4 January 2021
- Terminated
- 16 September 2021
- Reason
- Manufactured with incorrect material, which may result in the stent being stiffer and potentially susceptible to degradation could potentially lead to breakage/fragmentation.
- Distribution
- FL, KY, MD, NS, NY, PA, TX Foreign: Canada, Switzerland
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-0960-2021
- Cook Medical Kwart Retro-Inject Ureteral Stent Set 7.0Fr. 22-32cm -Used for retrograde injection during Extracorporeal Shock Wave Lithotripsy (E.S.W.L.) and leaving an indwelling ureteral stent Reference Part Number (RP
Terminated Class I Medical devices
- Recall started
- 24 November 2020
- Terminated
- 7 September 2022
- Reason
- There is an increased likelihood of the introducer sheath separating at the proximal bond site. If separation occurs during use, it could result in life-threatening adverse events. The potential adverse events that may occur include, but are not limited to increased procedural time, intervention to retrieve a separated segment, embolization occluding blood flow to a vital organ, vessel injury, and hemorrhage.
- Distribution
- US Nationwide Distribution including states of: AK, CT, IA, LA, MO, NH, OK, TN, WI, AL, DC, ID, MA, MS, NJ, OR, TX, WV, AR, DE, IL, MD, MT, NM, PA, UT, WY, AZ, FL, IN, ME, NC, NV, RI, VA, CA, GA, KS, MI, ND, NY, SC, VT, CO, HI, KY, MN, NE, OH, SD, WA and Global Distribution to: AU, SG, Austria, Belg
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-0598-2021, Z-0606-2021, Z-0599-2021, Z-0596-2021, Z-0608-2021, Z-0591-2021, Z-0610-2021, Z-0607-2021, Z-0605-2021, Z-0601-2021, Z-0600-2021, Z-0609-2021 and 9 more
- Flexor Check-Flo Introducer, RPN KCFW-6.0-35-55-RB-HFANL0-HC, Order Number G56231 Flexor Introducers and Guiding Sheaths are intended to introduce therapeutic or diagnostic devices into the vasculature, excluding cor
- Flexor Check-Flo Introducer, RPN KCFW-8.0-35-55-RB-HFANL0-HC, Order Number G56233 Flexor Introducers and Guiding Sheaths are intended to introduce therapeutic or diagnostic devices into the vasculature, excluding cor
- Flexor Check-Flo Introducer, RPN KCFW-6.0-35-55-RB-HFANL1-HC, Order Number G56236 Flexor Introducers and Guiding Sheaths are intended to introduce therapeutic or diagnostic devices into the vasculature, excluding cor
- Flexor Check-Flo Introducer, RPN KCFW-5.0-35-70-RB-HFANL0-HC, Order Number G56220 Flexor Introducers and Guiding Sheaths are intended to introduce therapeutic or diagnostic devices into the vasculature, excluding cor
- Flexor Check-Flo Introducer, RPN KCFW-8.0-35-70-RB-HFANL0-HC, Order Number G56223 Flexor Introducers and Guiding Sheaths are intended to introduce therapeutic or diagnostic devices into the vasculature, excluding cor
- Flexor Check-Flo Introducer, RPN KCFW-4.0-35-55-RB-HFANL0-HC, Order Number G56229 Flexor Introducers and Guiding Sheaths are intended to introduce therapeutic or diagnostic devices into the vasculature, excluding cor
- Flexor Tuohy-Borst Side-Arm Introducer (Shuttle Select), RPN KSAW-5.0-38-90-RB-SHTL-FLEX-HC, Order Number G31133 Flexor Introducers and Guiding Sheaths are intended to introduce therapeutic or diagnostic devices into
- Flexor Check-Flo Introducer, RPN KCFW-8.0-35-55-RB-HFANL1-HC, Order Number G49045 Flexor Introducers and Guiding Sheaths are intended to introduce therapeutic or diagnostic devices into the vasculature, excluding cor
- and 13 more product lines
Terminated Class I Medical devices
- Recall started
- 22 October 2020
- Terminated
- 2 September 2021
- Reason
- Affected product was dropped during manufacturing, which may result in a bent wire guide and/or the presence of loose foreign matter on the wire guide.
- Distribution
- US distribution to AZ, IA, IN, MA, and ND.
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-0399-2021
- Fixed Core Wire Guide, RPN TSCF-35-260-3, GPN G00517. The product is a curved wire guide with PTFE coating manufactured from stainless steel. To assist in inserting catheters, introducer sheaths, and other medical devic
Terminated Class II Medical devices
- Recall started
- 31 August 2020
- Terminated
- 1 July 2021
- Reason
- The indwell time for the Percutaneous Neonatal Pigtail Nephrostomy Set and Pediatric Nephrostomy Stent Set is decreasing from 4 months to 4 weeks. The current IFU is in process of being updated and will be provided with orders placed following implementation.
