Cardiac Assist, Inc recalls

7 recalls recorded since 2018, most recently on 30 June 2023. 2 are listed as ongoing. Status and classification are as published by the FDA.

Recalls7
Ongoing (FDA)2
Terminated or completed5
Class I (FDA)2

Ongoing Class II Medical devices

Recall started
30 June 2023
Reason
Epoxy used during manufacturing of the LifeSPARC Pump may have uncured epoxy. if undetectable curing may result in added risk of early pump failure,
Distribution
US Nationwide distribution in the states of IL, FL.
Initiated
Voluntary: Firm initiated
Recall number
Z-2465-2023

Terminated Class I Medical devices

Recall started
5 December 2022
Terminated
8 August 2023
Reason
A software update (v1.1.5) has been developed to address the issue of Critical Failure which can occur on the LifeSPARC Controller. The Critical Failure was first addressed in the firm's recall initiated July 21, 2022. During the Critical Failure, the software freezes or crashes and the screen does not display data.
Distribution
US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall number
Z-0926-2023

Terminated Class I Medical devices

Recall started
21 July 2022
Terminated
8 August 2023
Reason
Critical Failure of the LifeSPARC controller when the software freezes or crashes and the screen does not display data.
Distribution
US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall number
Z-1763-2022

Ongoing Class III Medical devices

Recall started
14 October 2021
Reason
Top label of the outer package of the collection kit Item 5820-3118 showed expiration date of 2022/05/01 (correct expiration date) while side label showed expiration date of 2021/05/01 (incorrect expiration date)
Distribution
US Distribution to states of: DC, MA, SC, and VA.
Initiated
Voluntary: Firm initiated
Recall number
Z-0311-2022

Terminated Class II Medical devices

Recall started
19 March 2021
Terminated
22 November 2021
Reason
On 02/26/2021, it was found that there were incorrectly packaged Dilator sets in Finished Goods. On 03/1/2021, it was confirmed that 14 kits containing the impacted products had been distributed to the field.
Distribution
US Nationwide distribution in the states of AL, AZ, IL, TN.
Initiated
Voluntary: Firm initiated
Recall number
Z-1486-2021

Terminated Class II Medical devices

Recall started
6 March 2020
Terminated
20 November 2020
Reason
Failure to prime due to an assembly error
Distribution
US Nationwide distributions.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1954-2020, Z-1950-2020, Z-1948-2020, Z-1951-2020, Z-1952-2020, Z-1955-2020, Z-1953-2020, Z-1949-2020

Terminated Class II Medical devices

Recall started
19 November 2019
Terminated
7 October 2020
Reason
The kit contained a 29Fr ProtekDuo cannula instead of a 24Fr Protek Solo Venous cannula.
Distribution
US Nationwide distribution in the states of AZ, AL, ME, LA, PA.
Initiated
Voluntary: Firm initiated
Recall number
Z-1106-2020

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.