Recall Record

College Station, TX, United States

CardioQuip, LLC recalls

4 recalls recorded since 2018, most recently on 24 October 2023. 4 are listed as ongoing. Status and classification are as published by the FDA.

Recalls4
Ongoing (FDA)4
Terminated or completed0
Class I (FDA)0

Ongoing Class II Medical devices

Recall started
24 October 2023
Reason
There is the potential that cooler-heater devices may contain bacterial contamination which could result in patient infection.
Distribution
US: California, Texas, Pennsylvania, Washington, Ohio, Nebraska OUS: None
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0566-2024, Z-0565-2024

Ongoing Class II Medical devices

Recall started
19 October 2023
Reason
There is a potential for the heating-chamber to malfunction resulting in melting/burning of the heater chamber components.
Distribution
US Distribution: TX, IN, IL, WI, FL, NJ, PA, KY, VA, WV, CT, CA, WA, AL, IA
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0439-2024, Z-0438-2024

Ongoing Class II Medical devices

Recall started
24 June 2021
Reason
There is the potential bacterial including Nontuberculous mycobacterial (NTM) to grow in the water system of heater-cooler devices which could result in contaminated water potentially aerosolized into the operating room during surgery which could result in patient infection.
Distribution
Worldwide distribution - US Nationwide distribution including in the states of TX, PA, FL, CA, HI, SC, OH, AK, NY, IL, AR, WI, MI, AL, KS, NJ, GA, AZ, KY, MS, TN, LA, DE, MA, VA, CO, NE, NC, NH, ME, WV, PA, MO, MN, IN, CT, NH, NV, UT and the country of Venezuela.
Initiated
Voluntary: Firm initiated
Recall number
Z-0069-2022

Ongoing Class II Medical devices

Recall started
12 May 2021
Reason
Labeling does not include guidance for proper water-quality maintenance and device inspection for heater cooler devices.
Distribution
Worldwide distribution - US Nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV and the countries of Germany
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0035-2022, Z-0038-2022, Z-0037-2022, Z-0036-2022

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.