- Distribution
- Worldwide distribution - US Nationwide.
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-0106-2021, Z-0105-2021
- Pediatric Nephrostomy Stent Set, Reference Part Numbers 080208-S6 and 080208-S7, Order Numbers G15034 and G15035. The set contains a radiopaque pigtail catheter and stent, retention disc with pull tie, and a 10 French,
- Percutaneous Neonatal Pigtail Nephrostomy Set, Reference Part Number 080106, Order Number G14508. The set contains a 22 gage, 5 cm Chiba Needle; an 18 gage, 5 cm trocar needle; an .038-inch (.97 mm) diameter, 30 cm PTFE
Terminated Class II Medical devices
- Recall started
- 11 August 2020
- Terminated
- 28 July 2021
- Reason
- May contain excess glue within the handle of the delivery system, resulting in an inability to fully deploy the graft via the standard method or troubleshooting deployment method provided in the IFU
- Distribution
- Foreign Only: Canada SWEDEN AUSTRIA Belgium CZECH REPUBLIC DENMARK ESPA¿A FINLAND France GERMANY GREECE HUNGARY IRELAND ISRAEL Italia Nederland NORWAY POLAND PORTUGAL REUNION SERBIA SLOVENIA SOUTH AFRICA SWEDEN UNITED ARAB EMIRATES UNITED KINGDOM AUSTRALIA Hong Kong J
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-2964-2020
- Zenith Alpha Abdominal Endovascular Graft - Distributed OUS only. Indicated for the endovascular treatment of patients with abdominal aortic or aorto-iliac aneurysms Reference Part Number (RPN): ZIMB-22-108 ZIMB-22
Terminated Class II Medical devices
- Recall started
- 30 July 2020
- Terminated
- 4 November 2021
- Reason
- Universa Soft Ureteral Stent Sets were incorrectly placed in the outer boxes labeled as Universa Firm Ureteral Stent Set ;pouch label reflects the packaged product.
- Distribution
- International distribution to the country of Australia.
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-2971-2020
- Cook Medical Universa Soft Ureteral Stent Sets-5 FR 26 cm Reference part Number: UFH-526 - Product Usage: Used for temporary internal drainage from the ureteropelvic junction to the bladder
Terminated Class II Medical devices
- Recall started
- 2 June 2020
- Terminated
- 5 May 2022
- Reason
- Identified products may contain a damaged bushing within the delivery system, which could potentially result in difficulty or an inability to fully deploy the graft via the standard method provided in the IFU.
- Distribution
- Foreign distribution to the countries of Canada, Chile, Argentina, Colombia, Great Britain, AUSTRIA Belgium. CZECH REPUBLIC DENMARK ESPA¿A France GERMANY GREECE HUNGARY ICELAND IRELAND ISRAEL Italia MARTINIQUE Nederland NORWAY POLAND PORTUGAL SERBIA SOUTH AFRICA SWEDEN UNITED KI
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-2546-2020
- Zenith Alpha Abdominal Endovascular Graft, Catalog Prefix ZIMB - Product Usage: is indicated for the endovascular treatment of patients with abdominal aortic or aorto-iliac aneurysms having morphology suitable for endov
Terminated Class II Medical devices
- Recall started
- 30 April 2020
- Terminated
- 29 March 2021
- Reason
- The Pressure Monitoring Set and Tray may contain excess coating material on the outside and/or interior of the needle component which could cause occlusion.
- Distribution
- Worldwide distribution including distribution to the following states: Alaska, Alabama, Arkansas, Arizona, California, Colorado, Connecticut, Washington DC, Delaware, Florida, Georgia, Hawaii, Iowa, Idaho, Illinois, Indiana, Kansas, Kentucky, Louisiana, Massachusetts, Maryland, Maine, Michigan, Minn
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-2203-2020, Z-2204-2020
- Pressure Monitoring Set, C-PMS, G02838
- Pressure Monitoring Tray, C-PMSY, G02854
Terminated Class II Medical devices
- Recall started
- 15 April 2020
- Terminated
- 15 April 2024
- Reason
- The firm received complaints on five devices from one specific lot where the stent dislodged from the balloon delivery system prior to deployment. Use of affected product may lead to increased procedural time and additional intervention to remove a stent that dislodged within an unintended location.
- Distribution
- Domestic distribution to OH, IA, VA, and NC. Foreign distribution to Canada.
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-1940-2020
- Formula 418 Renal Balloon-Expandable Stent, RPN FOR418-18-80-6-12, Order Number (GPN) G56611 Balloon-expandable stent made of 316L stainless steel with a slotted tube configuration. It is pre-mounted on an over-the-wi
Terminated Class II Medical devices
- Recall started
- 6 March 2020
- Terminated
- 18 December 2020
- Reason
- The printed expiration dates are incorrect.
