Centinel Spine, Inc. recalls

4 recalls recorded since 2018, most recently on 20 March 2026. 3 are listed as ongoing. Status and classification are as published by the FDA.

Recalls4
Ongoing (FDA)3
Terminated or completed1
Class I (FDA)0

Ongoing Class II Medical devices

Recall started
20 March 2026
Reason
Product labeling mix up. The prodisc C SK U.S. Implant Extra Large 6mm product was labeled as a 5mm and prodisc C SK U.S. Implant Extra Large 5mm product was labeled as a 6mm.
Distribution
US Nationwide distribution in the states of AZ, CA, GA, LA, MO, NY, TN, TX.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2050-2026, Z-2051-2026

Ongoing Class II Medical devices

Recall started
31 December 2025
Reason
Products were mislabeled as the 6mm product but included the 5 mm product.
Distribution
US Nationwide distribution in the states of AL, CA, CO, LA, NY, OR, TN, TX.
Initiated
Voluntary: Firm initiated
Recall number
Z-1426-2026

Ongoing Class II Medical devices

Recall started
18 July 2024
Reason
During the assembly of Prodisc C, a blue discoloration was observed on a Prodisc C inlay. The discoloration is considered a foreign contaminate that may not have been fully evaluated in the manufacturing validations.
Distribution
US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-3212-2024, Z-3213-2024

Terminated Class II Medical devices

Recall started
5 March 2021
Terminated
2 August 2021
Reason
Centinel Spine learned through five customer complaints that the product was missing the "UP" etching on the superior plate of the prodisc C device. Per the Surgical Technique, the "UP" is one of the methods used to indicate the direction in which the surgeon should place the prodisc C into intradiscal space. There are also "UP" indications in the packaging and on instrumentation as additional guides to help the surgeon properly implant the device that were present for the affected product.
Distribution
US Nationwide distribution in the states of CA, MD, DC, TN, GA, NC, PA, KY, NY, AZ, CO, OH, FL, MI.
Initiated
Voluntary: Firm initiated
Recall number
Z-1406-2021

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.