Canon Medical System, USA, INC. recalls

28 recalls recorded since 2018, most recently on 26 August 2026. 22 are listed as ongoing. Status and classification are as published by the FDA.

Recalls28
Ongoing (FDA)22
Terminated or completed6
Class I (FDA)0

Ongoing Class II Medical devices

Recall started
26 August 2026
Reason
Loosening of the bolts securing the longitudinal movement mechanism may cause the tabletop to shift, resulting in [positional/imaging/patient] misalignment and a longitudinal motion error
Distribution
US Nationwide distribution in the state of ID.
Initiated
Voluntary: Firm initiated
Recall number
Z-0108-2027

Ongoing Class II Medical devices

Recall started
4 August 2026
Reason
Due to the fixing screws of ceiling movement gear in ceiling suspended arm that are part of the X-ray interventional systems may become loose.
Distribution
U.S. Nationwide distribution in the states of CA, GA, MN, NC, NV, PA, OH, and TX.
Initiated
Voluntary: Firm initiated
Recall number
Z-3014-2026

Ongoing Class II Medical devices

Recall started
19 December 2025
Reason
There is a potential for formation of ice occurring in the venting system of the superconducting magnet of the MRI systems. if a quench occurs in this condition, helium gas may not be able to escape via the designed vent paths, and the pressure in the helium vessel may rise. This pressure build-up could eventually rupture the helium vessel, causing helium gas to be released into the MR scanning room.
Distribution
U.S. Nationwide distribution in the states of CA, CO, FL, IA, IL, IN, KS, MD, MO, MT, NV, NY, OH, OR, PA, PR, TX, VA, and WI.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1017-2026, Z-1018-2026

Ongoing Class II Medical devices

Recall started
19 December 2025
Reason
It has been discovered that there is a possibility that the fixing screws of the ceiling movement gear in the ceiling suspended arm that are part of the X-ray interventional systems may become loose. As a result, ceiling lateral movement may no longer be possible, abnormal noise may be generated from the suspension system and a sensor error may be displayed
Distribution
Worldwide distribution - US Nationwide and the country of Dominican Republic.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1177-2026, Z-1178-2026, Z-1179-2026, Z-1176-2026

Ongoing Class II Medical devices

Recall started
16 October 2025
Reason
Affected system contains components that do not meet the required distance between electrodes for compliance with the IEC 60601-1 standard for medical electrical equipment.
Distribution
US distribution to Ohio and Nevada
Initiated
Voluntary: Firm initiated
Recall number
Z-0327-2026

Ongoing Class II Medical devices

Recall started
2 September 2025
Reason
It has been found that when SPOT Fluoro is first turned on, the default ROI is selected and the collimator blades move to the boundary of the ROI, and the static image is overlayed around the ROI, covering the collimated area. If the customer changes the ROI, the collimator blades and static images are adjusted accordingly. However, when the acquisition program is changed, the collimator returns to the default ROI position which may be larger than the specified one, leaving the X-ray area partially covered by the static image. This is a software problem with the SPOT Fluoro program and will not occur if the function is not used.
Distribution
U.S.
Initiated
FDA Mandated
Recall number
Z-0037-2026

Ongoing Class II Medical devices

Recall started
17 September 2024
Reason
When PET-CT system is executing reconstruction, if PET acquisition for another patient is performed (or PET reconstruction for another patient is performed from raw data processing), PET reconstruction control processing may hang up and PET reconstruction may stop as a result, which may necessitate repeat PET/CT examinations.
Distribution
US: NV, AR, MN, IA, IL, NC, LA, KS, MS, CA, AL, MD, CO, IN, TN, AZ, WI, VA
Initiated
Voluntary: Firm initiated
Recall number
Z-0284-2025

Ongoing Class II Medical devices

Recall started
10 May 2024
Reason
The terminal block used to secure the power supply cable for the gradient coil of the MRI system was not secured as designed, which could cause cable disconnection, which could lead to arcing, followed by melting of the cable, ignition, smoke, and fire.
Distribution
US: PA, SC, IA, AL, MI, AR, SD, TX, ID, KY, NY, MN, MD, OH, WI, NV, CT, AZ, IL, OK, CA, FL, GA, OR, IN, NJ, CO, MT, KS, MO, WY, VA, VT, LA, NH, WV, TN, PR
Initiated
Voluntary: Firm initiated
Recall number
Z-2041-2024

Ongoing Class II Medical devices

Recall started
29 January 2024
Reason
For some MRI systems, it has been found that some of the maximum Spatial Field Gradient (SFG) values displayed in "System Information" of the operation window and some of the maximum SFG values described in the safety manual are lower than the correct values, which may cause a patient with an MR conditional device to feel some discomfort during the MR scan.
Distribution
Worldwide - US Nationwide distribution including in the states of MD, IL, WV, CA, AZ, PR, CO, GA, PA, WA, UT, OK, KY, IA, IN, TN, CT, WI, OH, AL, MT, NM, TX, MO, MN, OR, ID, WY, NV, AR, FL, SD, MI, NY, NJ, KS, LA, NH, VT, MS, SC, ME and the country of DO.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1249-2024, Z-1250-2024, Z-1248-2024