- Distribution
- The products were distributed to the following US states: CO, DE, MD, MN, NC, NV, and WV. The products were distributed to the following foreign countries: Canada.
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-1566-2020, Z-1565-2020
- Spectrum Central Venous Catheter Tray, Irrigation Catheter, Reference Part Number C-UDLMY-401 J-ABRM-HC-IHl-FST, Order Number G48053
- Pericardiocentesis Set, Irrigation Catheter, Reference Part Number C-PCS-700-TORONT0-042997, Order Number G10643
Terminated Class II Medical devices
- Recall started
- 16 January 2020
- Terminated
- 12 November 2020
- Reason
- The firm has identified factors which contributed to failures relating to thrombus formation and/or lumen occlusion reports, and is highlighting to customers key points of the IFU pertaining to the prevention of these factors. In addition, the firm has updated the Planning and Sizing Worksheet to include associated information.
- Distribution
- Foreign distribution only. Distributed worldwide.
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-1355-2020
- Zenith Alpha Spiral-Z Endovascular Leg, Catalog Prefix ZISL. The product is an endovascular graft.
Terminated Class II Medical devices
- Recall started
- 27 December 2019
- Terminated
- 23 November 2020
- Reason
- Specific lots of the Dawson-Mueller Drainage Catheter were not manufactured to specification, which could lead to leakage from the Mac-Loc hub assembly.
- Distribution
- Nationwide domestic distribution. Foreign distribution to Australia, Belgium, Canada, China, Colombia, Denmark, France, Germany, Italy, Korea, Netherlands, New Zealand, Peru, Russia, Spain, Sweden, and United Kingdom.
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-0815-2020
- Dawson-Mueller Drainage Catheter, Multipurpose Drainage Catheter, RPN UL T6.3-35-15-P-5S-CLDM-HC, UL T6. 3-35-25-P-58-CLDM-HC, UL T7.0-35-25-P-5S-CLDM-HC
Terminated Class I Medical devices
- Recall started
- 8 November 2019
- Terminated
- 1 February 2021
- Reason
- A manufacturing error may cause the radiopaque marker bands to be too loose on certain CXC3.0 CrossCath Support Catheters (compatible with 0.014 wire guides) and too tight on certain CXC3.4 CrossCath Support Catheters (compatible with 0.018 wire guides).
- Distribution
- Distribution to US states of AL, CA, CT, FL, GA, IL, MA, MD, MI, MO, NC, NJ, NY, OH, OK, PA, SC, TN, TX, and VA, and Brazil.
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-0588-2020, Z-0590-2020, Z-0589-2020, Z-0587-2020
- CrossCath Support Catheter, RPN CXC3.4-2.2-18-90-P-NS-0, Order Number G51575 For diagnostic and interventional vascular procedures.
- CrossCath Support Catheter, RPN CXC3.4-2.2-18-150-P-NS-0, Order Number G51577 For diagnostic and interventional vascular procedures.
- CrossCath Support Catheter, RPN CXC3.4-2.2-18-135-P-NS-0, Order Number G51583 For diagnostic and interventional vascular procedures.
- CrossCath Support Catheter, RPN CXC3.0-1.9-14-90-P-NS-0, Order Number G51572 For diagnostic and interventional vascular procedures.
Terminated Class II Medical devices
- Recall started
- 21 October 2019
- Terminated
- 30 June 2020
- Reason
- The potential compromise of the sterility of products from an undersealed chevron in the packaging pouches of Torcon NB Advantage Catheters.
- Distribution
- Worldwide distribution - US Nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, IL, IN, KS, LA, MA, MD, Ml, MN, MO, MS, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SD, TN, TX, UT, VA, WA, WI, WV, and PR. Countries of AU, BE, CA, CN, CZ, DK, FR, DE, GT, HK, IT, KR, LU, MX
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-0435-2020, Z-0471-2020, Z-0487-2020, Z-0483-2020, Z-0479-2020, Z-0456-2020, Z-0441-2020, Z-0472-2020, Z-0452-2020, Z-0426-2020, Z-0427-2020, Z-0439-2020 and 59 more
- Torcon NB Advantage Catheters - Product Usage: Intended for use in the peripheral and coronary vascular system, including the carotid arteries, in giographic procedures by physicians trained and experienced in angiograph
- Torcon NB Advantage Catheters - Product Usage: Intended for use in the peripheral and coronary vascular system, including the carotid arteries, in giographic procedures by physicians trained and experienced in angiograph
- Torcon NB Advantage Catheters - Product Usage: Intended for use in the peripheral and coronary vascular system, including the carotid arteries, in giographic procedures by physicians trained and experienced in angiograph
- Torcon NB Advantage Catheters - Product Usage: Intended for use in the peripheral and coronary vascular system, including the carotid arteries, in giographic procedures by physicians trained and experienced in angiograph
- Torcon NB Advantage Catheters - Product Usage: Intended for use in the peripheral and coronary vascular system, including the carotid arteries, in giographic procedures by physicians trained and experienced in angiograph
- Torcon NB Advantage Catheters - Product Usage: Intended for use in the peripheral and coronary vascular system, including the carotid arteries, in giographic procedures by physicians trained and experienced in angiograph
- Torcon NB Advantage Catheters - Product Usage: Intended for use in the peripheral and coronary vascular system, including the carotid arteries, in giographic procedures by physicians trained and experienced in angiograph
- Torcon NB Advantage Catheters - Product Usage: Intended for use in the peripheral and coronary vascular system, including the carotid arteries, in giographic procedures by physicians trained and experienced in angiograph
- and 63 more product lines
Terminated Class II Medical devices
- Recall started
- 3 October 2019
- Terminated
- 13 May 2020
- Reason
- Bent transfer catheter tips may cause difficulty or inability in inserting the transfer catheter into the guiding catheter. Potential adverse events that may occur if an affected product is used include increased procedural time, the need for a repeat embryo transfer procedure, or the need for a patient to undergo an additional IVF cycle.
