Recall Record

Vernon Hills, IL, United States

Carefusion 2200 Inc recalls

9 recalls recorded since 2018, most recently on 13 July 2023. 5 are listed as ongoing. Status and classification are as published by the FDA.

Recalls9
Ongoing (FDA)5
Terminated or completed4
Class I (FDA)0

Ongoing Class II Medical devices

Recall started
13 July 2023
Reason
Specific lots of the product have the potential to exhibit cracking or breaking of the rotation knob due to the incorrect use of an adhesive in the manufacturing process.
Distribution
US and Canada
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2421-2023, Z-2420-2023, Z-2428-2023, Z-2417-2023, Z-2418-2023, Z-2415-2023, Z-2422-2023, Z-2423-2023, Z-2427-2023, Z-2416-2023, Z-2424-2023, Z-2419-2023 and 3 more

Ongoing Class II Medical devices

Recall started
16 March 2023
Reason
A portion of the two products were incorrectly marked (etched) and packaged resulting in a product mismatch. The labels and laser etchings were switched on the two products.
Distribution
Distribution was made to AL, AR, AZ, CA, CO, CT, FL, HI, IA, IL, IN, KY, LA, MA, MD, MI, MN, MO, MT, NC, NE, NJ, NM, NV, NY, OH, OR, PA, SC, TN, TX, VA, WA, WI, and WV, including Puerto Rico. There was also government/military distribution. Foreign distribution was made to Canada.
Initiated
Voluntary: Firm initiated
Recall number
Z-1421-2023

Ongoing Class II Medical devices

Recall started
30 January 2023
Reason
Product did not meet shelf-life testing requirements resulting in a breach of the sterility of the Genesis container holding surgical equipment.
Distribution
Worldwide distribution - US Nationwide, including Puerto Rico. There was also government/military distribution. The countries of Australia, Canada, Denmark, Finland, Hong Kong, Ireland, Japan, Malaysia, Philippines, Poland, Saudi Arabia, Singapore, South Korea, Spain, Sweden, Taiwan, Thailand, Un
Initiated
Voluntary: Firm initiated
Recall number
Z-1236-2023

Ongoing Class II Medical devices

Recall started
2 November 2022
Reason
BD has determined that the Instructions For Use (IFU) provided by BD is missing certain content that is contained in the manufacturer's IFU. This missing content is related to the interface with the device power supply, as well as cleaning and maintenance instructions. Refer to the attached V. Mueller" Bipolar Forcep IFU.
Distribution
Nationwide, Austria, Canada, Egypt, Germany, Hong Kong, Israel, Philippines, Saudi Arabia, South Africa, South Korea, Thailand
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0391-2023, Z-0382-2023, Z-0377-2023, Z-0411-2023, Z-0415-2023, Z-0410-2023, Z-0393-2023, Z-0394-2023, Z-0397-2023, Z-0387-2023, Z-0404-2023, Z-0409-2023 and 33 more

Ongoing Class II Medical devices

Recall started
26 September 2022
Reason
Genesis STERRAD Reusable Rigid Sterilization Containers have not consistently met the requirements for aerosol challenge testing.
Distribution
US and Australia, Bahrain, Canada, Denmark, Hong Kong, Italy, Japan, Malaysia, Mexico, Norway, Philippines, Saudi Arabia, Singapore, South Africa, Sweden, Thailand, UK, United Arab Emirates
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0214-2023, Z-0211-2023, Z-0213-2023, Z-0210-2023, Z-0204-2023, Z-0217-2023, Z-0205-2023, Z-0203-2023, Z-0208-2023, Z-0209-2023, Z-0207-2023, Z-0219-2023 and 5 more

Terminated Class II Medical devices

Recall started
23 February 2022
Terminated
2 May 2023
Reason
The finished device package did not contain the IFU.
Distribution
Worldwide distribution - US Nationwide distribution in the states of AL, AZ, CA, DC, FL, GA, IA, IL, IN, LA, MA, MI, MN, MS, MT, NC, NY, OH, OK, PA, TN, TX, VA, and WI. The countries of Taiwan and Australia.
Initiated
Voluntary: Firm initiated
Recall number
Z-0833-2022

Terminated Class II Medical devices

Recall started
28 May 2020
Terminated
20 September 2024
Reason
The Universal Drainage Set (UDS) may contain valves that leak internally which may result in the introduction of up to 30cc of air into the patient.
Distribution
370 US customers and 13 distributors. Product distributed in: AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MD, MS, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY Customer listed, including Federal distribution
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2659-2020, Z-2652-2020, Z-2656-2020, Z-2657-2020, Z-2649-2020, Z-2651-2020, Z-2647-2020, Z-2650-2020, Z-2654-2020, Z-2648-2020, Z-2653-2020, Z-2658-2020 and 2 more

Terminated Class II Medical devices

Recall started
18 May 2020
Terminated
25 April 2021
Reason
Becton Dickinson and Company, is conducting a voluntary recall on the V. MuellerTM 3H Knife Handles catalog number listed in the table above. BD has been informed by our supplier that the external threads on some of the collets (Figure 1) contain a machining burr, potentially causing the collet to not fully seat in the knife handle (Figure 2), inhibiting complete grip of the inserted knife blade possibly resulting in a loose blade.
Distribution
US Nationwide distribution including in the states of CA, FL, VA and KY.
Initiated
Voluntary: Firm initiated
Recall number
Z-2400-2020

Terminated Class II Medical devices

Recall started
23 February 2017
Terminated
9 July 2019
Reason
The integrity of the sterile packaging is potentially compromised.
Distribution
Nationwide, Australia, Brazil, Canada, China, Japan, Kuwait, Malaysia, Mexico, Netherlands, New Zealand, Philippines, Singapore, South Africa, Taiwan, Thailand, United Arab Emirates, Vietnam
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1422-2018, Z-1426-2018, Z-1425-2018, Z-1423-2018, Z-1424-2018, Z-1421-2018

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.