Ongoing Class II Medical devices

Recall started
15 November 2023
Reason
VAREX changed the manufacturing of the internal board, and the detectors displayed abnormal images. Internal board of the X-ray flat panel detector may display abnormal images, such as horizontal stripes, which may lead to the inability to complete an ongoing exam, or the patient may need to be examined with another x-ray system.
Distribution
US: CA, LA
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1155-2024, Z-1154-2024

Ongoing Class II Medical devices

Recall started
22 June 2023
Reason
The C-arm, used to set the position and angle of the x-ray exposure field, may not function due to the tableside console PWB, which needs replacement. There may be a delay in the procedure, but procedures can continue using the operators console, the switches on the back of the column, switches on the side of the flat panel detector, and the emergency stop will continue to function normally.
Distribution
US Nationwide Distribution: NY, CO, IL, CA, MD, MN, WA
Initiated
Voluntary: Firm initiated
Recall number
Z-2392-2023

Ongoing Class II Medical devices

Recall started
29 September 2022
Reason
After completion, ultrasound system's Stress Echo may not terminate if user does not manually terminate the protocol. During next patient exam, an abnormal confirmation message will be displayed "Same view was saved in this phase. Do you use this image with StressEcho Review?" If "no", acquired images not saved, and if "yes" images assigned to the previous patient; could cause treatment delay.
Distribution
US: MT, MO, NC, SC, MN, WA, KY, TN, AR, IL, IA, MI, WV, KS, CA, HA, CO, LA, OK, ME, KY, NY, PA, GA, FL
Initiated
Voluntary: Firm initiated
Recall number
Z-0198-2023

Ongoing Class II Medical devices

Recall started
3 May 2022
Reason
CT operation may be restricted by an interlock which is a result of an system error and the CT system may stop operating properly after an attempt to cancel the error message has been initiated.
Distribution
U.S.: CA, CO, FL, NY, and WY O.U.S.: None
Initiated
Voluntary: Firm initiated
Recall number
Z-1289-2022

Ongoing Class II Medical devices

Recall started
31 January 2022
Reason
The fluoroscopic dose rate might exceed the conforming value during biplane fluoroscopy.
Distribution
US Nationwide Distribution
Initiated
FDA Mandated
Recall number
Z-1666-2022

Ongoing Class II Medical devices

Recall started
20 December 2021
Reason
Abnormal images such as horizontal stripes may be displayed and recovery may not not be possible by restarting the system. The worst case scenario is an operator who is working without a workable image may remove a catheter from the patient's blood vessel and damage the blood vessel.
Distribution
US Nationwide distribution in the states of DE, FL, LA, MT, NJ, OH, and WV.
Initiated
Voluntary: Firm initiated
Recall number
Z-0492-2022

Ongoing Class II Medical devices

Recall started
21 May 2021
Reason
The MRI system contains a component that does not meet the appropriate flame retardancy-rating which could result in smoke or fire.
Distribution
US Nationwide distributions in the states of AL, AR, AZ, CA, CO, CT, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, SC, TN, TX, UT, VA, VT, WA, WI, WV, WY.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2046-2021, Z-2044-2021, Z-2045-2021, Z-2047-2021

Ongoing Class II Medical devices

Recall started
5 February 2021
Reason
The x-ray irradiation field may shift with respect to image receiving surface displayed on the screen at some C-arm angles.
Distribution
Nationwide
Initiated
FDA Mandated
Recall numbers
Z-1175-2021, Z-1176-2021, Z-1174-2021

Ongoing Class II Medical devices

Recall started
8 December 2020
Reason
System table may tilt because the bolts that fasten the vertical column and the table top tilting drive unit may become loose.
Distribution
Worldwide distribution - U.S. Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MT, NC, NJ, NY, OH, OK, PA, SD, TN, TX, WA, WI and WV. The countries of Canada, China, Japan, Korea, and Singapore.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1271-2021, Z-1270-2021, Z-1269-2021

Terminated Class II Medical devices

Recall started
11 November 2020
Terminated
30 June 2023
Reason
A software problem has been identified which could result in the diagnostic imaging system not proceeding to the next actual scan even though automatic start of the next scan is specified. This could result in the diagnostic imaging system failing resulting in rescanning and reinjection of contrast medium.
Distribution
US Nationwide distribution in the states of NV, AR MN.
Initiated
Voluntary: Firm initiated
Recall number
Z-0142-2022

Terminated Class II Medical devices

Recall started
9 November 2020
Terminated
16 December 2022
Reason
A software problem has been identified which could result in the CT Scanner not proceeding to the next actual scan even though automatic start of the next scan is specified. This could result in the CT scan failing resulting in rescanning and reinjection of contrast medium.
Distribution
US Nationwide distribution. OUS: Unknown
Initiated
Voluntary: Firm initiated
Recall number
Z-0549-2021