- Distribution
- No US distribution. The products were distributed to the following foreign countries: Belgium, Denmark, Germany, Italy, Netherlands, Norway, Poland, Republic of Macedonia, Slovenia, Spain, and Switzerland.
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-0415-2020, Z-0416-2020
- Guardia Access Nano Embryo Transfer Catheter is use for transferring IVF embryos into the uterine cavity.
- Guardia Access Embryo Transfer Catheter is used for transferring IVF embryos into the uterine cavity.
Terminated Class II Medical devices
- Recall started
- 17 September 2019
- Terminated
- 21 May 2020
- Reason
- Lidstock perforation line may be over the sterile area of the tray, when it should be on the lip of the tray, potentially compromising the sterility of the device.
- Distribution
- US Nationwide distribution in the states of CA, FL, IN, MI, MT, NC, NY, OH.
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-0506-2020
- Pneumothorax Tray, Catalog No. C-UTPTY-1400-WAYNE-112497-IMH, GPN G56537 - Product Usage: The modified Wayne Pneumothorax Set is used for the relief of simple, spontaneous, iatrogenic, and tension pneumothorax.
Terminated Class I Medical devices
- Recall started
- 24 May 2019
- Terminated
- 31 October 2019
- Reason
- Multiple complaints were received for balloons bursting below the rated burst pressure. Potential adverse events that may occur if an affected product is used include a delay in the procedure, additional intervention, vessel injury, and balloon fragmentation in the patient.
- Distribution
- Worldwide Distribution - US Nationwide The affected product was distributed to the following states: FL, OH, WA. The products were distributed to the following foreign countries: Belgium, Denmark, Finland, Germany, Italy, Netherlands, Norway, Poland, South Africa, Spain, Sweden, Switzerland, Un
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-1758-2019, Z-1756-2019, Z-1757-2019
- Advance Enforcer 35 Focal Force PTA Balloon Catheter 6mm x 4cm, 5FR/135cm, Catalog Number ASB5-35-135-6-4, REF Number G35257 Product Usage: The Advance Enforcer 35 Focal-Force PTA Balloon Catheter is intended for per
- Advance Enforcer 35 Focal Force PTA Balloon Catheter 6mm x 4cm, 5FR/50cm, Catalog Number ASB5-35-50-6-4, REF Number G35248 Product Usage: The Advance Enforcer 35 Focal-Force PTA Balloon Catheter is intended for percu
- Advance Enforcer 35 Focal Force PTA Balloon Catheter 6mm x 4cm, 5FR/80cm, Catalog Number ASB5-35-80-6-4, REF Number G35252 Product Usage: The Advance Enforcer 35 Focal-Force PTA Balloon Catheter is intended for percu
Terminated Class II Medical devices
- Recall started
- 2 May 2019
- Terminated
- 23 August 2019
- Reason
- The products may have been manufactured without the bottom seal on the outer packaging, compromising the sterility of the product.
- Distribution
- No distribution in the US. Products were distributed to the following foreign countries: China.
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-1780-2019
- Neff Percutaneous Access Set, Catalog Number NPAS-100-RH-NT
Terminated Class II Medical devices
- Recall started
- 26 April 2019
- Terminated
- 8 November 2019
- Reason
- may contain a damaged gray safety lock knob, which could potentially result in difficulty or an inability to fully deploy the graft via the standard or troubleshooting method provided in the Instructions for Use (IFU). Potential adverse events that may occur if an affected product is used include a prolonged procedure and open surgical intervention.
