Capintec Inc recalls
2 recalls recorded since 2018, most recently on 3 March 2020. None are listed as ongoing. Status and classification are as published by the FDA.
Recalls2
Ongoing (FDA)0
Terminated or completed2
Class I (FDA)0
Terminated Class II Medical devices
- Recall started
- 3 March 2020
- Terminated
- 12 November 2021
- Reason
- It is possible that the collimators have screw hole locations that were insufficiently tapped. There is a potential for the collimator assembly to fall and come into contact with a patient or operator.
- Distribution
- Worldwide distribution. US Nationwide, Australia, Austria, Bangladesh, Brazil, Canada, Chile, Colombia, Costa Rica, Ecuador, France, Hong Kong, India, Israel, Italy, Korea, Malaysia, Pakistan, Philippines, Poland, Saudi Arabia, Spain, Sweden, Taiwan, Thailand, Turkey, UAE, UK, and Vietnam. (list u
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-1564-2020
- Captus 4000e Thyroid Uptake System, Model Numbers 5430-30151 and 5430-30152 - Product Usage: is intended to be used by trained nuclear Medicine Technologists or Nuclear Medicine Physicians to perform Thyroid Uptake Proce
Terminated Class II Medical devices
- Recall started
- 5 November 2018
- Terminated
- 30 June 2020
- Reason
- There is a potential for the spring arm failure as a result of a broken tension rod (a component inside the arm). A break in the tension rod can casue the collimator to fall downward to its lowest point of travel, approximately 25 inches from the ground.
- Distribution
- Worldwide distribution - US nationwide distributed in the states of AR, AZ, CA, CO, CT, DC, FL, IA, IN, LA, MD, MI, MN, NC, NJ, OH, OK, PA, SC, SD, TX, VA, WI, WV, and countries of Argentina, Brazil, Canada, China, Cyprus, Greece, Hong Kong, India, Japan, South Korea, Vietnam.
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-0551-2019, Z-0550-2019
- Captus 700 Thyroid Uptake System, Model # 5430-3137 and 5430-3138
- Captus 3000 Thyroid Uptake System, Model # 5430-0076 and 5430-0077