Terminated Class II Medical devices

Recall started
8 October 2020
Terminated
6 December 2022
Reason
There is a potential that the cable of the Flat panel detector may break during a procedure which may cause the x-ray device to stop before the examination is completed. This could result in a delay with procedure and/or potential damage to the patients blood vessels due to removal without visibility.
Distribution
US Nationwide distribution including in the states of LA, CA, CO, WV, GA, NH, FL, PA, WI, MN, ID.
Initiated
Voluntary: Firm initiated
Recall number
Z-0532-2021

Ongoing Class II Medical devices

Recall started
8 July 2020
Reason
During a procedure, when images were acquired, and these images were transferred to PACS /image archival system, the number of images displayed were not the same as originally acquired.
Distribution
US - CA, CT, ID, LA, MT, NV, OH, SC, WI and WV
Initiated
Voluntary: Firm initiated
Recall number
Z-2885-2020

Ongoing Class II Medical devices

Recall started
3 March 2020
Reason
Bolts holding the longitudinal movement of catheterization tables may be loose, which may lead to table surface/top tilting.
Distribution
US Nationwide distribution including in the states of PA, AR, OH, NC, CA, IA, GA, IL, AL, WV, LA, MO, FL, MS, TX, AZ, MA, NJ, MD, NY, NV, OK, WI, PR.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2327-2020, Z-2325-2020, Z-2326-2020

Ongoing Class II Medical devices

Recall started
25 March 2019
Reason
Scanning may be interrupted due to an error during execution of the eXam Plan and the system is unable to be shut down normally, resulting in the need to forcibly turn the system power off and reboot. Loss of the acquired raw data would occur.
Distribution
Distribution was made to AK, AR, CA, FL, IA, ID, IL, KS, MA, MD, MI, MN, MO, MT, NC, ND, NH, NY, OH, OK, OR, PA, TN, TX, VT, WA, WI, and WV. There was government and military distribution but no foreign distribution.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1726-2020, Z-1724-2020, Z-1723-2020, Z-1719-2020, Z-1725-2020, Z-1722-2020, Z-1721-2020, Z-1720-2020

Terminated Class II Medical devices

Recall started
8 February 2019
Terminated
7 June 2022
Reason
Table tilts without operator involvement as a result of liquid spilling into tilt console triggering a short circuit.
Distribution
Worldwide distribution - US Nationwide distribution in the states of AL, CT, FL, IL, OR, PA, and TX; and country of Netherlands.
Initiated
Voluntary: Firm initiated
Recall number
Z-0360-2020

Terminated Class II Medical devices

Recall started
16 October 2018
Terminated
20 July 2020
Reason
software malfunction; It was found when a user performs radiography using the wireless flat panel detector (FPD), a message window displays on the monitor stating image transmission was not completed and there was no image available. It also showed the "OK" button to reacquire image data from the FPD, and the "Cancel" button to cancel the acquisition . When the user selects the "OK" button, the same message window appears. This prompted the user to repeat the same operation several times and finally select the "Cancel" button to quit the reacquistion mode.
Distribution
US Nationwide Distribution - PA, UT, FL, NY, AR, LA, WV, NJ, OH, MI, WI, TN, TX, VT, GA, NC, IL
Initiated
Voluntary: Firm initiated
Recall number
Z-1958-2019

Ongoing Class II Medical devices

Recall started
30 May 2018
Reason
When using respiratory gating with Steady Mode ON in Spin Echo sequence scanning, the SAR (Specific Absorption Rate) value displayed on the console may be lower than the actual SAR value. The actual applied SAR values may exceed the FIRST LEVEL CONTROLLED OPERATING MODE.
Distribution
Domestic: AL, AR, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, LA, MA, MD, MI, MN, MO, MS, MT, NC, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VT, WA, WI, WV, and WY. International: Algeria, Australia, Bangladesh, Brazil, Canada, China, Colombia, Congo, Dominican Republic, Egypt, Guine
Initiated
Voluntary: Firm initiated
Recall number
Z-0199-2019

Terminated Class II Medical devices

Recall started
18 April 2018
Terminated
13 May 2021
Reason
It was found when an operator performs a radiography using the wireless flat panel detector (FPD), a message window was displayed on the monitor stating image transmission was not completed and there was no image. It also showed the OK button to reacquire image data form the FPD, and the Cancel button to cancel the reacquisition. When the operator selects the OK button, the same message window appears. The operator then repeated the same operation several times and finally selected the Cancel button to quit the reacquisition mode.
Distribution
Domestic: AR, FL, GA, PA, IL, LA, MI, NC, NJ, NY, OH, TN, TX, UT, VT, WI, and WV Foreign: Australia, Canada, The Netherlands, South Korea, and Malaysia
Initiated
Voluntary: Firm initiated
Recall number
Z-2964-2018

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.