- Distribution
- No US distribution. Product distributed to the following foreign countries: Australia, Austria, Belgium, Brazil, Canada, Canary Islands, Chile Columbia, Costa Rica, Czech Republic, Denmark, Finland, France, Germany, Greece, Ireland, Israel, Italy, Japan, Morocco, Netherlands, Norway, Poland, Port
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-1844-2019
- Zenith Alpha Abdominal Endovascular Graft, Reference Part Numbers: ZIMB-22-108, ZIMB-22-118, ZIMB-22-128, ZIMB-22-70, ZIMB-22-84, ZIMB-22-98, ZIMB-24-108, ZIMB-24-118, ZIMB-24-128, ZIMB-24-70, ZIMB-24-84, ZIMB-24-98, ZIM
Terminated Class II Medical devices
- Recall started
- 22 April 2019
- Terminated
- 5 May 2020
- Reason
- Kwart RetroInject Stent Set (Multi-Length) and Salle Intraoperative Pyeloplasty Stent Set Instructions for Use (IFU) did not include a warning for stent knotting.
- Distribution
- Nationwide Foreign: Australia Canada Austria Belgium Denmark Germany Spain Finland France Martinique French Polynesia Switzerland Ireland Italy Netherlands Poland Sweden United Kingdom Hong Kong New Zealand Chile Colombia Israel Russian Federation Czech Republic Greece Jo
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-1369-2019, Z-1370-2019
- Cook Kwart Retro-Inject Ureteral Stent Set: PRODUCT/CATALOG NUMBER/GPN: 4.7 Fr 003500 G14916; 6.0 Fr 003600 G14836; 7.0 Fr 003700 G14837; 8.0 Fr 003800 G14844; Set with AQ¿ Hydrophilic Coating PRODU
- Cook Salle Intraoperative Pyeloplasty Stent Set PRODUCT/CATALOG NUMBER/GPN: 4.0 Fr - SIPSF-040018-56-6 G18168 4.0 Fr - SIPSF-040018-59 G32773 4.7 Fr SIPSF-050018-59 G32774
Terminated Class II Medical devices
- Recall started
- 9 April 2019
- Terminated
- 28 April 2020
- Reason
- Affected products may have been manufactured with the catheter distal end hole too small
- Distribution
- Domestically: OH; Internationally: CANADA, SPAIN, ITALY, BELGIUM, GERMANY, UNITED ARAB EMIRATES, GREECE, SWITZERLAND, OMAN, MONACO, SAUDI ARABIA, AUSTRALIA, HONG KONG, INDIA & KOREA
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-1725-2019, Z-1724-2019, Z-1722-2019, Z-1723-2019
- Thoracentesis Set, RPN C-THS-850, Order # G03286
- Pericardiocentesis Catheter Tray, RPN C-PCSY-850-LHSC-061096, Order # G10030
- Pericardiocentesis Catheter Set, RPN C-PCS-850, Order # G03282
- Pericardiocentesis Catheter Tray, RPN C-PCYS-850, Order # G03284
Terminated Class I Medical devices
- Recall started
- 1 February 2019
- Terminated
- 30 June 2020
- Reason
- The products were manufactured without a back bevel on the needle tip. This could cause damage to the inside of the introducer during needle insertion.
- Distribution
- US nationwide and Canada. US distribution to the following states: AZ, FL, IL, IN, KS, MA, MI, MO, PA, SC, TN, and WA. The products were distributed to the following foreign countries: Canada.
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-0979-2019
- Cook Transseptal Needle, Reference Number TSNC-18-71.0, Order Number G02364. For transseptal left heart access in both diagnostic and interventional procedures.
Terminated Class II Medical devices
- Recall started
- 4 January 2019
- Terminated
- 29 May 2020
- Reason
- The wire guide may be incorrectly loaded into the wire guide holder, leading to the stiff tip of the wire guide being introduced into the patient instead of the flexible tip.
- Distribution
- Domestic distribution to nationwide US. Foreign distribution to Australia, Belgium, Bolivia, Canada, Canary Islands, Chile, China, Colombia, Croatia, Czech Republic, Denmark, France, French Polynesia, Germany, Guatemala, Honduras, Hungary, India, Ireland, Italy, Japan, Korea, Luxembourg, Malay
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-0977-2020, Z-0962-2020, Z-0966-2020, Z-0972-2020, Z-0970-2020, Z-0967-2020, Z-0975-2020, Z-0974-2020, Z-0964-2020, Z-0971-2020, Z-0963-2020, Z-0959-2020 and 8 more
- Universa Firm Ureteral Stent, Global Product No. G49864 G49865 G49866 G49867 G49868 G49869 G49870 G49874 G49875 G49877 G49879 G49881 G49882 G49887 G49888
- PTFE Wire Guide, Guidewire, Global Product Number 638813 635413-10
- Angled Tip Ureteral Catheter Set with Bentson PTFE Wire Guide, Ureteral Catheter, Global Product No. G19108 G14587 G15302 G14326
- Bander Ureteral Diversion Open-End Stent Set, Global Product No. G14822 G14823
- Urethral Dilation Balloon Catheter with Open Tip , Global Product No. G17844
- Percutaneous Entry Set, Nephrostomy Catheter, Global Product No. G14649
- Sof-Flex Double Pigtail Ureteral Stent Set, Global Product No. G15076 G14906 G14840 G14865 G17128
- Kwart Retro-Inject Ureteral Stent Set, Global Product No. G15016 G14836 G14885 G14887 G17151
- and 12 more product lines
Terminated Class II Medical devices
- Recall started
- 21 December 2018
- Terminated
- 30 April 2020
- Reason
- The products were manufactured with longer than specified sheaths.
- Distribution
- The products were distributed to the following US states: CA, CT, GA, MA, NY, PA, VA, and WI.
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-0851-2019
- Check-Flo Performer Introducer, Catalog Number RCFW-5.0-25-45-RB-CHB, GPN G11571
Terminated Class II Medical devices
- Recall started
- 19 December 2018
- Terminated
- 28 April 2020
- Reason
- There is a potential for difficulty in sliding down the retention mechanism on the sets manufactured with a specific extension spring lot.
- Distribution
- Worldwide Distribution: US (nationwide) to of states: AL, AZ, CA, GA, HI, IL, IN, KY, LA, MI, MN, MO, NC, NJ, NY, OH, PA, RI, TN, TX, and UT; and countries of: Austria, Belgium, Canada, Denmark, France, Germany, Great Britain, Hungary, Iceland, Italy, Netherlands, Norway, Spain, and Switzerland.
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-0968-2019, Z-0967-2019
- Entuit Secure Gastrointestinal Suture Anchor Set, Catalog Number GIAS-SRM-3, Order Number G35570 The Entuit Secure Gastrointestinal Suture Anchor Set is intended for anchoring the anterior wall of the stomach to the a
- Entuit Secure Gastrointestinal Suture Anchor Set, Catalog Number GIAS-SRM-2, Order Number G35569 The Entuit Secure Gastrointestinal Suture Anchor Set is intended for anchoring the anterior wall of the stomach to the a
Terminated Class II Medical devices
- Recall started
- 31 May 2018
- Terminated
- 11 April 2019
- Reason
- This lot of VentriClear failed endotoxin testing. Potential adverse events include immune responses ranging from non-specific febrile reaction to life-threatening systemic inflammatory response syndrome (SIRS).
- Distribution
- One distributor in Indiana; product was not further distributed to end users.
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-2791-2018
- VentriClear Ventricular Drainage Catheter Set, Cat. No. N-VVDC-01-ABRM Product Usage: The VentriClear Ventricular Drainage Catheter Set has been designed for obtaining access to a ventricular cavity of the brain for
Terminated Class II Medical devices
- Recall started
- 16 April 2018
- Terminated
- 27 June 2019
- Reason
- Incorrect product labeling. Product labeled as 6mm x 2cm balloon are packaged with a 4mm x 4 cm balloon label and vice-versa.
- Distribution
- Worldwide Distribution - US Nationwide in the states of AZ, DE, FL, GA, ID, IL, KY, MI, MO, NH, NJ, PA, RI, TX, UT and the country of Peru
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-1521-2018
- Advance¿ 35LP Low-Profile PTA Balloon Dilatation Catheter; Catalog numbers: PTA5-35-135-6-2 .0 and PTA5-35-135-4-4 .0; UDI: 10827002522647 and 10827002522524 Product Usage: The Advance¿ 35LP Low Profile PTA Balloo
Terminated Class II Medical devices
- Recall started
- 29 March 2018
- Terminated
- 9 April 2019
- Reason
- Cap-adapter assembly of the device may potentially not be adequately tightened, possibly leading to fluid leakage from or air aspiration into the connecting tube.
- Distribution
- Distributed to accounts in 7 states: AL, CA, GA, NE, NV, NY, and TX. International distribution to Japan, Korea, and Taiwan.
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-1752-2018, Z-1755-2018, Z-1756-2018, Z-1754-2018, Z-1753-2018, Z-1751-2018
- Connecting Tube - Female to Male Luer Lock, HPCT8.8-60-M-FM-BNS Product Usage: The connecting tube is intended to be a conduit between a fluid reservoir and a vascular catheter in the patient. The connecting tube may
- Connecting Tube - Female to Male Luer Lock, HPCT8.8-100-M-FM Product Usage: The connecting tube is intended to be a conduit between a fluid reservoir and a vascular catheter in the patient. The connecting tube may be
- Connecting Tube - Female to Male Luer Lock, HPCT8.8-120-M-FM Product Usage: The connecting tube is intended to be a conduit between a fluid reservoir and a vascular catheter in the patient. The connecting tube may be
- Connecting Tube - Female to Male Luer Lock, HPCT8.8-80-M-FM Product Usage: The connecting tube is intended to be a conduit between a fluid reservoir and a vascular catheter in the patient. The connecting tube may be
- Connecting Tube - Female to Male Luer Lock, HPCT8.8-30-M-FM Product Usage: The connecting tube is intended to be a conduit between a fluid reservoir and a vascular catheter in the patient. The connecting tube may be
- Connecting Tube - Female to Male Luer Lock, HPCT8.8-60-M-FM Product Usage: The connecting tube is intended to be a conduit between a fluid reservoir and a vascular catheter in the patient. The connecting tube may be
Terminated Class II Medical devices
- Recall started
- 22 March 2018
- Terminated
- 13 May 2020
- Reason
- Pump was found to be constructed from crimp connectors that do not meet the fixation wiring requirements and can cause failure of the device to operate or electric shock or burn to the user.
- Distribution
- Worldwide Distribution -- USA, Argentina, Bolivia, Brazil, Canada, Colombia, Guatemala, Mexico, Panama, Peru, Uruguay, and Venezuela.
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-1461-2018, Z-1460-2018
- Cook Vacuum Pump ( Cook Aspiration Unit) K-MAR-5200-US, GPN G49275. Intended for the aspiration of body fluids and cells, in particular oocyte aspiration.
- Cook Vacuum Pump (Cook Aspiration Unit) K-MAR-5200-US, GPN G51607. Intended for the aspiration of body fluids and cells; in particular oocyte aspiration.
Terminated Class II Medical devices
- Recall started
- 31 January 2018
- Terminated
- 17 April 2020
- Reason
- The interior of the product may not be sterilized to the appropriate Sterility Assurance Level when following the reprocessing instructions in the Instructions for Use.
- Distribution
- US Nationwide, OUS countries: AE, AT, AU, BE, BG, BR, CA, CH, CL, CN, CO, CR, CZ, DE, DJ, DK, DO, EC, EE, EG, ES, FI, FJ, FR, GB, GR, GT, HK, HN, HR, HU, ID, IE, IL, IN, IT, JO, KZ, LB, LT, LV, MA, MO, MX, MY, NI, NL, NO, NZ, PE, PH, PK, PL, PR, PT, QA, RO, RU, SA, SE, SG, SI, SK, SV, TH, TT, TW, U
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-1591-2018, Z-1596-2018, Z-1594-2018, Z-1590-2018, Z-1595-2018, Z-1593-2018, Z-1592-2018
- StoneBreaker Exhaust Cap, Catalog number: SBA-EC; GPN: G52599; UDI: (01)00827002525993
- StoneBreaker Single Use Probe, Catalog numbers: SBP-010500; SBP-010605; SBP-016500; SBP-016605; SBP-020425 GPN : G52607; G52605; G52606; G52609; G52608
- StoneBreaker Sterilization Cap, Catalog number: SBA-SC; GPN: G52602; UDI: (01)00827002526020
- StoneBreaker Pneumatic Lithotripter, Catalog number: SBL-KIT1; GPN: G52604 UDI:(01)00827002526044
- StoneBreaker C02 Cartridge, Catalog number: SBC-10; GPN: G52603; UDI: (01)10827002526034
- StoneBreaker Probe Cap, Catalog number:SBA-PC; GPN: G52601; UDI:(01)00827002526013
- StoneBreaker Exhaust Line, Catalog number: SBA-EL; GPN: G52600; UDI: (01)00827002526006
Terminated Class II Medical devices
- Recall started
- 12 January 2018
- Terminated
- 8 May 2020
- Reason
- The temperature of the incubation lids of the MINC Bench top Incubators may become too high or too low for proper gamete/embryo incubation due to fatigue failure of the flat flex lid heater.
- Distribution
- Nationwide (K-IMNC-100 -US G46022) . (K-IMNC-100 G20079 manufactured in Australia and distributed worldwide)
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-1378-2018, Z-1377-2018
- MINC Benchtop Incubator (Mini Incubator) catalog (K-MINC-1000) G20079 is a microprocessor controlled, gassed, humidified, incubator intended for use in cell culture.
- MINC Benchtop Incubator (Mini Incubator) catalog (K-MINC-1000 US) G46022 is a microprocessor controlled, gassed, humidified, incubator intended for use in cell culture.
Terminated Class II Medical devices
- Recall started
- 19 December 2017
- Terminated
- 9 April 2019
- Reason
- Certain lots of in vitro fertilization (IVF) products were distributed without completed finished device quality testing with potential adverse events that may include nonviable gametes and/or nonviable embryos.
- Distribution
- Worldwide Distribution - U.S. Nationwide in the states of: CA, CO, FL, IN, MA, NE, NJ, OH, TN, TX, and WI. and countries of Belarus, Belgium, Brazil, Canada, Chile, China, Egypt, Italy, Japan, Poland, Saudi Arabia, Spain, and United Kingdom.
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-1185-2018, Z-1186-2018, Z-1188-2018, Z-1187-2018
- Norm-Ject Tuberkulin Syringe, Catalog Number: K-ATS-1000, Global Product Number: G19871 Product Usage: Used to place in vitro fertilized (IVF) embryos into the uterine cavity. Intended for one-time use
- Echosight Jansen-Anderson Intratubal Transfer Set, Catalog Number: K-J-JITS-572900, Global Product Number: G17946 Product Usage: Used to place in vitro fertilized (IVF) embryos into the uterine cavity. Intended for
- 1. SMS Embryo Transfer Set, Catalog Number: K-J-SMS-571800, Global Product Number: G19182; 2. Guardia" Access Nano Embryo Transfer Catheter, Catalog Number: K-JETS-551910-L, Global Product Number: G24215; 3. Sydney IVF¿
- Marrs Laparoscopic GIFT Catheters, Catalog Number:K-J-MLC-503000, Global Product Number: G18022 Product Usage: Used to place in vitro fertilized (IVF) embryos into the uterine cavity. Intended for one-time use
Terminated Class II Medical devices
- Recall started
- 19 October 2017
- Terminated
- 16 April 2019
- Reason
- Cook has identified that the affected lots may contain wire guides that were loaded backwards into the wire guide holder.
- Distribution
- US Nationwide Distribution
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-0830-2018, Z-0831-2018, Z-0832-2018
- Mallinckrodt Institute Percutaneous Gastrostomy Set Product Usage: Intended for the percutaneous placement of a catheter for gastroenteric feeding
- Percutaneous Gastrojejunostomy Set Product Usage: Intended for the percutaneous placement of a catheter for gastroenteric feeding
- Wills-Oglesby Percutaneous Gastrostomy Set Product Usage: Intended for the percutaneous placement of a catheter for gastroenteric feeding
Terminated Class II Medical devices
- Recall started
- 13 October 2017
- Terminated
- 23 October 2019
- Reason
- Label does not state that the product is heparin-coated.
- Distribution
- Nationally
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-0954-2018, Z-0959-2018, Z-0948-2018, Z-0950-2018, Z-0951-2018, Z-0966-2018, Z-0961-2018, Z-0949-2018, Z-0956-2018, Z-0965-2018, Z-0953-2018, Z-0947-2018 and 8 more
- Coons Interventional Wire Guide
- Movable Core Wire Guide
- Amplatz Extra Stiff Wire Guide
- Amplatz Ultra Stiff Wire Guide
- Bentson Cerebral Wire Guide
- Femoral Artery Pressure Monitoring Set
- Reuter Tip Deflecting Wire Guide
- Amplatz Stiff Wire Guide
- and 12 more product lines
Terminated Class II Medical devices
- Recall started
- 31 July 2017
- Terminated
- 8 May 2019
- Reason
- Central venous catheters may not fit over guide wire potentially causing procedure delay.
- Distribution
- Worldwide distribution - US Nationwide in the states of: AL, CA, CO, FL, GA, IA, IL, KY, LA, MD, MI, MN, MO, NC, NY, OH, OK, OR, PA, RI, TN, TX, UT, VA, WA, WI; the District of Columbia and the countries of Belgium, India Italy, Spain, and United Kingdom.
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-0633-2018, Z-0632-2018, Z-0630-2018, Z-0631-2018, Z-0634-2018
- Central Venous Tray Triple Lumen Polyurethane Catheter, Catalog Identifier: C-UTLMY-501J-CCT Product Usage: The central venous catheter is designed for treatment of critically ill patients and is suggested for: 1. Co
- Central Venous Catheter Set Heparin Coated Triple Lumen Polyurethane, Catalog Identifiers: C-UTLM-501J-25-6.5-WCE-BH, C-UTLM-501J-BH Product Usage: The central venous catheter is designed for treatment of critically
- Central Venous Catheter Set Triple Lumen Polyurethane, Catalog Identifier: C-UTLM-501J Product Usage: The central venous catheter is designed for treatment of critically ill patients and is suggested for: 1. Continuo
- Central Venous Catheter Tray Triple Lumen Polyurethane, Catalog Identifier: C-UTLMYJ-501J Product Usage: The central venous catheter is designed for treatment of critically ill patients and is suggested for: 1. Conti
- Spectrum Central Venous Catheter Tray Minocycline/Rifampin Antibiotic Impregnated, Catalog Identifiers: C-UTLMY-501J-ABRM-HC-FST and C-UTLMY-501J-ABRM-HC-IHI-FST Usage: The Spectrum Central Venous Catheter is used fo
A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.